Electrical Measurement Module Kit with Docking Station
Specifications
13
Specifications
Operating Temperature .......................... 15 °C to 35 °C
Storage Temperature .............................. -20 °C to +70 °C
Electrical Measurement Module (EMM)
Connection ......................................... Banana Jack, CAN bus, RS232
Maximum 30 V dc
HART Communication and Functions
HART Modes ...................................... HART mA measurement with 24 V (Loop)
Automatic HART Detection ................. HART connected with automatic polling
HART Selectable Resistor .................. Built-in 250 Ω loop resistor. Selectable ON/OFF
HART Commands .............................. HART universal and common practice commands (no device specific commands)
Write Protection .................................. HART Write enable/disable
EM300 Modules
Specifications are valid from 18 °C to 28 °C. For temperatures from 15 °C to 18 °C and 28 °C to 35 °C, add 0.002 % FS/°C.
EM300 Module Measurement Specifications
All outputs are positive only.
DC Voltage
Range Resolution 1 Year Instrumental Uncertainty
30 V 1 mV 0.01 % of reading + 2 mV
DC Current
Range Resolution 1 Year Instrumental Uncertainty
24 mA 1 μA 0.01 % of reading + 2 µA
Operating Humidity ................................. 5 % to 95 % relative humidity, non-condensing
Weight ..................................................... 1080 g (2.4 lb)
Dimensions ............................................. 165 x 240 x 85 mm (6.5 x 9.5 x 3.4 in)
Safety
General ............................................... IEC 61010-1
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty
does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal
conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain
service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information,
then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may
not apply to you.