3. Depress the on/off switch to the “ON”
position. The internal oscillating pumps
will now cycle and fill the solution tanks
to their proper levels. When the “LOW
SOLUTION LEVEL” lamp has gone out,
the solution heater will start and the
transports will turn. DO NOT PROCESS
FILMS AT THIS TIME.
4. After 15 to 20 minutes, the “READY”
lamp will illuminate. This indicates that
the proper processing temperature has
been reached.
5. Film Processing may now begin.
Depress run/standby switch to the
“RUN” position. The processor
transports will now turn and the film
dryer will activate.
Note: GENDEX recommends the use of
a daily clean-up film (such as GENDEX
Roller Transport Cleanup film, cat no.
E4001) each morning prior to the
processing of any radiographs.
6. Check for “Endo” processing speed. If
the processor is in the “Endo” process
mode, this switch will flash on and off.
The “Endo” process is a 2 minute
cycle. It is not intended for archival
quality radiographs or extra-oral films.
If the “Endo” process is not desired,
press “Endo” switch once, the lamp will
extinguish and a normal five minute
process cycle will resume.
7. Lift the film loading door (see Page 2-1)
and begin loading film into the entrance
slot. When the last film has fully entered
the processor, close the loading door.
Room lights may now be turned on.
Note: Allow sufficient time for the film
to enter the processor before feeding
your next film – avoid film overlap.
8. The processed film will exit from the film
exit slot (see Page 2-1) in approximately
five minutes for normal process and two
minutes for “Endo” process.
9. When the last film has exited from the
processor, depress the run/standby
switch to the “STANDBY” position.
Note: DO NOT DEPRESS THE
ON/OFF SWITCH. It should remain
“ON” for the entire working day. As
long as the on/off switch is in the “ON”
position, several functions will occur
throughout the day.
•Approximately every four
minutes, the oscillating pumps
will cycle for several seconds.
This action will maintain a
uniform solution strength.
•The solution heater will
maintain a proper processing
temperature.
•The transports will intermittently
turn, allowing the solutions to wet
the rollers. This will prevent dried
solution deposits from forming.
10. At the end of your working day,
depress the on/off switch to the “OFF”
position.
10A. On GXP models with a solenoid valve
— turn the water inlet valve on the
plumbing line to the “OFF” position.
3-2