
Using the DisChargePlus:
Testing should be done in a cool, clean, dry, and well-ventilated
environment. Position the discharger on stone, brick, concrete or
grounded metal.
DANGER: To reduce the risk of fire, do not use the
discharger near flammable materials or vapors.
The purpose of using a discharge unit on battery sets is to
determine battery capacity, and to find defective cells or batteries
in the set.
Battery energy is measured in minutes obtained upon discharging
a fully charged 36 or 48 volt battery set. This standard is used by
all major battery manufactures’ of lead acid deep cycle marine,
golf car, or floor scrubber type batteries. For a 36 volt pack, the
discharger can be set to a load of 25, 56, or 75 amps. The default
shut off voltage is 1.75 volts per cell, or 31.5 volts. For 48 volts,
the discharger can be set to a load of 25, or 56 amps - and the
shut off voltage is 42. If the shut off voltage is not reached in 245
minutes, the discharger will time out and shut off.
1. Before the test, fully charge the battery pack. It is important
that all the batteries in the pack are fairly even. Monitor each
battery by voltage towards the end of the charge cycle. Any
battery(s) that are 3/4 – 1 volt lower than the rest of the pack
should be charged individually.
2. To start the test, observe polarity, red+ black –, and connect
the clamps to the pack.
IMPORTANT: Make sure the posts are clean, and the clamps are
tight. Bad connections will cause the unit to fault, and the cycle
to be interrupted with a “system failure” message.
3. The solenoid will close, and the system voltage will be
displayed, along with the rate of discharge.Press the
START/STOP/PAUSE button. The cycle will start, and the running
time and battery volts will be displayed.
4. The cycle may be interrupted by pressing the button, and
restarted by pressing it again. If it is not pressed again, the tester
will shut off. The fan will continue to run for about 3 minutes after
shut down.
DANGER: Never move the clamps while the red LED is lit, the
resulting arcing can cause injury and damage to personnel and
equipment.
In case of an over temp, “OT” will be displayed, and the cycle will
pause until the temperature comes down, then restart. To avoid
intermittent starting and stopping, tests should be conducted at
temperatures below 95 degrees Fahrenheit.
5. When complete, minimum voltage, actual voltage, and cycle
time can be viewed by using the MENU button. Cycle time refers
to the time it took to reach 1.75 volts per cell. To figure capacity,
you must know how many minutes the battery is rated for at the
discharge rate selected, you then divide the actual time by the
rated time to find the percent capacity.
Example:
An 18 cell battery set is rated at 107 minutes at a 75 amp
discharge rate. The actual time it took to reach 1.75 volts per cell
is 90 minutes therefore capacity is 84%.