F-750 Instruction Manual rev. 3/26/2014
1554 NE 3rd Ave, Camas, WA 98607, USA Phone (360) 833-8835 Fax (360) 833-1914
sales@felixinstruments.com http://www.felixinstruments.com
GPS Receiver: The GPS receiver menu shows the state of the GPS sensor as enabled or disabled. If the
GPS sensor is enabled, View Status will appear with the following parameters: latitude, longitude,
satellites in view and if a satellite lock is acquired.
Calibrate Instrument: If the Calibrate Instrument menu is selected, a warning will appear that
calibration should only be performed by authorized users. Press right to continue. Press left to abort.
Instruments should only be calibrated under direction of a Felix Instruments technician. The pixel
coefficients are the spectrometer calibration values given by manufacturer (C0, C1, C2, C3). The pixel
coefficients influence what wavelengths of light the spectrometer is responding too. These values are
instrument specific.
In the Calibrate Instrument menu, the scans to average allows the user to select the number of scans
(measurements) that are average together before a value is shown on the display. The options are: 1, 2,
5, 10, 20, 40, 80, or 160. The integration time for the calibration-collect spectra can be set to the
following: 50, 75, 100, 150, 250, 350, 450, or 500. The spectrometer zero offset can be set at: 0.3, .035,
0.4, 0.45, 0.5, 0.55, 0.6, or 0 .65. The lamp state option is used to turn the lamp on or lamp off. Press the
down/up to turn off/on. The lamp automatic gain, which is used to fine-tune how much the logic has
control of the gain, can be set to: 0.1, 0.2, 0.3, 0.4, 0.5, 0.7, 0.8, or 1.0. The shutter state works similar to
the lamp state. The options are shutter closed or shutter open (press down/up to open/close shutter).
The collect spectra option will use whatever values are set in calibration menu to collect spectra. After
rendering, the results appear graphically on the display (y=0-100%, x=nm raw reflectance).
Measurement Accuracy: The setup measurement accuracy option is used in conjunction with the white
Teflon calibration standard. Place the white Teflon calibration standard over lens when the F-750
prompts to put on Teflon block. Next the display will read “storing dark measurements,” “storing
reflectance measurements,” and finally show the standard deviation.
Language: The language on the F-750 can be set to English or Portuguese.
Factory Reset: The factory reset will erase the temporary cache files for the display screens and reset
the defaults from manufacturer.
F-750 Data
Data collected by the F-750 can be reviewed on the instrument by using the “Browse Records” menu.
The data is stored on the removable SD card. To remove, press into the unit and the card will eject. To
replace, insert metal side first with the label facing the back of the unit. In order to view the files on a
computer, an optional software, such as a command line tool or GUI (Graphical User Interface), is
required.