SPECIFICATIONS
Selectable Temperature Range:
–4° to 140°F (–20° to60°C)or 64° to 95°F (18° to 35°C)
Display: 7-digit LCD— Time ½ " 3-digit LCD— Temperature 9/16"
Temperature DisplayUpdate: 5 seconds
FRONT PANEL REFERENCE BUTTONS
HOUR- Advances hourdigit & advancestemperature in temperature setting mode
MIN- Advances minute digit
/D.R.- Hold down to settime-of-day or press to set programmed time settings to
active/inactive
TEMP.S/CLOCK- Pressto toggle the display between the TEMP SET temperature
and the current temperature
- (see "OutletOverride" section)
RESET- Clears all settings
G:OFF/O:ON- Outlet Indicator light
O:ON- Orange light indicates that power at the outlet module is on.
G:OFF- Green light indicates that power at the outlet module is off.
No light indicatesthat the outlet module is notreceiving powerfrom the wall outlet.
DESCRIPTION OF LCD PANEL
1. Time-of-Day controller settings
2. TOO HI- will appear when the temperature is 5° ormore above the TEMP SET
point
TOO LO- will appearwhen the temperature is 5° or below the TEMP SET point.
3. Low batterydisplay
4. Time-of-daydisplay hr/min/sec
5. Temperature display
6. Celsius/Fahrenheit indicator
7. Temperature setting mode indicator
BACK PANEL REFERENCE
1. Colder (up)/ Heater (down)
2. F° (up)/ C° (down)
3. Temperature Range
–20°C~60°C (up) 18°C-35°C (down)
(–4°F-140°F) (64°F-95°F)
DIP SWITCH SETTINGS
The dip switches are located in the lower rightcorner on the backof the display
module. See Fig.3 for illustrated schematic.
1. Switch #1 selects Colder or Heater setting.
a. Colder (up) selection turns the outlet ONwhen the temperature rises above
the TEMP SETpointwhile in the temperature controlling mode.(At
temperatures below the TEMP SET point the outlet is off.)
b. Heater (down) selection turnsoutlet ON when the temperature fallsbelow
the TEMP SETpointwhile in the temperature controlling mode.(At
temperatures above the TEMPSET the outlet is off.)
2. Switch #2 selects°F or°C for the temperaturedisplay.
3. Switch #3 selectsthe temperature range (wide or narrow see "Specifications").
TEMPERATURE CONTROLLING- HEATER MODE
Plug the electronic apparatus to be controlled into theOutlet Module. Turn on the
power switch of the electrical apparatus tobe controlled. Plug the OutletModule into a
poweroutlet.
1. Set dip switch #1 to HEATER (down).
2. Set dip switches #2 and#3 to thedesiredposition (see "DipSwitch Settings").
3. Pressthe RESET button. Wait 5 secondsfor the clock and current temperature
display to reappear.
4. Pressthe TEMP.S/CLOCK button, "TEMP SET" willappear in the lower right
corner of the LCD display.
5. Pressthe HOUR button to advance the temperature setting to the desired
temperature.
6. Pressthe TEMP.S/CLOCK button to return to the currenttemperaturedisplay.
7. In order to initiate the switching function, reduce the temperature at the probe to
at least 2°below the TEMP SET point.
8. Pressthe button to trigger the outletmodule on.(All subsequentON/OFF
operations will be automatic.)
9. Placeprobe in contact with the materialto be measured.
The outlet will now automatically turn ON when the temperature falls 2°below theSET
TEMP point and OFFwhen the temperature rises above theTEMPSET point.
TEMPERATURE CONTROLLER-COLDER MODE
Plug the electronic apparatus to be controlled into theOutlet Module. Turn on the
power switch of the electrical apparatus tobe controlled. Plug the OutletModule into a
poweroutlet.
1. Set dip switch #1 to COLDER (up).
2. Set dip switches #2 and#3 to thedesiredposition (see "Dip Switch Settings").
3. Pressthe RESET button. Wait 5 secondsfor the clock and current temperature
display to reappear.
4. Pressthe TEMP.S/CLOCK button, "TEMP SET" willappear in the lower right
corner of the LCD display.
5. Pressthe HOUR▲button to advance thetemperature setting to the desired
temperature.
6. Pressthe TEMP. S/CLOCK button to return to the current temperature display.
7. In order to initiate the switching function, raise the temperature atthe probe to at
least2° above the TEMP SET point.
8. Pressthe button to trigger the outletmodule on.(All subsequentON/OFF
operations will be automatic.)
9. Placeprobe in contact with the materialto be measured.
The outlet will now automatically turn ON when the temperature rises2° above the
SET TEMP point and OFF when the temperature fallsbelow the TEMP SETpoint.
TIME-OF-DAY CONTROLLING
1. Set dip switch #1 to "colder" (up) and switch #3 to -4°F~140°F (up).
2. Pressthe RESET button. Wait 5 secondsfor clockand current temperature
display to reappear.
3. Push the TEMP.S/CLOCK button, TEMPSET appearson the display.
4. Pressthe HOUR▲button until the display reads0.0°F (-18°C). (TEMP SET
display defaults to 82° F. Each press of the HOUR ▲button advances the
display. At 140° F it rolls over to –4° F.)