manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Fluke
  6. •
  7. Multimeter
  8. •
  9. Fluke 8010A User manual

Fluke 8010A User manual

:
|
8010A/8012A
P/N
491944
August
1978
Rev
2
1/85
@
1985,
John
Fluke
Mfg.
Co,,
Inc.
All
rights
reserved.
Litho
In
U.S.A.
Digital
Multimeters
Instruction
Manual
—
LF
ters
ime:
ital
Multi
igi
8010A/8012A
Di
vi
Section
1
Introduction
and
Specifications
1-1.
INTRODUCTION
1-2.
This
manual
contains
complete
operating
and
maintenance
instructions
for
the
Fluke
Models
8010A
and
8012A
Digital
Multimeters.
The
information
presented
in
this
manual
reflects
both
instruments
except
where
indicated
by
a
particular
model
number.
The
term
Multimeter
is
used
throughout
this
manual
to
indicate
both
the
8010A
and
the
8012A.
1-3.
The
Multimeters
are
portable
bench-type
digital
multimeters
(DMMs)
with
3-1/2
digit
liquid
crystal
displays
(LCDs).
The
Multimeters
have
the
following
industry-standard
features:
e
Voltage
measurements
from
100
pV
to
1000V
dc
and
10
mV
to
750V
true-rms
ac.
e
Current
measurements
from
100
nA
to
2A
deand
10
HA
to
2A
true-rms
ac
(up
to
10A
for
the
8012A).
e
Resistance
measurements
from
100
mN
(1
mQ
for
the
8012A)
to
20
MQ.
The
Multimeters
also
have
the
following
special
measurement
features:
e
Conductance
measurements
up
to
10,000
MQ
of
equivalent
resistance.
e
Resistance
ranges
that
supply
enough
voltage
to
turn
on
a
PN
junction
for
testing
diodes
and
transistors.
@
Automatic
polarity
indication
and
overrange
indication.
°
Protection
from
overloads
and
transients
up
to6
kV
for
10
microseconds.
©
Dual-slope
integration
a/d
conversion
to
ensure
noise-free
measurements.
e
Long
term
calibration
stability
(1
year).
14.
Your
Multimeter
is
warranted
for
a
period
of
one
year
upon
shipment
of
the
instrument
to
the
original
purchaser.
Conditions
of
the
warranty
are
given
at
the
front
of
this
manual.
1-5.
OPTIONS
and
ACCESSORIES
14.
The
use
of
your
Multimeter
can
be
enhanced
by
the
options
and
accessories
available
for
these
instruments.
The
options
and
accessories
are
listed
in
Table
1-1.
The
Multimeters
can
be
ordered
with
the
Option
8010A-01
or
8012A-01
Rechargeable
Battery.
Detailed
information
on
options
and
accessories
is
contained
in
Section
6
of
this
manual,
1-7.
SPECIFICATIONS
1-8.
Specifications
for
the
Multimeters
are
listed
in
Table
1-2.
Specifications
for
the
Option
8010A-01
or
Option
8012A-01
and
other
accessory
specifications
are
given
in
Section
6
of
this
manual.
INTRODUCTION
AND
SPECIFICATIONS
SPECIFICATIONS
Table
1-1.
8010A/8012A
Options
and
Accessories
INPUT
IMPEDANCE
MODEL
DESCRIPTION
MODEL
DESCRIPTION
C86
Ruggedized
Carrying
Case
80J-10
Current
Shunt
Y8205
Soft
Carrying
Case
80K-6
High
Voltage
Probe
M00-200-611
Offset
Mounting
Kit
80K-40
High
Voltage
Probe
M00-200-612
Center
Mounting
Kit
83-RF
RF
Probe
MO0-200-613
Dual
Mounting
Kit
85-RFE
RE
Probe
80T-H
Touch-Hold
Probe
Y8100
DC/AC
Current
Probe
80T-150C
Temperature
Probe,
Celsius
Y8101
AC
Current
Transformer
80T-150F
Temperature
Probe,
Fahrenheit
Y8133
Deluxe
Test
Lead
Set
80i-400
AC
Current
Probe
Y8140
Slim-Flex
Test
Leads
80i-600
AC
Current
Probe
8010A-01
Rechargeable
Battery
Option
for
8010A
8012A-01
Rechargeable
Battery
Option
for
8012A
Table
1-2.
8010A/8012A
Specifications
ELECTRICAL.
ss 2s
esis
iia
See
S55
swe
eats
The
electrical
specifications
given
apply
for
an
operating
temperature
of
18°C
to
28°C
(64.4°F
to
82.4°F),
relative
humidity
up
to
90%,
and
a
1-year
calibration
cycle.
Functions
.......:eccescesceveeetecetececn
DC
Volts,
AC
Volts,
DC
Current,
AC
Current
Resistance
and
Conductance.
DC
Volts
[
RANGE
RESOLUTION
ACCURACY
for
1-Year
£200
mV
100
uV
+2V
1mvV
+20V
10
mV
£(0.1%
of
reading
+
1
digit)
+200V
100
mV
+1000V
1V
NORMAL
MODE
REJECTION
RATIO
10
MQ,
all
ranges.
>90
dB
at
de,
50
Hz
and
60
Hz.
.
>60
dB
at
60
Hz
(at
50
Hz
with
50
Hz
Option).
COMMON
MODE
REJECTION
RATIO
....
(1
kQ
unbalanced)
AC
Volts
(True
RMS
Responding)
OVERVOLTAGE
PROTECTION
....
RESPONSE
TIME
.......-..ecse
eee
se
iaastidee
1
second.
10Q00V
dc
or
peak
ac
on
all
ranges.
ACCURACY
for
1-Year
(5%
of
Range
to
Full
Range)
RANGE
RESOLUTION
45
Hz
10
kHz
20
kHz
:
to
1
kHz
to
10
kHz
|
to
20
kHz
to
50
kHz
200
mV
100
nV
+
(5%of
2vV
1
mV
+.(1,0%
of
reading
+
3
20V
10
mV
+(0.5%
of
reading
+
2
digits}/
reading
+
2
digits)
LL
200V
O.1V
t
digits)
+
(0.5%
of
FE
Se
750V
WV
reading
+
2
|
digits)
L
INTRODUCTION
AND
SPECIFICATIONS
SPECIFICATIONS
Table
1-2.
8010A/8012A
Specifications
(cont)
AC
Volts,
continued
VOLT-Hz
PRODUCT
oe
eee
eee
eee
eee
107
max
(200V
max
@
50
kHz).
EXTENDED
FREQUENCY
RESPONSE
...
Typically
+3 dB
at
200
kHz.
COMMON
MODE
NOISE
REJECTION
RATIO
(1
kQ
unbalance)
...............0,
>60
dB
at
50
Hz
and
60
Hz.
CREST
FACTOR
RANGE
.............055
1.0
to
3.0.
INPUT
IMPEDANCE
.........c..cesee
eee
10
MQ
in
parailel
with
<100
pF.
OVERLOAD
PROTECTION
750V
rms
or
1000V
peak
continuous
not
to
exceed
the
volt-hertz
product
of
10’
(except
10
seconds
maximum
on
200
mV,
2V
ranges).
RESPONSE
TIME...
eee
cece
eee
e
eee
2
seconds
maximum
within
a
range.
DC
Current
RESOLUTION
ACCURACY
for'1-Year
BURDEN
VOLTAGE
+(0.3%
of
reading
+1
digit)
0.3V
max.
0.9V
max.
OVERLOAD
PROTECTION
...........065
2A/250V
fuse
in
series
with
3A/600V
fuse
(for
high
energy
sources).
AC
Current
(True
RMS
Responding
AC
Coupled)
—
=
ACCURACY:
from
5%
of
range
to
full-scale,
1-Year
“
BURDEN
E
RESOLUTION
BANG
Orr
VOLTAGE
200
pA
0.1
pA
t
2mA
1
uA
‘
.
2
£(1%
of
reading
+2
digits)
#(2%
of
reading
0.3V
rms
max.
20
mA
10
uA
;
+2
digits)
200
mA
100
pA
2000
mA
1mA
£(1%
of
reading
'
0.9V
rms
max.
+2
digits}
OVERLOAD
PROTECTION
..............
2A/250V
fuse
in
series
with
3A/600V
fuse
(for
high
energy
sources).
:
CREST
FACTOR
RANGE
........-...0005
1.0
to
3.0
High
Current
8010A
Only
ACCURACY:
for
1-Year
BURDEN
VOLTAGE
10A
de
+(0.5%
of
reading
+
1
digit)
0.5V
max.
10A
45
Hz
to
2
kHz
Trms
ac
One
+(1%
of
reading
+
2
digits}
DSON
EMS
CaN:
OVERLOAD
5.
posse
a8
5h
deca
ec
lnjo
bin
gueonns
ttewe
deat
12A
maximum
unfused.
1-3
INTRODUCTION
AND
SPECIFICATIONS
SPECIFICATIONS
Table
1-2.
8010A/8012A
Specifications
(cont)
1-4
aa
a
re
Resistance
FULL-SCALE
MAXIMUM
TEST
!
:
‘
RESOLUTION
ACCURACY:
for
1-Year
VOLTAGE
CURRENT
2kQ
wt
12
+(0.2%
of
reading
>
1.0V
1.3mA
+1
digit)
<0.25V
10
pA
200
kKQ
~ph
>
1.0V
35
uA
2000
kQ
+(0.5%
of
reading
<0.25V
0.10
A
+1
digit)
0.35
uA
OVERLOAD
PROTECTION
..............
300V
dc/ac
rms
on
all
ranges.
OPEN
CIRCUIT
VOLTAGE
...............
Less
than 3.5V
on
all
ranges.
RESPONSE
TIME
2.2...
ccc
eee
eee
eee
1
second
all
ranges
except
4
seconds
on
the
2000
kQ
and
20
MQ
ranges.
DIODE
TEST
-cicii
sie
sie’
su
iwitle
ate
ktnnes
These
three
ranges
have
enough
voltage
to
turn
on
silicon
>
junctions
to
check
for
proper
forward-to-back
resistance.
The
2
kQ
range
is
preferred
and
is
marked
with
a
larger
diode
symbol
on
the
front
panel
of
the
instrument.
The
three
non-diode
test
ranges
will
not
turn
on
silicon
junctions
so
in-circuit
resistance
measurements
can
be
made
with
these
three
ranges.
Low
Resistance
8012A
Only
FULL-SCALE
MAXIMUM
RANGE
RESOLUTION
|
ACCURACY:
for
1-Year
VOLTAGE
TEST
CURRENT
£(1%
of
reading
+2
digits)
ie
pee
202
(LO
Ohms}
+(0.5%
of
reading
+2
digits)
OVERLOAD
PROTECTION
..........-...
300V
de/ac
rms
on
ail
ranges.
RESPONSE
TIME...
«0...
ese
e
eee
eee
eee
1
second
maximum.
OPEN
CIRCUIT
VOLTAGE
............055
16V
maximum
on
both
ranges.
Conductance
OPEN
MAXIMUM
RESOLUTION
|
ACCURACY:
for
1-Year
CIRCUIT TEST
VOLTAGE
CURRENT
+(0.2%
of
reading
20
uS
10
nS
+
1
digit)
+(1%
of
reading
.
|
200
nS
i,
.1nS
Zz
+10
digits}
<1V
0.10
uA
OVERLOAD
PROTECTION
............4.
300V
dce/ac
rms
on
ail
ranges.
,
CONDUCTANCE
UNITS
........--
eevee
We
use
the
international
unit
of
conductance,
the
Siemen
=
S
=
1/Q.
Another
unit
of
conductance
is
the
mho.
Table
1-2
ENVIRONMENTAL
Temperature
Coefficient
Operating
Temperature
RELATIVE
HUMIDITY
Storage
Temperature
RELATIVE
HUMIDITY
GENERAL
Maximum
Common
Mode
Voltage
Power
Requirements
Weight
LINE
MODEL
.....
cece
eee
e
eens
BATTERY
MODEL
INTRODUCTION
AND
SPECIFICATIONS
;
SPECIFICATIONS
.
8010A/8012A
Specifications
(cont)
<0.1
times
the
applicable
accuracy
specification
per
°C
for
0°C
to
18°C
and
28°C
to
50°C
(32°F
to
64.4°F
and
50.4°F
to
122°F).
O°C
to
50°C
(32°F
to
122°F).
Oto
80%,
0°C
to
35°C
(32
to
95°F)
on
2000
kQ,
20
MQ
and
200
nS
ranges.
O
to
80%,
0°C
to
35°C
(32
to
95°F)
on
all
other
ranges.
0
to
70%,
35°C
to
50°C
(95
to
122°F).
(without
batteries):
-40°C
to
+60°C
(-40°F
to
+158°F)
(with
batteries):
-40°C
to
+50°C
(-40°F
to
+122°F)
0
to
90%,
-40
to
50°C
{-40
to
+122°F)
0
to
90%,
50
to
60°C
(122
to
140°F).
500V
de
or
peak
ac.
90
to
132V
or
200
to
264V
at
50
Hz
or
60
Hz.
Line
Model
2W.
(Battery
Model
3.5°W.)
(See
Figure
1-1.)
1.08
kg
(2
Ib
6
oz)
1.42
kg
(3
Ib,
6
0z)
8.55
in.
(21,72
cm)
ome
in.
(19,05
=
9.65
in.
(27,05
cm)
9.90
in.
(25,15
—
—
2.52
in.
(6,40
cm)
Figure
1-1.
8010A/8012A
Dimensions
1-5/1-6
dg
2-1.
INTRODUCTION
2-2.
This
section
describes
how
to
set
up
and
make
measurements
with
your
Multimeter.
Even
though
you
may
have used
a
multimeter
before,
we
recommend
that
you
read
the
entire
section
carefully
so
that
youcan
use
all
of
the
features
of
your
Multimeter.
2-3.
SETTING
UP
YOUR
INSTRUMENT
2-4.
Unpacking
2-5.
Your
Multimeter
is
shipped
in
a
special
protective
container
that
should
prevent
damage
to
the
instrument
during
shipping.
Check
the
shipping
order
against
the
contents
of
the
container
and
report
any
damage
or
short
shipment
to
the
place
of
purchase
or
the
nearest
Fluke
Technical
Service Center.
A
list
of
these
service
centers
is
located
in
Section
5.
The
container
should
include
the
following:
@
The
8010A
or
8012A
Multimeter
@
Two
test
leads
(one
red
and
one
black)
e
Line
power
cord
e@
The
8010A/8012A
Instruction
Manual
2-4.
Ifreshipment
of
the
instrument
is
necessary,
please
use
the
original
shipping
container.
If
the
original
container
is
not
available,
be
sure
that
adequate
protection
is
provided
to
prevent
damage
during
shipment.
We
recommend
that
the
instrument
be
surrounded
by
at
least
three
inches
of
shock-absorbing
material
on
all
sides
of
the
container.
2-7.
Remove
the
Multimeter
from
its
container
and
place
it
in
a
convenient
location.
The
carrying
handle
on
the
meter
can
be
used
as
a
prop-stand
or
positioned
out
of
the
way
(behind
the
Multimeter).
To
position
the
handle,
pull
outward
on
the
hubs
of
the
handle
and
rotate
the
handle
into
position.
Section
2
Operation
2-8.
AC
Line
Voltage
Requirements
2-9.
AC
line
voltage
requirements
for
your
Multimeter
are
listed
on
a
decal
attached
to
the
bottom
of
the
instrument.
Refer
to
Section
4
for
the
procedure
to
change
the
ac
line
voltage
setting.
If
your
Multimeter
has
the -01
Battery
Option,
refer
to
Section
6
for
information
on
changing
ac
line
voltages.
CAUTION
Bo
not
connect
the
power
cable
to
the
instrument
before
verifying
that
the
intended
source
matches
the
ac
line
configuration
of
the
instrument.
2-10.
Fuse
Replacement
2-11.
There
is
one,
user-replaceable
fuse
(F1)
in
your
Multimeter.
The
fuse
(Fl)
and
the
fuse
holder
form
an
integral
part
of
the
mA
input
connector
and
can
be
removed
with
ordinary
tools.
The
fuse
rating
is:
2A,
normal
blow
(recommended
part
AGX2).
2-12.
Use
the
following
procedure
to
replace
the
fuse,
Fl:
1.
Set
the
POWER
switch
to
OFF.
°
2.
Remove
the
input
power
cord
from
the
Multimeter.
WARNING
DO
NOT
REPLACE
THE
FUSE WITH
THE
INSTRUMENT
TURNED
ON OR
CONNECT-
ED
TO
LINE
POWER.
3.
Refer
to
Figure
2-1,
item
5
for
the
location
of
the
fuse
holder.
2-4
OPERATION
FRONT
PANEL
FEATURES
SE
OO
wm
4,
Using
a
coin
or
wide
blade
screwdriver,
push
in
while
turning
the
fuse
holder
in
the
direction
of
the
arrow
on
the
front
panel
decal.
5.
Pull
out
the
fuse
holder
and
replace
the
defective
fuse.
2-13.
FRONT
PANEL
FEATURES
2-14.
Before
using
your
Multimeter,
take
a
few
minutes
to
become
familiar
with
the
use
ofits
controls,
indicators,
and
connectors.
The
front
panel
features
are
shown
in
Figure
2-1
and
described
in
Table
2-1.
The
features
of
the
Liquid
Crystal
Display
(LCD)
are
also
described
in
the
following
paragraphs,
2-15.
LCD
DISPLAY
2-16.
The
features
of
the
Liquid
Crystal
Display
(LCD)
are
shown
in
detail
in
Figure
2-2.
The
position
of
the
floating
decimal
point
is
determined
by
the
range
selected.
The
maximum
measurement
value
that
can
be
displayed
is
one
count
less
than
the
range
selected
(e.g.,
maximum
—
measured
voltage
that
can
be
displayed
in
the
200
mV
range
would
be
199.9
mV).
2-17.
To
extend
the
life
of
the
LCD
and
to
ensure
that
the
display
will
be
ready
to
operate,
observe
the
following
precautions:
e
Do
not
store
or
use
the
instrument
in
temperatures
above
or
below
those
specified
in
Section
1.
e
Do
not
store
or
use
the
instrument
in
humidity
above
that
specified
in
Section
1.
NOTE
Low
temperatures
(within
the
specified
operating
limits)
will
cause
the
LCD
response
to
be
sluggish.
e
Avoid
prolonged
exposure
of
the
LCD
to
direct
sunlight
(ultraviolet).
750y
AC
}
200
mv¥
2
COC
}QM
vA
2)
O78)
(BaP
ee
"20
2
=
POM}
EI
Ee)
Ee)
sea)
Gee
Ems ae)
(ae
20
Soe)
Ca
2000S
mm)
Figure
2-1.
Controls,
Indicators,
and
Connectors
OPERATION
FRONT
PANEL
FEATURES
Table
2-1.
8010A/8012A
Controls,
Indicators,
and
Connectors
ITEM
NO.
NAME
FUNCTION
1
Display
2
AC/DC
Function
Switch
3
V/mA/kQ/S
Function
Switches
4
Range
Switches
5
mA
Input
Connector
6
COMMON
Input
Connector
7
V/KQ/S
Input
Connector
8
10A
Input
Connector
9
Low
Ohms
Input
Connector
/ZERO
Controi
10
POWER
Switch
3%-digit
LCD
display.
Indicates
measured
input
values
and
an
overrange
condition.
(Also
contains
an
annunciator
for
low
battery
charge,
if
the
Rechargeable
Battery
Option
is
installed.)
A
two-position
switch
(push
IN
and
push
OUT)
used
to
select
ac
(IN)
or
dc
(OUT)
for
current
or
voltage
measurement.
Interlocked
switches,
used
with
the
AC/DC
Function
switch
to
select
the
measurement
functions.
Pushing
one
switch
will
release
the
others.
The
conductance
function
is
selected
by
pushing
the
kQ
switch
and
one
of
three
pairs
of
Range
Function
switches.
The
Low
Ohms
feature
of
the
8012A
is
selected
by
pressing
the
V
and
mA
switches
simultaneously.
Interlocked
switches
that
select
the
measurement
ranges.
Pushing
a
switch
selects
the
corresponding
range
and
releases
a
depressed
switch(es).
A
fuse
protected,
test
lead
connector
for
current
measurements,
less
than
2A.
Fuse
is
accessible
from
the
front
panei.
Test
lead
connector
used
as
the
low
or
common
input
for
all
measurement
functions.
Test
lead
connector
used
as
the
high
input
for
ail
voltage,
resist-
ance,
continuity
and
conductance
measurement
functions.
Test
lead
connector
used
for
the
10A
Range
current
function
of
the
8010A.
Test
lead
connector
used
for
the
Low
Ohms
resistance
function
of
the
8012A.
ZERO
Control
used
to
compensate
for
test
lead
resistance.
Push-on/push-off
switch.
Used
for
energizing
and
de-energiz-
ing
the
instrument.
LOW
BATTERY
INDICATOR
(RECHARGEABLE
BATTERY
OPTION
ONLY)
POLARITY
SIGN
DISABLED
ae
agi
|
|
DURING
Vac,
mA/Aac,
and
KQ
FUNCTIONS
DISPLAY
ANNUNCIATORS
OVERRANGE
INDICATION
Position
of
decimal
point
dependent
on
range
selected.
NOTE:
Figure
2-2.
Liquid
Crystal
Display
2-3
i
Sah
Oe
OPERATION
SIGNAL
INPUT
LIMITS
2-18.
SIGNAL
INPUT
LIMITS
CAUTION
Exceeding
the
maximum
signal
input
limits
can
damage
the
instrument.
2-19,
Before
using
your
Multimeter,
it
is
important
to
note
the
maximum
input
limits
that
may
bg
applied
to
the
instrument.
Table
2-2
lists
the
maximum
signal
input
levels
for
each
function,
range,
and
input
connector.
WARNING
TO
AVOID
ELECTRICAL
SHOCK,
DO
NOT
CONNECT
THE
COMMON
INPUT
CON-
NECTOR
TO
ANY
SOURCE
MORE
THAN
500V
DC,
OR
500V
PEAK
AC
ABOVE
EARTH
GROUND.
2-20.
OPERATING
TECHNIQUES
2-21.
The
following
paragraphs
describe
how
to
operate
your
Multimeter
in
each
of
its
five
primary
measurement
functions.
Additional
operating
instructions
and
applications
are
given
in
the
paragraphs
on
Applications,
later
in
this
section.
2-22.
AC/DC
Voltage
(V)
2-23.,
Figure
2-3
describes
how
to
operate
your
Multimeter
for
ac or
de
voltage
measurements.
For
all
measurements,
select
the
highest
range
that
will
provide
the
required
resolution
of
the
measurement.
If
measuring
an
unknown
voltage,
set
the
DMM
on
the
highest
range,
then
(if
needed)
select
a
lower
range.
2-24.
AC/DC
Current
(mA)
2-25.
Figure
2-4
describes
how
to
operate
your
Multimeter
for
ac
or
dc
current
measurements
up
to
2A.
(The
10A
Range
current
measurement
feature
of
the
8010A
is
described
in
the
following
paragraph.)
Turn
off
power
to
the
circuit
being
measured
before
breaking
the
circuit
and
connecting
the
Multimeter
in
series
with
the
current
source.
To
minimize
common
mode
voltages,
break
the
circuit
on
the
ground
side
of
the
current
source.
The
mA
input
connector
contains
an
in-line
fuse.
If
the
Multimeter
does
not
respond
when
measuring
curtent
using
the
mA
input
connector,
check
the
fuse
(refer
to
the
fuse
replacement
procedure
earlier
in
this
section).
If
measuring
an
unknown
current,
set
the
Multimeter
on
the
highest
range,
then
select
a
lower
range
if
needed.
2-26.
AC/DC
Current
(10A
max,
8010A
only)
CAUTION
The
10A
input
connector
on
the
8010A
is
not
fused.
Take
extra
precautions
to
not
exceed
the
10A
maximum
current
handling
ability
of
the
8010A.
2-27.
Figure
2-5
describes
how
to
operate
the
8010A
for
ac
or
dc
current
measurements
up
to
10A.
The
10A
input
connector
on
the
8010A
is
not
fused.
Observe
the
Caution
given
above.
All
other
conditions
for
normal
ac or
de
current
measurements,
given
in
the
preceding
paragraph,
apply
to
the
LOA
Range
feature
of
the
8010A.
2-28.
Resistance
(Q)
2-29.
Figure
2-6
describes
how
to
operate
your
Multimeter
for
resistance
measurements.
(The
Low
Ohms
resistance
feature
of
the
8012A
is
described
in
the
following
paragraph.)
Erroneous
measurements
can
occur
if
power
is
present
in
the
resistance
being
measured.
Ensure
that
power
is
removed
and
the
circuits
are
discharged
before
measuring
in-circuit
resistances.
The
AC/DC
function
switch
has
no
effect
during
resistance
measurements.
2-30.
Low
Ohms
Resistance
(LO
RANGE
©,
8012A
only)
2-31.
Figure
2-7
describes
how
to
operate
the
8012A
for
low
ohms
resistance
measurements.
All
other
conditions
for
normal
resistance
measurements,
given
in
the
preceding
paragraph,
apply
to
the
LO
OHMS
feature
of
the
8012A.
Tabie
2-2.
Maximum
Input-Signal
Limits
FUNCTION
RANGE
INPUT
MAXIMUM
INPUT
SELECTED
SELECTED
TERMINALS
OVERLOAD
_|
DC
ALL
RANGES
1000V
de
or
peak
ac
Or
VIKQUS
or
20V,
200V,
750V
and
750V
rms
continous
or
107
V-Hz
dB
|
COMMON
2V,
200
mV
750V
rms
for
no
longer
than
15
seconds
or
107V-Hz
L___
-
+
nig
Double
f
d:
2A,
250V
fi
i
mA
|
or
ALL
RANGES
mA
and
COMMON
OR
eee
bea
cuean
ery
wee
series
with
a
3A,
6O0V
fuse
AC
V/KQUS
and
kQorS
ALL
RANGES
COMMON
SO00V
de
or
ac
rms
2-4
OPERATION
OPERATING
TECHNIQUES
VOLTAGE,
LINEAR
(V)
1.
SELECT
RANGE
2.
SELECT
FUNCTION
a
\
7
sf
me
&
ee
ies)
Fy
x
WO?
BEBEaO
0)
Gee
ZS
oon)
Coe
20
Sm
a
}
6
Max
A)
cena
DIGiTAL/MUBTIMETER
Figure
2-3.
AC/DC
Voltage
Operation
CURRENT
(ma)
;
1.
SELECT
RANGE
2.
SELECT
FUNCTION
©)
Max
Ay)
FONeR
a
Pam
©
3.
CONNECT
TEST
LEADS
4.
BREAK
CIRCUIT
5.
PLACE
8010A
IN
SERIES
Figure
2-4.
AC/DC
Current
Operation
OPERATION
OPERATING
TECHNIQUES
CURRENT
(mA)
1,
SELECT
RANGE
2,
SELECT
FUNCTION
}
200
mv
“7
2200
Ci
Ma
2
Dooogoo
=
0)
Came
Fn
ve)
(aed
Sem)
Ceo)
govelloigiTaL
MULTIMETER
3.
CONNECT
TEST
LEADS
4.
BREAK
CIRCUIT
5.
PLACE
800A
IN
SERIES
°
Figure
2-5.
AC/DC
Current
Operation,
10A
(8610A
only)
RESISTANCE
(kQ)
1.
DE-ENERGIZE
CIRCUIT
TO
BE
MEASURED
2.
SELECT
RANGE
3.
SELECT
FUNCTION
4.
CONNECT
TEST
LEADS
HIGH
(+)
LOW
(-)
aos
ee)
=i)
ae
Lay
6
Powe
Ecoee]
©
80108
OIGITAL
MUBTIMETER
Figure
2-6.
Resistance
Operation
_
LOW
OHMS
RESISTANCE
(LO
RANGE
9)
1.
DE-ENERGIZE
CIRCUIT
TO
BE
MEASURED
2.
SELECT
RANGE
(2
or
20Q)
>
3.
SELECT
FUNCTIONS
(BOTH
SWITCHES)
4.
CONNECT
TEST
LEADS
OPERATION
INITIAL
CHECKOUT
PROCEDURE
ta
ent
=)
a
J
LOW
(—)
J
GUORANGE
Om)
(S510)
Cae
2S
me)
Ce
Sem
}
Centr)
HIGH(+)
—+
5.
SHORT
TEST
LEADS
TOGETHER.
USING
THE
ZERO
CONTROL,
“ZERO”
THE
DISPLAY
READING.
6.
MEASURE
UNKNOWN
RESISTANCE
en
Ee
ee
Figure
2-7.
Resistance
Operation,
LO
Range
Q
(8012A
only)
2-32.
Conductance
(S=1/0)
2-33.
Figure
2-8
describes
how
to
operate
your
Multimeter
for
conductance
measurements.
When
S=1/0
is
selected,
three
ranges
of
measurements
are
available,
2
mS,
20
wS,
and
200
nS.
To
select
a
range,
press
both
range
switches
(above
the
grey-shaded
area)
simultaneously.
2-34.
Diode
Test
2-35.
Figure
2-9
describes
how
to
operate
your
Multimeter
for
diode
tests.
The
three
resistance
ranges
with
the
diode
symbol
beside
the
range
value
provide
a
measurement
voltage
sufficient
to
cause
a
silicon
junction
to
conduct.
These
ranges
(2
kQ,
200
kQ,
and
20
MQ)
can
be
used
to
check
silicon
diodes
and
transistors.
The
2
kQ.
resistance
range
is
the
preferred:
diode
and
transistor
testing
range
and
is
labeled
with
the
largest
diode
symbol
(P+).
For
a
silicon
diode,
the
typical
forward’
bias
voltage
(on
the
2
kQ,
+
range)
is
0.6V.
A
reversed
bias
silicon
diode
should
display
the
overrange
indicator
(on
the
2
kQ,
-®
range).
2-36.
INITIAL
CHECKOUT
PROCEDURE
2-37.
The
following
procedure.allows
you
to
verify
that
the
Multimeter
is
operating
correctly
for
most
functions.
The
only
test
equipment
required
is
a'set
of
test
leads
and
access
to
a
standard
wall
socket.
This
procedure
checks
for
general
operation
only
and
is
not
intended
to
verify
Pre
——_____
instrument
accuracy.
Performance
tests
and
calibration
adjustments
are
contained
in
Section
4
of
this
manual
for
the
purpose
of
festing
and
correcting
instrument
accuracy,
2-38.
Use
the
following
procedure
to
verify
that
most
of
the
functions
of
your
Multimeter
are
operating
correctly:
1.
Select
the
AC
V
function
on
the
Multimeter.
2.
Set
the
Multimeter
to
the
750V
range.
WARNING
THE
LOCAL
LINE
VOLTAGE
IS
BEING
MEASURED
IN
THE
FOLLOWING
STEP.
DO
NOT
TOUCH
THE
PROBE
TIPS
OR
ALLOW
THE
PROBE
TIPS
TO
COME
IN
CONTACT
WITH
EACH
OTHER
WHILE
PREFORMING
THE
FOLLOWING
STEP.
3.
Insert
the
probe
tips
into
a
standard
wall
socket.
Note
the
preceding
warning.
The
display
should
read
the
local
line
voltage.
4,
Momentarily
set
the
instrument
to
the
20V
range.
The
overrange
indicator
should
be
displayed.
5.
Remove
the
test
leads
from
the
wall
socket.
2-7
OPERATION
APPLICATIONS
CONDUCTANCE
(S
=
1/9)
1.
DE-ENERGIZE
CIRCUIT
TO
BE
MEASURED
2.
SELECT
RANGE
3.
SELECT
FUNCTION
4,
CONNECT
TEST
LEADS
HIGH
(+)
LOW
(—)
LUKE]
*
aod
DIGITAL
MUBTIMETER
‘Cae
©
Hos
oe
ow
LI
eat
Bee
ee
ee
ee
Figure
2-8.
Conductance
Operation
6.
Select
the
resistance
function.
The
overrange
indicator
should
appear
in
the
display.
7.
Set
the
instrument
to
the
20020
range
and
short
the
test
leads.
The
display
should
read
“00.0”.
8.
Select
the
S=1/OQ
(conductance)
function,
2
mS
range.
The
display
should
read
“.000”
+
5
counts.
9.
Short
the
test
leads.
The
overrange
indicator
should
appear
in
the
display.
10.
This
concludes
the
Initial
Checkout
procedure
for
your
Multimeter.
If
the
performance
of
the
instrument
is
in
question
refer
to
the
Performance
tests
in
Section
4
of
this
manual.
2-39.
APPLICATIONS
2-40.
The
following
paragraphs
contain
additional
information
and
measurement
techniques
for
the
five
primary
functions
of
your
Multimeter.
WARNING
TO
AVOID
ELECTRICAL
SHOCK
AND/OR
INSTRUMENT
DAMAGE,
DO
NOT
CON-
NECT
THE
COMMON
INPUT
TERMINAL
TO
ANY
SOURCE
OF
MORE
THAN
500
VOLTS
DC
OR
PEAK
AC
ABOVE
EARTH
GROUND.
2-41.
Circuit
Loading
Error
(Voltage)
2-42.
Circuit
loading
errors
occur
when
voltage
‘measurements
are
taken
on
high
impedance
circuits.
This
is
because
the
Multimeter
loads
the
source,
thus
changing
the
operating
voltage
of
the
source.
As
long
as
the
circuit
impedance
(source
impedance)
is
low
compared
to
the
input
impedance
of
the
Multimeter
this
error
may
be
insignificant.
For
example,
when
measuring
a
circuit
with
a
source
impedance
of
10
k0
or
less,
the
error
will
be
<0.!
%.
If
the
circuit
loading
error
is
significant,
use
the
appropriate
formula
contained
in
Figure
2-10
to
calculate
the
percentage
of
error.
WARNING
OPERATOR
INJURY
AND
INSTRUMENT
DAMAGE
MAY
RESULT
IF
THE
BACKUP
FUSE
(F2)
BLOWS
WHEN
CURRENT
IS
BEING
MEASURED
FROM
A
VOLTAGE
OF.
GREATER
THAN
600
VOLTS.
.
for
the
highest
range.
DIODE
TEST
(kQ,
+
)
1,
DE-ENERGIZE
CIRCUIT
TO
BE
MEASURED
2,
SELECT
Range
(2-+-
range
preferred}
3.
SELECT
FUNCTION
4.
CONNECT
TEST
LEADS
HIGH
(+)
LOW
(-}
OPERATION
APPLICATIONS
Cap)
(eS)
poral
oiciTaL
me
rimeTER
6
sy
Gwe
oe
EE
_
FORWARD
BIAS:
x
REVERSE
BIAS:
+
Figure
2-9.
Diode
Test
Operation
2-43.
Burden
Voltage
Error
(Current)
2-44.
When
a
multimeter
is
placed
in
series
with
a
circuit
to
measure
current,
the
voltage
drop
of
the
multimeter
induces
an
error.
This
voltage
is
called
the
burden
voltage.
The
maximum
full-scale
burden
voltages
for
your
Multimeter
are
0.3V
for
the
four
lowest
ranges
and
0.9V
2-45.
These
voltage
drops
can
affect
the
accuracy
of
the
current
measurement
if
the
current
source
is
unregulated
and
the
resistance
of
the
shunt
and
fuses
of
the
multimeter
exceeds
1/1000
of
the
source
resistance.
Ifthe
multimeter
burden
voltage
is
significant,
the
formula
in
Figure
2-11
can
be
used
to
calculate
the
burden
voltage
error.
2-46.
Test
Lead
Compensation
(Resistance)
2-47.
When
measuring
low
resistances,
the
effects
of
test
lead
resistance
may
add
a
significant
error.
This
error
is
compensated
for
by
measuring
the
lead
resistance
and
subtracting
it
from
the
resistance
measured
in
the
circuit.
The
test
lead
resistance
of
the
8010A
must
be
subtracted
manually,
by
the
operator,
to
compensate
for
this
error.
The
8012A
provides
a
ZERO
function,
for
Low
Ohms
resistance
measurements,
that
“zeros
out”
the
value
of
the
test
lead
resistance,
Figure
2-7
and
the
paragraphs
on
Low
Ohms
Resistance
describe
how
to-use
the
ZERO
function
of
the
8012A.
2-48.
Use
the
following
procedure
to
manually
compensate
for
test
lead
resistance:
1.
Setup
the
Multimeter
as
shown
in
Figure
2-6.
2.
Short
the
test
leads
together
(press
the
test
leads
together
firmly).
2-9
OPERATION
APPLICATIONS
3.
Record
the
Multimeter
reading
obtained
in
Step
2.
4.
Proceed
with
the
resistance
measurement
and
subtract
the
value
of
Step
3
from
the
Multimeter
reading.
1.
DC
VOLTAGE
MEASUREMENTS
Loading
Error
in
%
=
100
x
Rs
+
(Rs
+
107)
Where:
Rs
=
Source
resistance
in
ohms
of
circuit
being
measured.
2.
AC
VOLTAGE
MEASUREMENTS
First,
determine
input
impedance,
as
follows:
*
10”
Zin=
F
V1i+(27F
°
Rin
*
Cin)?
Where:
Zin
=
etfective
input
impedance
Rin
=
10”
ohms
Cin
=
100
x
107}?
Farads
F
=
frequency
in
Hz
Then,
determine
source
loading
error
as
follows:
*
Loading
Error
in
%
=
100
x
>So
Where:
Zs
=
source
impedance
Zin
=
input
impedance
(calculated)
*Vector
algebra
required
Figure
2-10.
Circuit
Loading
Error
2-49.
High
Resistance
Measurements
(Conductance)
2-50.
The
conductance
function
of
your
Multimeter
can
be
used
to
measure
high
resistive
(low
leakage)
components
(diodes
and
capacitors)
while
minimizing
noise
problems.
The
three
conductance
ranges
(2
mS,
20
pwS,
and
200
nS)
can
be
used
for
making
resistance
measurements
from
500Qto
1
MO,
50k
to
100
MQ,
and
5
MOQ.
to
10,000
MO.
Refer
to
Figure
2-12
for
a
list
of
conductance
to
resistance
conversions.
2-51.
Leakage
Resistance
Measurements
(Conductance)
.
2-52.
Use
the
conductance
function
for
leakage
testing
on
purely
resistive
components
(e.g.,
cables
and
pcb’s).
NOTE
Under
high
humidity
conditions,
fingerprints
and
other
residual
surface
contaminants
can
create
their
own
leakage
paths.
Clean
all
surfaces
and
use
clean
test
leads
to
minimize
the
effect
of
leakage
paths.
2-53.
Diode
Leakage
Tests
(Conductance)
2-54.
Diode
leakage
(Ir)
tests
require
that
the
diode
junction
be
reverse
biased
while
being
measured.
Connect
the
anode
of
the
diode
to
the
COMMON
input
connector
to
reverse
bias
a
diode
junction.
A
good
silicon
diode
will
produce
an
in-scale
display
reading
on
the
200
nS
range
when
reverse
biased.
IM
Es
S
Ri
Es
Es
=
Source
voltage
RI
=
Load
resistance
+
Source
resistance
jm
=
Measured
current
(display
reading
in
amps)
Eb
=
Burden
valtage
(calculated)
£b
=
meas.
current
[(200/current
range
in
mA}
+.35]
ERROR:
Error
in
%
=
100
x
Eb/(Es
-
Eb)
Error
in
A
=
(Eb
x
Im)/(Es
-
Eb)
ee
EXAMPLE:
Es
=
15V
RI
=
100
kQ
Im
=
148.51
yA
(.14851
mA)
Eb
=
148.51
x
10-®
x
[(200/.2)
+
35]
=
148.51
x
10-®
x
1000.35
=
148.56
mV
Max.
error
in
%
=
100
x
[148.56
MV/{15V
~
.14856V)]
=
1.0003%
Add
this
to
the
range
spec.
accuracy:
Max.
error
in
%
=
1.0003%
+(.2%
+
2
digits)
Max.
error
in
A
=
(148.56
mV
x
148.51
wA)/(15000
mv
-
148.56
mV)
=
1.486
pA
Add
1.486
wA
to
the
reading
for
correct current
Figure
2-11.
Calculating
Burden
Voltage
Error
2-10
Sam,
Interpolation
Table
(I/no.)
2
3
4
5
196
6
.167
|
.164
7
143)
.141
8
|
.125
1.123
9
.117:
4.110
*mS
to
kQ
*uS
to
MQ
*n$
to
MQ
2mS
RANGE
20
uS
RANGE
200
nS
RANGE
(1/ms
=
kQ)
(1/u8
=
MQ)
(1000/nS
=
MQ)
ms
KQ
us
mQ
ns
MQ
2.0
5
20
05
200
5
1.0
1
10
J
400
10
05
2 5
2
50
20
0.2
5
2
5
20
50
0.1
10
1 1
10
100
0.05
20
5
2 5
200
0.02
50
2
5
2
500
0.01
100
a
10
1
1900
0.005
200
05
20
0s
2000
0.002
500
02
50
0.2
5000
0.001
1000
01
100
0.1
10000
CONVERSION
SCALES
Use
the
following
procedure
to
convert
a
conductance
reading
(displayed
in
siemens)
to
equivatent
resistance
(in
ohms):
1.
On
the
Interpolation
Table
locate
the
most
significant
digit
of
the
conductance
display
under
the
DIGIT
column
heading.
2.
On
the
Interpolation
Table
locate
the
second
most
significant
digit
of
the
conductance
display
across
from
the
DIGIT
row.
8.
On
the
Interpolation
Table
locate
the
value
(of
resistance)
at
the
intersection
of
the
two
digits
in
Steps
1
and
2,
then
use
*S
=
siemens
=
1/Q
=
International
Unit
of
Conductance
Formerly
Known
as
the
mho.
the
appropriate
Conversion
Scale
to
determine
the
position
of
the
decimal
point.
For
Example:
A
reading
of
52.0
nS
is
displayed
on
the
Muttimeter.
The
interpolation
Table
shows
a
value
of
.192.
Using
the
Conversion
Scale,
under
the
200
nS
Range
heading,'52.0
nS
corresponds
to
approximately
20
MQ.
Therefore,
the
actual
equivalent
resistance
is
19.2
MQ
Figure
2-12.
Conductance
to
Resistance
Conversion
OPERATION
APPLICATIONS
ae
2-11
OPERATION
APPLICATIONS
2-55.
Transistor
Tester
2-56.
The
transistor
tester
described
in
the
following
paragraphs
provides
approximate
test
information.
Beta
is
tested
using
a
Vceof
2V
and
an
Icof
about
200
pA.
This
transistor
tester
is
useful
for
checking
the
proper
operation
of
transistors
and
approximate
beta
values
for
comparative
measurements.
2-57.
The
transistor
tester
fixture
is
described
in
Figure
2-13.
When
assembled
and
connected
to
the
V/kQ/S
and
the
COMMON
input
connector,
the
Multimeter
can
be
used
to
determine
the
following
information
about
transistors:
®
Transistor
type
(NPN
or
PNP)
e
Defective
transistors
(shorted
or
open)
®
Collector-to-emitter
leakage
(Iczs)
e
Beta
from
10
to
1000
in
a
single
range.
2-58.
Transistor
type
is
determined
by
setting
the
switch
on
the
test
fixture
to
BETA,
setting
the
Multimeter
to
the
2
mS
range,
and
observing
the
display
reading.
If
a
low
reading
(<
0.010)
is
displayed,
reverse
the
test
fixture
at
the
input
connectors.
If
the
collector
of
the
transistor
is
now
connected
to
the
COMMON
input
connector
the
transistor
is
a
PNP
type.
An
NPN
type
will
have
its
collector
connected
to
the
V/kQ/S
input
connector.
2-59.
DEFECTIVE
TRANSISTORS
2-60.
If
the
transistor
is
defective,
the
following
indications
will
appear,
regardless
of
transistor
type
or
test
position:
I,
An
open
transistor
will
produce
a
display
reading
of
0.001
or
less.
‘
2.
A
shorted
transistor
will
produce
an
overrange
indication
on
the
display.
SCHEMATIC
TRANSISTOR
UNDER
TEST TEST
FIXTURE
—_—_|_—
BETA
CONSTRUCTION
DETAIL
$1
TOGGLE
SWITCH
SPDT
4
Pt
R1
1
|
PLUG
INTO
TBOKS
|
COMMON
AND
|
V/KQ/S
INPUT
TERMINALS
—_______-l_»
J1
TRANSISTOR
SOCKET
5%
1
3/16”
HOLE
WIRE
TO
BASE
CONTACT
OF
TRANSISTOR
SOCKET
Figure
2-13,
Transistor
R1
750
kQ
/4w
Pi
BANANA
PLUG
CONNECTOR
0.75”
SPACING
GENERAL
RADIO
TYPE
274
MB
WIRE
TO
SWITCH
ARM
Beta
Test
Fixture
2-61.
TRANSISTOR
LEAKAGE
TEST
2-62.
Use
the
following
procedure
to
test
transistors
for
leakage
(Ices):
1.
Install
the
transistor,
and
connect
the
test
fixture
to
the
Multimeter
(see
preceding
paragraphs).
2.
Set
the
switch
on
the
test
fixture
to
ICES.
3.
Select
the
conductance
function,
2
mS
range
on
the
Multimeter.
4,
Areading
of
more
than
0.0020
(6
uA)
indicates
a
faulty
transistor
(silicon).
2-63.
TRANSISTOR
BETA
TEST
2-64.
Use
the
following
procedure
to
test
the
beta
of
a
transistor:
1.
Install
the
transistor
and
connect
the
test
fixture
to
the
Multimeter
(see
preceding
paragraphs).
2.
Set
the
switch
in
the
test
fixture
to
BETA.
3.
Select
the
conductance
function,
2
mS
range
on
the
Multimeter.
4.
Note
the
display
reading
on
the
Multimeter,
then
shift
the
decimal
point
three
places
to
the
tight.
This
will
be
the
beta
of
the
transistor.
NOTE
Beta
is
a
temperature-sensitive
measurement.
Allow
sufficient
time
for
each
tested
transistor
to
stabilize.
Avoid
touching
the
transistor
case
with
your
fingers
while
making
beta
measurements.
2-65.
True-RMS
Measurements
2-66.
One
of
the
most
useful
features
of
the
Multimeters
is
the
direct
measurement
of
true-rms
ac
voltages
and
ac
current.
Mathematically,
rms
is
defined
as
the
square
root
of
the
mean
of
the
squares
of
the
instantaneous
voltages.
In
physical
terms,
rms
is
equivalent
to
the
de
value
that
dissipates
the
same
amount
of
heat
in
a
resistor
as
the
original
waveform.
True-rms
is
the
effective
value
of
any
waveform
and
represents
the
energy
level
of
the
signal.
It
is
used
directly
in
the
relationships
of
Ohm’s
Law
and
provides
a
reliable
basis
for
comparisons
of
dissimilar
waveforms.
2-67.
Most
multimeters
in
use
today
have
average-
responding
ac
converters
rather
than
true-rms
converters
like
the
8010A
and
8012A.
Usually
the
gain
in
average-
responding
meters
is
adjusted
so
that
the
reading
gives
the
rms
value,
provided
the
input
signal
is
a
harmonic-free
sinusoid.
However,
if
the
signal
is
not
sinusoidal,
the
average-responding
meter
does
not
give
a
correct
rms
reading.
OPERATION
APPLICATIONS
2-68.
Your
Multimeter’s
ac
converter
calculates
the
rms
value
through
analog
computation.
This
results
in
accurate
rms
values
for
mixed
frequencies,
modulated
signals,
square
waves,
sawtooths,
10%-duty-cycle
pulses,
etc,
when
these
signals
are
measured
with
your
Multimeter.
2-69.
Waveform
Comparison
(RMS
vs
Averaging
Meters)
2-70.
Figure
2-14
shows
the
relationship
between
common
waveforms
and
their
displayed
value,
as
they
appear
on
the
8010A
or
8012A,
compared
to
average-
responding
meters.
Figure
2-14
also
illustrates
the
relationship
between
ac
and
de
measurements
for
ac-
coupled
meters.
For
example,
the
first
waveform
(in
Figure
2-14)
is
a
sine
wave
with
a
peak
voltage
of
1.414V.
Both
Fluke
Multimeters
(8010A
and
8012A)
and
the
average
responding
meters
display
the
correct
rms
reading
of
1.000V
(the
de
component
equals
0).
The
1.414V
(peak)
rectified
square
wave
also
produces
a
correct
de
reading
(0.707V)
on
all
the
multimeters,
but
only
the
Fluke
Multimeters
correctly
measure
the
ac
component
(0.707V).
The
average
responding
meter
measures
the
ac
component
of
the
rectified
square
wave
as
0.785V,
which
is
an
error
of
5.6%.
2-71.
Waveform
Crest
Factors
2-72.
The
crest
factor
of
a
waveform
is
the
ratio
of
the
peak
to
rms
voltage.
In
waveforms
where
the
positive
and
negative
half-cycles
have
different
peak
voltages,
the
higher
voltage
is
used
in
computing
the
crest
factor.
Crest
factors
start
at
1.0
for
a
square
wave
(peak
voltage
equals
rms
voltage).
2-73.
Your
Multimeter
can
measure
signals
with
a
crest
factor
of
3.0
or
less,
at
full
scale.
Figure
2-15
illustrates
some
typical
signals
and
their
crest
factors.
The
waveforms
in
Figure
2-15
show
that
a
signal
with
a
crest
factor
of
greater
than
3.0
is
not
common.
2-74.
To
ensure
that
a
signal
measured
with
your
Multimeter
has
a
crest
factor
below
3.0,
measure
the
peak
value
with
an
ac
coupled
oscilloscope.
If
the
peak
value
is
not
more
than
three
times
the
true-rms
reading
of
your
Multimeter,
then
the
crest
factor
of
the
signal
is
3.0
or
less.
Another
method
of
verifying
the
error
caused
by
the
crest
factor
of
a
signal
is
to
compare
the
reading
of
your
Multimeter
with
a
reading
on
the
next
higher
range
of
your
Multimeter.
The
crest
factor
capability
of
your
Multimeter
increases
(from
3.0)
for
readings
less
than
full-scale.
The
crest
factor
capability
of
your
Multimeter
is
shown
by
the
following
equation:
Crest
Factor
Capability
=
3
(Ranse)
Input
The
error
caused
by
exceeding
the
crest
factor
of
3.0
at
full
scale,
will
be
reduced
significantly
on
the
next
higher
measurement
range
of
your
Multimeter.
The
crest
factor
capability
at
1/10
scale
approaches
10.
2-13

Other manuals for 8010A

3

This manual suits for next models

1

Other Fluke Multimeter manuals

Fluke 233 User manual

Fluke

Fluke 233 User manual

Fluke Ti30 User manual

Fluke

Fluke Ti30 User manual

Fluke 8846A/C User manual

Fluke

Fluke 8846A/C User manual

Fluke 28 II Ex Use and care manual

Fluke

Fluke 28 II Ex Use and care manual

Fluke 8050A User manual

Fluke

Fluke 8050A User manual

Fluke 1587 FC User manual

Fluke

Fluke 1587 FC User manual

Fluke 183 User manual

Fluke

Fluke 183 User manual

Fluke 8505A User manual

Fluke

Fluke 8505A User manual

Fluke a3000 FC Use and care manual

Fluke

Fluke a3000 FC Use and care manual

Fluke 91 Series II User manual

Fluke

Fluke 91 Series II User manual

Fluke 101 User manual

Fluke

Fluke 101 User manual

Fluke 8010A User manual

Fluke

Fluke 8010A User manual

Fluke 867B User manual

Fluke

Fluke 867B User manual

Fluke 77 User manual

Fluke

Fluke 77 User manual

Fluke 116 User manual

Fluke

Fluke 116 User manual

Fluke CNX 3000 User manual

Fluke

Fluke CNX 3000 User manual

Fluke 8808A Configuration guide

Fluke

Fluke 8808A Configuration guide

Fluke 175 User manual

Fluke

Fluke 175 User manual

Fluke 8024B User manual

Fluke

Fluke 8024B User manual

Fluke 115C User manual

Fluke

Fluke 115C User manual

Fluke 287 User manual

Fluke

Fluke 287 User manual

Fluke 8808A Configuration guide

Fluke

Fluke 8808A Configuration guide

Fluke 8021B User manual

Fluke

Fluke 8021B User manual

Fluke 8021B User manual

Fluke

Fluke 8021B User manual

Popular Multimeter manuals by other brands

PeakTech 3355 operating manual

PeakTech

PeakTech 3355 operating manual

HP 3457A operating instructions

HP

HP 3457A operating instructions

Sperry instrument DM-2000 operating instructions

Sperry instrument

Sperry instrument DM-2000 operating instructions

Urrea UD18 User manual and warranty

Urrea

Urrea UD18 User manual and warranty

LOVATO ELECTRIC DMK16 quick start guide

LOVATO ELECTRIC

LOVATO ELECTRIC DMK16 quick start guide

Kusam-meco 5040 Operation manual

Kusam-meco

Kusam-meco 5040 Operation manual

Amprobe AU92 user manual

Amprobe

Amprobe AU92 user manual

Major tech MTC50 instruction manual

Major tech

Major tech MTC50 instruction manual

Weidmüller 1037 operating instructions

Weidmüller

Weidmüller 1037 operating instructions

AEMC 1030 user manual

AEMC

AEMC 1030 user manual

Unitest Telaris 0100 instruction manual

Unitest

Unitest Telaris 0100 instruction manual

Beha-Amprobe AM-555-EUR user manual

Beha-Amprobe

Beha-Amprobe AM-555-EUR user manual

PCE Instruments PCE-IT 100 manual

PCE Instruments

PCE Instruments PCE-IT 100 manual

VOLTCRAFT 12 44 55 VC135 operating instructions

VOLTCRAFT

VOLTCRAFT 12 44 55 VC135 operating instructions

DMMetering PRO-75 SERIES user manual

DMMetering

DMMetering PRO-75 SERIES user manual

Aktakom ACM-2353 user manual

Aktakom

Aktakom ACM-2353 user manual

Daiichi Electronics SMLC-110L instruction manual

Daiichi Electronics

Daiichi Electronics SMLC-110L instruction manual

Velleman INIA011 user manual

Velleman

Velleman INIA011 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.