Carefully clean all parts and remove any medicine residues and possible impurities after each
treatment. The compressor (1) and the air tube (11) must be wiped clean with a clean damp cloth.
Always wash hands thoroughly before cleaning and disinfecting the accessories.
Wash/sterilize every part of the nebulizer.
Do not expose the compressor to water or heat.
Replace the air tube when passing to a new patient or in case of impurities.
• Clean all of the device’s parts () under hot running water for about
5 minutes, adding, a small amount of neutral detergent if necessary, following the dosage
and limits specied by the detergent manufacturer.
• Rinse thoroughly, taking care to remove all residues, then leave to dry.
• All parts () can be disinfected with chemical disinfectants, using
doses and limits as specied by the manufacturer of the disinfectant.
• Disinfectants are generally available from pharmacies.
• All of the components () can be steam-sterilized up
to a temperature of 100°C (15-20 min.).
• After sterilizing, always leave the components to cool down to ambient temperature
before using again.
Do not repeat the sterilization cycle when the components are still hot.
• All of the components () can be immersed in boiling
water for no longer than 1 minute.