5
When using a lift to place the user in
the bed and the head end of the bed is
raised, make sure that the user is not
placed too high on the mattress.
Otherwise, there is a risk of shear.
When the alternating or pulsating
program is used and the head end
of the bed is raised, make sure that
the user and/or the mattress is not
moving downwards due to the
movement in the mattress. Also
raise the foot end of the bed.
3Common operations
Following operations apply to the Cirrus system
regardless of which control unit that is used.
3.1 Sitting positioning in bed
When raising the head end of the bed into a
sitting position, always secure the user´s
position. To ensure the product functionality,
we always recommend to perform a hand check
(see section 4 or 5 depending on the pump and
mattress of interest). This function is
recommended to use for short periods only. For
additional support, positioning pillows can be
used.
3.2 CPR (Cardipulmonary resuscitation)
In case of an emergency where CPR
(Cardiopulmonary resuscitation) is necessary,
remove the connection (marked ‘CPR’) from the
control unit
and leave the
lid open in
order to
empty the
mattress of air
quickly.
3.3 Transport function
If the user needs to be moved in bed, either:
Unplug the CPR connection, close the lid, place
the CPR connection at the end of the bed and
remove the control unit from the bed. The
mattress will retain air for at least 12 hours.
Or
Remove the power supply from the wall socket
and leave the control unit hanging on the bed
during transport. The mattress will retain air for
at least 12 hours.
We recommend using this function for short
periods only.
3.4 Restart
If a restart is
required, set the
On/Off switch on
the side of the
control unit to 0 (Off). Wait for approx. 10
seconds and restart the control unit.
3.5 Maximum pressure notification
When the function Maximum pressure
has been used for a long time, the
Maximum pressure diode will blink. If the
use is intentional, ignore the notification.
3.6 Power failure
In the event of a power failure, unplug the CPR
connection, close the lid, place the CPR
connection at the end of the bed and remove
the control unit from the bed. The mattress will
retain air for at least 12 hours. Perform a hand
check to make sure the pressure of the mattress
is not too hard or too soft.
3.7 Automatic Setting
The mattress system regulates the mattress's
internal pressure independently and without
manual adjustment to different values based on
the user's weight and position. No manual action
is required to affect the internal pressure of the
mattress. This function works in the following
three ways:
1. The control unit carries out an automatic setting
immediately after switching the system on.
2. During usage, if any significant change occurs,
the control unit will perform an automatic setting.
3. When in use, the control unit will carry out
automatic settings at fixed intervals to ensure
correct control of the pressure in the mattress at
all times.
Once the automatic setting has completed, the
control unit will start in the Pulsating mode.