eXact Pool EZ User manual

1
Sensafe.com
QUICK START GUIDE
Pool
ADVANCED PHOTOMETER SYSTEM
DIGITAL ACCURACY FOR BETTER POOL CARE

2
©2016, Industrial Test Systems, Inc. Printed in USA
WELCOME TO YOUR NEW POOL EXACT
®
EZ
PHOTOMETER.
Thank you for your Pool eXact
®
EZ purchase! This guide will quickly
walk you through the technical details of your new photometer. After
initial set-up, test procedures, and tips, you will be on your way to
digital water testing! Each test will require the use of one or more of
the testing methods outlined in this manual.
YOUR POOL EXACT
®
EZ COMES WITH:
•Cleaning Brush
•Quick Start Guide (this booklet)
•48 foil packet strips (6 of each): Free Chlorine (DPD-1),
Combined Chlorine (DPD-3), pH, Total Alkalinity, Calcium
Hardness, Phosphate, Chloride, and Copper
WHAT YOU WILL NEED TO GET STARTED:
•Four (4) AAA batteries
•#4 Phillips head screwdriver
Note:
This system has been calibrated for use with only our
eXact®Micro reagents.
CONTENTS
Welcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Pool eXact®EZ Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Install “AAA” Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Fill, Dip, Read
Total Alkalinity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
pH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Free Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Combined Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11
eXact®Strip Micro CL Interferences . . . . . . . . . . . . . . . . . . . . 11
Phosphate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-13
Calcium Hardness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-15
Cyanuric Acid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-17
Copper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19
Chloride (Salt). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-22
For Best Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pool eXact®EZ Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
About
Built-in Sample Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Warranty (2 years) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Method Verification Ready Snap®3. . . . . . . . . . . . . . . . . . . . . 25
Compliance Testing (USEPA) . . . . . . . . . . . . . . . . . . . . . . . . . 26
Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
eXact® Micro Tests & Reagents . . . . . . . . . . . . . . . . . . . . . . . . 28
Patent Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

3
Sensafe.com
POOL EXACT®EZ OVERVIEW
YOUR NEW POOL EXACT EZ PHOTOMETER IS IDEAL FOR
TESTING AND MAINTAINING YOUR POOL AND SPA WATER.
SAMPLE CELL - 4 ML
Built-in plastic 4mL cell
for collecting your water
sample
LCD DISPLAY
Displays results
and test name
LIGHT SOURCE
Uses 525nm
wavelength and
11mm path-length
for testing
READ BUTTON
Starts test timer
MENU BUTTON
Cycles through
available tests
BASE
Install/replace batteries here
(IP67 rated waterproof)
RESULT
UNITS
LOW
BATTERY
LCD DISPLAY
ZERO/ON BUTTON
Turns the meter
on and creates a
baseline for your
water testing
CELL COVER
Covers the cell for
mixing and bright
light situations
Protected against the effects
of immersion in water to depth
between 15cm and 1m
Part no. 486201
SCREW
Unscrew to
remove base
INSTALL “AAA” BATTERIES (NOT INCLUDED)
1. Use a #4 Phillips head screwdriver to remove the screw
from the base of your Pool eXact®EZ
.
2. Remove the base.
3. Install Four (4) new AAA batteries as illustrated inside
your photometer’s battery compartment. We recommend
using high quality batteries.
4. Replace the base firmly with pressure while tightening the
screw. The meter will turn on automatically.
5. Tighten the screw with #4 Phillips head screwdriver. Be
sure not to over tighten.

4
©2016, Industrial Test Systems, Inc. Printed in USA
TOTAL ALKALINITY
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows AL1.
ZERO METER
Press and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
4
FILL, DIP, READ
In bright light situations use cell cover for zeroing and reading.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

5
Sensafe.com
TOTAL ALKALINITY
REMOVE STRIP
Remove one eXact
®
Strip Micro AL Total Alkalinity
and set in a dry, convenient place. Replace cap
on bottle.
5
READ RESULTS
READ result displayed as Total Alkalinity. This result is
automatically stored in the AL1 menu. After testing is
complete, rinse the sample cell immediately.
7
FILL, DIP, READ
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
Extender Strip (part no. 486665) until a numerical value is
displayed; each strip adds 130 ppm to the final result.
TIP
For water temperatures above 95°F/35°C (hot tubs), remove
the strip when the timer displays “10”, or use the specially
calibrated SPA eXact
®
EZ photometer.
TIP

6
©2016, Industrial Test Systems, Inc. Printed in USA
PH
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows PH2.
ZERO METER
Press and the LCD Display reads 0.0PH,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
FILL, DIP, READ
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

7
Sensafe.com
PH
FILL, DIP, READ
REMOVE STRIP
Remove one eXact
®
Strip Micro PH and set in a dry,
convenient place. Replace cap on bottle.
5
READ RESULTS
READ result displayed as pH. This result is
automatically stored in the PH2 menu. After testing is
complete, rinse the sample cell immediately.
7
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6

8
©2016, Industrial Test Systems, Inc. Printed in USA
FREE CHLORINE AND BROMINE
FILL, DIP, READ
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows CL3.
ZERO METER
Press and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2

9
Sensafe.com
FREE CHLORINE AND BROMINE
REMOVE STRIP
Remove one eXact
®
Strip Micro CL (DPD-1) Free
Chlorine and set in a dry, convenient place. Replace
cap on bottle
5
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6
READ RESULTS
READ result displayed as Free Chlorine (for result
as Bromine, multply value by 2.6). This result is
automatically stored in the CL3 menu. After testing is
complete, rinse the sample cell immediately.
7
Important! Do not empty the CELL if you are also
testing for Combined Chlorine. You will need the sample.
FILL, DIP, READ
If result is 6.0 or greater, repeat steps 5-7 above
using same sample and a fresh strip.
TIP

10
©2016, Industrial Test Systems, Inc. Printed in USA
COMBINED CHLORINE
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays ‘1’. Remove and discard the strip.
10
This procedure is a continuation from Free Chlorine Test Method
on previous pages.
DO NOT DISCARD SAMPLE FROM FREE
CHLORINE TEST.
REMOVE STRIP
Remove one eXact®Strip Micro CL (DPD-3) Combined
Chlorine and set in a dry, convenient place. Replace
cap on bottle.
9
8
ZERO METER
With the Free Chlorine sample still in the CELL,
press and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
FILL, DIP, READ
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

11
Sensafe.com
COMBINED CHLORINE
11
READ RESULTS
READ result displayed as Combined Chlorine. This
result is automatically stored in the CL3 menu. After
testing is complete, rinse the sample cell immediately.
eXact
®
Strip Micro CL (DPD-1, DPD-3, DPD-4) Interferences
Interfering
Substance
Interfering Levels & Treatments
Acidity If sample has acidity above 150mg/L CaCO3
test may not develop full color. Neutralize to
pH 6.0 to 7.0 with 0.5N Sodium hydroxide.
Alkalinity If sample has alkalinity above 200mg/L
CaCO3 test may not develop full color.
Neutralize to pH 6.0 to 7.0 with 0.5N
Sulfuric acid.
Bromine &
Bromamines, Br
2
Color similar to free chlorine reaction at all levels.
Chlorine Dioxide, ClO2
Color similar to free chlorine reaction at all levels.
Copper, Cu
+2
Color development is reduced above 10
ppm (mg/L).
Iodine, I
2
Color similar to free chlorine reaction at all levels.
Manganese, Oxidized
(Mn
+4
, Mn
+6
)
or
Chromium,
Oxidized (Cr
+6
)
See AWWA procedure 4500-CL F, 1(d) for
removal of interferences.
Monochloramines
(NH
2
Cl)
(applies to DPD-1 only)
Monochloramine interferences are known
to occur in free chlorine
DPD methods. This interference is
dependent on temperature and
monochloramine concentration.
Ozone, O
3
Color similar to free chlorine reaction at all levels.
Peroxides Interference is possible.
pH Typical pH samples of potable water with
a pH of 6.0 to 9.0 are OK. If outside this
range adjust to pH 6.0 to 7.0 using acid
(0.5N Sulfuric acid) or base (0.5N Sodium
hydroxide).
FILL, DIP, READ

12
©2016, Industrial Test Systems, Inc. Printed in USA
PHOSPHATE
FILL, DIP, READ
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows PO4.
ZERO METER
Press and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
Clean CELL with 0.1N HCl, Distilled Vinegar (5%), or Muriatic
Acid before filling the meter with the sample to be tested.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

13
Sensafe.com
PHOSPHATE
REMOVE STRIP
Remove one eXact
®
Strip Micro PO
4
Phosphate and set
in a dry, convenient place. Replace cap on bottle
5
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip. The display
will automatically start to count up for 120 seconds.
6
READ RESULTS
READ result displayed as Phosphate. This result is
automatically stored in the PO4 menu. After testing is
complete, rinse the sample cell immediately.
7
FILL, DIP, READ

14
©2016, Industrial Test Systems, Inc. Printed in USA
CALCIUM HARDNESS
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows CA5.
ZERO METER
Press and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
4
FILL, DIP, READ
In bright light situations use cell cover for zeroing and reading.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

15
Sensafe.com
CALCIUM HARDNESS
REMOVE STRIP
Remove one eXact
®
Strip Micro CA Calcium
Hardness and set in a dry, convenient place. Replace
cap on bottle.
5
READ RESULTS
READ result displayed as Calcium Hardness
(CaCO3). This result is automatically stored in the CA5
menu. After testing is complete, rinse the sample cell
immediately.
7
FILL, DIP, READ
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6

16
©2016, Industrial Test Systems, Inc. Printed in USA
CYANURIC ACID
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows CY6.
ZERO METER
Place cell cover on sample cell before zeroing.
Press and the LCD Display reads 0PPM,
indicating the meter is ready for testing.
4
FILL, DIP, READ
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

17
Sensafe.com
CYANURIC ACID
SHAKE BOTTLE & ADD DROPS
Tilt the meter to discard about 0.2mL water to leave
room for liquid reagent. Shake the bottle of eXact
®
Reagent CY to mix the suspension in the bottle. Add
5 drops of the eXact
®
Reagent CY to the sample cell.
Keep reagent bottle vertical while adding drops.
5
READ RESULTS
READ result displayed as Cyanuric Acid. This result is
automatically stored in the CY6 menu. After testing is
complete, rinse the sample cell immediately.
7
FILL, DIP, READ
CAP & PRESS READ
Place the cell cover onto the sample cell. Press
to initiate a 20 second countdown. Place thumb or
finger over cap, and mix the sample by turning the
meter upside-down repetitively during the 20 second
countdown. When timer displays 1, place the meter on a
flat surface (upright). The display will automatically start
to count up for 60 seconds.
6

18
©2016, Industrial Test Systems, Inc. Printed in USA
COPPER
FILL, DIP, READ
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact®EZ.
3
SELECT TEST
Press and re-press the button until the
display shows CU7.
ZERO METER
Press and the LCD Display reads 0.00PPM,
indicating the meter is ready for testing.
4
In bright light situations use cell cover for zeroing and reading.
TIP
FILL CELL
Rinse the cell 3 times with the water sample to
be tested and FILL to the top to begin test.
2
Use the included cleaning brush to clean the cell before/
after each test to prevent reagent interferences and
achieve the most accurate result.
TIP

19
Sensafe.com
COPPER
REMOVE STRIP
Remove one eXact
®
Strip Micro CU Copper and set in a
dry, convenient place. Replace cap on bottle
5
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously DIP the eXact® Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip. The display
will automatically start to count up for 120 seconds.
6
READ RESULTS
READ result displayed as Copper. This result is
automatically stored in the CU7 menu. After testing is
complete, rinse the sample cell immediately.
7
FILL, DIP, READ

20
©2016, Industrial Test Systems, Inc. Printed in USA
CHLORIDE (Salt)
1
PREPARE SAMPLE
Prepare sample for testing using Mini Dilution Kit
II #487202.
1. Rinse 50mL graduated conical tube with
distilled or deionized (salt-free) water.
2. Rinse the 3.0mL syringe with saltwater sample
to be tested. Finally, fill the 3.0mL syringe to the
2.0mL line very precisely (plunger ring should
line up at the 2.0mL line and little or no air
bubble should be present).
3. Add the syringe content (2.0mL salt water
sample) to a clean 50mL graduated conical
tube by pushing plunger all the way down to
expel sample.
4. Now, fill the graduated conical tube to the 40mL
line with distilled or deionized (salt-free) water.
Cap graduated conical tube.
5. Mix content of graduated conical tube by turning
up side down at least three times. 1:20 Dilution
Sample is ready for testing.