Fullwat LIR18650-26-CIT User manual

www.fullwat.com
LIR18650-26-CIT • High discharge battery

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
1. Scope
This specication describes the type, dimension, performance, technical characteristics, warnings and cautions of
the lithium ion rechargeable cell. The specication only applies to LIR18650-26-CIT cell supplied by FULLWAT®.
2.Denition
2.1Standardchargemethod
At 25 ± 2ºC, the cell is charged to 4.2V at constant current of 0.5C (1250mA), then charged at constant voltage
of 4.2V until the current tapers to 0.01C (25mA) followed by resting for 5min.
2.2Standarddischargemethod
Under 25 ± 2ºC, the cell is discharged to 2.75V at a constant current of 0.2C (500mA).
2.3Nominalcapacity
Nominal capacity, signed as capacity and using mAh as unit, is obtained by discharging a cell via standard
discharge method after it is standard charged.
3.Cellmodelanddimensions
3.1Descriptionandmodel
Description: Cylindrical Li-ion rechargeable cell.
Model: LIR18650-26-CIT.
3.2Celldimensions
Physical dimensions of the cell shown in the following gure (unit: mm), with tube.
Item Specication
DMax. 18.55
L Max. 70.00
units:mm
D
L

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
4.Cellcharacteristics
Unless otherwise specied, the cell is fresh cell and tested by standard charge and standard discharge.
Item Specications
Capacity
Nominal capacity 2600mAh by standard charge and discharge
Min. capacity 2500mAh by standard charge and discharge
Nominal Voltage 3.60V
Charge cut-off voltage 4.20V
Discharge cut-off voltage 2.75V
Max charge current 1C (25ºC, not for cycle life)
Max discharge current 3C (25ºC, not for cycle life)
Max. discharge current 8000mA pulse<5s at 25±3ºC
Storage temperature 1 year: -20~25ºC
3 months: -20~45ºC
1 month: -20~60ºC
Humidity range 0 ~ 60% RH (non-condensing)
Internal resistance ≤30 mΩ(AC Impedance, 1000Hz)
Weight ≤48g
5.Technicalrequirements
5.1Celloperatingconditions
Charge temperature: 0 ~ 45ºC
Discharge temperature: -20 ~ 60ºC
5.2Celltestingconditions
Unless otherwise specied, all tests stated should be done at 25 ± 2ºC
5.3Requirementofthetestingequipment
The voltage measurement device: not less than 0.5 grade.
The current measurement device: not less than 0.5 grade.
AC Impedance: 1000Hz
Temperature meter: precision ≤ 0.5ºC
The size measurement device: ±0.1%
The quality measurement device: ± 0.1%

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
5.4Electrochemicalcharacteristics
Unless otherwise specied, the cell should be fresh cell and tested by standard charge and discharge.
NO.
Item
Testmethodandconditions Criterion
5.4.1
Rate
discharge
capability
(1) Charge: 0.5C constant current
charge to 4.2 V followed by 4.2V
constant voltage charge to cut-off
current 0.01C;
(2) Discharge: 0.2C, 0.5C, 1C, 2C, 3C
constant current discharge to cut-off
voltage 2.75 V.
Current Relative capacity
0.2C 100%
0.5C ≥92%
1C ≥92%
2C ≥92%
3C ≥92%
5.4.2
RT cycle life
(1) Charge: 0.5C constant current
charge to 4.2 V followed by 4.2V
constant voltage charge to cut-off
current 0.01C;
(2) Discharge: 1C constant current
discharge to cut-off voltage 2.75V.
Discharge capacity of 801th cycle
Original discharge capacity ≥80%
(1) Charge: 0.5C constant current
charge to 4.15V followed by 4.15V
constant voltage charge to cut-off
current 0.05C;
(2) Discharge: 1C constant current
discharge to cut-off voltage 3.0V.
Discharge capacity of 1001th cycle
Original discharge capacity ≥80%
5.4.3
High-low
temperature
discharge
performance
(1) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C at 25±2ºC;
(2) Discharge: 0.2C constant current
discharge to cut-off voltage 2.75V at
25ºC, 60ºC, 45ºC, 10ºC, 0ºC, -10ºC,
-20ºC.
Temperature Relative capacity
25ºC 100%
60ºC ≥100%
45ºC ≥100%
10ºC ≥90%
0ºC ≥80%
-10ºC ≥75%
-20ºC ≥65%

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
NO.
Item
Testmethodandconditions Criterion
5.4.4
Storage
performance
at 25ºC
(100% SOC)
(1) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C;
Discharge:0.2C constant current
discharge to cut-off voltage 2.75V, to
obtain the initial capacity;
(2) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C;
(3) Stored at 25±21C for 28 days;
(4) Discharge: 0.2C constant current
discharge to cut-off voltage 2.75V, to
obtain the residual capacity;
(5) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C;
Discharge: 0.2C constant current
discharge to cut-off voltage 2.75V, to
obtain the recovery capacity.
Capacity Relative
capacity
Residual capacity ≥95%
Recovery capacity ≥97%
Initial capacity 100%
5.4.5
Storage
performance
at 60ºC
(100% SOC)
(1) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C at 25±2ºC;
Discharge: 0.2C constant current
discharge to cut-off voltage 2.75V at
25±2ºC, to obtain the initial capacity;
(2) Charge: 0.5C constant current
charge to 4.2V followed by 4.2V
constant voltage charge to cut-off
current 0.01C at 25±2ºC;
(3) Stored at 60 ± 2ºC for 7 days;
(4) Kept at 25 ± 2ºC for 5 hours;
(5) Discharge: 0.2C constant current
discharge to cut-off voltage 2.75V
at 25±2ºC, to obtain the residual
capacity;
(6) Charge: 0.5C constant current
charge to 4.2 V followed by 4.2V
constant voltage charge to cut-off
current 0.01C at 25±2ºC; Discharge:
0.2C constant current discharge to
cut-off voltage 2.75V at 25±2ºC, to
obtain the recovery capacity.
Capacity Relative
capacity
Residual capacity ≥90%
Recovery capacity ≥95%
Initial capacity 100%

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
5.5Environmentalandsafetycharacteristics
NO. Item Criterion Testingmethod Standard
5.5.1 Vibration No re,
No explosion,
No leakage,
The maximum
mass loss
≤0.1%
After standard fully charge, cell shall be attached to a
vibration table directly and subjected to vibration that
consists of 10Hz to 55Hz to 10Hz at the speed of 1Hz/
min in 90~100mins.The total excursion of the vibration
is 0.8mm (0.060 inches). The cell shall be vibrated in
each direction along axis of the cylinder and the vertical
directions of axis of the cylinder.
UL1642
5.5.2 Drop test No re,
No explode
The cell is charged following the standard charge
method; The cells is dropped three times from a height
of 1.0 m onto a concrete oor, to obtain impacts in
random orientations, at 25ºC±2ºC; Observation of 1 h.
IEC62133
5.5.3 Overcharge
test
No re,
No explode
The cell is discharged following the standard discharge
method. Apply a 8.4V power supply and a 2C charge
current for 24hrs.
UN38.3
5.5.4 Forced
discharge
test
No re,
No explode
Cell shall rst be discharged according to standard
discharge method, then the cell is subjected to a reverse
charge at 1C for 90 min at 25±2°C
IEC62133
5.5.5 130ºC hot
oven test
No re,
No explode
The cell is charged following the standard charge
method. After charging the cell is put in the oven.
And then the oven temperature will be ramped at
5ºC per minute to 130ºC and held at 130ºC. When
the temperature of the cell reach 130ºC, the cell is
maintained in the 130ºC oven for a maximum of 30
minute or until a re or explosion is obtained, whichever
comes rst.
GB31241
5.5.6 Crush test No re,
No explode
After charging a cell following the standard charge
method, the cell shall be crushed between two at
surfaces. The direction of the crushing force shall be
vertical to axis of the cylinder. The crushing force is to
be applied by a hydraulic ram with a 32mm diameter
piston. Crushing force is approximately 13KN. Once
the maximum pressure has been obtained it is to be
released.
UL1642
5.5.7 Short circuit
test at room
temperature
No re,
No explode
Cell shall rst be charged according to standard
charge method, and then cell is to be short-circuited by
connecting the positive and negative terminals of the
cell with copper wire having a maximum resistance load
of 80±20mΩ. This test is done at room temperature.
Monitor the cell temperature while testing. The cell is
continuously discharged until the cell case temperature
has returned to be 10ºC less then peak
temperature.
UL1642

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
NO. Item Criterion Testingmethod Standard
5.5.8 Short circuit
test at 55ºC
No re,
No explode
Cell shall rst be charged according to standard
charge method, and then cell is to be short-circuited by
connecting the positive and negative terminals of the
cell with copper wire having a maximum resistance load
of 80±20mΩ. This test is done at 55±5ºC. Monitor the
cell temperature while testing. The cell is continuously
discharged until the cell case temperature has returned
to be 10ºC less then peak temperature.
UL1642
5.5.9 Thermal
cycling
No re,
No explode
No leakage
After fully charged according to the standard charge
method, the cell is put in an oven. Then:
1. Raising the chamber temperature from Room
temperature to 70ºC within 30 min and keep the cell
under 70ºC for 4hrs.
2. Reducing the chamber temperature from 70ºC to
20ºC within 30 min and maintaing this temperature for
2hrs.
3. Reducing the chamber temperature from 20ºC to
-40ºC within 30 min and keep the cell under -40ºC for
4hrs.
4. Raising the chamber temperature from -40ºC to 20ºC
within 30min.
5. Repeat the sequence for a further 9 cycles.
Afterwards, the cell is observed for a minimum of 24hrs.
UL1642
5.5.10 Low pressure
test
No re,
No explosion,
No leakage,
The maximum
mass loss
≤0.1%
Cell shall rst be charged according to standard
charge method, and then cell is to be stored at an
absolute pressure of 11.6kPa for six hours at ambient
temperature.
UL1642
Note: All above safety tests will be conducted at 25ºC±2ºC except where specied differently. Use proper
ventilation with protective equipment.

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
6.Shipment
The cell shall be shipped in voltage range of 3.5~3.8V or in accordance with customers’ requirement. The remaining
capacity before charging shall be changed depending on the storage time and conditions.
7.Warranty
The warranty is specied in our warranties section of Terms of Sales. If the product is to be stored for more than
three months it is necessary to perform the appropriate maintenance to ensure the good condition of the batteries.
Consult our annex to the Terms of Sales on the recommended maintenance.
8.Storageandshipmentrequirement
Item Conditions Permissibletime
Storage environment -20ºC~ 60ºC, 60% RH Max Less than 1 month
-20ºC~ 45ºC, 60% RH Max Less than 3 months
-20ºC~ 25ºC, 60% RH Max Less than 1 year
Aboutlongtimestorage:
If the cell needs to be stored for a long time, the cell’s storage voltage should be 3.5 ~ 3.8 V. Also, it is recommended
to charge the cell every six months.
9.Warningandcautionsinhandlingthelithium-ioncell
Lithium-ion rechargeable cells subjected to abuse can cause damage to the cell and/or personal injury. Please read
and observe the standard cell precautions below before utilization.
Note1: The customer is required to contact FULLWAT® in advance, if and when the customer needs other
applications or operating conditions not described in this document.
Note2: FULLWAT® will take no responsibility for any accident when the cell is used under other conditions not
described in this document.

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
10.PCMspecication-Ref.model:PCM-L01S05-F45(1S)
NO. Item Condition Specication
1 Voltage
Charging voltage (CC/CV)- DC4.2V CC/CV
Cell-balance voltage ---
2 Current
Cell-balance current ---
Current consumption on operation for single cell ≤10μA
Maximal continuous charging current 3A
Maximal continuous discharging current 3A
3Overcharge
(single cell)
Overcharge detection voltage 4.25±0.05V
Overcharge delay time 0.5—2S
Overcharge release voltage 4.15±0.05V
4Overdischarge
(single cell)
Over discharge detection voltage 2.50±0.1V
Over discharge delay time 10—200mS
Over discharge release voltage 2.50±0.1V
5Discharge
overcurrent
Discharge overcurrent protection 5.5±1A
Discharge overcurrent delay time 5ms—60ms
Release condition Cut-off-load,reset automatically
6Short-circuiting
protection
Detection condition Exterior short circuit
Load short-circuiting detection delay time 200-800us
Release condition Cut-off-load, reset automatically
7 Internal resistance Proection circuit (MOSFET) ≤50mΩ
8 Temperature
Wide range of operation temperature -40~+85ºC
Storage Temperature Range -40~+125ºC
10KNTC
P+=B+=Charge+/Discharge+
P-=Charge-/Discharge-
Size: 117*17*2 mm (L*W*T)
Battery

www.fullwat.com
LIR18650-26-CIT •High discharge batterty
Warnings!
To prevent the possibility of the cell from leaking, heating, explosion, please observe the following precautions:
(It should be indicated especially in manual or instruction for users.)
• Do not use and leave the cell near a heat source such as re or heater.
• Do not use or leave the cell under the blazing sun (or in heated car by sunshine).
• Do not use or leave the cell at very high temperature conditions (e.g., strong direct sunlight or a vehicle in
extremely hot conditions). Otherwise, it can overheat or catch re or its performance will be degenerate and its
service life will be decreased.
• Do not short circuit, over-charge or over-discharge the cell.
• Don’t immerse the cell in water and seawater. Please put it in cool and dry environment if no using.
• Don’t reverse the positive and negative terminals
• Do not disassemble or modify the cell.
• Do not transport or store the cell together with metal objects such as necklaces, hairpins, coins, etc.
• Make sure the cell is not with conspicuous damage or deformation.
• Don’t connect the cell to an electrical outlet directly.
• If the cell leaks and the electrolyte splashes into the eyes, rinse the eyes with clean running water immediately
for at least 15 minutes, and go to hospital for treatment if necessary.
• Mixed use of cells of different types is not allowed.
• Keep the cell away from babies.
• Do not directly solder the cell and pierce the cell with a nail or other sharp object.
• Do not strike, throw or trample the cell.
• Use the cell charger specically for that purpose when charging.
• Please separate cells of different electrochemical systems from one another when disposing of secondary cells.
• Clean the terminals with a dry cloth before use if the cell terminals are dirty. Otherwise power failure or charge
failure may occur due to the poor connection with the instrument.
• Cells should be removed from the device or charger immediately and not used again if they are over heat, give
off odor, discolor or deform, or appear abnormally in any way during use, charging and storage.
• The cell replacement shall be done only by either cell supplier or device supplier instead of the user.
• Please tape the terminals to insulate cells before discarding them in case of re and explosion.
• Do not use cells in strong electrostatic and magnetic occasions, otherwise, it can cause safety problems easily.
• Use of damaged cells is not permitted.
• Make sure package designing will not cause cell damages.
• Cell packing should be conducted strictly according to level range, any misuse of different levels should not be
permitted.
• Disassembling cells from pack or module is not permitted unless under the guidance of professional technicians.
11.Restrictionofhazardoussubstances
This model of lithium-ion cell is in accordance with our company’s request of “ The hazardous substances and
material management standard ” or customer’s requirements.
Table of contents
Popular Batteries Pack manuals by other brands

Ravpower
Ravpower RP-PB31 user manual

Treasure Garden
Treasure Garden P-AKZPBATTERY Assembly and operation guide

Sony
Sony NP-F950/B operating instructions

King Innovation
King Innovation Insta-Light 75500 quick start guide

Tektronix
Tektronix THSBAT instructions

LG Chem
LG Chem RESU Series installation manual