In the step-by-step instructions, you are given the action of the three buttons mainly
used to program data. The normal procedure at each step is to push on a
number of times to make changes and push on ENTER to validate changes and
advance to the next programming step. At the end of a complete menu sequence the
meter returns to the run mode saving changes in memory.
In general the following actions can be made during the program mode.
ENTER validate changes and advance to next step
ESC discard changes and go to the run mode
select among a list of available options / shift to next digit to the right
increment di
it value
The programming instructions are composed by a general description and a series of step-by-step instructions to be followed
sequentially. Each menu step is represented by an illustration of the display and keyboard module with indicators (display and
LED's), reference [page number . figure number] and a text describing the action of each key at that step.
12
TARE
HOLD LIMIT MA
MIN DAT
DSP2INP
FL
STORE
DSP1
INP1
1
4
3
TARE
RESET LIMIT MAX/MI
ENTE
ES
DAT
PROG
TEAC
PRG
RUN
[page nº/figure nº] Mnemo
With respect to the figures in the step-by-step instructions, the display indications may have the
following meanings :
1./ The display shows one of the available options with filled-out segments. That means
that the display shows the choice made previously. The use of allows to select from
available options.
2./ A series of black "8" also represents the display indication of a previous choice, with
the difference that it cannot be changed in the current step. If it is already the desired
parameter, you may exit from the menu by a push of ESC without making changes or,
if wanted to modify it, a push of ENTER advances the meter to the next step where
changes are allowed.
3./ A series of white "8" represents any numerical value that is programmed by using keys
(increment digit value) and (advance to the next digit).
Program
module and
menu step
indicators
32
Nº Function Description Activation
13 ANA GROSS Makes the analog output follow the gross value (measured value + tare) Fixed level
14 ZERO ANA Puts the analog output to the zero state (0 V for 0-10 V, 4 mA for 4-20 mA) Fixed level
15 ANA PEAK Makes the analog output follow the peak value Fixed level
16 ANA VALLEY Makes the analog output follow the valley value Fixed level
13 to 16 : FUNCTIONS ASSOCIATED WITH THE ANALOG OUTPUT
17 to 23 : FUNCTIONS FOR USE WITH A PRINTER VIA THE RS OUTPUTS
Nº Function Description Activation
17 PRINT NET Prints the net value. Pulse
18 PRINT GROSS Prints the gross value. Pulse
19 PRINT TARE Prints the tare value. Pulse
20 PRINT SET1 Prints the setpoint1 value and its output status. Pulse
21 PRINT SET2 Prints the setpoint2 value and its output status. Pulse
22 PRINT SET3 Prints the setpoint3 value and its output status. Pulse
23 PRINT SET4 Prints the setpoint4 value and its output status. Pulse
24 to 25 : FUNCTIONS ASSOCIATED WITH THE SETPOINTS
Nº Function Description Activation
24 FALSE SETPOINTS Exclusively for instruments WITHOUT relays/transistors control outputs card.
Allows programming and operation of 4 setpoints. Fixed level
25 RESET SETPOINTS Exclusively for instruments with 1 or more setpoints programmed as "latched setpoints" (That is, the
setpoints that once energized remain on the ON status althoug the alarm condition disappears).
Desactivates the setpoints output.
Pulse
26 to 31 : SPECIAL FUNCTIONS
Nº Function Description Activation
26 S&H SETPOINTS Grants the setpoints run during the S&H function activation Fixed level
27 SAMPLE & HOLD On activating this function, the measuring values peak,valley, peak-peak and actual
value are hold on display as well as analog output, rs output and setpoints execpt if
function 26 is active. View details on page.34. Fixed level
28 ASCII SEND Transmits the four last digits of the display to a remote ASCII indicator.
By holding the input to a low level, transmission takes place every second. Pulse or fixed level
29 PEAK-PEAK Display the peak-peak value Pulse
30 ANALOG PEAK-PEAK The analog output follows the peak-peak value. Fixed level
31 FAST RS Sends by serial output RS2 or RS4 the display value at 200 per second. View Page. 54. Fixed level