P/N 460018 Rev. D 5
SECTION 4: WARNINGS AND CAUTIONS
EVER PUT YOUR HEAD INTO THE
RY
CID
ISSOLUTION
NIT
ANK
HIS MAY RESULT IN OPERATOR
N THE EVENT OF A FOREIGN OBJECT FALLING INTO THE TANK DURING ANY ACTIVE OPERATION
THE
OPERATION SHOULD BE PAUSED. IF NEEDED,USE STEP MODE TO PLACE UNIT INTO A DRAIN OPERATION TO EMPTY
THE TANK.IF CONCENTRATE SOLUTION NEEDS TO BE DISPOSED,MAKE SURE TO COMPLY WITH LOCAL,STATE AND FEDERAL
REQUIREMENTS.AFTER DRAIN OPERATION,UNPLUG DRY ACID DISSOLUTION UNIT FROM ITS POWER SOURCE.AREACH
TOOL IS RECOMMENDED FOR REMOVING FOREIGN OBJECT OUT OF TANK.HOWEVER,SHOULD THE FOREIGN OBJECT BE
UNATTAINABLE,THEN SEE SECTION 13: DISPOSAL OF CONCENTRATE SOLUTION,PAGE 43.
F THE DRY ACID DISSOLUTION UNIT IS RELOCATED
THEN A SAMPLE MUST BE DRAWN FROM THE FIRST BATCH
OF CONCENTRATE MADE AND IT MUST BE ANALYZED FOR CORRECT SOLUTION MIX BEFORE THE CONCENTRATE CAN BE
USED.PRIOR TO RELOCATION,REQUEST TWO (2) EMPTY SAMPLE BOTTLES (P/N G83-535-02) THEN SEE APPENDIX D:
FIRST BATCH VERIFICATION FOR FURTHER INSTRUCTIONS ON WHEN AND WHERE TO SEND SAMPLE OF THE FIRST BATCH OF
HE USE OF EYE PROTECTION AND GLOVES IS RECOMMENDED WHEN HANDLING DRY ACID PRODUCT
F
CONTACT WITH EYES,RINSE IMMEDIATELY FOR 15 MINUTES.IF CONTACT WITH SKIN,FLUSH WITH PLENTY OF SOAP AND
WATER.SEE MATERIAL SAFETY DATA SHEETS (MSDS) FOR THE DRY ACID PRODUCT BEING USED FOR FURTHER PERSONAL
PROTECTIVE EQUIPMENT (PPE)OR EMERGENCY REQUIREMENTS/INSTRUCTIONS.
WARNING! SHOULD THE DRY ACID DISSOLUTION UNIT FAIL TO COMPLETE ANY OPERATION WITHIN ANY CYCLE DO NOT
PROCEED THROUGH CYCLE USING STEP MODE.CALL TECHNICAL SERVICE FOR FURTHER ASSISTANCE (1-800-227-2572).
WARNING! FAILURE TO INSTALL,OPERATE,AND MAINTAIN THIS EQUIPMENT ACCORDING TO THE MANUFACTURER’S
INSTRUCTIONS MAY CAUSE PATIENT DEATH OR INJURY.
WARNING! THIS DRY ACID DISSOLUTION UNIT MAY BE SUSCEPTIBLE TO ELECTROMAGNETIC INTERFERENCE (EMI).
DEVICES EMITTING ELECTROMAGNETIC RADIATION SUCH AS ANALOG PORTABLE PHONES,RADIO EQUIPMENT (WALKIE-
TALKIES,ETC.), RADIO TRANSMITTERS,AND LIKE EQUIPMENT,SHOULD NOT BE USED IN THE VICINITY OF THIS EQUIPMENT.
THE OPERATOR SHOULD MONITOR THE FUNCTION OF THE MACHINE AND REMOVE UNNECESSARY EQUIPMENT FROM THE
TREATMENT AREA SHOULD THESE EVENTS OCCUR.
WARNING! THE ELECTRICAL SOURCE MUST BE SINGLE PHASE,THREE-CONDUCTOR TYPE PROVIDED WITH A GROUND
FAULT INTERRUPTER AT 120VAC,60 HZ.THE PROPER POLARITY AND GROUND INTEGRITY MUST BE INITIALLY CHECKED
AND MAINTAINED.FAILURE TO DO SO MAY RESULT IN ELECTRICAL SHOCK OR BURN TO THE OPERATOR.
HOCK HAZARD
O NOT REMOVE COVERS
EFER SERVICING TO QUALIFIED PERSONNEL
EPLACE FUSES
ONLY WITH THE SAME TYPE AND RATING.
AILURE TO FLUSH THE TRANSFER HOSE WITH
GALLONS PRIOR TO MEASURING SPECIFIC GRAVITY MAY
RESULT IN AN INCORRECT SPECIFIC GRAVITY READING,AND/OR SUPPLY OF AN INCORRECT CONCENTRATE FORMULATION TO
THE HEMODIALYSIS MACHINE,WHICH MAY RESULT IN PATIENT INJURY OR DEATH.SEE SECTION 7.1, #5, PAGE 17.
OLLOWING DISINFECTION
TWO
COMPLETE
MUST BE PERFORMED AND THE UNIT
TESTED TO ENSURE THE ABSENCE OF RESIDUAL BLEACH (SEE SECTION 8.3 Sodium Hypochlorite (Regular Bleach)
OLLOW THE
ANUFACTURER
S INSTRUCTIONS WHEN MEASURING THE SPECIFIC GRAVITY OF A SAMPLE OF
THE MIXED FRESENIUS MEDICAL CARE DRY ACID PRODUCT (SEE FORM 1FOR GRANUFLO®PRODUCT,PAGE 32,OR FORM
2FOR CITRASATE DRY PRODUCT,PAGE 34)WITH A HYDROMETER CALIBRATED AT 60°F FOLLOWING ASTM STANDARD
E126 PRIOR TO TRANSFERRING TO A DIALYSIS MACHINE OR HOLDING TANK.
WARNING! CONNECT WATER INLET ACCORDING TO THE SPECIFICATIONS FOR THE DRY ACID DISSOLUTION UNIT.THE
WATER USED MUST MEET ANSI/AAMI OR ISO STANDARDS FOR WATER USED IN HEMODIALYSIS (CURRENTLY ANSI/AAMI
RD62, OR ISO 13959). THE CORRECT IONIC CONCENTRATION AND BACTERIAL QUALITY CAN GENERALLY BE ACHIEVED IN
THE DIALYSATE ONLY WITH TREATED WATER.BE SURE THAT ALL SPECIFICATIONS ARE SATISFIED.THE WATER SOURCE
MUST BE MONITORED PERIODICALLY TO DETECT FLUCTUATIONS IN WATER COMPOSITION AND QUALITY THAT COULD HAVE
AN ADVERSE EFFECT ON THE PATIENT,HEMODIALYSIS MACHINE OR DRY ACID DISSOLUTION UNIT.PARTICULAR ATTENTION
MUST BE TAKEN FOR CHEMICALS SUCH AS ALUMINUM
CHLORINE
AND CHLORAMINES
AS THESE CHEMICALS CAN CAUSE