Measurement rise time: < 450 ms typical Stimulus current range (supramax and manual):
10 to 70 mA with 5 mA steps
Airway pressure (D-lite)
Accuracy: +/-(2 vol% + 2% of reading) Measurement range: –20 to +100 cmH2O
Gas cross effects: < 2 vol% anesthetic agents Stimulus current accuracy: 10% or ±3mA
(whichever is greater)
Accuracy*: ±1 cmH2O
Anesthetic agent (AA) Airway pressure (Pedi-lite) Max load: 3 kΩ
Measurement rise time: < 400 ms typical Measurement range ***: –20 to +100 cmH2O Max voltage: 300 V
Halothane, Isoflurane, Enflurane
Gas cross effects: < 0.15 vol% N2O
Measurement range: 0 to 6%
Accuracy: +/- 0.2 vol%
Flow (D-lite) Regional block mode
Measurement range: 1.5 to 100 l/minute Stimulation mode: single twitch (ST)
Flow (Pedi-lite) Intervals: 1 second, 2 seconds, 3 seconds
Measurement range: 0.25 to 25 l/minute
Sevoflurane
Measurement range: 0 to 8%
Accuracy: +/- 0.2 vol%
Stimulus pulse: square wave, constant current
Pulse width: 40 µs
Compliance (D-lite and Pedi-lite)
Stimulus current range: 0 to 5.0 mA with
0.1 mA steps
Measurement range: 4 to 100 ml/cmH2O
Desflurane
Measurement range: 0 to 20%
Accuracy*:
0 to 5 vol%: +/- 0.2 vol%
Airway resistance (D-lite and Pedi-lite) Stimulus current accuracy: 20% or 0.3 mA
(whichever is greater)
Measurement range: 0 to 40 cmH2O/l/second
Sensor specifications (D-lite) Recorder
5 to 10 vol%: +/- 0.5 vol% Dead space: 9.5 ml
10 to 20 vol%: +/- 1.0 vol% Resistance at 30 l/minute: 0.5 cmH2O Principle: thermal array
Agent identification Print resolution:
Vertical: 8 dots/mm (200 dots/inch)
Horizontal: 32 dots/mm (800 dots/inch) at
speed of 25 mm/second and slower
Sensor specifications (Pedi-lite)
Identification threshold*: 0.15 vol% Dead space: 2.5 ml
Patient Spirometry Resistance at 10 l/minute: 1.0 cmH2O
Detection through D-lite™ or Pedi-lite™ flow
sensor and gas sampler with following
specifications: NMT Paper width: 50 mm; printing width 48 mm
Traces: selectable; 1, 2, or 3 traces
Stimulation modes:
Train of four (TOF)
Double burst (3.3) (DBS)
Single twitch (ST)
50 Hz tetanic + post-tetanic count (PTC)
Tidal volume (D-lite) Print speed: 1, 6.25, 12.5, 25 mm/second
Measurement range: 150 to 2000 ml
Accuracy*: ± 6% or 30 ml * Typical value
Tidal volume (Pedi-lite) ** Alarm limits and their adjustment range
may vary depending on the mode used.
Measurement range: 15 to 300 ml Measurement intervals:
Accuracy*: ± 6% or 4 ml TOF and DBS: manual; 10 seconds,
12 seconds, 15 seconds, 20 seconds,
1 minute, 5 minutes, 15 minutes
*** The user may set the SpO2low alarm
limit lower than 80% and Ppeak alarm
limit higher than 50 cmH2O during
normal use of the monitor. To guarantee
patient safety, the user cannot save such
limits as user default values.
Minute volume (D-lite)
Measurement range: 2 to 20 l/minute
Accuracy*: ± 6% ST: manual; 1 second, 10 seconds,
20 seconds
Minute volume (Pedi-lite)
Measurement range: 0.5 to 5 l/minute Stimulator
Accuracy*: ± 6%
Stimulus pulse: square wave, constant current
Pulse width: 100, 200 or 300 µs
– 4 –