Fluke 80i-410 User manual

-.
I .
,
..
INTRODUCTION
The Fluke 80i-410
DC/AC
Current Probe is a battery-powered
· clamp-on current probe that is used with a voltmeterto measure de
oraccurrentsupto400 amps. Each jawofthe probecontai
ns
halfof
a magnetic core. The jaws clamp around a conductor allowing
current measurements without breaking the circuit. A Hall sensor
t-
---=--
-11
~.1nt.tedao.uiinu:c:iµg::t.taat:p~in--tbe
~n~
_
nu:ignoticfialdproducad
by
the currentbeing measured.
An
amplifiercircuitgenerates
an
output
signal of 1
mV
per amp. Demagnetizing (degaussing) is accom-
plished
by
manually snapping the jaws.
SPECIFICATIONS
ELECTRICAL
Specified Current Range: 5 to 400A de or ac
Usable Current Range: 2 to 600A de or ac
Output Signal: 1
mV
per amp de or ac
Accuracy (uncertainty): ±
(5%
.of reading + 2.
5A
),
5 to 400A, de
to
62
Hz, for 1 year, 18 to
28
°C (
64
to
82°F), probe demagnetized, conductor
centered
Frequency Response: -
7%
at 400
Hz,
-3 dB at 1200
Hz
typical
Working Voltage: 660V rms maximum at input
Maximum Conductor Sizes: 1
ea
30
mm
(1
.18") diameter
2
ea
25
mm
(0.98
") diameter
Load Impedance: > 1
MO,
100
pf
(The probe's acti
ve
circuit is
capable of driving impedances
as
low as 1
kO
(chart recorders etc.
).
However,the battery test
indicati
on
will be incorrect (low) if the
rr.
r:her
input impedance
is
less than 1
MO.
GENERAL
Temperature Coefficient: ±
(0.1
o/o
+.
2A)
per °C
(0
to 18°,
28
to
50
°)
Operating Temperature: 0 to 50°C
(32
to 122°F)
S'orage Temperature: -40 to 60°C (-40 to 14 .°F/
Relative Humidity: 0 to
90%
(0
to
35
°
C)
0 to
70%
(35
to 50°C)
PIN 795203 JULY 1986
®1986, John Fluke Mfg. Co
..
Inc. All rights reserved. Litho in U.S
.A.
Instruction Sheet
:.·
SOi-410
DC/AC Current Probe
Batteries:
QUANTITY/TYPE: two 1.5V AAA
LIF.E
: 600 hrs typical
.,
LOW INDICATION: Output voltage is between 0.0 and -
99
.9
mV
with the power switch off (load
impedance >1
MO)
VOLTMETER COMPATIBILITY
The80i-41O
DC/
AC
CurrentProbe is compatible with any voltmete
r,
multimeter, or other voltage-measuring instrument that has
th
e
following features:
• InputjacksthatacceptFlukesafety-designed shrouded
bararia plugs.
• Rang
"'
and resolution capable of displaying 1
mV
of
output per amp of measured
curre11t.
• Voltmeter accuracy (uncertainty) of 0.75% or better to
take full advantage of the accuracy of the probe.
• Input impedance of 1 MO, 100
pF
or greater.
(Se
e the
Load Impedancespecifications for operation with lower
impedance instn
1T1
en
ts.)
When the probe is making
c.
measurement, the current-carrying
conductor is not brokan and remains electrically isolated from the
probe's output.
As
a result, the probe's output common may be
either floated (isolated) or gr
C'unde
ri.
BATTERY INSTALLATION
The
80
i-410
DC/AC
Current Probe requires two 1.
5V
, AAA size
batteries for operation. Install the batteries as follows:
1. Use your thumbnail to raise the latch and open the
battery compartment (see
FiQure
1)
.
2.
Observe the polarity man
-:
ings on the inside of the
batterydoor. Notethatthe polarityforeach batterv isthe
reverse of the other.

3. Install the batteries
so
that the polarity markings
on
the
batteries match the markings
on
the battery door.
4.
Snap the battery door shut.
NOTE
If
both batteries are not installed with the correct
polarity, thebatterytestmayindicate thatthebatteries
are good, but the probe will not operate.
OPERATION
Battery Test
' \ I I
CT]
I/
\ \
{
-9~mv
} •
~
00.0mV
O.K.
~
~~
+
Figure 1.
Use the following procedure to test the batteries:
1.
Connectthe probe output leads to the voltmeter inputs:
black to COMMON or
r.
ow
,and red to
VetT
.
2.
Setthe voltmeterfunction to
de
millivolts
(de
mV==),
and
turn the voltmeter power
on.
3.
Select a range capable of displaying 170
mV
de.
4.
Verify that the power switch on the probe is in the
OFF
position.
5.
Observe the voltmeter reading.
If
the reading falls
between -100.0 and -170.0
mV,
the probe is ready for
use.
(New batteries should read approximately -150
mV.)
If the reading falls between
0.0
and -99.9
mV,
the
batteries should be replaced. If the voltmeter reading is
positive, either the test leads are reversed or the
batteries are installed incorrectly.
Demagnetizing
When the 80i-410
DC/
AC
Current Probe has been used for a
currentmeasurementand then removed from the conductor,asmall
amo_unt
of residual magnetism often remains
in
the core. This
residual magnetism will generate a de offset at the probe output.
In
otherwords, the voltmeter will show asmall de reading even though
there is no current passing through the jaws of the probe. The
residual magnetism should not cause a problem for ac current
measurement because the ac voltage function
in
most multimeters
is ac coupled; therefore, the de offsetwill not
be
displayed. However,
for de current measurements, the
de
offset caused by the residual
magnetism will contributeto reading errors. The residual magnetism
can
be
minimized
as
follows
(see
Figure
2):
1.
Use
the thumb and forefinger of one hand to grasp the
probe
in
the middle where the two screw heads are
exposed.
2. With your other hand, compress the handles together,
then release them suddenly. The jaws should come
together with a snapping action.
3.
Repeat the above procedure twice.
Figure 2.
NOTE
The
snapping action will not damage the jaws
if
the
mating surfaces are kept clean. ·
Alternate Demagnetizing Procedures
Ifaconductor carrying ac current is available, clampthe jawsofthe
probe around the conductor, wait a few seconds, then slowly
remove the probe.
If
a conductor carrying de current is available,
clampthe jawsofthe probe around the conductor.Then removethe
probe and clamp itaround the conductor again so that the current
flows through the jaws
in
the opposite direction.
Making Measureme·nts ·
Use the following procedure to make current measurements
(see
Figure
3):
~
~
..CC).Qll~~~)ioiba-.OU:~lt.J.i~~~~~~~:..&Apw.I~
~~~-
(black to COMMON or
LOW
and
red
to
VOLTS)
.
2.
Selectthe appropriateac orde function on thevoltmeter.
3.
Turn on the voltmeter and select a voltage range to
display the expected value
(1
mV
per
amp).
4.
Turn the probe ON/OFF switch located on the probe
handle to the
ON
position.
5.
Observe the voltmeter reading. If necessary, demag-
netize the probe.
6.
Clampthejawsofthe probe around the conductorto
be
measured.Position the probeso thattheconductoris
in
the center of the jaw opening for the most accurate
reading.
7.
Observe the voltmeter display. The reading
in
millivolts
is equivalent to the current
in
amps
(i.e.,
5
mV
=SA). A
positive polarity reading indicates thatthe de current is
flowi
ng
in
th
e direction of the arrow
on
the probe.
Figure 3.

Low Level Measurement
The 80i-41O
DC
/
AC
CurrentProbe isspecified to measurecurrents
of
SA
orgreate
r.
Currents below
SA
can
be
accurately measured
by
looping the inputwirethrough thejawsofthe probe so thatthe sum
of the amp-turns through the jaws
is
greater than
SA.
The actual
current can then
be
calculated
by
dividing the
me
~
er
reading by the
number of turns looped through the jaws.
TROUBLESHOOTING
If
the 80i-
41
O
DC/
AC
Current Probe doesnotperform properly, use
the following check steps to help isolate the problem:
1.
Inspect the jaw mating surface for cleanliness.
If
any
foreign material is present, the jaws will not close
properly and errors will result.
2.
Visually inspect the batteries to make sure thay are
installed with the proper polarity.
(Re
fer to the Battery
Installation procedure.)
3. Test the batteries.(Refer to the Battery Test procedure.)
4. Verify that the probe leads are connected to the VOLT
input of the meter. ,
S.
Verifythatthefunction selection on the meter is correct,
i.e.
, de
mV
for de current or ac
mV
for ac current.
CALIBRATION PROCEDURE
Required Equipment
The equipment required to calibrate the 80i-410 is listed in Table
1.
Table
1.
Required
Equipment
- ·
---
-----
·-
--·
--
--
--
0-MOD&L
···-. --.. -
DC/
AC
Voltage Calibrator Fluke Model
S1
OOB
Transconductance Amplifier Fluke Model
S220A
Digital Multimeter Fluke Model
80SOA
Magnet Wire Coil 15
tu
.rns of
#1
Osingle film coated
copper wire diameter 1S" (380
mm)
minimum
Small insulated screwdriver
Calibration Procedure
Use the following procedure to calibrate the 80i-410:
1. Connect the
DC/
AC
Voltage Calibrator output to the
Transconductance Amplifier input.
2. Turn both instruments
on
and allow sufficient warm-up
time for stable operation.
3. Connectthe 80i-410 to the Digital Multimeter input and
perform the Battery Test procedure given under the
heading OPERATION of this instruction sheet.
4. Slide the 80i-410 power switch to the
ON
position.
5.
Perform the demagnetizing procedure given under the
heading OPERATION ofthisinstruction sheet. Snap the
jaws at least three times to ensure the residual mag-
netism is as low as possible.
NOTE
The
probe should be at room temperature before
proceeding to the next step.
6.
Locate the
ZERO
adjustment in the end of the probe
handle Figure
4).
Use
an
insulated screwdriverto adjust
this control fora multimeter reading as close
as
possible
to
0.0
mV.
7. Set the calibrator and amplifer controls so that the
calibrator delivers 8.00V de to the amplifier and the
amplifier drives 8.00A de
in
the 15-turn coil.
8.
Clamp the jawsof the probe around the coil. Adjustthe
position of the probe so that the coil passes through the
center of the jaw opening.
9. Locate the
GAIN
adjustment control in the end of the
probe handle (see Figure
4).
Use
an
insulated screw-
driver to adjust this control for a multimeter reading as
close
as
possible to 120.00
mV
de.
10. Set the calibrator and amplifier controls so that the
calibratordelivers 8.00Vacat
60
Hz
to theamplifier, and
the amplifier drives 8.00A ac in the 15-turn coil.
11
. Set the multimeter function switch to
ac.
_
__
_.._
_.
_
2,
~
v....,e...,,.dfyJY'
that,the
multimataEJead1Ag.faUs.betweeB-
1-1:5.00
and 125.
00
mV
ac.
ARROW
FOR
POSITION
CURRENT PROBE
FOR
DC
MEASUREMENTS.
BATTERY
COMPARTMENT
Figure
4.
CLEANING AND STORAGE
ON/OFF
SWITCH
Clean the mating surfaces of the core at regular intervals. Apply a
light coating of dripless oil to preventcorrosion.
If
the probe will not
be used for periods of longerthan
60
days,the batteries should be
removed and stored separately.

REFERENCE
DESIGNATOR
BT1
,BT2
C1
C2,C3
01
R1
R2
R3
R4
RS
R6
R7
RB
R9
S1201,202
U1
U2
01C7
.CfilJ
0
8~~~
!
R7
:
=~
·
1
.
._
__
___
__
-
--
r=B~~
C3
0
r-----
-
-----
-
---
-
--
----------------
--
------
---
----1
2 I
s2e
:
5
+
1.
SV
BT1
R3
3 6
+
1.
SV
BT2
R6
RS
2
U1B
R4
RB
1mA
±0.1 mA
R9
I
I
:
1'
I
I
HI
mV Output
LO
"'-----
-
--------
- -
---
---
--
---
--------------
-
---
--
_J
Replaceable Parts
DESCRIPTION
BATIERY
, 1.5V AAA
1000 pF CAP, 160V, 10% CER, GEB605 FS
220 pF CAP, 500V, 10%, CEA, GEB 605
FS
NPN TRANSISTOR, BC 237 B
200n
RES,
MF, 1%,
YeW,
50 PPM
10
kn,
POTENTIOMETER
59
kn
RES,
MF, 1%,
YeW
, 50 PPM
1
Mn
RES,
MF, 1%,
YeW
, 50 PPM
59
kn
,
RES
, MF, 1%,
YeW
, 50 PPM
1
Mn
,
RES,
MF, 1%,
YeW
, 50 PPM
10
kn
, POTENTIOMETER
4.02
kn,
RES,
MF, 1%,
YeW,
50 PPM
137
kn,
RES, MF, 1%,
.YeW
, 50 PPM
SWITCH, BIPOLAR
(2
POSITIONS) SLS 220
LINEAR AMPLIFIER +VOLTAGE REFERENCE
LM 10 CLN 8
HALL EFFECT CELL
WARRANTY
ONE YEAR LIMITED WARRANTY
Fluke
warrants
your
accessory
to
be
free from defects
in
material
and
workmanship
under
normal use
a~d
service
for
1 YEAR from
date
of
purchase
to
the
original purchaser.
It
does
not
apply
to
batteries
or
fuses
or
when
the
accessory
has been misused, altered
or
damaged
by
accident
or
abnormal
conditions
of
operation. . _
For
warranty
service,send
the
accessory,
with
a
description
of
the
difficulty,
postage
prepaid
,
to
a
Fluke
Service Center.
Fluke
assumes
no
risk
for
damage
In transit.
Fluke
will, at
our
option, repair
or
replace the defective
accessory
free
of
charge.
However,·
if
we
determine
tt;at the failure was
caused
by
misuse, alteration,
accident,
or
abnormal
condition
of
operation,
you
will
be
billed
for
the
repair
. The repaired
accessory
"\Y..
lll
be
returned,
transportation
prepaid
.
FLUKE
MAKES
NO WARRANTY OTHER THAN THE LIMITED WARRANTY
STATED ABOVE.
ALL
WARRANTIES, INCLUDING IMPLIED WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE,
ARE LIMITED TO A PERIOD OF 1 YEAR FORM THE DATEOF PURCHASE.
FLUKE
SHALL
NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT, TORT, OR
OTHERWISE.
Note(USAonly): Somestates
do
not
allow
limitation
of
implied
warranties,
or
the
exclusion
of
I
ncidenta
l
or
consequential
damages,
so
the
above
limitations
or
exclusions may
not
apply
to you. This warranty gives you
specific
legal ri
ghts
and
you
may
have
other
rights
which
vary from state
to
state.
For
application
or
operation assistance
or
information on Fluke products,
call:
800-426-0361 in
most
of
U.S.A.
206-356-5400 from
AK
and
WA
206-356-5500 from other countries
John
Fluke Mfg. Co.,
Inc
P.O. Box C9090
Everett, WA 9820e
Other Fluke Measuring Instrument manuals

Fluke
Fluke 945 User manual

Fluke
Fluke i100 User manual

Fluke
Fluke 7 Series User manual

Fluke
Fluke 368 User manual

Fluke
Fluke ScopeMeter 225C User manual

Fluke
Fluke 741B User manual

Fluke
Fluke 2700G User manual

Fluke
Fluke 481 User manual

Fluke
Fluke DSX-600 CableAnalyzer User manual

Fluke
Fluke 5790A User manual

Fluke
Fluke 789 Owner's manual

Fluke
Fluke a3001 FC User manual

Fluke
Fluke 381 Use and care manual

Fluke
Fluke 1620A DEWK User manual

Fluke
Fluke 971 User manual

Fluke
Fluke SigmaPace 1000 User manual

Fluke
Fluke i1000s Configuration guide

Fluke
Fluke CableAnalyzer DSP-4000 Series User manual

Fluke
Fluke CNX a3000 Use and care manual

Fluke
Fluke CO-220 User manual