4
Do not use compressed air or a brush to clean the filter! HEPA filters are very easily damaged by the use of compressed air or by brushing. Misuse
of CleanSpace HALO by cleaning the filter may result in overexposure to contaminants and lead to sickness or death.
If the blocked filter alarm is triggered (2 beeps, repeated every second, red LED flashes), leave the contaminated area immediately and replace the
filter. Operating the respirator after the blocked filter alarm has sounded can cause the flow to fall below the manufacturer’s minimum designed flow,
which may result in overexposure to contaminants and lead to sickness or death.
When fitting a new filter the blocked filter alarm shall be tested before the respirator is put back into service. See the heading “Testing the Blocked
Filter Alarm” below.
Use only CleanSpace™filters with the CleanSpace HALO™respirator. Use of other filters may result in overexposure to contaminants and lead to
sickness or death.
CleanSpace HALO is not intrinsically safe. Do not use in flammable or explosive atmospheres. Doing so may result in injury or death.
Always correctly use and maintain the internal lithium-ion battery packs. Failure to do so may result in fire or explosion or could adversely affect
respirator performance and result in injury, sickness or death. Do not charge the on-board battery with unapproved chargers, in enclosed cabinets
without ventilation, near flammable liquids or gasses, or near sources of high heat. Do not immerse the device in water. Do not use, charge or store
the device outside the recommended temperature limits.
The Power Off State is considered to be an abnormal situation. Although respiratory protection will still be provided the level of protection will be
reduced in comparison to the Power On State.
If the battery alarm sounds (3 beeps, repeated every second), leave the contaminated area immediately and re-charge the battery. Operating the
respirator after the low battery alarm has sounded can cause the flow to fall below the manufacturer’s minimum designed flow, which may result in
overexposure to contaminants and lead to sickness or death.
In the extremely rare circumstance that the battery is damaged and electrolyte comes in contact with eyes, flush with water immediately and seek
urgent medical attention.
The Steri-Plus and HALO BIO Exhalation Filters (CS3039 and CS3027) are electrostatic filters and their performance will be degraded by exposure
to certain organic solvents including isopropanol (IPA), xylene and toluene. Always remove the Exhalation Filter before cleaning the mask. If there is
a possibility that the Exhalation Filter has become contaminated with any solvent, replace it.
Never dry the mask or exhalation valve with a cloth that may leave behind lint. Lint contamination of the exhalation valve may cause it to leak,
resulting in overexposure to contaminants and lead to sickness or death.
Using a full-face mask with the rain cover raised can cause the exhalation valve to vibrate, which you may notice as a buzzing noise. Lower the rain
cover for correct operation.
The power unit contains a battery, sensitive electronics and a motor. Always use the CS3011 Cleaning Plug when using anything wetter than a
damp cloth to clean it.
You must recalibrate the internal pressure sensor any time that your CleanSpace HALO is exposed to changes in temperature of more than 20°C. It
is best practice to also re-calibrate if the unit has been in storage, particularly if the storage temperature is not known.
To check that the test port plug is fitted correctly (after using the on-board fit test ports), a negative pressure Seal Check shall be performed –see
page 19.
European users should be aware that all markings on the HALO filters refer to the standard EN 12942 (1998) + A2 (2008), Class III, TM3 and should
not be confused with any other standard.
5. Special or Critical User Instructions
This respirator contains a system for synchronising with your breathing and regulating mask pressure. This system
requires re-calibration whenever it experiences a change in temperature of more than 20°C. It is best
practice to also re-calibrate if the unit has been in storage, particularly if the storage temperature is not
known.
For the most accurate battery charge indication, the respirator must be plugged into its charger. To be sure it is fully
charged, plug in the charger and wait for all 3 LEDs to light solidly without flashing. If the 3rd LED is flashing
rapidly, the battery is 95% charged.
The CS3038 Steri-Plus Exhalation Filter Case and the CS3039 Steri-Plus Exhalation Filter shall only be used in
combination with a HALO Half Mask (CS3003, CS3004 or CS3005) fitted with the PAF-1111 Standard Exhalation
Valve Cover. The Steri-Plus does NOT contribute to the system’s respiratory protection.