2 CS 7600 System Overview
Introduction
CS 7600 is a scanner system for intraoral X-ray exposed imaging plates.
The CS 7600 is intended for digital dental radiography using an imaging plate (storage phosphor
plate) for radiographic diagnostic intraoral images. The CS 7600 system is used to scan and review
intraoral dental X-ray images. When scanning the X-ray exposed imaging plate a digital image is
previewed on the scanner’s LCD and saved to the scanner’s internal memory. After scanning, the
scanner erases the imaging plate and ejects it. The imaging plate is ready for re-use.
The exam acquisition’s scanned image is sent over the network to the workstation’s image Acquisition
interface. Using the image Acquisition interface the image is processed and reviewed.
The image Acquisition interface sends the images to the Kodak Dental Imaging Software, which
archives the acquired images.
The scanner connects directly to the network, or is connected peer-to-peer to the workstation.
The scanner can be optionally placed on a desk/counter top or mounted on a wall, using a special shelf
(adapter).
The system is composed of the following elements:
•CS 7600 scanning device (page 4)
•Smart imaging plates (page 6)
•CS 7600 image Acquisition interface (page 11)
•Scan & Go device (optional, page 9)
Note: The CS 7600 is intended for dental professionals. You are required
to have training to use the CS 7600.
Note: The scanner is supplied with a separate viewing and archiving
software package: Kodak Dental Imaging Software, approved by
Carestream Health.
Note: The CS 7600 User Guide is also provided as a PDF file on the
supplied CD-ROM.

4 Chapter 2 CS 7600 System Overview
CS 7600 Components Overview
Scanner Overview
The CS 7600 scanner performs the functions necessary for scanning and acquiring images from the
imaging plates.
Figure 1 CS 7600 Front View
13.5" QVGA color LCD (320 X 240 pixels, landscape matrix)
2Function buttons for scanner use and menu navigation
3On/Off button + LED power indicator.
On/Off button functions:
•Turn scanner On: press for 1 second
•Turn scanner Off (close application and files and then turn off): press for 2 seconds.
4Base
5Intraoral imaging plates tray
6Detachable insertion panel. The CS 7600 makes use of two insertion panel sizes according to
the inserted imaging plates size (0–3 or 4).
7Front cover
8Back covers
2
7
8
3
6
1
4
5

CS 7600_User Guide (8J4071)_Ed01 5
Figure 2 CS 7600 Rear View
Figure 3 Cables Compartment in Detail
1Service door
2Base
3Cables compartment for power supply connection and interface ports (RJ-45 Ethernet port)
4Back cover
1
4
3
2
1
3
2
4
1Power supply inlet
2RJ-45 Ethernet port (Ethernet
activity LEDs: yellow indicates
speed, green indicates link status)
3Service ONLY ports
4Compartment cover

6 Chapter 2 CS 7600 System Overview
Smart Imaging Plates Overview
The CS 7600 works with five imaging plate sizes: 0 to 4. The scanned image file size depends on the
imaging plate size and the scanning resolution set by the user.
The smart function of the imaging plates is utilized when using the Scan & Go device.
Table 1 Intraoral Imaging Plate Sizes
Size Illustration Application
0 Periapical exam (Pediatric dentistry)
1 Bitewing exam (Pediatric dentistry)
2 Periapical exam (Adult)
3 Bitewing exam (Adult)
4 Occlusal exam

CS 7600_User Guide (8J4071)_Ed01 7
The imaging plate has two sides: the active side for capturing the X-ray image and the inactive side.
The active side of the imaging plate includes an orientation mark ‘c’ to facilitate accurate positioning
within the patient’s mouth. The inactive side has a printed dot that corresponds to the location of the
orientation mark on the active side. This dot is visible through the transparent hygienic sheath and is
also helpful for positioning the imaging plate correctly in the patient’s mouth.
Hygienic Sheaths
For each imaging plate size there is a corresponding one-time use, disposable, hygienic sheath for
preventing cross contamination. The transparent side allows you to see the orientation mark for
positioning the imaging plate correctly in the patient’s mouth while the opaque side protects the active
Inactive side
Active side
C
Important: Use only Carestream imaging plates and hygienic
sheaths!

8 Chapter 2 CS 7600 System Overview
side of the inserted imaging plate from light damaging effects. The foam strip is a stopper which
prevents inserting the hygienic sheath by mistake into the scanner (as long as the insertion direction is
correct).
Figure 4 Hygienic Sheath (front view)
1 Adhesive strip (for sealing the hygienic sheath)
2 Foam stopper
3 Sheath, transparent on one side and with a light protection barrier on the other side
3
1
2

CS 7600_User Guide (8J4071)_Ed01 9
Scan & Go Overview
Scan & Go is an optional device that enables you to link the imaging plate to a specific exam
acquisition (for example, patient) prior to acquisition. This information is retrieved by the scanner and
the imaging plate is scanned automatically according to this information (resolution, target workstation,
etc.). Scanned images are then routed automatically to the correct workstation, associated to the
correct exam acquisition and placed in the correct frame in the CS 7600 image Acquisition interface.
Figure 5 Scan & Go device (Before June 2012)
1USB 2.0 port
2On/Off button
3On/Off/Recording LED status indicator (blue)
Turning On/Off, 3 short flashes
Standby (intermittent flashing)
Recording data to the imaging plate (successive short flashes)
4Battery LED status indicator (amber):
Charging (static light)
Low battery (successive short flashes)
(light Off) battery charged
5Recording surface for imaging plate
6Mounting adapter
5
1
4
3
2
6

10 Chapter 2 CS 7600 System Overview
Scan & Go Overview (for Scan & Go models released from June 2012)
A new Scan & Go device is available from June 2012. The following improvements have been made to
the new Scan & Go device:
•Faster scanning - Approximately 1/4 sec instead of 1 sec
•No battery - Therefore no need for charging.
•No on off switch - Turns on automatically when the USB is connected.
Figure 6 Scan & Go device (After June 2012)
1USB 2.0 port
2On/Off/Recording LED status indicator (blue)
Power Up: One Flash & Beep (Internal FW)
Beacon: Flashes blue every few seconds in idle mode (Not during Tagging)
Tagging OK: One flash (long) & 1 bleep (long)R
Tagging Fail: One Red flash (long) & 1 bleep (long)R
3Connected to USB: LED status indicator (green):
4Recording surface for imaging plate
5Mounting adapter
4
1
3
2
5

Carestream Health, Inc.
150 Verona Street
Rochester, NY 14 608, USA
0086
Reference:
HT022_20120424_EN
ATHT0296_20120424_EN

Note: To use this Scan & Go, you need to install Acquisition Interface V1.0 Rev 3
(or higher).
Unlike previous versions, this Scan & Go does not use batteries.
Note: If the Scan & Go LED lights up green and the USB icon appears red, there
is a possibility that the image acquisition version needs updating to V1.0.Rev 3

Note: This picture is for demonstration only and the Scanner version may
vary.
Note: If the scanner is already updated with Version 1.30, and you have
clicked Update Scanner, a message appears The scanner is already
updated then, click OK, close the Scanner Manager window.

Note: In the rare case that the Update Failed message appears, restart
the scanner, press and verify that the scanner version is 1.30.
Note: For more information, see the CS 7600 User Guide provided in the original
CD delivered with the system.

Intraoral Imaging Plate System
CS 7600
Safety, Regulatory & Technical
Specifications User Guide

Notice
The Regulatory Information & Technical Specifications User Guide for the CS 7600 includes
information on the safety instructions, regulatory information and the technical specifications of the
device. Please read and observe all warnings and instructions in this Guide and those marked on the
unit.
The information contained in this Guide may be subject to modification without notice, justification
or notification to the persons concerned. Make sure you have the most current version.
No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.
U.S. Federal law restricts this device to sale by or on the order of a dentist or physician.
This document is originally written in English.
Guide Name: Intraoral Imaging Plate System CS 7600 System Safety, Regulatory and Technical
Specifications User Guide
Part Number: 8J4070
Revision Number: 02
Print Date: 2012-06
The Brand names and logos reproduced in this Guide are copyright.
Manufacturer
Authorized Representative in the European Community
Carestream Health France
1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
FRANCE
0086
Carestream Health, Inc.
150 Verona Street
Rochester, NY 14 608, USA
EC REP

CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed02 iii
Contents
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Indications for Use . . . . . . . . . . . . . . . . . . . . . .1
Conventions in this Guide . . . . . . . . . . . . . . . . . . .1
Warning and Safety Instructions . . . . . . . . . . . . . . . .2
General Notice . . . . . . . . . . . . . . . . . . . . . . . .4
Laser Safety Instructions . . . . . . . . . . . . . . . . . . .5
Hygiene and Disinfection . . . . . . . . . . . . . . . . . . .5
Marking and Labeling Symbols . . . . . . . . . . . . . . . .6
IEC Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Label Locations . . . . . . . . . . . . . . . . . . . . . . . .8
Declaration of Conformity . . . . . . . . . . . . . . . . . . .9
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . 11
General Regulatory Information . . . . . . . . . . . . . . . 11
Compliance with International Regulations . . . . . . . . . . 18
Compliance with FCC Rules . . . . . . . . . . . . . . . . . 18
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . 19
Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . 19
Model . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CS 7600 Technical Specifications . . . . . . . . . . . . . . 19
System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Imaging Plate Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
RF Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
CS 7600 Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . 21
Imaging Plates Environmental Requirements . . . . . . . . . . . . . . . . . 22
Minimum Computer System Requirements . . . . . . . . . . 23

iv

CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed02 1
1 Safety Information
Indications for Use
The CS 7600 is intended for digital dental radiography using an imaging plate
(storage phosphor screen) for radiographic diagnostic intraoral images.
Conventions in this Guide
The following special messages emphasize information or indicate potential risk to
personnel or equipment.
WARNING: Warns you to avoid injury to
yourself or others by following the safety instructions
precisely.
Important: Alerts you to a condition that might
cause problems.
Tip: Provides extra information and hints.
Note: Emphasizes important information.

2 Chapter 1 Safety Information
Warning and Safety Instructions
WA R N I NG S
Scanner
•Read and understand this safety information before using the CS
7600.
•To ensure safety, read all user guides carefully before using the
system and observe all Warnings, Important and Notes located
throughout the guides.
•Keep this guide with the equipment.
•You are responsible for the operation and maintenance of this device.
You are required to have training to use the CS 7600. When
necessary, have a trained authorized service technician carry out
inspection and maintenance operations.
•The CS 7600 is Class I, continuous operated stationary equipment
without applied parts and has one signal input/output part. Product
is provided with ordinary protection against the harmful ingress of
water.
•DO NOT use this device in the presence of a flammable anesthetics
mixture with air or with oxygen or with nitrous oxide.
•DO NOT remove or open system covers or plugs. Internal circuits
use high voltage capable of causing serious injury. Fuses blown
within 36 hours of being replaced by a qualified technician may
indicate malfunctioning electrical circuits within the system. Have
the system checked by qualified service personnel. Do not attempt to
replace any fuse. Fluids that seep into the active circuit components
of the system may cause short circuits that can result in electrical
fires. Therefore, do not place any liquid or food on any part of the
system. Access to the rear inspection cover is allowed, as an
exception, according to the instructions for Retrieval of the imaging
plate in the Troubleshooting section of the user guide.
•DO NOT operate the device if there is the threat of an earthquake.
Following an earthquake, ensure that the device is operating
satisfactorily before using it again. Failure to observe this precaution
may expose patients to hazards.
•DO NOT place objects within the field of operation of the device.
•Connect this equipment ONLY to a mains power supply with
protective ground to avoid any risk of electric shock.

CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed02 3
•The appliance coupler/inlet of the flexible power cord is used as a
disconnecting device from the mains.
•To dispose of the device or its components, contact a service
technician.
•No modification of this equipment is allowed.
•DO NOT block air circulation around the unit. Always maintain at
least 15 cm clearance around the unit to prevent overheating and
damage to the system.
•The scanner should be positioned so that there is always easy access
to the mains power supply socket.
Imaging Plates:
•To prevent damage to the imaging plates and the possibility of image
artifacts, avoid contact between the imaging plates and the following
materials/solutions/solvents: Isopropyl alcohol, hydrogen peroxide
and other peroxides, citrus-based cleaners, hand lotions and
water-less hand sanitizers, as well as surfactants and lubricants.
•The imaging plate contains Barium and should be considered
hazardous or special waste in specific conditions at the end of its
useful service life. For disposal or recycling information, contact
your local authorities.
•Do not soak the imaging plate in any cleaning or disinfecting
solutions. Do not autoclave; autoclaved imaging plates must be
discarded.
•Imaging plates should be stored in their original packing or storage
box when they are not in use. Always store imaging plates in a dark
and dry place.
•Do not expose the imaging plates to light for long periods as this can
have a degrading effect.
•Do not store imaging plates in hot or moist conditions.
•Do not fold, crease, or bend the imaging plates.
•Avoid touching the imaging side of the imaging plates and be careful
not to drag the imaging side of the imaging plate across any surface
as this will damage the imaging plate.
•Do not leave imaging plates where they can become damaged by
liquid or chemical spills.
•Read and follow the instructions in Material Safety Data Sheets
(MSDS) for the CS Screen Cleaner.

4 Chapter 1 Safety Information
•If a commercially prepared equivalent solution of diluted bleach is
used, it should be used according to its manufacturer’s instructions.
Computer:
•DO NOT place the computer and the peripheral equipment
connected to it in the immediate vicinity of the patient in the unit.
Leave at least 1.83 m distance between the patient and the unit. The
computer and the peripheral equipment must conform to the
IEC60950 standard.
•See your computer installation guide for details of the data
processing system and screen. Leave a sufficient amount of clear
space around the CPU to ensure that it is properly ventilated.
•To obtain maximum image quality and visual comfort, position the
screen to avoid direct light reflections from internal or external
lighting.
•In order to guarantee medical-grade leakage current, the computer
connected to the system should be a medical-grade computer or
connected to the mains through medical-grade power supply. This is
not relevant if the system is connected to a laptop by USB through an
isolator or network cable.
General Notice
WA R N I NG S
If the product does not operate properly or fails to
respond to the controls as described in the product
accompanying documentation:
•Follow the safety precautions as specified in this guide.
•Stop using the unit and prevent any changes to it.
•Immediately contact the service office, report the
problem, and await further instructions.
Other manuals for CS 7600
1
Table of contents
Popular Medical Equipment manuals by other brands

Carroll
Carroll Spirit Series User instruction manual

Ulvac
Ulvac MDA-015A user manual

Veridian Healthcare
Veridian Healthcare ProKit Combo quick start guide

Medline
Medline Alterra MAXX Operation and maintenance manual

Sirona
Sirona galileos comfort plus operating instructions

inhealth
inhealth Blom-Singer Instructions for use

Mindray
Mindray BeneHeart D3 Operator's manual

Respironics
Respironics Trilogy 100 technical information

Balanced Body
Balanced Body Clinical Reformer Assembly instructions

NDC
NDC Pro Advantage P548440 Standard Use, care & maintenance

Dräger
Dräger Fabius Tiro Supplement

Tandem
Tandem AutoSoft 90 Instructions for use