dacell DN520N User manual

Digital Weighing Indicator
Instruction Manual
DN520N
DACELL CO.,LTD
TEL : +82-43-260-2242
FAX : +82-43-260-2242
WEBSITE: http://www.dacell.com
EMAIL : info@dacell.com

PAGE 1/ 53
CONTENTS
1. Before Installation
----------------------------------
2 Page
2. Introduction
----------------------------------
3 Page
3. Specification
----------------------------------
4 Page
4. Installation
----------------------------------
11 Page
5. Set-Up
----------------------------------
12 Page
6. Interface
----------------------------------
34 Page
7. Error and Treatment
----------------------------------
50 Page
8. Warrantee
----------------------------------
54 Page

PAGE 2/ 53
1. BEFORE INSTALLATION
1-1. Caution / Warning Marks
This mark warns the possibility to arrive death or serious injury in
case of wrongly used.
This mark cautions the possibility to arrive serious human body
injury or product lose in case of wrongly used.
1-2. Other Marks
Warning for Electric Shock or Damage.
Please do not touch by hand
Protective Ground(Earth) terminal
Prohibition of Operation process
1-3. Copy Rights
1). All Right and Authority for this Manual is belonged to DACELL co.,Ltd.
2). Any kinds of copy or distribution without DACELL co.,Ltd..’s permission will be prohibited.
1-4. Inquiries
If you have any kinds of inquiries for this model, please contact with your local agent or Head
Office.
Head Office : DACELL co.,Ltd.
Website : http://www.dacell.com
Email : info@dacell.com
Warning
Caution

PAGE 3/ 53
2. INTRODUCTION
2-1. Introduction
Thank you for your choice, this “DN520N” Industrial Digital Weighing Indictor..
This “DN520N” model is control purpose application usage Digital Weighing Indicator, with
powerful communication performance.
With 6pcs control relay outputs and High Speed A/D conversion performance will lead you to
precise weighing process.
This “DN520N” Weighing Indicator is control purpose application model, and it can be used for
most kinds of control applications.
Please review this instruction Manual and learn more about information about “DN520N”.
Enjoy your process efficiency with “DN520N” Weighing Indicator..
2-2. Cautions
1). Don’t drop on the ground or avoid serious external damage on item.
2). Don’t install under sunshine or heavy vibrated condition.
3). Don’t install place where high voltage or heavy electric noise condition.
4). When you connect with other devices, please turn off the power of item.
5). Avoid from water damage.
6). For the improvement of function or performance, we can change item specification without
prior notice or permission.
7). Item’s performance will be up-dated continuously base on previous version’s performance.
2-3. Features
1). All Modules and Option Cards are isolated to maximize accuracy and performance.
2). External input terminal inside.
3). By using “Photo-Coupler” on each module(Option, Analog board, In/Out), we improved
“Impedance problem”, “Isolation ability among inputs”, “Leading power problem”, and
“Noise covering function”.
4). Data back-up function, when the sudden power off
5). Polycarbonate film panel, strong against dust and water
6). RS-232C (Com. Port1) is standard installed.
8). Variable options(Order in advance)
2-4. Box Contents
1). Power Cable(1pcs) /Load cell Connector(1pcs) / Manual(1pcs)
Cautions
!

PAGE 4/ 53
3. SPECIFICATION
3-1. Analog Input & A/D Conversion
Input Sensitivity
0.2㎶/ Digit
Load Cell Excitation
DC 10V ( - 5V ~ + 5V )
Max. Signal Input Voltage
Max.32mV
Temperature Coefficient
[Zero] ±20PPM/℃
[Span] ±20PPM/℃
Input Noise
±0.6㎶P.P
Input Impedance
Over 10㏁
A/D Conversion Method
Sigma-Delta
A/D Resolution(Internal)
520,000 Count(19bit)
A/D Sampling Rate
Max. 200times / Sec
Non-Linearity
0.01% FS
Display Resolution(External)
1/30,000
3-2. Digital Part
Display
Parts
Specification
Display
Main Display
7Segments, 6digits Red color FND
Size :20.0(H) ×13.0(W)mm
Sub Display
7Segments, 14digits Red color FND
Size :9.2(H) ×4.8(W)mm
Min. Division
×1, ×2, ×5, ×10, ×20, ×50
Max. display value
+999,950
Under Zero value
"-" (Minus display)
Status lamp
Steady, Zero, Tare,
Hold, Low, Hi, Finish,
RTxD
Green color Condition display Lamp (8pcs)
K e y
Number, Function Key
Number Key, Function (12pcs)

PAGE 5/ 53
3-3. General Specification
Power Supply
SMPS Free Voltage Power Supply(AC86~265V)
Operating Temperature Range
-5℃~ 40℃
Operating Humidity Range
Under 85% Rh (non-condensing)
External Dimension
193mm(W)x100mm(H)x140mm(L)
Net Weight(kg)
About 1.5kg
Gross Weight(kg)
About 2.0kg
3-4. Option Card
Option No.1
Analogue Output (0~10V)
Option No.2
Analogue Output (4~20mA)
Option No.3
Serial Interface : RS422 / RS485
Option No.4
BCD Input
Option No.5
BCD Output
※Serial Interface (RS-232C) or Current Loop is Standard installed.
3-5. Front Panel (Display & Key pad)

PAGE 6/ 53
3-5-1. Status Lamp (ANNUNCIATORS) : Green Color Lamp is “ON”.
Steady
When the weight is Steady, Lamp is turn on.
Zero
When the current weight is Zero, Lamp is turn on.
(Displayed weight is Zero, Lamp is turn on.)
Tare
Tare function is set, Lamp is turn on.
(Tare Reset Lamp is turn off.)
Hold
Hold Function is Activated, Lamp is ON.
Low
Low Relay output Lamp
Hi
Hi Relay output Lamp
Finish
Weighing process is Finished, Lamp ON
RTxd
When indicator transfers or receives data from other devices, Lamp is turn
on. (If the Lamp is off although there is some data transference, please
check communication settings).
3-5-2. Key Pad Function
Make Weight value as Zero.
Under F08, you can set the Zero key operation range, as 2%, 5%, 10%,
20% or 100% of Max. Capacity.
※Under “Tare” key input, Zero key will not be activate within operation
range.
Make Weight value as Zero, including Tare Weight.
Under F09, you can set the Tare key operation range, as 10%, 20%, 50%,
or 100% of Max. Capacity.
※Whenever pressing “Tare” key, you can set the Tare continuously.
Start Weighing process, under Packer Mode, only.
You can set each weighing process as a certain P/N.
And you can call certain P/N with pressing this key.
P/N save : Select P/N and Enter key input.
P/N call : P/N + Number key + Enter
Activate Hold Function, Hold display until Hold Reset input.

PAGE 7/ 53
Stop weighing process, under Packer Mode, only.
Set the Low value or Check the current value.
- Press key and enter new set value with keypad, and press enter to save
Set the Hi value or Check the current value.
- Press key and enter new set value with keypad, and press enter to save.
Set “Free Fall” value for “HI” set point
If you set the “Free Fall”, “HI” relay output will be “ON” in advance,
according to “Free Fall” set value.
Manual Printer
- Key input, print output.
Calibration mode
- Digit setting
Whenever pressing “0”key, digit will be change 1, 2, 5, 10, and 50.
1. Modify the set value during setting process.
2. Calibration mode
- Move back to previous step.
3. F-function setting mode
- Change F-function No.
F-function no.(number key) + Clear directly move
1. Save set value during setting process.
2. Calibration mode
- Save current setting and move to next step.
3. F-Function mode
- Save current F-function setting, and move to next F-function

PAGE 8/ 53
※Function Keys (Combined Key functions :key + other keys)
Time set value check or Change
Date set value check or Change
Code value check or Change
Grand-total Data Delete
Sub-total Data Delete
Serial No. Check or Change
Low Set value Display or Change
Hi Set value Display or Change
Grand-total Data Delete
Sub-total Data Delete

PAGE 9/ 53
3-6. Rear Panel
①POWER
-Power ON/OFF Switch
-Fuse : AC 250V 10A
-AC IN : AC86~265V Power In
②OPTION 1,2
- OPTION BOARD install slot.
- ANALOG out, Serial I/F, etc
③LOAD CELL
CONNECTOR
(N-16)
-EXC + (+5V) PIN1 (RED)
-EXC - (-5V) PIN2 (WHITE)
-SIG+ PIN3 (Black or BLUE)
-SIG- PIN4 (GREEN)
-SHIELD PIN5 (SHEILD)
④Digital Input
- Digital Input Signal terminal
Refer to “F-function 11”.
⑤Output
Terminal
-RS-232C/CURRENTLOOP (Standard Installed)
( GND,TXD1,CL1,CL2,RXD,GND,TXD )
⑥ISP
(Digital Lock Pin)
- Insert “Lock Pin Header”, to protect “F-function”data and other
settings from Electric Noise effect.
- To change the setting, please remove the “Lock Pin Header”.
⑦Relay Output
6pcs Relay output terminal
- According to “F21-XX”setting, relay will be output.
Power
Digital Input
Output
ISP
Option
Slot
Load Cell
Connector
Comm.

PAGE 10 / 53
4. INSTALLATION
4-1. External Dimension & Cutting Size
(External Dimension) (unit : mm)
Cutting Size

PAGE 11 / 53
Chapter 5. Set Up
5-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of
Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
5-2. Test Weight Calibration (span Calibration) –Mode 1.
- Applicable model : DN500N,510N,520N,530N,540N series
Prepare at least 10% of Max. capacity of your weighing scale.
Step 1. Enter Calibration Mode
Step 1. Enter Calibration Mode
Turn on the Power + with pressing display
Press key display
Press key to start “Calibration Mode”.
Remarks : Go to next step with save key / Back to previous step key
Step 2. Digit(Division)setting ( display)
Press key and stop the optimal Digit value.
(Whenever pressing key, digit value will be changed, like 01-02-05-10-20-50)
Press key and save change and move to next step.
Step 3. Max. Capacity Setting ( display)
Input Max. Capacity of Scale with No. keys.
Input Capacity and press key, and move to next step.
※Caution
(Max. capacity value / division value) can not be over 30,000.(as Indicator resolution is 1/30,000).
XXXXXX
CAPA
d _1
SET-CAL
TEST

PAGE 12 / 53
Step 4. Zero Balance setting ( display)
Make empty the scale part, and press key.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Span value calculation ( )
Input prepared Test weight value with No. keys.
And press key.
Then, display will show and then, load prepared test weight unit on the scale.
After a few seconds(to remove the vibration effect), press key.
Then, indicator will calculate Span value and move the next step.
※Caution
For the precise Span calibration, please prepare Test weight unit, at least 10% of Max. capacity
of Scale.
Step 6. Check Span Value and Finish the Calibration( )
Check the Calculated Span value.
And after 3sec, will displayed automatically and move to weighing Mode.
C-END
C-END
XXXXXX
C_UP_
C_UP_
XXXXX
SPAN
CAL _0_

PAGE 13 / 53
5-3. Simulation Calibration Mode (Without Test Weight) –Mode 2.
- This calibration Method will be useful to make calibration more than 10ton capacity setting.
- Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
Step 1. Enter to the “SET-CAL”mode
Please turn on the Power + with pressing display
Press key display.
Press key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save key / Back to previous step key Step 2.
Step 2. Digit (Division)setting ( display)
Press key and stop the optimal Digit value.
(Whenever pressing key, digit value will be changed, like 01-02-05-10-20-50)
Press key and save change and move to next step.
Step 3. Max. Capacity of Load cell ( display)
Input Max. Capacity of Scale with No. keys.
- Under this step, input Total sum of each load cell’s Max. Capacity. (Not weighing Scale)
- The Max. Capacity of load cell is stated on “Test report” or “Label”.
- If you installed 4 load cells, and each load cell’s Max. Capacity is 500kg, then you have to input
2,000kg, as a Max. Capacity.
Input Capacity and press key, and move to next step.
XXXXXX
CAPA
d _ 1
CAL_ _2
_1
SET-CAL
TEST

PAGE 14 / 53
Step 4. Measure/Adjustment optimal Zero balance of Scale ( display)
Make empty the scale part, and press key.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Input the Rate Output (mV/V) value of load cell
Input Max. Output Rate(mV/V) value of load cell with No. keys.
- Under this step, input Max. Output rate(mV) of load cell.
- If you installed a few pieces of load cells, the connection will be parallel, so the rated output of
a few load cells are as same as single load cell’s rated output.
-The Output rate is stated on “Calibration certificate” or “Label”
And press key.
Step 6. End Calibration and Auto Reset
- Calculated Span value will be displayed and automatically reset and move the normal
weight indicating mode.
CAL _0_

PAGE 15 / 53
5-4. Function Setting –Mode 1.
- Applicable model : MI –1000/2000/3000/4000 series
To make more accuracy performance of Digital Weighing Indicator, through this Function setting.
Step 1. Enter to Function setting mode.
Turn on the Power + with pressing display
Press key display
Press key to start “F-function Mode”.
Step 2. Change the F-Function No.
To change the F-function No., press key. whenever pressing key, Function No. will
be changed
If you want to move certain function No. directly, press function No. with keypad and press
key.
Step 3. Change the Set value.
Input new set value with keypad, and press key to save new setting.
If you don’t press key, after changing the set value, the new set value will not be saved.
Step 4. Exit from Function setting mode.
Press + + key to exit function mode.
SET-CAL
TEST

PAGE 16 / 53
5-5. Function List
Function
No.
Contents
Remark
F01
Decimal point setting
Setting range : 0~3
F02
Back up mode selection
Setting range : 0, 1
F03
Motion Band setting
Setting range : 0~9
F04
Zero Tracking setting
Setting range : 0~9
F05
Auto Zero Range setting
Setting range : 00~99
F06
Digital Filter setting
Setting range : 00~49
F07
Zero / Tare key activating setting
Setting range : 0, 1
F08
Zero key operating range setting
Setting range : 0~4
F09
Tare key operating range setting
Setting range : 0~3
F10
Hold Function setting
Setting range : 0~4
F11
Digital Input setting
Setting range : 0~8
F12
Code No. Setting
Setting range : 0~2
F14
Hold Off time setting
Setting range : 0.0~9.9sec
F21
Weighing Mode Selection
Setting range : 1~7
F22
Weighing Finish Relay “ON”delay time setting
Setting range : 0.0~9.9sec
F23
Weighing Finish Relay “ON”Duration time setting
Setting range : 0.0~9.9sec
F24
Weighing Judge Relay “ON”delay time setting
Setting range : 0.0~9.9sec
F25
Weighing Judge Relay “ON”Duration time setting
Setting range : 0.0~9.9sec
F28
Weighing NG Relay “ON”Duration time setting
Setting range : 0.0~9.9sec
F30
Serial I/F Parity Bit setting (Port No.1)
Setting range : 0~2
F31
Serial I/F Communication Speed setting Port No.1)
Setting range : 0~9
F32
Serial I/F Mode setting Port No.1)
Setting range : 0~2
F33
Serial I/F Transference Method setting Port No.1)
Setting range : 0~5
F34
ID Number setting
Setting range : 1~99
F35
Transferred Data Format Port No.1)
Setting range : 0~2
F36
BCC selection mode
Setting range : 0, 1
F37
Data Transferring count setting Port No.1)
Setting range : 0~6
F40
Serial I/F Parity Bit setting (Port No.2)
Setting range : 0~2
F41
Serial I/F Communication Speed setting (Port No.2)
Setting range : 0~9
F42
Serial I/F Mode setting (Port No.2)
Setting range : 0~2
F43
Serial I/F Transference Method setting (Port No.2)
Setting range : 0~5
F45
Transferred Data Format (Port No.2)
Setting range : 0~2
F47
Data Transference count setting (Port No.2)
Setting range : 0~6

PAGE 17 / 53
F50
Weight Unit Selection (Printer)
Setting range : 0~2
Function
No.
Contents
Remark
F51
When Automatically print, Data output selection
Setting range : 0, 1
F52
Print format selection
Setting range : 0, 1
F53
Sub-Total Data delete Selection
Setting range : 0, 1
F54
Paper withdraw rate Selection
Setting range : 0~9
F55
Print Line interval Selection
Setting range : 0~9
F56
Sub-Total Print Mode Selection
Setting range : 0, 1
F57
Print Language Selection
Setting range : 0~3
F58
Print Delay time selection
Setting range : 0.0~9.9sec
F59
Auto Print Setting
Setting range : 0, 1
F60
BCD output Selection
Setting range : 0, 1
F63
Average Display setting
Setting range : 00~99
F64
Steady LED Status Lamp Delay time setting
Setting range : 0.0~9.9sec
F65
Tension and Compression setting
Setting range : 0, 1
F80
Empty Range
Setting range : 0~Max. Capa
F81
Zero Range Setting
Setting range : 0~Max. Capa
F83
Analogue output setting
Under option installed
F89
Span Value check
F90
Date check / change
F91
Time check / change

PAGE 18 / 53
5-6. Function List detailed information.
Decimal Point Setting
F01
0
No Decimal point
●
1
1st place under Zero (0.0)
2
2nd place under Zero (0.00)
3
3rd place under Zero (0.000)
Back up mode selection
F02
●
0
Normal mode
1
Back up mode
Motion Band Range setting
F03
5
0
∫
9
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and
reduce the vibration effect on weighing process.
0 : Weak vibration
∫
9 : Strong Vibration
Zero Tracking Compensation Range setting
F04
5
0
∫
9
Due to external causes(Temperature, wind, and dust), there are small
weight difference, indicator will ignore the weight difference and
display Zero.
For this compensation function, indicator will estimate the weight
difference is over the set range during fixed time period.
If there is large weight difference over set range within fixed time
period, the “Zero” is breaking and will find new zero point.
Auto Zero Range setting
F05
00
00
∫
99
Within the “Auto Zero” range, weighing part is steady, indicator will
display current weight as “Zero”
If the weighing part is not “Steady”, indicator will display current
weight.
(Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
15
00
∫
49
Small set value for weak vibration
Large set value for strong vibration
Small set value more
sensitive
Zero /Tare key Operation mode selection
F07
●
0
Activate when “Steady” condition, only
1
Always activated

PAGE 19 / 53
Zero key Operation Range selection
F08
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
2
Activated within 10% of Max. Capacity
●
3
Activated within 20% of Max. Capacity
4
Activated within 100% of Max. Capacity
Tare key Operation Range selection
F09
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
2
Activated within 50% of Max. Capacity
●
3
Activated within 100% of Max. Capacity
“Hold” Mode selection
F10
●
0
Peak Hold : Measure Max. weight value and hold on display.
1
Sample Hold : Hold current weight until “Hold Reset”.
2
Average Hold : Make average during 3sec, and hold display
3
Average Hold : Make average during 5sec, and hold display
4
Average Hold : Make average during 8sec, and hold display
External Input Selection
F11
Set Value
Input 1
Input 2
Input 3
Input 4
●
0
RUN
STOP
TARE
TARE RESET
1
RUN/STOP
TARE/RESET
ZERO
PRINT
2
ZERO
TARE/RESET
JUDGE
PRINT
3
ZERO
TARE/RESET
HOLD
HOLD RESET
4
ZERO
TARE
TARE RESET
PRINT
5
ZERO
SUB-TOTAL
GRAND-TOTAL
PRINT
6
ZERO
TARE/RESET
SUB TOTAL
PRINT
7
ZERO
PRINT
SUB TOTAL
SUB TOTAL
DELETE
8
ZERO
PRINT
GRAND-TOTAL
GRAND TOTAL
DELETE