1. Overview
TheG9205 areportabletrueRMS AC/DCclampmeters with
automaticrange. TheyaredesignedaccordingtoEN61010-1
CAT II600V/CAT III300Vsafetystandards,andcomewithfull-function
protectionwhichensuresusers a safeandreliablemeasurement
experience. Aside from basic measurement functions, they also have
highprecisioncurrentscaleandhighvoltagefrequencymeasurement
extension.
G9205
600A AC/DC Clamp Meter
User Manual
Preface
Thank you for purchasing the new . In order to use this
product safely and correctly, please read this manual thoroughly,
especially the part.
After reading this manual, it is recommended to keep the manual at
an easily accessible place, preferably close to the device, for future
reference.
Limited warranty and liability
GAZELLE guarantees that the productis freefrom any defect in material
andworkmanshipwithinoneyearfromthepurchasedate. Thiswarranty
does not apply to damages caused by accident, negligence, misuse,
modification, contamination and improper handling. The dealer shall
notbeentitledtogiveanyotherwarrantyonbehalfofGAZELLE.If
you need warranty service within the warranty period, please contact
your seller directly.
GAZELLE will not be responsible for any special, indirect, incidental
emossA.ecivedsihtgnisuybdesuacssolroegamadtneuqesbusro
countriesorregionsdonotallowlimitationsonimpliedwarrantiesand
incidentalorsubsequent damages, theabovelimitationof liabilitymay
not apply to you.
2. Features
True RMS measurement
Audio visual NCV electric field detection
High voltagefrequencyrange:10Hz~60kHz; low voltagefrequency
range: 60Hz~10MHz
G9205 AC/DCcurrentrange:60A, 600A,frequencyresponse:
45Hz~400Hz, live/neutral wire detection
ACA/DCA mode memory function for current measurement
Largecapacitance(G9205:60mF)andtemperature measurement
LargeLCDandfastrefreshrate(3times/s),capacitancemeasurement
response time:
≤1mF: less than 3s; ≤10mF: about 6s; ≤60mF: about 8s
Full-functionfalsedetectingprotectionforupto600V(3.6kVA)
energy surge; overvoltage and overcurrent alarm
Thepowerconsumptionwithoutbacklightisabout1.8mA. The
circuit has an automatic power saving function. The consumption
in sleep state is <11uA, which effectively extends the battery life to
400 hours.
Please readthe safety andwarningparts in this manual
thoroughly.
Caution:Pleasereadthesafetyinstructionscarefullybeforeuse.
3. Standard Accessories
Open the package and check the below items, if any is missing or
damaged, please contact your supplier immediately:
User manual ----------------------- 1 pc
Test leads --------------------------- 1 pair
K-type temperature probe ------ 1 pc
Cloth bag ---------------------------- 1 pc
4. Safety Instructions
6. General Specifications
Max LCD display counts: 6099 (G9205)
Polarity display: Auto
Overload display: “OL” or “- OL”
Low battery indication: “ ” means low battery
Low battery shutdown prompt: The “Lo.bt” interface appears on LCD
and lasts for about 10s, the buzzer beeps 3 times, and the meter
automatically shuts down.
Test position error: If the source under test is not placed at the center
of the clamp jaws when measuring current, ±1.0% additional error in
reading may be produced.
Drop protection: 1m
The maximum size of jaw opening: 28mm in diameter
Battery: AAA 1.5V battery × 2
Auto power off (adjustable): The meter will auto power off if there is no
operation for 15 min.
Dimensions: 215mm × 63.5mm × 36mm
Weight: About 235g (including batteries)
Altitude: 2000m
Operatingtemperatureandhumidity:0°C~30°C (≤80%RH),30°C~40°C
(≤75%RH), 40°C~50°C (≤45%RH)
Storage temperature and humidity: -20°C~+60°C (≤80%RH)
Electromagneticcompatibility:RF=1V/m,overallaccuracy=specified
accuracy+5% of range; RF>1V/m, no specified calculation
7. External Structure (Picture 1)
1) NCV sensing end
2) Clamp jaws
3) Hand guard
4) LED indicator
5) Jaw opening trigger
6) Function scale knob
7) LCD display
8) Function buttons
9) Signal input jack (red and positive +)
10)COM input jack (black and negative -)
8. Button Description
SELECT Button
In composite scale, press this button to switch between the corresponding
functions or ranges;
HOLD/BACKLIGHT Button
Shortpressthisbuttontoenter/exit thedataholdmode,andlongpress
(about 2s) this button to turn on/off the backlight.
MAX/MIN Button
(valid for ACV/DCV, ACA/DCA, “C/”F, resistance and capacitance scales)
Shortpressthisbuttontoenterthemaximum/minimummeasurementmode
and long press this button to exit.
REL Button
(valid for ACV/DCV, DCA, “C/”F and capacitance scales).
Press this button to store the current reading as a reference for future
readings. When the LCD display value is reset to zero, the stored reading
will be subtracted from the future readings. Press this button again to exit
the relative value mode.
9.Operating Instructions
9.1 AC/DC Current Measurement (Picture 2)
Select the corresponding current range.
Pressthetriggertoopentheclampjaws,andfullyencloseoneconductor.
Only one conductor can be measured at a time, or the measurement
reading will be wrong.
Caution:
Do not insert thetestingleadsduring current measurement to avoid
electric shock.
The current measurement must be taken with safeguard protection.
PresstheRELbuttontoreturntozerobeforetheDCcurrent
measurement,andmeanwhilethecenterholeofjawshouldbe
perpendicular to the current direction to ensure accuracy.
The open circuit zeroing reading may be relatively large after (high)
noitcetedtnerrucCAehtmrofrepesaelP.tnemerusaemtnerrucCD
againtocounteracttheremanencesignalbyalternatingelectricfield.
9.2 AC/DC Voltage and Voltage Frequency (% duty cycle)
Measurement (Picture 3)
Select the corresponding function scale.
Insert the red test lead into the “positive” jack, and the black into
the “COM” jack.
Connect test leads with both ends of the measured objects.
Caution:
Do notinputvoltageabove600Vtopreventelectricshockordamage.
Theinputimpedanceofeachrangescaleis10MΩ, thisloadeffect
in high resistance measurementmaycause error. Ifthe input
impedanceislowerthan10kΩ, theerrorcanbeignored (≤0.1%).
Be cautious to avoid electric shock when measuring high voltage.
Please checkthe functionsbyapplying a known voltage before use.
9.3 Continuity Test/Resistance/Diode/Capacitance
Measurement (Picture 4)
Select the corresponding function scale.
Insert the red test lead into the “ ” jack, and the black
into the “COM” jack.
Connect test leads with both ends of the measured object.
Caution:
Do not input voltage above DC 60V or AC 30V to avoid personal
injury.
Pleasedisconnectalltheotherpartsofthecircuittoavoidinaccuracy.
Before the resistance online measurement, please do disconnect
allthepowerandfullydischargeallcapacitorstoavoidinjuryor
device damage.
If theresistanceisover0.5Ωwhenthetest leadsareshort-circuited,
please check the test leads for looseness or other abnormalities.
Ifthemeasuredresistorisopenortheresistanceexceedsthe
maximum range, the LCD will display “OL”.
Measuredvalue = displayedvalue – probeshortcircuitvalue
It isrecommendedtouse“REL”measurementmodeforcapacitance
less than 100nF.
9.4 Temperature Measurement (Picture 5)
Select the temperature measurement function scale.
eborperutarepmetehtxif,retemehtotelpuocomrehtepyt-KtresnI
on the measured object, and read after the value is stable.
5. Symbols
Current Measurement
Picture 2
Capacitance
Continuity
Diode
Resistance
Picture 4
(G9205)
Picture 5
Picture 3
Red Black
Red Black
Red
Black
WARNING
To ensure safe operation and service of the tester, follow these
instructions. Failure to observe these warnings can result in severe
injury or death.
ThemeterisdesignedaccordingtoEN61010-1,61010-2-032/033
andelectromagneticradiationprotectionEN61326-1safetystandards,
and conforms to double insulation,CAT II 600V, CAT III 300V and
pollution grade II. Incasethe meter is not used properly as instructions,
the protection provided may be weakened or lost.
Beforeeachuseverifytesteroperationbymeasuring a knownvoltage.
Before use, please check if there is any item which is damaged or
behaving abnormally. If any abnormal item (such as bare test lead,
damaged meter casing, broken LCD, etc.) is found, or if the meter
is considered to be malfunctioning, please do not use the meter.
Do notusethemeteriftherearcoverorthebatterycoverisnot
covered up, or it will pose a shock hazard!
Keep fingersbehind the fingerguardsand awayfromthe metal
probe contacts when making measurements.
The function switch should be placed in the correct position before
measurement. It is forbidden to change the position during
measurement to avoid damage to the meter!
Do not apply voltage over 600V between any meter terminal and
earth ground to prevent electric shock or damage to the meter.
Use caution when working with voltages above AC 30Vr.m.s,
42Vpeak or DC 60V. Such voltages pose a shock hazard.
Never input voltage or current which exceeds the specified limit.
Maximum range should be selected if measured value is unknown.
Beforemeasuringresistance, diodeandcontinuity, pleasedisconnect
all power and fully discharge all capacitors to avoid inaccuracy.
Whenthe “ ” symbolappearsontheLCD,pleasereplacethe
batteriesintimetoensuremeasurementaccuracy. Batteriesshould
be removed If the meter is long-term idle.
Do not change the internal circuit of the meter to avoid damage to the
meter and user!
Do notexposethemeterinhightemperature/humidity, flammable,
explosive or intense magnetic environments.
Cleanthemetercasingwith a softclothandmilddetergent.Donotuse
abrasives or solvents!
Description
Caution, possibility of electric shock
Alternating current
Direct current
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
Earth (ground) TERMINAL
Warning or Caution
Symbol
It is applicable to test and measuring circuits connected
directly to utilization points (socket outlets and similar
points) of the low-voltage MAINS installation.
It is applicable to test and measuring circuits connected
to the distribution part of the building'slow-voltage
MAINS installation.
CAT III
CAT II
Picture 1
REL/ZERO
REL/ZERO
REL/ZERO
G9205
60.00A
600.0A
0.01 A
0.1 A
±(2%+5)
G9205
600.0mV
6.000V
60.00V
600.0V
0.1mV
0.001V
0.01V
0.1V
±.0 7 3
)
± .0 5 2
)
G9205
6.000V
60.00V
600.0V
0.001V
0.01V
0.1V
0.01Hz 0.01kHz
±.1 0 5
)
± .0 8 5
)
±( . %+ )0 5 2
Voltage Frequency
10Hz 60KHz
0 1.
0 001V.
G9205
0 1.
0 001K.
0 01K.
0 1K.
0 001M.
0 01M.
±( . +1 0 2)
±( . +0 8 2)
±%+ )2 5 5
G9205
60.00nF
600.0nF
6.000uF
60.00uF
600.0uF
6.000mF
60.00mF
0.01nF
0.1nF
0.001mF
0.01mF
±( %+ )4 5
± %10
10Hz 10MHz 0.01Hz 0.01MHz ±(0.1%+4)
10.9 NCV
NCV
EFLo
EFHI
-40 40
40 500
500 1000
-40 104
104 932
932 1832
±4
±(1.5%+5
)
±(2.0%+5)
±6
±(2.0%+6)
±(2.5%+4)
1℃
1℉
9.5 NCV AC Electric Field Sensing (Picture 6a)
The electric field sensing sensitivity is divided into two levels
(“EFHI” and “EFLo”). The meter defaults to “EFHI”. Users can select
different sensitivity levels according to the intensity of the measured
electric field. Select "EFHI" of NCV when electric field is around 220V
AC .Bring the NCV sensing end close to a charged electric
field (socket, insulated wire, etc.). The LCD will display the segment
derusaemehtfoytisnetniehtsA.gnihsalfDELderdnaspeebhtiw"-"
electric field increases, the more segments (----) display , the higher
frequency of buzzer beeps and LED flashes. Select “EFLo” when the
electric field is around 110V
Caution:
Use the NCV sensing end to approach measured electric field,
otherwise the measurement sensitivity will be affected. When the
ehtrehtehwevresbo,CAV001revosiegatlovdleifcirtcelederusaem
conductor is insulated to avoid personal injury.
9.6 Others
Auto power off: The meter will automatically power off to save power
if there is no operation for 15 min. You can wake it up by pressing
any button or restart it after turning the switch to OFF.
Press and hold the SELECT button in off state and then turn on the
meter again to disable the auto power off function. Restart the meter
after shut it down to resume this function.
Buzzer: When any button is pressed or the function switch is turned,
if it is valid, the buzzer will make one beep (about 0.25s). The buzzer
will also beep intermittently to indicate the over range during the
voltage or current measurement.
Low battery detection: The battery voltage will be automatically
detected as long as the meter is on. If it is lower than 2.6V, the LCD
will display the “ ” symbol.
Low battery shutdown function: When the battery voltage is lower
than 2.5V, the LCD displays the “ ” symbol, the “Lo.bt” interface
appears and lasts for about 10s, the buzzer makes consecutive
beeps 3 times, and then the meter automatically shuts down (no
is displayed).interface
11. Technical Specifications
Accuracy: ± (%reading + counts), the calibration period is 1 year.
Ambient temperature and humidity: 23°C±5°C; ≤80%RH.
Temperature coefficient: the accuracy assured temperature condition
is 18°C-28°C, the range of ambient temperature fluctuation is stable
within ±1°C. When the temperature is less than 18°C or over 28°C,
the additional temperature coefficient error is 0.1 x (specified
10.1 Current Measurement
AC Current
Range
Resolution Accuracy
Caution:
Accuracy guarantee range: 5%~100% of range
When the measured current reaches the warning value, there
will be an alarm sound (G9205: 610A)
With DC current DCA mode, LCD may display non-zero value
in open circuit state, users can press "REL" button to clear display
to zero before each measurement.
10.2 Voltage Measurement
DC Voltage
Range Resolution Accuracy
AC Voltage/Voltage Frequency
Range
Resolution Accuracy
Caution:
G9205 : long press "SELECT" to t the Hz function, the input
range over 5V.
The input impedance is about 10MΩ
value
Accuracy guarantee range: 1%~100%
AC crest factor of non-sinusoidal wave can reach 3.0 at 4000 counts
while can only reach 1.8 at 6000 counts, the additional error should be
added for the corresponding crest factor as follows:
A. Add 3% when the peak factor is 1 ~ 2
B. Add 5% when the peak factor is 2 ~ 2.5
C. Add 7% when the peak factor is 2.5 ~ 3
10.3 Continuity/Diode Measurement
Function Range Resolution Accuracy
≤
The median: uncertain
The open circuit voltage is
about 4V
For the silicon PN junction
diode, the voltage value is
generally about 0.5~0.8V.
10.4 Resistance Measurement
Range Resolution Accuracy
Caution:
Measured resistance value = displayed value – resistance value
of short circuited test leads
Open circuit voltage is about 1V
Overload protection: 600Vrms
10.5 Capacitance Measurement
Range Resolution Accuracy
Caution:
Measured value = displayed value - open circuit value of the test leads
(For capacitance ≤100nF, “REL” mode is recommended, open circuit
has residual reading). The guaranteed accuracy is 1%~100%.
Overload protection: 600Vrms
10.6 Frequency Measurement
Range Resolution Accuracy
Caution:
Measurement sensitivity:
≤100kHz: 200mVrms≤ input range≤30Vrms;
>100kHz~1MHz: 600mVrms≤ input range≤30Vrms;
>1MHz~10MHz: 1Vrms≤ input range≤30Vrms.
Duty ratio is only applicable to ≤10kHz square wave measurement
with a range of 1Vp-p:
If frequency≤1kHz, duty cycle will be 10.0%-95.0%;
If frequency>1kHz, duty cycle will be 30.0%-70.0%.
10.8 Temperature Measurement (G9205)
Range Resolution Accuracy
Caution:
The meter displays "OL" after startup, it is only suitable for K-type
thermocouple (Nickel-Chromium ~ Nickel-Silicon temperature
F.
The formula for Celsius to Fahrenheit is °F =1.8°C + 32.
Range
Electric field
sensing
sensitivity level
Accuracy
The electric field sensing sensitivity
is divided into two levels (“EFHI” &
“EFLo”). The meter defaults to “EFHI”.
a) AC voltage above 24V±6V can be
sensed. "EFLo" mode is recommended
when the power frequency voltage is
110V.
b) "EFHI" can be set in 220V condition.
AC voltage above 74V±12V can be
sensed with getting close to wires,
and identify whether the main socket
wire of socket according to the intensity
of sensing.
Note: Test results may be affected by
different socket designs or wire I
nsulation thickness.
G9205
60.00A
600.0A
0.01 A
0.1 A
±(2%+5)
DC Current
Range Resolution Accuracy
11. Maintenance
Warning: Before opening the rear cover of meter , remove test
leads to avoid electric shock.
11.1 General Maintenance
Clean the meter casing with a soft cloth and mild detergent. Do
not use abrasives or solvents!
Do not use the tester or test leads if they appear to have any
abnormality.
The maintenance and service must be implemented by qualified
professionals or designated departments.
11.2 Battery Replacement (Picture 7)
When the “ ” symbol appears on the LCD, please replace the
batteries in time to ensure measurement accuracy. Batteries
specification:
2 standard AAA 1.5V batteries.
Operation
Turn off the meter and remove the test leads from the input terminals.
Unscrew the screw of the battery compartment, remove the battery
cover, and take out the used batteries as shown.
Replace the 2 standard AAA batteries according to the polarity
indication.
Picture 7
Picture 6a
10.Probe usage
TESTING IN CAT III MEASUREMENT LOCATIONS
Ensure the test lead shield pressed firmly in place. Failure to use
the CAT III shield increases arc-flash risk.
TESTING IN CAT II MEASUREMENT LOCATIONS
CAT III shields may be removed for CAT II locations. This will allow
testing on recessed conductors such as standard wall outlets.
Take care not to lose the shields.
0.1 99% 0.1% ±(3%+5)
Range Resolution Accuracy
10.7 Duty Ratio Measurement
Made in China

Other Gazelle Measuring Instrument manuals

Gazelle G9204 - II User manual

Gazelle

Gazelle G9204 - II User manual

Gazelle G9405-Log User manual

Gazelle

Gazelle G9405-Log User manual

Gazelle G9407-Log User manual

Gazelle

Gazelle G9407-Log User manual

Gazelle G9406-Log User manual

Gazelle

Gazelle G9406-Log User manual

Popular Measuring Instrument manuals by other brands

Levenhuk Ermenrich Verk LY80 user manual

Levenhuk

Levenhuk Ermenrich Verk LY80 user manual

Agilent Technologies E444xAU Option H37 installation guide

Agilent Technologies

Agilent Technologies E444xAU Option H37 installation guide

Tecsis E3906 operating manual

Tecsis

Tecsis E3906 operating manual

Horiba Scientific LAQUAact ES-71 instruction manual

Horiba Scientific

Horiba Scientific LAQUAact ES-71 instruction manual

Levitronix LEVIFLOW LFS-SU user manual

Levitronix

Levitronix LEVIFLOW LFS-SU user manual

Allsheng OZYME Fluo-100 Series Operation manual

Allsheng

Allsheng OZYME Fluo-100 Series Operation manual

GREISINGER GLF-100 operating manual

GREISINGER

GREISINGER GLF-100 operating manual

LaserLine GL3000PMC owner's manual

LaserLine

LaserLine GL3000PMC owner's manual

LaserLiner CondenseSpot Pro manual

LaserLiner

LaserLiner CondenseSpot Pro manual

Scarlet Tech WR-3 user guide

Scarlet Tech

Scarlet Tech WR-3 user guide

LaserLiner SuperSquare-Laser manual

LaserLiner

LaserLiner SuperSquare-Laser manual

Presys DMY-2030-CC Technical manual

Presys

Presys DMY-2030-CC Technical manual

REC ACTIVE BMS 4S manual

REC

REC ACTIVE BMS 4S manual

Endress+Hauser Cerabar S PMC71 Functional safety manual

Endress+Hauser

Endress+Hauser Cerabar S PMC71 Functional safety manual

Teledyne Lecroy M21 quick start guide

Teledyne Lecroy

Teledyne Lecroy M21 quick start guide

Rover Instruments CNG 90 user guide

Rover Instruments

Rover Instruments CNG 90 user guide

LINSHANG LS201 user manual

LINSHANG

LINSHANG LS201 user manual

SIKA E2 operating manual

SIKA

SIKA E2 operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.