eschmann Little Sister UltraSONICA m User manual

Powered by TCPDF (www.tcpdf.org)
1
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
ULTRASONIC CLEANER
Instructions for use

Powered by TCPDF (www.tcpdf.org)
2
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
TABLE OF CONTENTS
Section 1 ............................................................................................................................................... 3
Introduction to Ultrasonic Cleaning ............................................................................................. 3
Section 2 ............................................................................................................................................... 3
Warranty ......................................................................................................................................... 3
Safety ................................................................................................................................................ 4
Symbols ............................................................................................................................................ 4
Section 3 ............................................................................................................................................... 4
Electrical connection ....................................................................................................................... 4
Before operating the device ............................................................................................................ 5
Recommended Cleaning Solution .................................................................................................. 6
........................................................................................................................................................... 6
Preventing damage to the device ................................................................................................... 6
Important when using the Ultrasonic Cleaner ............................................................................. 7
Section 4 ............................................................................................................................................... 7
Ultrasonic Cleaner Figures ............................................................................................................ 7
Section 5 ............................................................................................................................................... 8
Operating instructions .................................................................................................................... 8
Pre-Use (If Ultrasonic Cleaner has been drained) ........................................................................ 8
During Use .................................................................................................................................... 9
After Use ....................................................................................................................................... 9
Section 6 ............................................................................................................................................. 10
Accessories ..................................................................................................................................... 10
Section 7 ............................................................................................................................................. 10
Useful tips and suggestions ........................................................................................................... 10
Section 8 ............................................................................................................................................. 11
Technical Specification ................................................................................................................. 11
Section 9 ............................................................................................................................................. 11
Maintenance and Repair .............................................................................................................. 11
Section 10 ........................................................................................................................................... 12
Product Disposal ........................................................................................................................... 12

Powered by TCPDF (www.tcpdf.org)
3
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
SECTION 1
Introduction to Ultrasonic Cleaning
Ultrasonic: what is it?
Ultrasonic Cleaners employ the use of sound waves at very high (inaudible) frequencies
(approximately 39 kHz) to agitate a cleaning medium, which in turn results in the removal of debris
and contaminants from the object being processed.
The Ultrasonic Cleaner’s tank is made up of an ultrasonic generator and two or three piezoelectric
transducers (depending on the model) that are placed on the external bottom of a stainless steel tank
containing the cleaning solution. The electronic generator produces a continuous signal at the
frequency of 39 kHz and drives the piezoelectric transducers that transform the electrical signal into
a mechanical vibration. This vibrational energy is transmitted to the cleaning solution within the tank
at a frequency of 39,000 oscillations per second. These pressure and depression oscillations create an
enormous quantity of microbubbles inside the cleaning solution that, by imploding in rapid
succession, create immense impact energy between the cleaning solution and the surface to be
cleaned. This is called "cavitation" and enables an efficient and safe cleaning system with reduction
in time.
Sweep System Technology
The Little Sister Ultrasonic Cleaner contains a brand new generator which incorporates a special
ultrasonic oscillator called Sweep System Technology. With this technology, the output frequency
of the ultrasonic generator is modulated around a central frequency, so the transducers working at a
frequency of 39 kHz are modulated at a frequency between 38 and 40 kHz. This frequency modulation
offers the following advantages:
• Decreases cleaning times;
• Prevents the damage of delicate parts to be cleaned;
• Decreases the effects of stationary waves increasing the distribution of ultrasonic energy in
the cleaning solution;
• Improves cleaning results and facilitates the cavitation in liquids that are difficult to cavitate
with traditional ultrasonic systems.
SECTION 2
Warranty
This product and purchased accessories are warranted for a period of 12 months to be free from
defects in materials and workmanship at the time of delivery.
Eschmann will be under no liability for any defect arising from fair wear and tear, negligence, wilful
damage, misuse, abnormal working conditions, failure to follow the manufacturer’s instructions,
unauthorised alteration or repair of hardware, unauthorised or accidental alteration of software or
configuration, lost profits, commercial loss, economic loss, or loss arising from personal injury. We
may, at our discretion, raise a charge for any faults repaired that fall outside the warranty cover.
Where charges are necessary, replacement parts will be charged at manufacturers’ list prices and
labour will be charged at the prevailing hourly rate. Repairs performed by Eschmann carry a 3-month
parts and labour warranty.

Powered by TCPDF (www.tcpdf.org)
4
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
Safety
To preserve the original safety features of the product, the Customer must not replace any part or
make any unauthorised changes.
This symbol focuses attention on a procedure (s) whose failure or partial compliance may produce
partial or total damage to the product or lead to injury to the operator. Before performing the
procedures reported after this symbol, be sure to fully understand and comply with the specified
conditions.
Symbols
Consult the instructions for use
Manufacturer
Serial Number
Catalogue Number
CE mark
Indicates the need for the user to consult the instructions for use for
important cautionary information such as warnings and precautions that
cannot, for a variety of reasons, be presented on the medical device itself
WEEE recycling: Do not dispose of product at end of life with other waste
SECTION 3
Electrical connection
Before connecting the Ultrasonic Cleaner to power, check that the voltage indicated on the rating
plate attached to the rear of the unit (Fig.2 - 5), corresponds to that of your workplace.
ATTENZI
ONE
ATTENZI
ONE
ATTENZI
ONE
ATTENZI
ONE
ATTENZI
ONE
ATTENZI
ONE
WARNING

Powered by TCPDF (www.tcpdf.org)
5
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
The earthing of the device is compulsory according to law. It is prohibited to intentionally
interrupt the earthing conductor inside or outside the device or remove the earth pin from the
plug, as it makes the use of the device dangerous.
The manufacturer assumes no liabilitytowards people or property resulting from failure to observe
this rule. The electrical safety of this device is ensured only when it is properly connected to an
effective earthing system, as required by the electrical safety regulations in force.
THIS DEVICE REQUIRES EARTHING.
If you do not have an electrical system with regular earthing, do not connect the device to the power
outlet and consult a qualified electrician as soon as possible.
Before operating the device
Before connecting the Ultrasonic Cleaner to power fill the tank with recommended cleaning
solution up to a maximum level of 3 cm from the top. Refer to Recommended Cleaning Solution
heading below. If the Ultrasonic Cleaner is equipped with a manual drainage tube (Fig. 3 - 9), check
that the blue valve on the green tap is properly closed (Fig. 3 - 8) and held in place using the clip (Fig.
3 - 7) provided before filling the tank and during use. This will prevent accidental spillage of liquid
substances which may cause a short-circuit or electrocution and therefore be dangerous for the
operator.
Make sure that the device is not damaged. Do not use damaged devices; if in doubt, ask your
supplier.
Install the ultrasonic cleaner on a flat and stable surface, able to support the weight of the device,
instruments and cleaning solution that are placed in the Ultrasonic Cleaner. Handle it with care.
Install the Ultrasonic Cleaner away from heat sources. In addition, be careful not to install it near
sources of moisture, wet work surfaces, or near sources of dust.
Make sure that the Ultrasonic Cleaner’s feet are in the correct position to ensure air circulation.
Do not use the Ultrasonic Cleaner if: the power cord or plug is damaged or does not work properly.
This may lead to electrical shock, fire or other accidents. Do not attempt to personally service the
device. Contact only qualified technical personnel or the supplier.
Do not immerse the power cord or the plug in water. Keep the power cord away from hot surfaces.
Do not let the cord hang over the edges of tables or furniture.
Do not leave or do not use this device outdoors.
Do not tamper with, for any reason, the electronic system of the Ultrasonic Cleaner "Electric shock
hazard". For any repairs, contact the supplier.
Do not replace the power cord. If the cable is damaged due to wear, or other reasons, turn off the
device immediately and contact the supplier.
Do not lift and carry the device when it is full of cleaning solution or connected to the power outlet.
Lift and carry the device only when it is empty and disconnected from the power outlet.
WARNING
WARNING

Powered by TCPDF (www.tcpdf.org)
6
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
Recommended Cleaning Solution
The Little Sister Ultrasonic Cleaner should only be used with the cleaning solution recommended
by the manufacturer.
To clean any instrument, you should principally use clean running water and dissolve in it a quantity
of detergent or solution depending on the type of material residue to be removed from the instrument
itself. Use only solutions that are appropriate to the type of work to be performed. Only use undiluted
detergent if it is explicitly specified in the product's instructions for use. Use only the amount of liquid
needed to clean. Always follow the instructions for use of the solution.
Do not use any of the following solutions:
• Acidic or strong alkaline solutions such as sodium hypochlorite because these substances,
when in direct contact with the tank, cause a series of microscopic holes with irreparable and
especially dangerous consequences for the operation of your device.
• All acidic or alkaline substances that can release corrosive compounds such as chlorine or
other chemical compounds, or glutaryaldehyde disinfectants, used together with the energetic
ultrasonic cavitation activity, can cause accelerated corrosion of stainless steels.
• Substances such as petrol, benzene, benzol or other harmful, explosive or flammable solvents.
Any acid particles, micronised by intense ultrasonic cavitation activity, are dispersed in the working
environment causing corrosion to the device and also damaging health.
Preventing damage to the device
To prevent damage to the device:
• Periodically replace the solution. It is recommended that the cleaning solution is replaced at
the end of every session, or as a minimum daily, unless local guidance recommends a more
frequent schedule or if the solution is heavily contaminated.
• Do not operate the device without the recommended cleaning solution.
• Do not put parts or containers directly in contact with the bottom of the tank, use the stainless
steel basket provided.
• When the Ultrasonic Cleaner is in operation, make sure that the cleaning solution does not fall
below 2/3 of the total height of the tank.
Failure to follow such instructions may cause damage to the transducers, invalidating the warranty.
WARNING

Powered by TCPDF (www.tcpdf.org)
7
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
Important when using the Ultrasonic Cleaner
Before connecting the Ultrasonic Cleaner to power, if the system has
been drained, fill the tank with recommended cleaning solution up to a
maximum level of 3 cm from the top. Refer to Recommended Cleaning
Solution on page 6.
If the Ultrasonic Cleaner is equipped with a manual drainage tube (Fig.
3 - 9), check that the blue valve on the green tap is properly closed (Fig.
3 - 8) and held in place using the clip (Fig. 3 - 7) provided before filling
the tank and during use.This will prevent accidental spillage of liquid
substances.
If new solution has been added to the Ultrasonic Cleaner, the system
must first be de-gassed. To de-gas the Ultrasonic Cleaner turn the
Timer/Switch knob clockwise (Fig. 1-1) and run for 10 minutes without
any instruments or basket inside the tank.
Always pay attention to the liquid level so that it does not fall below
2/3 of the total height of the tank. If this occurs, the ultrasonic
transducers and electronic circuitry may be seriously damaged.
Do not overload the basket and ensure the instruments within the
basket are not touching.
Do not immerse your hands in the tank or basket during operation.
Use this Ultrasonic Cleaner only for the use described in this manual and
for the purpose for which it was designed. This device is designed to clean
and detach excess materials from approved dental and medical
instruments. Always refer to the instrument manufacturers cleaning
instructions.
SECTION 4
Ultrasonic Cleaner Figures
Fig. 1: FRONT
1. Timer/Switch knob
2. ‘0’ position (cleaning time expressed
in minutes)
3. ∞ position (infinite cleaning time)
4. Power indicator light

Powered by TCPDF (www.tcpdf.org)
8
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
Fig. 2: REAR LABEL
5. Serial number plate and device
identification
6. Power cord input
Fig. 3: MANUAL DRAINAGE TUBE (OFF)
7. Clip (holds manual drainage tube in
place)
8. Blue valve in the CLOSED position
9. Manual drainage tube
10. Green tap
Fig. 4: MANUAL DRAINAGE TUBE (ON)
11. Blue valve in the OPEN position, in
preparation for draining the tank
SECTION 5
Operating instructions
PRE-USE (IF ULTRASONIC CLEANER HAS BEEN DRAINED)
1. Before connecting the Ultrasonic Cleaner to power, if the system has been drained, fill the
tank with recommended cleaning solution up to a maximum level of 3 cm from the top. Refer
to Recommended Cleaning Solution on page 6. The Ultrasonic Cleaner is equipped with a
manual drainage tube (Fig. 3-9). If the Ultrasonic Cleaner is equipped with a manual drainage
tube (Fig. 3-9), check that the blue valve on the green tap is properly closed (Fig. 3-8) and
held in place using the clip (Fig. 3-7) provided before filling the tank and during use. This
will prevent accidental spillage of liquid substances.

Powered by TCPDF (www.tcpdf.org)
9
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
2. If new solution has been added to the Ultrasonic Cleaner, the system must first be de-gassed.
To de-gas the Ultrasonic Cleaner turn the Timer/Switch knob clockwise (Fig. 1-1) and run for
10 minutes without any instruments or stainless steel basket inside the tank.
DURING USE
3. Place the instrument(s) into the stainless steel basket and into the Ultrasonic Cleaner tank. Do
not overfill the Ultrasonic Cleaner, refer to Useful tips and suggestions on page 10.
4. Place the stainless steel cover onto the Ultrasonic Cleaner (unless using the stainless steel
container for small parts).
5. Set the cleaning time by turning the Timer/Switch knob (Fig.1 - 1) clockwise. The power
indicator light (Fig.1 - 4) will light up on the front panel to indicate the actuation of the unit.
To stop the operation of the Ultrasonic Cleaner at any time, simply turn the Timer/Switch
knob (Fig.1 - 1) anti-clockwise to the "0" position (Fig.1-2).
6. At the end of the set time the machine will turn off automatically.
To set a cleaning time of more than 15 minutes, turn the Timer/Switch knob (Fig.1- 1) anti-
clockwise to the "∞" position (Fig.1 – 3). To stop the operation of the Ultrasonic Cleaner at
any time, simply turn the Timer/Switch knob (Fig.1- 1) clockwise back to the "0" position
(Fig.1- 2). This will allow the user to set the cleaning time manually.
AFTER USE
7. After the cycle has completed, instruments must be removed and rinsed in clean running water
to remove any residue or contaminant that might be in the Ultrasonic Cleaner or on the
instrument. This should be performed before the instrument is further processed
8. It is recommended that the cleaning solution is replaced at the end of every session, or as a
minimum daily, unless local guidance recommends a more frequent schedule or if the solution
is heavily contaminated. To drain the Ultrasonic Cleaner:
a) Drain the solution by turning the blue valve to the OPEN position (Fig.4 - 11) on the green
tap. The solution must be drained into an appropriate disposal area as per local guidance.
b) Once the tank is empty, turn the blue valve to the CLOSED position (Fig.3 - 8) on the
green tap and position the drainage tube using the clip (Fig.3 - 7) provided.
If the Ultrasonic Cleaner does not come equipped with a manual drainage tube, empty tank by tipping
the tank (with stainless steel basket removed) into an appropriate disposal area as per local guidance.
When the tank is drained, always ensure it is left clean (i.e any signs of contamination are removed)
using recommended cleaning/disinfection wipes.

Powered by TCPDF (www.tcpdf.org)
10
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
SECTION 6
Accessories
Stainless steel cover
Used to cover the Ultrasonic Cleaner during use.
Stainless steel basket
The stainless steel basket optimises the cleaning of instrument(s)
and then drains them into the tank.
Stainless steel container for small parts
Useful accessory for small instruments including burrs.
Note: Do not place this item directly into the tank. This must be
placed into the stainless steel basket and then into the tank.
SECTION 7
Useful tips and suggestions
The type of cleaning solution to use. To clean any instrument, you should principally use clean
running water and dissolve in it a quantity of detergent or solution depending on the type of material
residue to be removed from the instrument itself. Only use undiluted detergent if it is explicitly
specified in the product's instructions for use. Always follow the instructions for use of the solution.
Do not overfill the Ultrasonic Cleaner. Always place the instruments in a logical manner at the
bottom of the stainless steel basket, place the basket in the tank and proceed with the appropriate
cleaning. Too many instruments cleaned simultaneously will reduce the effectiveness of cleaning and
if touching, could lead to damage.
When objects are very dirty, use longer cleaning times.
When to replace the cleaning solution. It is recommended that the cleaning solution is replaced at
the end of every session, or as a minimum daily, unless local guidance recommends a more frequent
schedule or if the solution is heavily contaminated.

Powered by TCPDF (www.tcpdf.org)
11
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
Shape and size. There are no special recommendations for the shape of the instruments to be cleaned,
however avoid placing in the tank particularly heavy objects of very large dimensions. The cleaning
solution must always cover the entire object.
Cleaning time. The cleaning time always depends on the type of instrument and level of
contamination.
SECTION 8
Technical Specification
Little Sister
UltraSONICA m Model
2200M – 3L
2400M – 4.5L
3200M – 6L
Power supply voltage
230/240V - 50/60Hz
Consumption
130 W
130 W
180 W
Solution volume (Litres)
3
4.5
6
Weight in kg. (approx.)
3.8
3.8
4.5
External dimensions (in mm)
270x170x260
325x175x260
325x270x260
Tank dimensions (in mm)
240x140x100
300x150x100
300x240x100
Number of transducers
2
2
3
Environmental conditions
Temperature from 5 to 40°C; relative humidity 80% up to 31°C
with linear decrease up to 50% at 40°C.
Installation conditions
Class II according to EN 61010-1
Information for use as "Medical Device"
Name: Ultrasonic Cleaning Device UMDNS-nomenclature (ECRI) 14-263
Field of application: cleaning and disinfection of surgical and dental instruments.
Classification (Medical Devices Directive 2007/47/EC, Appendix IX): Class I; active,
non-invasive, non-implantable medical device.
SECTION 9
Maintenance and Repair
Always disconnect the plug before carrying out any maintenance. To always ensure the safety of
the device, carry out a daily check to ensure that the external frame, tank and power cord of the device
are not damaged.
Cleaning is the only maintenance normally required. It must be carried out with the device
disconnected electrically. Only use neutral detergents to clean the inside of the tank and a soft cloth
to clean the outside such as the frame and the command panel.
Do not leave dirt deposits inside the tank. If the liquid outlet hole is obstructed by dirt residues,
clean with a non-pointed flexible stick, making sure that the rubber hose is not punctured.

Powered by TCPDF (www.tcpdf.org)
12
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
If the device does not work, perform the following checks:
• Check that the plug is properly plugged into the socket.
• Check the level of liquid in the tank.
Maintenance and repair should only be carried out by technicians trained by the manufacturer. It is
dangerous for any other person to carry out repairs. If assistance is required, contact your supplier.
SECTION 10
Product Disposal
The crossed out wheelie bin symbol on the equipment or its packaging indicates that the product at
the end of its useful life must be collected separately from other waste.
Separate collection of this equipment at the end of its life is organised and managed by the supplier.
The user shall contact the supplier and follow the method that has been implemented for disposing of
the equipment at the end of its life. The correct disposal technique of the used equipment, which must
be recycled, treated and disposed of in an environmentally friendly way, helps to avoid possible
negative effects on the environment and on health; it also facilitates the recycling of the materials the
equipment is composed of.
Incorrect disposal of the product by the user involves application of the administration sanctions
according to the laws in force.
The Declaration of conformity according to 2012/19/EU Directive must be required by sending an e-
mail to [email protected]. In accordance with EN 13306 standard, the lifetime is the period of
time which starts at a given instant and it ends when the failure rate is unacceptable, or when the
product cannot be repaired following a failure or other relevant elements.
The Little Sister Ultrasonic Cleaner has a lifetime of 6 years.
WARNING

Powered by TCPDF (www.tcpdf.org)
13
Instructions for Use
Little Sister UltraSONICA m Ultrasonic Cleaner / Rev.00 – Dated 09/12/2019
ESCHMANN TECHNOLOGIES Ltd
Eschmann House – 15 Peter Road
Lancing Business Park
Lancing
West Sussex – BN15 8TJ
Tel. +44 (0)1903 875787
Email: [email protected]
www.eschmann.co.uk
SOLTEC S.r.l.
Via Guglielmo Röntgen 16
20136 Milano (MI) - ITALY
Table of contents
Popular Cleaning Equipment manuals by other brands

Ultrawave
Ultrawave Neon Series instruction manual

Boule
Boule Medonic Series Instructions for use

BAUMOT
BAUMOT BA5000 operating manual

Fanset
Fanset Fantom Professional PROCART 450 Assembly manual

Bissell
Bissell CleanView ALLERGEN PET REWIND 3402 Series manual

Viavi
Viavi CleanBlastPRO FCLP-ACC-KIT1 user guide