
______________________________________ Bronkhorst® ______________________________________
________________________________________________________________________________________________
9.17.119A MASS-STREAMTM Instruction Manual D-6400
3. Operation
After correct installation of the MASS-STREAM D-6400 and when all safety precautions have been taken into account,
the instrument can be used for measuring and/or controlling mass flow in the system.
3.1.Powering up and powering down
3.2.First use
3.3.Mass flow measurement and control
When powering up, the instrument needs a couple of seconds to
start up the electronics. As soon as the start-up sequence has
finished (green LED glows continuously), the instrument is ready to
measure mass flows, however, optimal accuracy is only reached
after warming up (see Powering up and powering down).
After powering up, the control valve closes (normally open) or stays
closed (normally closed). The valve stays closed until the instrument
receives a setpoint from the active setpoint source. The internal PID
controller then immediately opens the control valve, until the
measured flow rate matches the setpoint. It maintains the resulting
flow rate until another setpoint is given.
•It is recommended to turn on power before applying fluid pressure and to switch off power only
after relieving fluid pressure.
•For best performance, allow the device to warm up and stabilize for at least 30 minutes before
starting measurement and/or control. This may be done with or without media flow.
When applying pressure, avoid pressure shocks and bring the fluid system gradually up to the level of
the specified operating conditions; open the fluid supply gently.
In systems for use with corrosive or reactive media, purging for at least 30 minutes with a dry, inert gas (like
Nitrogen or Argon) is absolutely necessary before use. After use with corrosive, reactive or hazardous
media (e.g. toxic or flammable), purging is also necessary before the fluid system is exposed to air.
If the instrument is mounted in a position with upward or downward flow, adjusting the zero point is
advised before using the instrument for the first time. See Adjusting zero point for background
information and instructions.
After changing to another fluid set, the instrument must be zeroed at the process conditions. See
Adjusting zero point for background information and instructions.
MASS-STREAM D-6400
instruments are most accurate at the specified inlet/outlet pressure,
temperature and process gas conditions. However, the instrument will function properly in a wide range
of varying conditions. It is strongly advised to use the FlowTune™ software available with the instrument
to set the correct process conditions if the actual process conditions differ from the conditions for which
the instrument is set (see Changing fluid set).
Although MASS-STREAM D-6400 instruments have good temperature stability, the best accuracy is
achieved when temperature gradients across the instruments are avoided. Make sure that the gas
temperature matches the ambient temperature as good as possible and mount the instruments on a
rigid (heat conducting) surface.
MASS-STREAM D-6400 instruments handle pressure shocks in the system well but are not insensitive
to pressure fluctuations. For optimum control stability, provide a stable (pressure controlled) inlet
pressure with sufficient buffer volume between the pressure regulator and the instrument and avoid
installing multiple instruments or control valves in close proximity to another with small volume piping in