
REAL CALIBRATION (WITH SAMPLE WEIGHTS)
After having performed the THEORETICAL CALIBRATION and TARE WEIGHT
ZERO SETTING, this function allows correct calibration to be done using sample
weights of known value and if necessary, any deviations of the indicated value from the
correct value to be corrected.
Load a sample weight onto the weighing system that is at least 50% of the maximum
quantity to be weighed.
By confirming the message the flashing value of the weight currently on the
system is displayed. In this phase all of the LEDs are off. Adjust the value on display by
using the arrow keys if necessary. After confirming, the new set weight will appear with
all the LEDs flashing. After an additional confirmation, the message will be
restored and by repeatedly pressing the key the weight will once again be displayed.
Example: for a system of maximum capacity 1,000 kg and 1 kg division, two sample
weights are available, one of 500 kg and the other one of 300 kg. Load both weights onto
the system and correct the indicated weight to 800. Now remove the 300 kg weight, the
system must show 500, then remove the 500 kg weight too and the system must read
zero. If this does not happen, it means that there is a mechanical problem affecting the
system linearity.
WARNING: identify and correct any mechanical problems before repeating the
procedure.
-If theoretical full scale and recalculated full scale in real calibration are equal,
it means that the theoretical calibration is currently in use. Otherwise, the real
calibration based on sample weights is in use.
-If the correction made changes the previous full scale for more than 20%,
all the parameters with settable weight values are reset to default values.
+
LINEARISATION OPTION ON MAX 5 POINTS:
It is possible to perform a linearisation of the weight repeating the above-described
procedure up to a maximum of five points, using five different sample weights. The
procedure ends by pressing the button or after entering the fifth value; at this point it
will no longer be possible to change the calibration value, but only to perform a new real
calibration. To perform a new calibration, return to the weight display and then re-enter
the calibration menu.
By pressing after confirming the sample weight that has been set, the full scale appears
recalculated according to the value of the maximum sample weight entered, with reference
to the cell sensitivity set in the theoretical calibration ( ).
Interface Inc, www.interfaceforce.com
INF1 Installation & User Manual 15-352