Enable Lifecare Aria + Operator's manual

Aria+
USER INSTRUCTION MANUAL
Product name
to go here
Product description to go here
USER MANUAL
enablelifecare.com.au
Aria
Pressre Care System
+

Aria+2
Aria+ is a pressure relieving mattress suitable for use with patients at VERY HIGH RISK of pressure
ulcer damage. Oering high levels of patient comfort, this unique system has the facility to “step
up” to that of a dynamic mattress when clinically required. Similarly, the mattress’s function can be
downgraded as the patient’s condition improves. These features make it particularly beneficial for use
within the patient’s home or palliative care environment and help reduce logistic and decontamination
costs. The clinical benefits of a single system are equally applicable to those of a modern hospital
setting. A higher maximum weight capacity, up to 39 stone / 250kg, allows the product to meet the
modern challenges of those heavier clients. All component parts are interchangeable and replaceable,
maximising product life and reducing environmental impact.
The Mattress consists of a foam head cell and series of 7 pairs of transverse air cells, each containing
a unique foam profiled insert, which are in turn held within a foam U Core, all protected by a vapour
permeable waterproof cover. The single pillow end consists of foam only. The transverse cells are
arranged into alternate pairs of A and B cells which are filled and emptied in sequence.
In Static Mode, the mattress attains the pressure reducing properties of the Gio-Form Care Foam
mattress (details available on request), whilst in Alternating Mode the mattress is able to oer similar
properties to a pressure relieving dynamic system.
The digitally controlled Power Unit controls a pump that allows air to flow into, or out of the air cells
as required according to the operating mode selected. It also maintains the air pressure within the
mattress at the required level and controls the action of the audible/visual Audible Warning system in
the event of mains supply failure or over or under inflation pressure. A CPR Valve located at the pump
end of the umbilical hose permits the rapid deflation of the Mattress in an emergency.
1. Introduction
A
A
B
B
A B A B

Aria+3
NOTE: Please ensure that all securing straps on the base of
the mattress are secured onto the NON MOVING PARTS of
the bed frame. For shut down procedure, see 4.2 Power Unit
(Pump) section.
Note: Please ensure that all securing straps on the base of
the mattress are secured onto the
NON MOVING PARTS of the bed frame.
For shut down procedure, see 4.2 Power Unit (Pump)
section.
Power Switch Audible Warning Reset
The power switch simply switches the mains power to
the pump on and off. When the pump detects an Audible
Warning condition, this can be silenced as below and re-
set by switching the pump off and then back on again.
CPR Valve
Please ensure that the CPR connector is always placed
fully home, prior to inating the mattress.
NB: The mattress will NOT inate properly should this not
be the case. The CPR connector is only to be used in the
event of a clinical emergency for priority use. However,
disconnecting this function will cleverly deate air rapidly
from the mattress in readiness for transport / static mode.
Warning Modes
There are two red LEDs which inform the user of potential
problems with the mattress system.
The yellow warning triangle denotes a pressure issue in
the mattress.
Low Pressure Alarm:- The red LED will ash slowly whilst
an audible alarm also sounds. Check all hose connections
High Pressure Alarm:- The red LED ashes rapidly whilst
an audible alarm also sounds.
System Failure:- The red LED next to the spanner will
ash and an audible alarm also sounds.
Pressure Settings
The +/- button is used to adjust the pressure from Low
pressure to High pressure. The left hand LED denotes
Low pressure selected. The right hand LED denotes High
pressure selected.
The power switch simply switches the mains power to the
pump on and o. When the pump detects an Audible Warning
condition, this can be silenced as below and reset by switching
the pump o and then back o againi.
Please ensure that the CPR connector is always placed fully
home, prior to inflating the mattress.
NB: The mattress will NOT inflate properly should this not be
the case. The CPR connector is only to be used in the event of
a clinical emergency for priority use. However, disconnecting
this function will cleverly deflate air rapidly from the mattress
in readiness for transport / static mode.
There are two red LEDs which inform the user of potential
problems with the mattress system.
The yellow warning triangle denotes a pressure issue in the
mattress.
Low Pressure Alarm: The red LED will flash slowly whilst an
audible alarm also sounds. Check all hose connections
High Pressure Alarm: The red LED flashes rapidly whilst an
audible alarm also sounds.
System Failure: The red LED next to the spanner will flash and
an audible alarm also sounds.
The +/- button is used to adjust the pressure from Low pressure
to High pressure. The left hand LED denotes Low pressure
selected. The right hand LED denotes High pressure selected.
Power Switch Audible Warning Reset
CPR Valve
Warning Modes
Pressure Settings

Aria+
4.1. Mattress
4.2. Power Unit (Pump)
4
Aria+ can be used as a pressure reducing mattress for patients at High Risk of pressure ulcer damage
without the need to attach the pump.
If/When required, the Aria+ can be used as an alternating mattress system by attaching the Aria+
pump system. No other system should be attached to the mattress as the design settings and internal
air pressure properties of the Aria+ pump are specific to this mattress only.
The Aria+ is a replacement mattress system and should NOT be placed on top of any existing mattress.
Hang the Power Unit (Pump) onto the foot board. The mounting hooks swivel to suit the thickness
of the foot board or rail. Connecting the Umbilical Hose to the Power Unit (Pump), place the 3-pin
electrical plug into the wall outlet and switch on:
Attach the Blue Umbilical Hose to the Power Unit (Pump) by connecting the air connector at the end
of the Umbilical Hose to the air inlet connector at the bottom left hand side of the pump. Ensure that
the Red CPR Release button is located on top of the Air Inlet connector after connection is complete.
Attach the mains cable to the pump by inserting the “kettle” type connector into the recess located
on the left hand side of the pump. The mains cable has been designed specifically as a removable
part to aid in easy replacement should it become damaged in use.
The mains plug should be turned o and removed from wall socket as a means of isolation.
Plug the mains cable into a suitable 230v mains socket and switch on the Power Unit using the on/
o switch.
After the pump has been turned on both the “Hi “and the “Lo” lights will flash together intermittently
until the pump has attained its initial operating pressure. Once the pump has attained its initial
operating pressure the “Lo” light will stay on constantly and the mattress is ready for use.
Place the Aria+ directly on to the bed platform ensuring that the Blue multi-stretch waterproof cover
is on top and that the umbilical hose is located at the left hand corner at the foot end of the bed. Only
Dartex material is used to construct the top covers.
Cover the Mattress with a loose fitting sheet.
The start up time from static to dynamic mode is immediate.
Shut down is the reverse of above.
Static Mattress Use
Alternating Mattress Use
5. Operation

Aria+5
5.1. Lo / Hi Settings
5.3. Troubleshooting
5.2. CPR Deflation
The Aria+ Mattress, in Alternatinig Mode, has two pressure settings. The initial setting that the pump
will revert to upon set up is “Lo”. The “Lo” comfort setting is ideal for the lighter patient or those
who feel discomfort when on a normal alternating air type mattresses system. However, for patients
with existing pressure damage or those at Very High Risk, it is recommended that dependant on
the clinical judgement of the clinician, the “Hi” setting is activated by pressing the +/- button once,
which is located on top of the pump. In “Hi” Mode the pump attains more of the characteristics
of an alternating air mattress system whilst still utilising the advantages of the static foam insert.
Repeatedly pressing the ‘mode’ button enables the Lo & Hi modes to be selected in turn.
For assistance (if needed) in setting up, using or maintaning the Aria+, or to report unexpected
operation or events, please contact Enable Lifecare on the contact details on the reverse of this
manual.
The CPR system consists of a manually operated button located on the Air Inlet connector attached
to the pump. By pressing the Red Button, which will release the connector locking system, the user
can remove the connector unit which will deflate the mattress air system back to that of a static foam
mattress.
NOTE: After a short period as the Mattress deflates the ‘Low Pressure’ Audible Warning is activated
and can be cancelled by switching the Power Unit o.
To change the location of the mattress, remove the Umbilical cord and allow the mattress to return
to its Static Mattress form. Switch o the Power Unit (Pump) using the on/o switch and disconnect
the electrical supply cable from the mains socket. The mattress can now be moved to a new location
where it must immediately be reconnected to the mains electrical supply and the Power Unit (Pump)
switched back on. Once the Mattress has been refilled, the ‘Alternating’ mode will automatically revert
back to the Lo setting and should be reselected to Hi should this be desired by the clinician.
6. Transportation
The Mattress will not ‘alternate’ when disconnected from the Power
Unit (Pump) and /or the mains electrical. Also refer to environmental
conditions section at rear of this manual.
WARNING
!

Aria+6
Audible Warning conditions are indicated by a flashing red display accompanied by an audible
warning. In each case the user should respond by turning the Power Unit’s switch o investigating
the cause.
7. Audible Warnings
8. Maintenance procedures
7.1. High Pressure Audible Warning
7.2. Low Pressure Audible Warning
7.3. Mains Failure Audible Warning
8.1. Safety Warning
8.1.1 Servicing
This condition could be caused, for example by a kinked Umbilical Hose or visitors, and others, sitting
suddenly on the Mattress.
This condition could be caused, for example, by incorrect fitting of the air inlet connector, opening of
the CPR Valve or a leak in the Mattress due to a cut or puncture.
If mains power is lost the all Mode lights will turn o. This Audible Warning condition will only be
audible. The red Audible Warning light will not flash.
Only qualified technicians trained or formally approved by Enable Lifecare in the operation and
maintenance of Enable Lifecare products may carry out maintenance, modification or repair work on
the equipment. Unqualified personnel attempting to work on Enable Lifecare Power Units risk serious
injury to themselves and others and possibly death by electrocution. Inlet fuse NOT to be replaced by
operator or patient, to be replaced by service personnel only.
Enable Lifecare recommend that the Power Unit (Pump) should be serviced every year. The unit
contains no user serviceable parts and should only be carried out by persons as described in section
8.1. Enable Lifecare will make available on request service manuals, component parts lists and other
information necessary for any suitably qualified person (As in 8.1) to carry out repair or service the
system. For Service, maintenance and any questions regarding this please contact Enable Lifecare.
Do not modify this equipment without authorisation of Enable
Lifecare.
WARNING
!

Aria+7
8.2. Cleaning Procedures
8.3. Warning – Cleaning the Power Unit (Pump)
8.3.1 Warning
8.4. Warning – Cleaning the Mattress
Before cleaning the System make sure that the Power Unit (Pump) is
disconnected from the mains electricity supply.
Do not immerse the Power Unit (Pump) in water or other fluids.
Do not autoclave, nor use phenol for cleaning.
Do wash hands before commencing the cleaning process.
Wear appropriate protective clothing such as gloves, apron and a
mask.
Ensure all work surfaces are cleaned before and after contact with
the mattress.
WARNING
!
The power Unit can be cleaned by wipping a cloth dampened with a detergent solution or Hypochlorite
solution.
Also referto symbol chart.
Ensure the Gio-Form Hybrid G4 System is not exposed to:
1. Excessive heat sources e.g. fires, radiators etc
2. Water, particularly immersion of the pump.
1. Cleaning should take place after use or between patients.
2. With cover left on the Mattress disconnect the Mattress from the Power Unit (Pump).
3. Clean the surface of the wash down table with Hypochlorite solution or equivalent disinfectant.
4. Wash Mattress top using hot water (60 degrees C) containing detergent – dry with a paper towel.
5. For heavy contamination use a Hypochlorite solution 1,000 parts per million available chlorine.
6. Using suitable brush, hot water, detergent or Hypochlorite solution, clean Umbilical Hose and CPR
Valve. Dry with paper towel.
7. If required, the Mattress Cover may be removed and machine washed at a temperature of 80
degrees C, for not less than 10 minutes. The individual Air Cells can be wiped down with established
disinfectants.
8. To avoid shrinkage of the cover line dry in an indoor clean environment or tumble dry on a low
heat setting not exceeding 40 degrees C and not for longer than 10 minutes. Covers must be
thoroughly dried before re-fitting to the mattress.

Aria+8
9. Technical data
Serial Number As per label on rear of pump
Electrical Supply 220-240 volt, 50 Hz
Power Consumption 10 watts
Fuses TA1H 250V
Protection against shock Class 2
Noise Level Approx. 25 dB (A)
Dimension 235 x 180 x 80 mm
Weight 1.7 kg
Service Interval 12 months
Expected life 5 years
Shelf life of parts 5 years
Warranty ( If serviced annually by Enable Lifecare
or Authorised Dealer) 12 months
Serial Number Label on inside of mattress cover
Number of Air Cells 10 Air Cells
Dimension 2000 x 880 x 180mm (Nominal)
Weight 20 kg
Expected life of Mattress 5 years
Shelf life of Mattress parts 5 years
9.1. Power Unit (Pump)
9.2 Mattress

Aria+9
EMC GUIDANCE AND DECLARATION
This guidance and manufacturer’s declaration pertains to the Gioco range of pressure relief equipment
which is listed at the back of this User Manual
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical
electrical equipment. This device complies with IEC EN 60601-1-2:2007.
• All medical electrical equipment must be installed and put into service in accordance with the EMC
information provided in this document.
• Portable and mobile RF communications equipment can aect the behaviour of medical electrical
equipment.
• Aria+ pump units comply with all applicable and required standards for electromagnetic interference.
• It does not normally aect nearby equipment and devices.
• It is not normally aected by nearby equipment and devices.
• It is not safe to operate Aria+ pumps in the presence of high-frequency surgical equipment.
• It is good practice to avoid using Aria+ pumps in extremely close proximity to other equipment.
The Pulse Press range of pumps are intended for use in the electromagnetic environment explained
below. The user should assure that they are used in such an environment.
EMC Compliance
Electromagnetic Emissions
EMISSIONS TEST COMPLIANCE ENVIRONMENT GUIDANCE
RF emmisions
CISPR 11
Group 1 Pumps emmision are very low and unlikely
to cause interference.
RF emmisions
CISPR 11
Class B Pumps are intended for use by medical
proesionals only.
Harmonic Emmisions
EN 61000-3-2
Class B If this equipment causes inference to near by
devices then try to relocate
or re-orient the device.
Voltage Fluctuations
EN 61000-3-3
Complies

Aria+10
Aria+ pump are intended for use in the electromagnetic environment explained below. The user
should assure that they are used in such an environment.
The device is intended for use in an electromagnetic environment in which the HF disturbances are
controlled. The user can contribute to the avoidance of electromagnetic disturbances by adhering
to the minimum distance required between portable and mobile HF telecommunications equipment
(transmitters) and the device – depending on the output power of the communications device, as
indicated below.
For transmitters whose maximum output power is not indicated in the above table, the recommended
safety distance in meters (m) can be determined by using the formula that belongs to the same
column, with P being the maximum output power rating of the transmitter in watts (W) according to
the manufacturer’s specifications.
Electromagnetic Immunity
IMMUNITY TEST 60601 TEST COMPLIANCE ENVIRONMENT GUIDANCE
Electrostatic discharge
IEC 61000-4-2
+/- 4kV Contact
+/- 8kV Air
+/- 4kV Contact
+/- 8kV Air
Floor should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2kV power
supply lines
+/- 1kV input
output lines
+/- 2kV power
supply lines
+/- 1kV input
output lines
Mains power quality should be
that of a typical commercial or
hospital.
Voltage Surge
IEC 61000-4-5
+/- 2kV
common mode
+/- 1kV
dierential mode
+/- 2kV
common mode
+/- 1kV
dierential mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips and
Interruptions
IEC 61000-4-11
30% Dip 25 cycle
60% Dip 5 cycle
100% Dip 0.5 cycle
100% Dip 250 cycle
30% Dip 25 cycle
60% Dip 5 cycle
>95% Dip 0.5 cycle
>95% Dip 250 cycle
RATED OUTPUT POWER 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5Ghz
P of transmitter in Watts d=1.17 √P d=1.17 √P d=2.33 √P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.4
100 11.7 11.7 23.3
RECOMMENDED SEPARATION DISTANCE
Conducted RF
IEC 61000-4-6
3 Vrms
150kHz to 80MHz
3 Vrms d=(1.17) √P 0.15 to 80 MHz
Radiated RF 3 V/m 3 V/m d=(1.17) √P 80 to 800 MHz
d=(2.33) √P 0.8 to 2.5 GHz

Aria+
At Enable we source quality,
aordable and innovative
products globally and have
them available for fast delivery.
But we do more than just
provide products.
We give the knowledge and
support to help our customers
grow their business and deliver
better life care.
Table of contents
Other Enable Lifecare Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Konica Minolta
Konica Minolta AeroDR Portable UF Unit Operation manual

OERTLI
OERTLI OS 4 Service manual

Top shelf Orthopedics
Top shelf Orthopedics TR Post-Op PCL Instructions for use

Optiflux
Optiflux F18NR quick start guide

orfit
orfit RAYCAST Instructions for use

Imetec
Imetec TERMOFORO E0801 Instructions for use