gefran 40B 48 User manual

99
10
45
45
48
70
70
63
48
105
10
108
96
48
44,5
92
70
115
Triac output
(option, only for 96 format)
20...240Vac ±10%, 3A max
Snubberless, inductive and resistive load (I2t = 128A2s)
Transmitter / Sensor Power Supply 1.2 Vdc for potentiometer > 100 Ω
5 Vdc, 10 Vdc max. 120mA, (for strain-gauge)
15 Vdc, 24 Vdc max. 50mA, (for transmitter)
Fault settings Alarm states can be configured in probe fault
condition
Analog retransmission (option) 4 to 20mA, max. 150 Ωload
Logic input Ri = 5,6KΩ(24V, 4mA), 1500V isolation
Logic input functions configurable for alarm memory reset, hold, flash,
zero, selection of max., min. peak value, peak-peak
Power supply (switching)
40B 48 (std) 100...240Vac ±10%, 8VA
(optional) 20...27Vac/dc ±10%, 8VA
40B 96 (std) 100...240Vac/dc ±10%, 10,5VA
(optional) 11...27Vac/dc ±10%, 8VA
For correct and safe
installation, follow the
instructions and observe
the warnings contained
in this manual.
1 • INSTALLATION
• Dimensions and cut-out: Panel mounting
Panel mounting:
Fix the device with the bracket provided before making any electrical connections.
To mount two or more devices side by side, use the cut-out dimensions shown above.
CE MARKING: EMC (electromagnetic compatibility) conformity to EEC Directive
89/336/CEE with reference to the generic Standard EN50082-2 (immunity in industrial
environments) and EN50081-1 (emission in residential environments). BT (low voltage)
conformity to Directive 73/23/CEE as modified by Directive 93/68.
MAINTENANCE: Repairs must be done out only by trained and specialized personnel.
Cut power to the device before accessing internal parts.
Do not clean the case with hydrocarbon-based solvents (Petrol, Trichlorethylene, etc.).
Use of these solvents can reduce the mechanical reliability of the device. Use a cloth
dampened in ethyl alcohol or water to clean the external plastic case.
SERVICE: GEFRAN has a service department. The warranty excludes defects caused
by any use not conforming to these instructions.
2 •TECHNICAL SPECIFICATIONS
FUNCTION CABLE LENGTH USED
Input 1 mm23 m
Power supply cable 1 mm21 m
Relay output cables 1 mm23,5 m
EMC conformity has been tested with the following connections
!
40B 48 / 40B 96
POSITION, FORCE and PRESSURE INDICATOR-INTERCEPTOR
with INPUT for STRAIN-GAUGE and POTENTIOMETER
SOFTWARE VERSION 1.0x / 2.0x
code 81651C / edition 09 - 03/08
USER’S MANUAL
Faceplate protection IP65
Working / Storage temperatures 0 to 50°C / -20 to 70°C
Relative humidity 20 to 85%, non-condensing
Installation Panel mounting, extractable from front
Weight
160g (mod. 48); 320g (mod. 96) for the complete version
Relay contact NO (NC) 5A, 250V
Logic output 11Vdc, Rout = 220Ω(6V/20mA)
Fuse (inside device, not
operator serviceable) 100 to 240Vac/dc - type T - 500mA - 250V
11 to 27Vac/dc - type T - 1.25A - 250V
Resolution
function of settable sample time
>13-bit, s.t. 120 msec. with probe power supply control
>12-bit, s.t. 30 msec.
(60msec with probe power supply control)
>11-bit, s.t. 15 msec.
(30msec with probe power supply control)
Linear scale ranges
-1999 to 9999 (with 4 digit display)
-999 to 999 (with 3 digit display - only for model 96)
Configurable decimal point position, possible 32
segment linearization
Alarms (set points) Maximum of three configurable alarms:
absolute, deviation, symmetrical deviation.
Adjustable hysteresis
Alarm masking
ability to:
- exclude on power-up
- latch reset from key and/or external contact
- insert delay filter (DON, DBI, DOF, DPO)
- set minimum intervention time
Display
3, 4 digit red LED’s
mod. 48 digit height 10mm (4 digits)
mod. 96 digit height 20mm (3 digits),
digit height 14mm (4 digits)
Keys 3 mechanical keys (Raise, Lower, F)
Accuracy
0.2% f.s. at 25°C ambient temperature, ts=120msec
Main input
Differential input:
- from strain-gauge 350Ω(for pressure, force, etc.),
sensitivity 5mV/V with strain-gauge max power
15V, (7.5mV/V with max power 10V - 15mV/V with
max power 5V), positive or symmetrical
polarization, calibration with automatic calibration
of sensitivity, interrupted probe power signal
- from potentiometer with 1.2V ≥100 Ωpower
supply
1
81651C_MHW_40B48-96_0308_ENG

• Input for Strain Gauge sensor (6 wires) (only for version
SW2.0x) for connection to Melt transducers
Analog output for
retransmission
4 to 20mA, Rmax. 150Ω
Generic user-
configurable output
- relay 5A/250Vac Out3
Isolated digital input
1500V
Ri = 5,6KΩ
(24V, 4mA)
6
5+
-
External
supply
6
5
6
5+
-
R
“Raise” and “Lower” keys:
These keys are used for any operation that requires a
numerical parameter to be raised or lowered. ••The
speed of change is proportional to the time the key is
pressed. •• The operation is not cyclic: once the
maximum (minimum) limit is reached, there will be no
further increase (decrease) of the value, even if the
key remains pressed.
The keys can be configured to perform reset, hold,
display of the peak value, etc. as determined by the
‘t.U.’and ‘t.d.’parameters on the ‘In’menu.
3 •DESCRIPTION OF FACEPLATE
Function key:
Gives access to different configuration stages ••
Confirms any parameter changes
Label with engineering units
Indication of output states:
OUT 1 (Alarm 1); OUT 2 (Alarm 2);
OUT 3 (Alarm 3)
Standard:
100 to
240Vac/dc±10%
Optional:
11 to 27Vac/dc ±10%
50/60Hz
6
5
4
3
2
1
7
8
9
10
11
12
18
17
16
15
14
13
19
20
21
22
23
24
•Power supply
11
10
~
~
!
•Logic input / Retransmission output / Relay output
•Inputs
PWR
Indication of output states:
OUT 1 (Alarm 1); OUT 2 (Alarm 2);
OUT 3 (Alarm 3)
1145 108397261
1219 18 131520 141621 1722
Standard:
100 to 240Vac
±10%
Optional:
20 to 27Vac/dc ±10%
50/60Hz
•Power supply
23
24
~
~
PWR
!
Label with engineering units
•Outputs
Generic user-
configurable
outputs
- relay 5A/250Vac
- logic 6V/20mA,
Rout = 220Ω
19
21
20
22
-
+
Out2
-
+
Out1
4 •CONNECTIONS
Digital input
1500V isolation
1500V
Ri = 5,6KΩ
(24V, 4mA)
6
5+
-
•Logic Input
Analog
retransmission
output
4...20mA,
Rmax. 150Ω
8
7+
-
•Retransmission output
•Strain-gauge probe input (4 wires)
4
2
3
1
- IN
+ IN
4
2
3
1
•Potentiometer input
+
-
Pot
+-
Strain-gauge
- IN
+ IN
+
-
R
1,2V
External
supply
PV display: Indication of process variable
•• Indication of ‘HI’or ‘Lo’out of range
•• Indication of open circuit (br) or short circuit (Er)
•• Display of configuration and calibration messages
•• Interrupted probe excitation signal (Eb)
PV display: Indication of process variable
•• Indication of ‘HI’or ‘Lo’out of range
•• Indication of open circuit (br) or short circuit (Er)
•• Display of configuration and calibration messages
•• Interrupted probe excitation signal (Eb)
Generic user-
configurable outputs
- relay 5A/250Vac
for Out1, Out2
- logic 6V/20mA,
Rout = 220Ω
- Triac
20...240Vac ±10%
3A max
•Outputs
12
14
13
15
Out3
-
+
Out2
16
17
-
+
Out1
~
~
Line
Load
TRIAC
4
2
3
1
- IN
+ IN
+
-
+-
Strain-gauge
5
6
12
13
CAL
CAL
mod. 48mod. 96 (*) used for internal calibration
(*)
GREEN E
WHITE E
BLACK or
YELLOW S
RED S
BROWN or
ORANGE
BLUE
281651C_MHW_40B48-96_0308_ENG

Device structure: identification of boards
-999 ... +999
scale points
•Information display
5 •PROGRAMMING and CONFIGURATION
(*) For deviation alarms the setpoint range is -999 ... 999
If the Inc, Dec, F keys are not pressed within 15 sec.,
the display returns to the P.V. value
I F
P.V. Process variable
Alarm Setpoint Output 2 (*)
Alarm Setpoint Output 3 (*)
Password
C F
P A
PA = 99
P r
U.C.
Custom linearization
User calibration
YES
NO
Alarm Setpoint Output 1
LEVEL 1 DISPLAY
L n
Hysteresis parameters Protection code
Information display
I F
Software version
Display CPU type
(potentiometer/strain-gauge)
and number of outputs available
b3= with 3 outputs
Information display
I n Input settings
O u Output settings
•Configuration parameters
C F
Hysteresis for set point 1
Hysteresis for set point 2
Hysteresis parameters
-999 ... +999
scale points
-999 ... +999
scale points
Hysteresis for set point 3
FPressed for
approx. 2 sec.
Keep the F key
pressed to browse
the menus.
Release the F key
to enter the
displayed menu.
Press the F key to
access the
parameters.
Keep the F key
pressed to exit any
menu at any time.
S1 A B
S5
S1 = Status of Out 3 relay
A = Direct
B = Inverse
S5 = ON with digital input
CPU BOARD
CPU BOARD
POWER SUPPLY
BOARD DISPLAY BOARD
S8
S2
S9
S1
S1
S3
S2
B
A
B
A
B
A
CPU BOARD + POWER
SUPPLY BOARD
OUTPUT BOARD
S1 = Status of Out 1 relay
S2 = Status of Out 2 relay
S3 = Status of Out 3 relay
A = Direct
B = Inverse
CPU BOARD + POWER
SUPPLY BOARD
OUTPUT BOARD
DISPLAY BOARD
S1 S2 S8 S9
1V OFF OFF OFF ON
5V ON OFF OFF OFF
10V OFF ON OFF OFF
15V OFF OFF ON OFF
24V OFF OFF OFF OFF
Sensor supply
POWER SUPPLY BOARD
24V
15V
10V
5V
1,23V
R20
S2
N.B. : you can keep the OUT1 relay energized at
power-up by inserting jumper S2 and removing
resistance R20.
3
81651C_MHW_40B48-96_0308_ENG

•Output parameters
0 None
1 Zero
2 Hold
3 Flash
4 Max. peak display
5 Min. peak display
6 Delta peak display
7 Peak memory reset
8 Zero + peak memory reset
9 Alarms reset
10 Peak + alarms reset
11 Zero + alarms reset
12 Peak + zero + alarms reset
13 Output status OUT1 / check CAL (*)
14 Output status OUT2 / check CAL (*)
15 Output status OUT3 / check CAL (*)
d. i. - t. u. - t. d.
Offset correction
of main input
Select function of
digital input
High limit for the alarms
set point
-999 to 999
scale points
L.S to H.S
Raise key function
(active only in
P.V.selection)
Low limit for the alarms
set point L.S to H.S
Lower key function
(active only in
P.V.selection)
Value Direct Absolute Normal
(max.) Relative to Symmetrical
Inverse previous (window)
(min.)
0 Direct Absolute Normal
1 Inverse Absolute Normal
2 Direct Relativa Normal
3 Inverse Relativa Normal
4 Direct Absolute Symmetrical
5 Inverse Absolute Symmetrical
6 Direct Relativa Symmetrical
7 Inverse Relativa Symmetrical
+64 output for calibration of 6-lead sensor
(from version SW 2.0x)
Output settings
Number of outputs
Alarm type 1
0 to 3
1. t - 2. t - 3. t
Alarm type 2
Alarm type 3
+8 to disable on power-up until first alarm
+16 to memorize
+32 to filter with F.O. mode (output filter mode)
O u
min…max scale
of input selected in t.P
Maximum limit of main
input scale and analog
output
delay for F.O. 0 to 99 min or sec
Minimum output pulse 0 to 99 sec
Disabled by setting value 0
Displayed if associated with at least one output
Fault action (in case of
damaged sensor)
Er, br, Eb
Value Output 1 Output 2 Output 3
0 OFF OFF OFF
1 ON OFF OFF
2 OFF ON OFF
3 ON ON OFF
4 OFF OFF ON
5 ON OFF ON
6 OFF ON ON
7 ONONON
Output filter mode 0 not active: calculated status
is sent directly to relay
1 On delay (DON)
2 delay in activation of
output after output is
turned off (DBI)
3 Off delay (DOF)
4 Delay for activation
only at power-up (DPO)
+ 8 time base max. 99 min (default = 99 sec)
Decimal point position for
main input scale
dP Format
0 XXXX
1 XXX.X
2 XX.XX
3 X.XXX
Digital filter on
main input
Digital filter on process
variable display
0.0 to 20.0 sec
0.0 to 9.9
scale points
Minimum limit of main
input scale and analog
output min…max scale of input
selected in t.P
•POTENTIOMETER / STRAIN-GAUGE
input parameters
Typ PROBE Signal 4 DIGIT 3 DIGIT
TYPE polarization + sign
Max Range Max Range
0 Potentiometer positive -1999/9999 -999/999
(ex. 0 / 1V)
1 Potentiometer positive Custom Custom
custom linear (ex. 0 / 1V) linear linear
2 Strain-gauge positive -1999/9999 -999/999
(ex. 0 / 10mV)
3 Symmetrical -1999/9999 -999/999
(ex. -10 / +10mV)
Input settings
Type of probe,
signal and scale of
main input
I n
Select sampling time
(resolution)
0 120ms > 13bit probe power supply control
1 120ms > 13bit; 8000 divs
2 60ms > 12bit; 4000 divs
3 30ms > 11bit; 2000 divs
+4 to disable filter (average of the last eight values sampled)
+8 disables Eb (sampling time is halved)
N.B.:maximum sampling frequency and minimum intercept time is
obtained with code 15 (15 msec, 11-bit resolution, filter off)
(*) for version SW 2.0x
481651C_MHW_40B48-96_0308_ENG

•Protection
Protection code Value Displayed Modifiable
parameters parameters
0 o.1, o.2, o.3 o.1, o.2, o.3
1 o.1, o.2 o.1, o.2
2 o.1 o.1
3 o.1 none
+4 to disable In and Ou pages
+8 to disable Cf page
+16 to enable maintenance of reset memory at shutdown
+32 base configuration (the following parameters will not be displayed:
In: Ft, Fd, Of, L_L, H_L
Ou: On [forced to no. outputs present], rE)
P r
.......
•Custom Linearization
Custom Linearization of
main input
Step 0
(beginning of scale value) Display limits
(-1999 to 9999 for 4 digit display)
Step 32
(full scale value) Display limits
(-1999 to 9999 for 4 digit display)
L n
the n step value corresponds to input:
mV beginning scale + n * ∆mV
∆mV = (mV full scale - mV beginning scale) / 32
•User Calibration
U.C. CPU B function
1 analog retransmission
output
2 potentiometer
3 strain-gauge
positive polarization
4 strain-gauge
symmetrical polarization
U. C.
Minimum
calibration (*)
Maximum
calibration (*)
(*) Press keys ∆∇to
calibrate the analog
output
if U.C. = 1
Zero acquisition phase,
potentiometer with cursor
in minimum voltage
position
Maximum acquisition
phase, potentiometer
with cursor in maximum
voltage position
if U.C. = 2 if U.C. = 3, 4
Tare zero acquisition phase with strain-
gauge discharged (no weight or
pressure)
Acquisition phase for automatic definition
of sensitivity and full scale, load strain-
gauge with sample reference or
automatic activation of configured output
to check 6-lead sensor (only for version
SW 2.0x). 80% f.s. is set as default. The
value can be changed to set the value in
engineering units corresponding to the
sample used.
Zero recalculation phase, unload strain-
gauge by removing sample reference.
P.V. Process variable
(LEVEL 1)
.......
enabled only if t.P = Linear Custom (only for version SW.2.0x)
Note: between the calibration phases some seconds could be requested to elaborate data.
5
81651C_MHW_40B48-96_0308_ENG

The standard device is produced with parameter C.I.= 8,
with sampling time of 120 msec and Eb function disabled.
The Eb function lets you detect if the probe power supply is interrupted. This function is valid for probe currents > 20mA (8mA
in version 2.0x)
Example:
- probe voltage 10V
- strain-gauge resistance 350Ω
V 10
- current = = ≅ 28mA
R 350
or for three probes connected in parallel
- probe voltage 10V
- strain-gauge resistance 350Ω
V 10 30
- corrente = = = ≅85mA
R 350 350
3
•Eb Function
6 •ALARMS
time
AL1 + H1
AL2 + H2
AL2
AL1
alarm 1
alarm 2
(*)
For AL1 inverse absolute alarm (min.) with positive H1, 1 t = 1
(*) = OFF if disabling on power-on exists
For AL2 direct absolute alarm (max) with negative H2, 2 t = 0
For AL1 inverse absolute, symmetrical alarm with hysteresis H1, 1 t = 5
For AL1 direct absolute, symmetrical alarm with hysteresis H1, 1 t = 4
Normal absolute alarm Symmetrical absolute alarm
inverse
direct
AL1
AL1 + [ H1 ]
AL1 - [ H1 ]
time
For AL1 direct absolute alarm (max) with negative H 1, 1 t = 0
For AL2 direct relative alarm (max) with negative H2, 2 t = 2 For AL1 direct absolute alarm (max) with negative H1, 1 t = 0
For AL2 symmetrical deviation alarm H2, 2 t = 6
time
AL1+AL2
AL1
alarm 1
alarm 2
AL1+AL2
AL1
alarm 1
alarm 2
time
AL1 + AL2 + H2
Normal deviation alarm
(AL1 absolute, AL2 relative) Symmetrical deviation alarm
(AL1 absolute, AL2 relative)
AL1-AL2
AL1 + H1
AL1+AL2+H2
AL1+H1
The input value and alarms are frozen while the logic input is active.
With the logic input active, a reset turns OFF both the relay outputs and the alarms latch.
Input value is sampled; state of alarms is not transferred to outputs; outputs are “frozen”.
When the logic input is active the input value is “frozen”and the outputs are updated according to the calculated alarms
state, including the ones latched.
•HOLD function
•FLASH function
681651C_MHW_40B48-96_0308_ENG

•UCAL: strain-gauge calibration
48x48 format 96x48 format
•RS323 interface cable for configuration
WSK-0-0-0 Interface cable + CD Winstrum
•ORDER CODE
N.B.: the PC configuration cable is supplied
with the programming software.
WARNING: make the connection with the
device powered and with inputs and outputs
disconnected.
time
Alarm
setpoint
Output
Alarm
F.0 = 1
DON = Delayed activation F.0 = 2
DBI = Delay in turning on output after output is
turned off
time
time
time
F.0 = 3
DOF = Delayed deactivation F.0 = 4
DP0 = Delayed activation only at power-on
Variable
rA rA t < rA
Alarm
setpoint
Output
Alarm
Variable
rA t < rA
t > rA
Alarm
setpoint
Output
Alarm
Variable
Alarm
setpoint
Output
Alarm
Variable
rA rA rA rA
The diagrams refer to a normal absolute alarm with hysteresis H = 0
•Filter - outputs with reference to parameters F.0 and r.A
a) Positive signal polarization
Suppose we test a probe (load cell) with sensitivity of 2mV/V, powered at 10V.
The input signal range is 0 to 20mV. From unloaded to loaded cell you should see a value from 0 to 1000.
Set scale limits: L.S. = 0; H.S. = 1000. Calibration procedure: U.C. = 3.
Phase F1: unload the cell (equal to applying input voltage of 0mV). Wait 2 seconds for the signal to stabilize. Press key F.
Phase F2: load the cell with a sample weight: for example, 80% of total weight (equal to applying input voltage of 80% of
20mV (2mV*10V) = 16mV). On display, set a value equal to 80% of H.S. = 800.
Press key F.
Phase F3: unload the cell (equal to applying input voltage of 0mV). Wait 2 seconds for the signal to stabilize. Press key F
(calibration finished).
b) Suppose we test a probe (force sensor) with sensitivity of 2mV/V, powered at 10V.
The input signal range is -20mV to 20mV. From maximum compression to maximum traction you should see a value from -
1000 to 1000. Set scale limits: L.S. = -1000; H.S. = 1000.
Calibration procedure: U.C. = 4.
Phase F1: do not apply any traction or compression to the sensor (equal to applying input voltage of 0mV).
Wait 2 seconds for the signal to stabilize. Press key F.
Phase F2: apply compression equal to 80% of total compression (equal to applying input voltage of 80% of 20mV
(2mV*10V) = 16mV). On display, set a value equal to 80% of H.S. = 800.
Press key F.
Phase F3: do not apply any traction or compression to the sensor (equal to applying input voltage of 0mV).
Wait 2 seconds for the signal to stabilize. Press key F (calibration finished).
7
81651C_MHW_40B48-96_0308_ENG

Read the following warnings before installing, connecting or using the device:
•follow instructions precisely when connecting the device.
•always use cables that are suitable for the voltage and current levels indicated in the technical specifications.
•the device has no ON/OFF switch: it switches on immediately when power is turned on. For safety reasons, devices permanently connected to the power
supply require a two-phase disconnecting switch with proper marking. Such switch must be located near the device and must be easily reachable by the
user. A single switch can control several units.
•if the device is connected to electrically NON-ISOLATED equipment (e.g. thermocouples), a grounding wire must be applied to assure that this
connection is not made directly through the machine structure.
•if the device is used in applications where there is risk of injury to persons and/or damage to machines or materials, it MUST be used with auxiliary alarm
units. You should be able to check the correct operation of such units during normal operation of the device.
•before using the device, the user must check that all device parameters are correctly set in order to avoid injury to persons and/or damage to property.
•the device must NOT be used in inflammable or explosive environments. It may be connected to units operating in such environments only by means of
suitable interfaces in conformity to local safety regulations.
•the device contains components that are sensitive to static electrical discharges. Therefore, take appropriate precautions when handling electronic circuit
boards in order to prevent permanent damage to these components.
Installation: installation category II, pollution level 2, double isolation
•power supply lines must be separated from device input and output lines; always check that the supply voltage matches the voltage indicated on the
device label.
•install the instrumentation separately from the relays and power switching devices
•do not install high-power remote switches, contactors, relays, thyristor power units (particularly if “phase angle”type), motors, etc... in the same cabinet.
•avoid dust, humidity, corrosive gases and heat sources.
•do not close the ventilation holes; working temperature must be in the range of 0...50°C.
If the device has faston terminals, they must be protected and isolated; if the device has screw terminals, wires should be attached at least in pairs.
•Power: supplied from a disconnecting switch with fuse for the device section; path of wires from switch to devices should be as straight as possible; the
same supply should not be used to power relays, contactors, solenoid valves, etc.; if the voltage waveform is strongly distorted by thyristor switching units
or by electric motors, it is recommended that an isolation transformer be used only for the devices, connecting the screen to ground; it is important for the
electrical system to have a good ground connection; voltage between neutral and ground must not exceed 1V and resistance must be less than 6Ohm; if
the supply voltage is highly variable, use a voltage stabilizer for the device; use line filters in the vicinity of high frequency generators or arc welders;
power supply lines must be separated from device input and output lines; always check that the supply voltage matches the voltage indicated on the
device label.
•Input and output connections: external connected circuits must have double insulation; to connect analog inputs (TC, RTD) you have to: physically
separate input wiring from power supply wiring, from output wiring, and from power connections; use twisted and screened cables, with screen connected
to ground at only one point; to connect adjustment and alarm outputs (contactors, solenoid valves, motors, fans, etc.), install RC groups (resistor and
capacitor in series) in parallel with inductive loads that work in AC (Note: all capacitors must conform to VDE standards (class x2) and support at least 220
VAC. Resistors must be at least 2W); fit a 1N4007 diode in parallel with the coil of inductive loads that operate in DC.
GEFRAN spa will not be held liable for any injury to persons and/or damage to property deriving from tampering, from any incorrect or
erroneous use, or from any use not conforming to the device specifications.
!WARNING: this symbol indicates danger.
It is seen near the power supply circuit and near high-voltage relay contacts.
44
Triac, None T 0 Kindly contact GEFRAN for information on available codes.
100 to 240Vac1
Power supply
20 to 27Vac/dc0
Digital input / Retransmission output
(alternative to output 3)
None0Digital input
Retransmission output 4...20mA on max 150Ω
1
2
N°Digits
44
40B
Output 3
(alternative to digital input / retransmission output)
None0RelayR
5Vdc 05
Sensor power supply
1.2Vdc for potentiometer 01
24Vdc, 50mA (transmitter) 24
10Vdc, 120mA
15Vdc (transmitter) 10
15
•WARNINGS
ORDER CODE
Relay, Static D2 R D
Output 1, Output 2
Relay, Relay R R
48 4
100 to 240Vac/dc1
Power supply
11 to 27Vac/dc0
Digital input / Retransmission output
None0Digital input
Digital input +
Retransmission output 4...20mA on max 150Ω
1
3
N°Digits
3 + sign 3
40B
Output 3
None0RelayR
5Vdc 05
Sensor power supply
1.2Vdc for potentiometer 01
24Vdc, 50mA (transmitter) 24
10Vdc, 120mA
15Vdc (transmitter) 10
15
Relay, Static D2 R D
Output 1, Output 2
Relay, Relay R R
96
881651C_MHW_40B48-96_0308_ENG
This manual suits for next models
1
Table of contents
Other gefran Measuring Instrument manuals

gefran
gefran KS Series User manual

gefran
gefran 2308 User manual

gefran
gefran KH-SIL2 Series User manual

gefran
gefran SENSORMATE QE2008-W SET User manual

gefran
gefran 4T72 User manual

gefran
gefran QE1008-W SET User manual

gefran
gefran 40A 48 User manual

gefran
gefran 4A 48 User manual

gefran
gefran 40T 72 User manual

gefran
gefran 40F 96 User manual