EN Error Codes
16
Wile moisture meter is equipped with error messages to guide the user to get the
most reliable results, as well as to ensure that the meter is in good condition.
Error code Cause Solution
Critical Error The meter must be submitted to an autho-
rized service center
7Weight not detected Repeat test / Do weight test
8Taring result is too high Do weight test and check the scale
9Weight measurement uncertain /
meter shook during the measure-
ment
Keep the meter stable,
measurement uncertainty has increased
10 Moisture measurement result stabi-
lize slowly Happens to very wet samples
11 Weight measurement too uncertain
/ meter shook during the measure-
ment
Keep the meter stable, measurement uncer-
tainty has increased
12 Temperature of the sample lower
than 2 C / Measurement uncertainty
is high
The tester is most accurate when Rice and
tester temperature is between 16C and 32C
13 Temperature of the sample above
50 C / Measurement uncertainty is
high
The tester is most accurate when Rice and
tester temperature is between 16C and 32C
14 Large temp. diff. between the Rice
and test cell / Measurement uncer-
tainty has increased
For better accuracy, Rice and meter test cell
should be at similar temperatures
15 Not on a suffciently level surface,
meter warns with sound alarm Keep meter stable during measurement
16 Weight result is over 330 g Check that leveling handle did not stay
above the test cell/ Do weight test
17 Weigh result is under 50 g Check that leveling handle did not stay
above the test cell/ Do weight test
18 Hectoliter weight is 20% higher than
average
Measurement uncertainty has increased, Do
weight test
19 Hectoliter weight is 20% lower than
average
Measurement uncertainty has increased,
Do weight test
20 Test cell is not cleared before the
measurement Clear test cell