P/N 460017 Rev. D 5
SECTION 4: WARNINGS AND CAUTIONS
EVER PUT YOUR HEAD INTO THE
RY
CID
ISSOLUTION
NIT
ANK
HIS MAY RESULT IN OPERATOR
FALLING INTO TANK
N THE EVENT OF A FOREIGN OBJECT FALLING INTO THE TANK DURING ANY ACTIVE OPERATION
THE OPERATION
SHOULD BE PAUSED. IF NEEDED,USE STEP MODE TO PLACE UNIT INTO A DRAIN OPERATION TO EMPTY THE TANK.IF
CONCENTRATE SOLUTION NEEDS TO BE DISPOSED,MAKE SURE TO COMPLY WITH LOCAL,
REQUIREMENTS.AFTER DRAIN OPERATION,UNPLUG DRY ACID DISSOLUTION UNIT FROM ITS POWER SOURCE.AREACH
TOOL IS RECOMMENDED FOR REMOVING FOREIGN OBJECT OUT OF TANK.HOWEVER,SHOULD THE FOREIGN OBJECT BE
UNATTAINABLE,CALL TECHNICAL SERVICE FOR FURTHER ASSISTANCE (1-800-227-2572).
F THE DRY ACID DISSOLUTION UNIT IS RELOCATED A SAMPLE MUST BE DRAWN FROM THE FIRST BATCH OF
CONCENTRATE MADE,WHICH MUST BE ANALYZED FOR CORRECT SOLUTION MIX BEFORE THE CONCENTRATE CAN BE USED.
PRIOR TO RELOCATION,REQUEST TWO (2) EMPTY SAMPLE BOTTLES (P/N G83-535-02) THEN SEE APPENDIX BFOR
FURTHER INSTRUCTIONS ON WHEN AND WHERE TO SEND A SAMPLE OF THE FIRST BATCH OF CONCENTRATE.
HE USE OF EYE PROTECTION
AND GLOVES IS RECOMMENDED WHEN HANDLING DRY ACID PRODUCT
F
CONTACT WITH EYES,RINSE IMMEDIATELY FOR 15 MINUTES.IF CONTACT WITH SKIN,FLUSH WITH PLENTY OF SOAP AND
WATER.SEE MATERIAL SAFETY DATA SHEETS (MSDS) FOR THE DRY ACID PRODUCT BEING USED FOR FURTHER PERSONAL
PROTECTIVE EQUIPMENT (PPE)OR EMERGENCY REQUIREMENTS/INSTRUCTIONS.
HOULD THE
RY
CID
ISSOLUTION
NIT
AIL TO COMPLETE ANY OPERATION WITHIN ANY
YCLE DO NOT
PROCEED THROUGH CYCLE USING STEP MODE.CALL TECHNICAL SERVICE FOR FURTHER ASSISTANCE (1-800-227-2572).
WARNING! FAILURES TO INSTALL,OPERATE,AND MAINTAIN THIS EQUIPMENT ACCORDING TO THE MANUFACTURER’S
INSTRUCTIONS MAY CAUSE PATIENT DEATH OR INJURY.
HIS
RY
CID
ISSOLUTION
NIT MAY BE SUSCEPTIBLE TO ELECTROMAGNETIC INTERFERENCE
DEVICES EMITTING ELECTROMAGNETIC RADIATION SUCH AS ANALOG PORTABLE PHONES,RADIO EQUIPMENT (WALKIE-
TALKIES,ETC.), RADIO TRANSMITTERS,AND LIKE EQUIPMENT,SHOULD NOT BE USED IN THE VICINITY OF THIS EQUIPMENT.
THE OPERATOR SHOULD MONITOR THE FUNCTION OF THE MACHINE AND REMOVE UNNECESSARY EQUIPMENT FROM THE
TREATMENT AREA SHOULD THESE EVENTS OCCUR.
HE ELECTRICAL SOURCE MUST BE SINGLE PHASE
THREE
CONDUCTOR TYPE PROVIDED WITH A GROUND FAULT
INTERRUPTER AT 120 VAC,60 HZ.THE PROPER POLARITY AND GROUND INTEGRITY MUST BE INITIALLY CHECKED AND
MAINTAINED
AILURE TO DO SO MAY RESULT IN ELECTRICAL SHOCK OR BURN TO THE OPERATOR
WARNING! SHOCK HAZARD.DO NOT REMOVE PANELS.REFER SERVICING TO QUALIFIED PERSONNEL.REPLACE FUSES
ONLY WITH THE SAME TYPE AND RATING.
AILURE TO FLUSH THE TRANSFER HOSE WITH
GALLONS PRIOR TO MEASURING SPECIFIC GRAVITY MAY
RESULT IN AN INCORRECT SPECIFIC GRAVITY READING,AND/OR SUPPLY OF AN INCORRECT CONCENTRATE FORMULATION TO
THE HEMODIALYSIS MACHINE,WHICH MAY RESULT IN PATIENT INJURY OR DEATH.SEE SECTION 7.1, #5, PAGE 16.
WARNING! FOLLOWING DISINFECTION,TWO (2) COMPLETE RINSE CYCLES MUST BE PERFORMED AND THE UNIT TESTED
TO ENSURE THE ABSENCE OF RESIDUAL BLEACH.SEE SECTION 8.3, PAGE 23.
OLLOW THE MANUFACTURER
S INSTRUCTIONS WHEN MEASURING THE SPECIFIC GRAVITY OF A SAMPLE OF
THE MIXED FRESENIUS MEDICAL CARE DRY ACID PRODUCT (SEE FORM 1FOR GRANUFLO®PRODUCT OR FORM 2FOR
CITRASATE DRY PRODUCT)WITH A HYDROMETER CALIBRATED AT 60°F FOLLOWING ASTM STANDARD E126 PRIOR TO
TRANSFERRING TO A DIALYSIS MACHINE OR HOLDING TANK.
ONNECT WATER INLET ACCORDING TO THE SPECIFICATIONS FOR THE
RY
CID
ISSOLUTION
NIT
HE
WATER USED MUST MEET ANSI/AAMI OR ISO STANDARDS FOR WATER USED IN HEMODIALYSIS (CURRENTLY ANSI/AMMI
RD62, OR ISO 13959). THE CORRECT IONIC CONCENTRATION AND BACTERIAL QUALITY CAN GENERALLY BE ACHIEVED IN
THE DIALYSATE ONLY WITH TREATED WATER.BE SURE THAT ALL SPECIFICATIONS ARE SATISFIED.THE WATER SOURCE MUST
BE MONITORED PERIODICALLY TO DETECT FLUCTUATIONS IN WATER COMPOSITION AND QUALITY THAT COULD HAVE AN
ADVERSE EFFECT ON THE PATIENT,HEMODIALYSIS MACHINE OR DRY ACID DISSOLUTION UNIT.PARTICULAR ATTENTION
MUST BE TAKEN FOR CHEMICALS SUCH AS ALUMINUM,CHLORINE,AND CHLORAMINES,AS THESE CHEMICALS CAN CAUSE
COMPLICATIONS IN DIALYSIS PATIENTS.THE CHEMICAL QUALITY OF THE TREATED WATER USED FOR DIALYSIS SHOULD BE
ANALYZED A LEAST ONCE A YEAR TO ENSURE IT MEETS THE REQUIREMENTS OF ANSI/AMMI
OR