Fluke i430-Flexi-TF-II User manual

PN 4426307 (4822 872 30932 – Rev.3)
December 2011, Rev.3, 9/15
© 2011-2015 Fluke Corporation. All rights
reserved. Specifications are subject to change
without notice. All product names are
trademarks of their respective companies.
i430-Flexi-TF-II
6000 A Flexible
AC Current Probe
Instructions
Introduction
The i430-Flexi-TF-II (the Probe or Product) is a flexible AC current probe
for use with the Fluke 430 series I and II Power Quality Analyzers and
other dedicated test tools. The i430-Flexi-TF-II is optimized for current
measurement on thick and hard to reach conductors. Check the user
documentation of your test tool for compatibility before you use this
current sensor.
Available versions are:
•i430-Flexi-TF-II 61 cm (24 inch)
•i430-Flexi-TFII-48 122 cm (48 inch)
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
•Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
•Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
•Canada:
1-800-36-FLUKE (1-800-363-5853)
•Europe: +31 402-675-200
•Japan:
+81-3-6714-3114
•Singapore:
+65-6799-5566
•Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
To locate an authorized service center, go to www.fluke.com.

Safety
Read this section carefully. It will familiarize you with the most important
safety instructions for handling the Product.
A Warning identifies hazardous conditions and procedures that are
dangerous to the user.
A Caution identifies conditions and procedures that can cause damage
to the Product or the equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
•Read all safety information before you use the Product.
•Carefully read all instructions.
•The Probe is only to be used and handled by qualified
personnel.
•Do not work alone.
•Always connect the Product to the test tool before it is installed
around a conductor.
•Use the Product only as specified, or the protection supplied by
the Product can be compromised.
•Comply with local and national safety codes. Use personal
protective equipment (approved rubber gloves, face protection,
and flame-resistant clothes) to prevent shock and arc blast
injury where hazardous live conductors are exposed.
•Limit operation to the specified measurement category, voltage,
or current ratings.
•De-energize the circuit or wear personal protective equipment in
compliance with local requirements before you apply or remove
the Product.
•Do not touch voltages >30 V ac, 42 V ac peak, or 60 V dc.
•Do not use the Product around explosive gas, vapor, or in damp
or wet environments.
•Do not use the Product if it operates incorrectly.
•Do not use a current measurement as an indication that a circuit
is safe to touch. A voltage measurement is necessary to know if
a circuit is hazardous.
•Do not use the Product if it is damaged.
•Disable the Product if it is damaged.
•Do not use the Product if it has damaged insulation,
exposed metal, or if the wear indicator is visible. Do not use
if the flexible jaw wear indicator shows wear indicated by a
contrasting color on the inner insulation. Examine the
output lead for damaged insulation, exposed metal, or wear
indication.
•Before each use, examine the Product. Look for cracks or
missing pieces of the clamp housing or output cable
insulation. Also look for loose or weakened components.
Carefully examine the insulation around the jaws.

•Do not exceed the Measurement Category (CAT) rating of
the lowest rated individual component of a Product, probe,
or accessory.
•High voltages and currents can be present in adjacent
circuits under test.
•Do not use the Product above its rated frequency.
•Use the Product only on insulated conductors. Use caution
around bare conductors or bus bars. Do not touch the
conductor.
Symbols
The table below is a list of the symbols used on the Probe and in the
instructions.
Symbol Description
WWARNING - RISK OF DANGER. Consult user
documentation.
XWARNING. HAZARDOUS VOLTAGE. Risk of
electric shock.
-Do not apply around or remove from uninsulated
hazardous live conductors without taking additional
protective measures.
TDouble Insulated
Measurement Category II is applicable to test and
measuring circuits connected directly to utilization
points (socket outlets and similar points) of the low-
voltage MAINS installation.
Measurement Category III is applicable to test and
measuring circuits connected to the distribution part
of the building’s low-voltage MAINS installation.
Measurement Category IV is applicable to test and
measuring circuits connected at the source of the
building’s low-voltage MAINS installation.
PConforms to European Union directives.
)Certified by CSA Group to North American safety
standards.
Conforms to relevant Australian safety and EMC
standards.
~
This product complies with the WEEE Directive
marking requirements. The affixed label indicates
that you must not discard this electrical/electronic
product in domestic household waste. Product
Category: With reference to the equipment types in
the WEEE Directive Annex I, this product is classed
as category 9 "Monitoring and Control
Instrumentation" product. Do not dispose of this
product as unsorted municipal waste.

Instrument Compatibility
The i430-Flexi-TF-II series is used with the Fluke 430-series and
Fluke 430-II series Power Quality Analyzers specifically or with other test
tools that explicitly state compatibility. Refer to user documentation of the
test tool for information about compatibility.
Maximum measurement range on all models is 6000 A.
Minimum measurement range on Fluke 433/434/435 is 30 A.
Unpack
These items are included in the shipment:
•i430-Flexi-TF-II Current Probe(s)
•Instructions (this paper)
Check the contents of the shipment for completeness. If something in the
box is damaged or missing, contact your distributor or the nearest Fluke
sales or service office.
How to Use the i430-Flexi-TF-II
To use the i430-Flexi-TF-II, follow these instructions:
1. Connect the BNC connector of the i430-Flexi-TF-II to the input on
the Fluke 430 series Power Quality Analyzer.
2. Wear protective equipment or de-energize the circuit before you
connect the flexible probe around the conductor. See Figure 1.
C
B
A
1/31/31/3
D
hcf157.eps
Figure 1.
3. Re-energize the circuit if not energized.
4. Center the conductor perpendicularly inside the flexible probe area.
If this is not possible, an additional measurement error of
±2 % of reading can occur. See Electrical Specifications.
5. Avoid taking measurements close to other current-carrying
conductors if possible.
6. Make sure that the arrow marked on the probe coupling points
toward the correct orientation for correct phase display on the
instrument. See also the setup screen of the instrument.
7. Keep the probe coupling more than 2.5 cm (1 inch) away from the
conductor.
8. Observe the current value and waveform on the instrument display.

9. Observe the safety instructions when you connect and use the
i430-Flexi-TF-II.
10. Wear protective equipment or de-energize the circuit before you
disconnect the flexible probe.
WWarning
To prevent false and misleading readings on the instrument,
the setup for the current probe in the 430-series Power
Quality Analyzer must be the i430 Flex (on Fluke 430 series I)
or i430-Flexi-TF-II (on Fluke 430 series II).
Maintenance
Before each use to assure continued safety, inspect the Probe and its
latching system for any damage. Pay particular attention to the insulation
surrounding the Probe. A Probe under warranty will be promptly repaired
or replaced (at Fluke's discretion) and returned at no charge.
XW Warning
To prevent possible electrical shock, fire, or personal
injury:
•Remove the input signals before you clean the
Product.
•Have an approved technician repair the Product.
The flexible probe and its latching system require no special care.
For routine maintenance:
1. Ensure that no foreign body obstructs the latching mechanism.
2. Clean the Probe with a damp cloth and a mild detergent. Do not use
abrasives, solvents, or alcohol.
3. Do not immerse the Probe in liquids.
If the i430-Flexi-TF-II does not work
If the Probe does not perform properly, use the following steps to help
isolate the problem:
1. Inspect the coupling system for any damage.
If any foreign material is present, the coupling system will not close
properly and errors will result.
2. Inspect the cable between the flexible probe and the Power Quality
Analyzer for any damage.
3. Check if the i430-Flexi-TF-II is the selected current probe on the
Power Quality Analyzer (Clamp i430-Flexi-TF-II).
4. Verify that the function and range selection on the Power Quality
Analyzer is correct.

Specifications
SAFETYWThe i430-Flexi-TF-II complies with:
•IEC 61010-1: Pollution degree 2
•IEC 61010-2-032: CAT IV 600 V / CAT III 1000 V
Electrical Specifications
Measuring range 0.5 A ac to 6000 A ac
Maximum non-destructive
current 100 kA (50/60 Hz)
Output signal 86.6 mV at 1000 A/50 Hz (on 1 MΩ)
85.0 mV at 1000 A/50 Hz (on 50 kΩ)
Basic accuracy ±1 % of reading at 25 °C, 50 Hz with
conductor placed at the center of the probe
Linearity ±0.2 % of reading at 10 % to 100 % of range
Noise <1 mV rms at 10 Hz to 10 kHz
Additional errors:
•temperature coefficient
over operating
temperature range
-for 24 inch model 0.02 % of reading / °C (0.011 % / °F)
-for 48 inch model 0.035 % of reading / °C (0.02 % / °F)
•with position of
conductor in the probe
window (see Figure 1):
Zone A
Zone B
Zone C
•0.5 to 30 A
±0.5 % of reading
±1.0 % of reading
±2.0 % of reading
±1 count or 0.1 A (whichever is greater)
External magnetic field
rejection in reference to
external current
40 dB (with cable >100 mm from the head
coupling and r-coil)
Phase shift (fundamental
frequency) <±1°
Bandwidth (-3 dB) 10 Hz to 10 kHz
Frequency derating I x f ≤385 kA Hz

General Specifications
i430-Flexi-
TF-II TF-II-48
Transducer length 61 cm (24 in) 122 cm (48 in)
Weight 153 g 200 g
Transducer diameter 7.5 mm (0.3 in)
Minimum bending radius 38 mm (1.5 in)
Cable length from
transducer to BNC 2.5 m (98.4 in)
Temperature, operating
-20 °C to +70 °C (-4 °F to +158 °F)
Temperature of conductor under test
shall not exceed 80 °C (176 °F)
Temperature, non-operating -40°C to + 80 °C (-40 °F to +176 °F)
Relative Humidity, operating 15 % to 85 %, noncondensing
Altitude, operating
2 000 m (6 500 ft); for use up to 4000 m
derate to 300 V CAT IV, 600 V CAT III,
1000 V CAT II.
Altitude, storage 12 km (40 000 ft)
Ingress protection IP40
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one
year from the date of purchase. This warranty does not cover fuses, disposable
batteries, or damage from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not authorized to extend
any other warranty on Fluke’s behalf. To obtain service during the warranty period,
contact your nearest Fluke authorized service center to obtain return authorization
information, then send the product to that Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH
AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED.
FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR
THEORY. Since some states or countries do not allow the exclusion or limitation of
an implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
Table of contents
Other Fluke Measuring Instrument manuals

Fluke
Fluke 1730 Use and care manual

Fluke
Fluke 700G Series Use and care manual

Fluke
Fluke 714C User manual

Fluke
Fluke VT Plus HF User manual

Fluke
Fluke 9140 User manual

Fluke
Fluke CableAnalyzer DSP-4100 User manual

Fluke
Fluke i30s User manual

Fluke
Fluke DP120 User manual

Fluke
Fluke i400s User manual

Fluke
Fluke i430-Flexi-TF User manual
Popular Measuring Instrument manuals by other brands

Teledyne Lecroy
Teledyne Lecroy M41h quick start guide

Growatt
Growatt TPM-CT-E-EU quick guide

Blue-White industries
Blue-White industries Sonic-Pro S6A Series quick start guide

Tektronix
Tektronix TLA7000 Series installation manual

Circutor
Circutor DCP-96 Aac instruction manual

Sealey
Sealey DBG5010.V2 quick start guide