1 Exposure to CO2 concentraons smaller than 400 ppm can aect the
accuracy of the sensor if the automac self-calibraon funcon is used.
2 Deviaon to a high-precision reference. Accuracy is fullled by >90% of the
sensors aer calibraon. Rough handling or shipping reduces the accuracy
of the sensor. Accuracy is restored with the automac recalibraon feature.
Accuracy is based on tests with gas mixtures having a tolerance of ±1.5%.
3 Time for achieving 63% of a respecve step funcon under test condions.
Response mes can depend on the operang environment in the
nal applicaon.
4 For proper automac calibraon the device has to be exposed to air with
CO2 concentraon 400 ppm regularly. Maximum accuracy dri per
year esmated from stress tests is ± (5 ppm + 2 % of reading). Higher
dri values may occur if the device is not handled according to
the instrucons.
www.lascarelectronics.com
Issue 1 06/2022
EL-IOT-CO2
Carbon Dioxide WiFi Cloud-Connected Data Logger
Your data, anytime, anywhere
OPERATING CONDITIONS
The performance of the humidity sensor can
be aected by long-term exposure to operang
condions at the extents of the logger’s range.
The sensor shows best performance when
operated within the recommended normal
temperature and humidity range of 5 to 55°C and
20 to 80%RH, respecvely. Long-term exposure to
condions outside the normal range, especially
at high humidity, may temporarily oset the RH
signal (e.g. +3%RH aer 60h kept at >80%RH).
Aer returning into the normal temperature and
humidity range the sensor will slowly come back
within the calibraon state by itself. Prolonged
exposure to extreme condions may also
accelerate ageing.
When tracking changes in ambient condions,
the response me of the humidity sensor in
your data logger is approximately 20 minutes to
reach 90% of the reading. However, if you are
measuring step changes in humidity (for example
if calibrang the product) it is advised that you
leave the unit for up to four hours to ensure that
it has enough me to sele at the new level.
It is worth remembering that the value of relave
humidity is of course sensive to temperature
variaon. As an example, at a relave humidity of
~90%RH at ambient temperature, a variaon in
temperature of 1°C will result in a change of up
to -5%RH. Therefore when comparing mulple
devices or calibrang them, any temperature
variaons must be considered.
The humidity measuring element in the humidity
data loggers can be contaminated through
exposure to a variety of compounds. These
products should not be kept in proximity to
volale chemicals such as solvents and other
organic compounds. Generally speaking, if a
material or compound emits a strong odour
you should not keep your humidity data logger
in close proximity to it. If you would like more
informaon, please contact your local Lascar
Electronics oce.
High levels of pollutants may cause permanent
damage to the internal sensor.
Typical tolerance of RH over Temperature:
SENSOR ACCURACY AND INFORMATION
Typical and maximal tolerance for the relave
humidity sensor at 25°C:
Typical and maximal tolerance
for the temperature sensor in °C:
Parameter Conditions Value
CO2measurement range1- 0 to 40,000 ppm
CO2measurement accuracy2400 to 2,000 ppm ±(50 ppm +5% of reading)
CO2measurement accuracy2400 to 5,000 ppm ±(40 ppm +5% of reading)
Repeatability Typical ±10 ppm
Response time3T63%, typical 60 s
Accuracy drift per year with
automatic self-calibration4Typical ±(5 ppm +0.5% of reading)
Default condions of 25°C, 50% RH and ambient pressure of 1013 mbar apply to values in the table below.
CO2SENSOR