Certification
PosiTector SPG probes include a Certificate of Calibration showing
traceability to a National Metrology Institution. For organizations with
re-certification requirements, instruments may be returned at regular
intervals for calibration. DeFelsko recommends that customers
establish calibration intervals based upon their own experience and
work environment. Based on our product knowledge, data and
customer feedback; a one year calibration interval from either the date
of calibration, date of purchase, or date of receipt is a typical starting
point.
Calibration, Verification & Adjustment
Calibration, Verification & Adjustment
The PosiTector SPG probes measure the peak-to-valley height of the
surface profile of abrasive blast cleaned surfaces.
Three steps ensure best accuracy...
1. Calibration - typically done by the manufacturer. All probes include a
Certificate of Calibration
2. Verification - typically done by the user on known reference
standards such as included metal shim
3. Adjustment - zero calibration adjustment on smooth flat surface
(included glass plate)
Calibration
Calibration is the high-level, controlled and documented process of
measuring traceable calibration standards over the full operating range
of the probe, and verifying that the results are within the stated
accuracy of the probe. Calibrations are performed by the manufacturer,
their authorized agent, or by an accredited calibration laboratory in a
controlled environment using a documented process.
PosiTector SPG probes are shipped with a Certificate of Calibration
showing traceability to a National Metrology Institution. For
organizations with re-certification requirements, probes may be
returned at regular intervals for calibration. DeFelsko recommends that
customers establish calibration intervals based upon their own
experience and work environment. Based on DeFelsko’s product
knowledge, data and customer feedback, a one year calibration interval
from either the last date of calibration, date of purchase, or date of
receipt is a typical starting point.
Written Calibration Procedures are available online at no charge.
http://www.defelsko.com/quality/calibration_procedures.htm