9.17.0932
SCOPE OF THIS GUIDE
min i CORI -FLOW™ instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or
gases, independent of fluid properties. These smart Coriolis instruments offer multiple process values as input- or output
parameters. Many parameters can be read and/or changed using analog or digital interfaces. Output parameters are: mass flow,
density, temperature, totalized mass flow, alarms. Input parameters are: setpoint (desired mass flow rate for controllers), reset
alarm/counter.
This manual will help you start-up your mini CORI-FLOW™ ML120 in only 10 steps and contains:
1. Instrument functional properties
2. Check safety properties
3. Check piping
4. Mount/install instrument
5. Leak check
6. Electrical connection
7. Operation
8. Multi-functional switch
mini CORI-FLOW™ ML120 instruments have modular instruction manuals consisting of:
- Instruction manual mini CORI-FLOW™ ML120 (document nr. 9.17.097)
Info about e.g. : sensors, valves, liquid dosing systems, maintenance, tooling, calibration, Kv-value calculation, troubleshooting.
- Hookup diagram mini CORI-FLOW™ and CORI-FLOW™ (general) (document nr. 9.16.132)
- FlowPlot Manual (document nr. 9.17.030)
Depending on optional fieldbus interface:
Fieldbus/interface description Manual Hookup diagram
- RS232 Interface with FLOW-BUS protocol 9.17.027 9.16.132
- FLOW-BUS Interface 9.17.024 9.16.133
- PROFIBUS DP Interface 9.17.025 9.16.134
- DeviceNet™ Interface 9.17.026 9.16.135
- Modbus interface 9.17.035 9.16.136
- EtherCAT interface 9.17.063 9.16.137
- PROFINET 9.17.095 9.16.146
These docum ents can be downloaded from the website:
http://www.bronkhorst.com/qrcoriolis or can be sent by e-mail on request.
Notes for temperature considerations
After ha ving use d the mini C ORI -F LOW™ ML120 the first time at low temperature tighten the fluid adapter screws again in
order to prevent any leakage! Please note: if you do not tighten, the leaking adapter / fitting can cause damage! After the
first shrinking and tightening of the screws, no further precaution is necessary.
Note: The maximum allowed temperature in the housing of the mini CORI -FLOW™ ML120 is 70 °C. To check this, the
internal temperature sensor can be used. Temperature can be readout digitally via FLOW-DDE/E-8000 (FlowDDE par. 142) or
BRIGHT (local readout/control module). Make sure the temperature value readout here (=actual temperature in housing)
will not exceed 70 °C.
Check mini CORI-FLOW™ ML120 functional properties
Before installing your Mass Flow Meter/Controller it is important to
read the attached label and check:
·Flow rate
·Fluid to be measured
·Up- and downstream pressures
·Input/output signal
·Temperature
·Valve type (if controller)