③.Before carrying the test, please weigh the specimen according to the test requirements.
④.Remove the load weight, lift the lever, put the specimen into the specimen’s holder.
⑤.Lay down the lever, press load pressure head on the surface of the specimen, adjust pressure
head according to the samples height to make the lever balance, hang on load weight.
⑥.According to the test requirements, corresponding label grinding materials well-distributed inside
test track on the grinding wheel.
⑦.Set the revolutions on the counter, default setting is 22. If need to change to other revolutions,
please press the button “ ” to make the corresponding digit flashes, then through pressing
the button “ ” to increase and decrease numbers, without other operations, wait 10 seconds
to save automatically.
⑧.Then press the button “ ” to clear the numbers to zero, press the button ”start”, the test starts.
⑨.After reaching the set revolutions, the grinding wheel will stop automatically. If need to quit the
test during the test process, please press the button “stop; if need to stop the equipment in
emergency, please press the button “emergency stop” or the button “stop”.
⑩.When test finished, remove the load weight, lift the lever, take out the specimen.
⑪.If not need anymore, please clear the grinding wheel and turn off the power, lid the machine
clothing.
IV. Common faults and troubleshooting
1 Main engine cannot start up
Power connection error Connect power as required
Frequency converter error Check the frequency converter
2 Motor cannot start up
Emergency stop button close Unscrew Emergency stop button
Counter alarm Press the counter resetting key
3 Inaccurate counter
Proximity switch install location moved Adjust proximity switch position
and fasten
Electromagnetic interference Far away interference source