Air leakage through Inner door gasket.
Check installation of the gasket ensure it is
properly secure to the cabinet and that there
is no gap.
Check the joining ends of the gasket to make
there is no gap. Use silicon sealant to seal the
joins if there is gap.
Check gasket profile is not deformed or torn.
Otherwise replace new door gasket.
Humidity level is below standard
humidity level (For units with RH display
options installed)
Possible Cause
There is no water or not enough water in
the humidity pan.
Ensure there is enough water in the
humidity pan. For normal operating
temperature, humidity pan should be half
filled.
Cleaning & Decontamination
Cleaning Procedure
1. Transfer all samples to another CO₂ incubator or store them in a safe place.
2. Turn off and unplug the device. If needed; mark that the unit is deactivated or being
serviced.
3. Strip the unit for cleaning:
Remove the ULPA filter. See section 2.2.3 ULPA Filter Installation.
Remove the shelf system. See section 2.2.4 Shelves System Installation.
Remove the sparger and tubing (for suppressed O2 model) and the humidity pan.
4. Start cleaning process:
Thoroughly remove dirt and residues onto the surfaces of the work space and on the
accessories.
Spray disinfectant onto work space surfaces and all dismantled parts or wipe with a
cloth dampened with disinfectant.