10 | USER MANUAL | XXL-Rehab Alternating Mattress
Product functions
Pump unit
Power switch
The switch is on the right side of the unit. Turn ON/OFF the
power, the unit will start/stop operation.
Pressure-selector touch panel
There are seven pressure levels on the touch panel, provid-
ing options of pressure value.
Cycle time-selector touch panel
Selects a desired alternating cycle time from the touch
panel.
The cycle time value can be 10, 15, 20 or 25 minutes.
There are two LEDs for each time value.
Alarm reset
The alarm will turn on when the low pressure LED is on.
To mute the alarm, simply press the reset button. The alarm
LED will flash.
Re-press the reset button to reactivate the audible alarm.
Static/alternating control
Sets the air mattress in static mode or alternating mode.
Auto firm
Sets the air mattress to auto-inflation mode.
Mattress
The XXL-Rehab Alternating Mattress consists of Nylon/PU
cells with a cell height of 22 cm.
User costomisation
1. Place the mattress flatly on the bed frame. The inflation
tube should be towards the foot end.
2. Hang the pump unit on the foot end of the bed frame
and make sure it is stable.
3. Connect the inflation tubes from the mattress to the
pump’s inflating nozzles. Make sure they are properly at-
tached.
Make sure the air hoses are not kinked or tucked under
the mattress.
Also check to ensure the CPR valve is properly attached.
4. Make sure the CPR valve is correctly connected and
CLOSED.
5. Plug the power cord into an electrical outlet with
grounded AC power. Before inserting the plug into
the outlet, make sure the voltage is compatible. Also
make sure this product is well grounded.
THE USER SHOULD NOT LIE ON THE MATTRESS IN THE
INITIAL PHASE.
6. Turn on the power by pressing the power switch on the
right side of the pump unit. A beep sound will begin the
operation.
7. The pump will enter static mode. It will inflate the mat-
tress to 30 mmHg and then enter alternating mode.
If 30 mmHg cannot be reached within 30 minutes, the
unit will stay in static mode and keep inflating. If 30
mmHg cannot be reached within another 15 minutes, the
alarm will beep and its LED will come on. Contact the
dealer.
8. The alternating cycle time and pressure level are preset to
the middle settings. Select from the touch panel to adjust
the cycle time and pressure level.
9. When the mattress is ready (normal pressure LED is on),
place a sheet on the mattress.
During normal operation, the unit will monitor pressure. If
the pressure is lower than set pressure, the pump will auto-
matically inflate the mattress to the set level and then stop.
The alarm will beep and its LED will come on. Turn off the
pump, wait 2 minutes and then turn on again. If the alarm
starts again within 20 minutes, press the alarm reset button
to mute the alarm and contact the dealer.
When activating the Auto Firm button the mattress will au-
tomatically inflate to the maximum level (above 70 mmHg),
and then returns to the alternating level.
For suitable pressure make a hand check
Check to see if suitable pressure is selected by sliding one
hand between the air mattress and the foam base (or bed
frame if there is no foam base) to feel the user’s buttock.
We recommend a distance between user and mattress of
2 - 4 cm.
When the user does not use the mattress, it is pumped to
achieve the set back pressure. When the user then lies down
on the mattress the pump releases the excess air.