9000-615-32/02 1808V002 5
Important information
Observe the EMC rules concerning medical
devices
iThe appliance is designed for the use in health
care establishments (in accordance with IEC
60601-1-2). If the appliance is operated in an-
other environment, potential effects on elec-
tromagnetic compatibility must be taken into
account.
iDo not operate the unit in the vicinity of HF
surgical instruments or MRT equipment.
iMaintain a minimum distance of at least 30
cm between the unit and other electronic de-
vices.
iKeep a minimum distance of 30cm between
the unit and mobile radio devices.
iNote that cable lengths and cable extensions
have effects on electromagnetic compatibility.
The following accessories can have an effect on
the electromagnetic compatibility:
Mains cable . . . . . . . . . . . . . . . . . 9000100846
Flexible foot switch cable . . . . . 9000-119-130E
NOTICE
Negative effects on the EMC due to
non-authorised accessories
iUse only Dürr Dental parts or acces-
sories specifically approved by Dürr
Dental.
iIf other accessories are used, note
any negative consequences to the
function of the unit.
2.7 Only use original parts
iOnly use Dürr Dental parts or accessories and
special accessories specifically approved by
Dürr Dental.
iOnly use only original wear parts and replace-
ment parts.
Dürr Dental accepts no liability for dam-
ages or injury resulting from the use of
non-approved accessories or optional
accessories, or from the use of non-origi-
nal wear parts or replacement parts.
The use of non-approved accessories,
optional accessories or non-genuine
wear parts / replacement parts
(e.g.mains cable) can have a negative
effect in terms of electrical safety and
EMC.
2.4 General safety information
WARNING
Contraindication
Ultrasound vibrations can interfere with
the function of heart pacemakers and
defibrillators.
iPatients with heart pacemakers or de-
fibrillators should not be treated with
these instruments.
iAlways comply with the specifications of all
guidelines, laws, and other rules and regula-
tions applicable at the site of operation for the
operation of this unit.
iCheck the function and condition of the unit
prior to every use.
iDo not convert or modify the unit.
iComply with the specifications of the Installa-
tion and Operating Instructions.
iThe Installation and Operating Instructions
must be accessible to all operators of the unit
at all times.
2.5 Qualified personnel
Operation
Unit operating personnel must ensure safe and
correct handling based on their training and
knowledge.
iInstruct or have every user instructed in han-
dling the unit.
Installation and repairs
iInstallation, readjustments, alterations, up-
grades and repairs must be carried out by
Dürr Dental or by qualified personnel specifi-
cally approved and authorized by Dürr Dental.
2.6 Electrical safety
iComply with all the relevant electrical safety
regulations when working on the unit.
iNever touch the patient and unshielded plug
connections on the device at the same time.
iReplace any damaged cables or plugs imme-
diately.
EN