HD2109 - 11 - V2.4
Hold function is enabled, the instrument performs the measurement and when it stabilizes it
goes in HOLD mode. To update the display indication, press FUNC/ENTER.
In the following figure you can see an example of the measurement process with the Auto-
Hold function enabled. A probe is immersed into a liquid at concentration C1and, to per-
form the measurement, the FUNC/ENTER key is pressed: The concentration measurement
raises progressively reaching the final value. The HOLD symbol blinks. In the stretch indi-
cated by 1, the measurement remains stable for 10 seconds, within two digits: at the end of
this interval (point 2), the instrument goes into HOLD mode, presenting the final stable
value.
2 digit
10s
t
Ο
2
C
1
3) SAL mg/l (Setting the salinity): the message "SAL mg/l – UNT=RSET" is scrolled in the
comment line. The display central line shows the current salinity value, and the secondary
line the corresponding dissolved oxygen concentration value. Use the arrows to enter this
value, and confirm using ENTER. To disable the salinity compensation, set the value to
zero: pressing the °C/°F key rapidly sets the value to zero without the need to use of the ar-
rows.
Note: the dissolved oxygen concentration depends on the measured liquid salinity. In con-
trast, salinity has no effect on the saturation index.
4) Management of memorized data (only HD2109.2): the message "LOG_DUMP_or_E-
RAS" (Transfer data or erase) is scrolled in the comment line. The center figure reports
the number of free memory pages (FREE). All memory data are permanently erased by
pressing SERIAL/EraseLOG. By pressing LOG/DumpLOG, the data transfer of the logged
data on the serial port is started: the "BAUD-RATE" must have previously been set to the
maximum value (please see the menu items described below and the paragraph "STORING
AND TRANSFERRING DATA TO A PERSONAL COMPUTER" on page 29).
5) Sleep_Mode_LOG (Automatic turning off during recording) (only HD2109.2): this
function controls the instrument's automatic turning off during logging, occurring between
the capture of a sample and the next one. When the interval is lower than 60 seconds, the in-
strument will always remain on. With intervals greater than or equal to 60 seconds, it is pos-
sible to turn off the instrument between loggings: it will turn on at the moment of sampling
and will turn off immediately afterwards, thus increasing the battery life. Using the arrows
select YES and confirm using ENTER in order to enable the automatic turning off, select
NO and confirm to disable it and keep the instrument on continuously.
Note: even if Sleep_Mode_LOG=YES is selected, the instrument does not turn off for less
than one minute intervals.
6) Print and log interval: sets the interval in seconds between two loggings or data transfers
to the serial port. The interval can be set at 0, 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min),
300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour).
If the value 0 is set, SERIAL works on command: the sending of data to the serial port
is performed each time the key is pressed. Recording (LOG) is performed with one sec-