Description
Display–Maininterfacetoviewcurrentsensorreadingsandmenu
state.Willblink“STRT”uponapplyingpowerwhilethesensorwarms
upandstabilizes.
LEDs–areinforlowerrightcorneroftheunitonthemainPCBand
arelabeledS,EandP.
SLED–StatusLED.Blinkseverytimeacompletemeasurementcycleis
performedandthevalueisupdated.Serve’sasaheartbeatfor
transmitteroperation.
ELED–ErrorLED.Willlightifanyabnormalconditionsaredetected
withinthedevice,primarilythesensorfallingoutofitsrecommended
operatingspecificationswilltriggerthislamp.
PLED–PowerLED.Willbeoncontinuouslywhenpowerisappliedto
theunit.
Buttons‐TheunithasthreesurfacemountbuttonsabovetheLEDson
thePCB.Theycanbegentlyandcarefullypressedwithyourfinger.
Thesebuttonsallowyoutoaccessallaspectsofsensorsoperationand
testorcalibratedifferentfunctionalmodules.
Standard Operation
TheO2transmitterscomesfactorycalibratedforeasyinstallation.
Duringitslifecyclecalibrationmaybecomeperiodicallynecessary.We
recommendperformingaspanO2calibrationeverymonthfor
guaranteedoperationandensuredaccuracy.Noothercalibrationis
necessaryduringthelifecycleofthesensor.
iSense Oxygen Transmitter – Manual CO2Meter.com
Rev. 27 October 2014 Pg. 12 of 21
Entering the Service Menu
Toentertheservicemenuthebuttonsmustbepressedinafactory‐set
order.Thiscodecanbecustomizedbeforeshipmentbutbydefaultwill
be1‐2‐3.Thebuttonsmustbepressedwithinahalfsecondofeach
other,elsetheunitwillreturntostandardoperatingmode.Whenin
theservicemenutheunitwillnotupdatethecurrentO2readings.
Onceyouenterthemenuthedisplaywillread“2222”.
Performing an Atmospheric Span Calibration
1. Allowsensortostabilizeforatleast1hour
2. Ensurethesensorisexposedtofreshair
3. Entertheservicemenu
4. Press2toentertheOxygensubmenu
5. Press2againtoperformtheatmosphericcalibration
Performing a 0% Oxygen Zero Point
Calibration
1. Allowsensortostabilizeforatleast1hour
2. Screwonthecalibrationadapter
3. ApplyN2calibrationgas
4. Wait5minutesforstablereadings
5. Entertheservicemenu
6. Press2toentertheOxygensubmenu
7. Press1tocalibrate