Gentec-EO INTEGRA Series User manual


INTEGRA User Manual Revision 1.1 ii
WARRANTY
First Year Warranty
All Gentec-EO devices carry a one-year warranty (from date of shipment) against material and/or workmanship
defects, when used under normal operating conditions. The warranty does not cover damages related to battery
leakage or misuse.
Gentec-EO Inc. will repair or replace, at Gentec-EO Inc.’s option, any device that proves to be defective during
the warranty period, except in the case of product misuse.
Any attempt by an unauthorized person to alter or repair the product voids the warranty.
The manufacturer is not liable for consequential damages of any kind.
In case of malfunction, contact your local Gentec-EO distributor or nearest Gentec-EO Inc. office to obtain a
return authorization number. The material should be returned to:
Contacting Gentec Electro-Optics Inc.
Gentec Electro-Optics, Inc.
445, St-Jean-Baptiste, Suite 160
Québec, QC
Canada G2E 5N7
Tel: (418) 651-8003
Fax: (418) 651-1174
Email: service@gentec-eo.com
Website: www.gentec-eo.com
CLAIMS
To obtain warranty service, contact your nearest Gentec-EO agent or send the product, with a description of the
problem, and prepaid transportation and insurance, to the nearest Gentec-EO agent. Gentec-EO Inc. assumes
no risk for damage during transit. Gentec-EO Inc. will, at its option, repair or replace the defective product free
of charge or refund your purchase price. However, if Gentec-EO Inc. determines that the failure is caused by
misuse, alterations, accident or abnormal conditions of operation or handling, it would therefore not be covered
by the warranty.

INTEGRA User Manual Revision 1.1 iii
SAFETY INFORMATION
Do not use a Gentec-EO device, if the monitor or the detector looks damaged, or if you suspect that the device is
not operating properly.
Appropriate installation must be done for water-cooled and fan-cooled detectors. Refer to the specific
instructions for more information. Wait a few minutes before handling the detectors after power is applied.
Surfaces of the detectors get very hot and there is a risk of injury if they are not allowed to cool down.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses, and can
radiate radio frequency energy. If not installed and used in accordance with the instructions, it
may cause harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off
and on, try to correct the interference by taking one or more of the following steps:
•Reorient or relocate the receiving antenna.
•Increase the distance between the equipment and receiver.
•Connect the equipment to an outlet that is on a different circuit than the receiver.
•Consult the dealer or an experienced radio/TV technician for help.
Caution: Changes or modifications not expressly approved in writing by Gentec-EO Inc. may void the
user’s authority to operate this equipment.
SYMBOLS
The following international symbols are used in this manual:
Refer to the manual for specific Warning or Caution information to avoid any
damage to the product.
DC, Direct Current

INTEGRA User Manual Revision 1.1 iv
TABLE OF CONTENTS
1. INTEGRA.........................................................................................................................1
1.1. Introduction.................................................................................................................................................1
1.2. Specifications.............................................................................................................................................1
2. Quick Start Procedure...................................................................................................3
3. User Interface.................................................................................................................4
4. USB Serial Communication ..........................................................................................4
4.1. Description .................................................................................................................................................4
4.2. Setting up Communication to the INTEGRA..............................................................................................4
4.2.1. Verify the COM Port............................................................................................................................4
4.2.2. Connect the INTEGRA........................................................................................................................4
4.2.3. To Echo Commands ...........................................................................................................................5
4.2.4. Test the Connection............................................................................................................................5
4.2.5. HyperTerminal Settings Shortcut........................................................................................................5
4.3. Serial Command Format............................................................................................................................5
4.3.1. Serial Protocol Rules: .........................................................................................................................5
4.3.2. Text Mode Rules:................................................................................................................................5
4.4. Binary Mode OUTPUT Format...................................................................................................................6
4.4.1. Description..........................................................................................................................................6
4.4.2. Codification .........................................................................................................................................6
4.5. List of Serial Commands for the INTEGRA (Summary).............................................................................8
4.6. List of Serial Commands for the INTEGRA (Complete).............................................................................9
4.6.1. Display ................................................................................................................................................9
01 - Set Scale.................................................................................................................................................................9
02 - Set Scale Up.........................................................................................................................................................10
03 - Set Scale Down.....................................................................................................................................................10
04 - Get Current Scale Index........................................................................................................................................10
05 - Set Autoscale........................................................................................................................................................10
06 - Get Autoscale........................................................................................................................................................11
07 - Display Valid Scale ...............................................................................................................................................11
08 - Set Trigger Level...................................................................................................................................................11
09 - Get Trigger Level ..................................................................................................................................................12
10 - Get Mode Display..................................................................................................................................................12
4.6.2. Data Acquisition............................................................................................................................... 12
11 - Query Current Value.............................................................................................................................................12
12 - Send Continuous Transmission of Data................................................................................................................13
13 - Send Continuous Value with Period......................................................................................................................13
14 - Sent Current Value with Frequency ......................................................................................................................13
15 - Stop the CAU Command.......................................................................................................................................14
16 - Query New Value Ready.......................................................................................................................................14
17 - Get Laser Frequency ............................................................................................................................................14
18 - Set Binary Joulemeter Mode.................................................................................................................................15
19 - Get Binary Joulemeter Mode.................................................................................................................................15
4.6.3. Setup................................................................................................................................................ 15
20 - Set Personal Wavelength Correction ....................................................................................................................15
21 - Get Wavelength ....................................................................................................................................................16
4.6.4. Control.............................................................................................................................................. 16
22 - Set Anticipation.....................................................................................................................................................16
23 - Get Anticipation Status..........................................................................................................................................16
24 - Noise Suppression................................................................................................................................................17
25 - Set Zero Offset......................................................................................................................................................17
26 - Clear Zero Offset...................................................................................................................................................17
27 - Get Zero Offset .....................................................................................................................................................17
28 - Set Diode Zero Offset ...........................................................................................................................................18
29 - Set User Multiplier.................................................................................................................................................18

INTEGRA User Manual Revision 1.1 v
30 - Get User Multiplier ................................................................................................................................................18
31 - Set User Offset......................................................................................................................................................18
32 - Get User Offset.....................................................................................................................................................19
33 - Set Single Shot Energy Mode...............................................................................................................................19
34 - Set Attenuator.......................................................................................................................................................19
35 - Get Attenuator.......................................................................................................................................................20
4.6.5. Instrument and Detector Information ............................................................................................... 20
36 - Query Version.......................................................................................................................................................20
37 - Query Status.........................................................................................................................................................20
38 - Query Extended Status.........................................................................................................................................22
4.7. Error Messages....................................................................................................................................... 23
5. USB Driver Installation................................................................................................24
6. Maintenance.................................................................................................................25
6.1. Free Software Upgrade........................................................................................................................... 25
6.2. Troubleshooting....................................................................................................................................... 25
7. Declaration of Conformity...........................................................................................26
Appendix A: WEEE Directive...............................................................................................27

INTEGRA User Manual Revision 1.1 1
1. INTEGRA
1.1.INTRODUCTION
The INTEGRA is a series of All-in-One detectors that combine a detector and a meter in one convenient product.
The small but powerful meter of the INTEGRA Series presents a direct USB connection so you can plug it into
your PC. Simply use the PC-Gentec-EO software supplied with your product and be ready to make power or
energy measurements within seconds! Each detector of the INTEGRA Series offers the same incredible
performance as the usual detector and meter combination, from pW to kW and from fJ to J. And the good news
is that all our most popular products are available with the INTEGRA option.
1.2.SPECIFICATIONS
The following specifications are based on a one-year calibration cycle, an operating temperature of 18 to 28ºC
(64 to 82ºF) and a relative humidity not exceeding 80%. INTEGRA must be stored in an environment between
10oC to 45oC (50 to 113oF) and a relative humidity not exceeding 90%. Photodiodes are sensitive to
temperature, especially at longer wavelength. It is the best to keep the temperature in the range of 25 oC (77oF)
close to the calibration temperature.
Power Meter Specifications
Power Range
4 pW to 30 kW
Power Scales:
Photo Detectors
Thermal Detectors
21 scales: 300 pW to 3 W
16 scales: 300 µW to 30 kW
Pyroelectric in Power Mode
(UM-B)
8 scales : 100 µW, to 300 mW
Accuracy1
±2.5%15 µV best scale1
Energy Meter Specifications
Energy Range
2 fJ to 30 kJ
Energy Scales:
Photo Detectors
Pyroelectric Detector
22 scales: 300 fJ to 30 mJ
15 scales: 3 mJ to 30 kJ
Accuracy 2
3.0%250 µV < 5.2 kHz best scale.
Trigger Level:
Default
Software
2 %
0.1% to 99.9%, 0.1% resolution
Data Transfer Rate
No missing points throughput serial frequency(power and energy):
5200 Hz
Frequency Measurement
Accurate frequency measurement up to 5.2 kHz
0-5200 Hz: ±1%
Software Specifications
Please refer to the PC-Gentec-EO manual (202322)
1
Including linearity, detector accuracy and is detector dependent.
2
Excludes non-linearity.

INTEGRA User Manual Revision 1.1 2
1.1.OUTLINE DRAWING

INTEGRA User Manual Revision 1.1 3
2. QUICK START PROCEDURE
1. Install the PC-Gentec-EO software on your PC.
2. Install the power or energy detector head on its optical stand.
3. Connect the INTEGRA device to the PC with the appropriate USB cable.
4. Start the PC-Gentec-EO software.
5. Choose the appropriate display for your measurement:
a. Real time with bar graph
b. Scope
c. Needle
d. Averaging
e. Histogram
f. Statistics
6. Press the Connect button on the top left corner of the main window and change the measurement and
display options in the appropriate tabs.
a. Power heads will default to power measurement; energy heads will default to energy
measurement. The display will default to a dual display in real time and scope (Fig 2-1) in auto
range mode.
b. Remove the head’s protective cover and start the laser.
Put the detector head into the laser beam path. Leave it there for a few minutes, until the detector
has reached an equilibrium temperature. The entire laser beam must be within the sensor aperture.
Do not exceed maximum specified densities, energies or powers. For the most accurate
measurement, spread the beam across 60% to 80% of the sensor area.
Warning: Power detectors can be used with both CW and pulsed lasers.
Energy detectors can only be used with pulsed lasers.
7. Adjust the Zero:
The power read by PC-Gentec-EO when no laser beam is incident on the detector may not be exactly
zero. For power measures, this is because the detector is not thermally stabilized OR there was a heat
source in the field of view of the detector when you connected the PC-Gentec-EO. As for photo
detectors, zeroing will remove the detector’s offset.
Thermal and Pyroelectric Detectors: Block all laser radiation on the detector. To reset the zero, wait
until the reading has stabilized and click Set Zero in the Main menu.
Photo Detectors: You must block all form of radiation (cover the detector). Click Set Zero in the Main
menu. In some software versions, a message will appear requesting you to put the black cover over
your photo detector. Press the OK button after you have done so. The software passes through all the
scales to determine the zero for each scale.
See Notes

INTEGRA User Manual Revision 1.1 4
Notes:
a. Refer to specific power detector documentation for complete installation and operating instructions.
b. Power detectors are thermal sensors and are thus sensitive to temperature variations.
c. For high-precision measurements, it is recommended to:
i. Allow the power detector to thermally stabilize before zeroing the software.
ii. Touch only the stand when handling the power detector. Do not touch the detector irtself. This is
especially true for highly sensitive detectors.
iii. Do not adjust the zero for energy detectors, such as the QE series.
iv. Avoid forced airflow or drafts around the detector.
3. USER INTERFACE
Please refer to the PC-Gentec-EO manual for more information concerning the user interface. The manual can
be downloaded on our website at https://gentec-eo.com/downloads/specsheets-manuals.
4. USB SERIAL COMMUNICATION
4.1.DESCRIPTION
The INTEGRA has two Communication Modes: the Binary Mode for fast data acquisition and the ASCII mode.
Both modes will require text input commands which must follow rules stated in section 4.3. The output can be in
binary mode or in ASCII mode. Section 4.6 describes all the commands in ASCII output mode, but keep in mind,
it is also valid for Binary Mode as described in section 4.4.
The USB class used by INTEGRA is a CDC, or Communications Device Class. This means it shows up in the
host PC as a COM port, but it is not a COM port, rather a true full speed USB port. You can talk to it like as if it
were an RS232 port, but very fast when it comes to speed. Follow the Windows Prompts to install the USB
drivers. The USB drivers are fully tested and digitally signed by Microsoft.
Open the appropriate port in your software with standard COM port tools. None of the port settings matter since
they are not used, so leave them at whatever default they are in. It’s a real USB connection.
Use the standard COM port writes and reads to control the INTEGRA.
4.2.SETTING UP COMMUNICATION TO THE INTEGRA
4.2.1. Verify the COM Port
To verify the USB installation and find the COM port number, click:
Start →Settings → Control Panel → System → Device Manager
Scroll down to Ports (COM & LPT) and double click that line. One of the options should be
USB-to-Serial Port (COM#)
Note the COM port number, you will need it for the next step.
4.2.2. Connect the INTEGRA
You may use any serial communications software that you are familiar with. Our instructions are for
HyperTerminal because it is widely available on PCs with Windows.

INTEGRA User Manual Revision 1.1 5
Select:
Start → Programs → Accessories → Communications → HyperTerminal
To save communication settings, enter a name for the connection. n the drop down menu for “Connect using”
select the COM port that the USB driver was installed on (Section 4.2.1). Select OK.
Input the following settings into the communications parameter window that appears next.
INTEGRA COM Port Settings
Bits per second
Any settings will work
Data bits
Any settings will work
Parity
Any settings will work
Stop bits
Any settings will work
Flow control
Any settings will work
Click OK to begin entering serial commands in the HyperTerminal window.
4.2.3. To Echo Commands
The commands you type will not appear in the HyperTerminal window unless you setup the HyperTerminal to do
so. Only the response from the INTEGRA will be displayed. If you prefer to see the commands you are typing on
the HyperTerminal window click the File menu and execute the following sequence:
File → Properties → Settings → ASCII setup → “Echo typed characters locally” → OK
4.2.4. Test the Connection
In the HyperTerminal window, type *VER. If the response is the version of your INTEGRA, you are successfully
connected and ready for serial command action.
4.2.5. HyperTerminal Settings Shortcut
When you end the session, HyperTerminal asks if you want to save your settings. To avoid inputting the
communication parameters again in the future, save by clicking Yes. The next time you execute the string of
commands, the name of your session will appear after HyperTerminal. Clicking on the session name will open
the connection using the saved settings. To avoid re-entering the string of commands, put a shortcut to this file
on your desktop: search for the file name, select the file. Right click and select Shortcut in the drop down menu.
4.3.SERIAL COMMAND FORMAT
4.3.1. Serial Protocol Rules:
Commands are sent as text strings. The response will either be data or an empty string.
4.3.2. Text Mode Rules:
All text commands must begin with a trig character (*). You do not need to end with a line-feed and/or a carriage-
return. Parameters must NOT be separated by spaces. Characters do not have to be capitals, mixed upper and
lower cases is ok. Replies to all text mode commands are also in text mode, and end with a carriage-return and
a line-feed.

INTEGRA User Manual Revision 1.1 6
In case of an error, the reply string is in the following format:
Error X: reason [enter]
Where Xis the error code, and reason is an explanation. See Error Codes at the end of this
section.
Because all Text Mode replies end with a carriage return <CR> or line-feed <LF> (or both), a text reply contains
tabulations when many elements need to be separated in the string. This is useful when exporting data to a
spreadsheet.
4.4.BINARY MODE OUTPUT FORMAT
4.4.1. Description
The resolution of both is 12-bit in Joulemeter mode. A 14-bit value is sent for compatibility with other meters, but
the two LSB’s are not significant.
Only the Joulemeters support the binary mode. Thermopiles in energy mode, thermal heads in standard mode
and photo detectors are coded in ASCII.
The value in binary mode is coded in two bytes.
4.4.2. Codification
When retrieving a measurement with either *cau or *cvu command the output is decoded as follow:
Byte 1
OXXX XXXX
Byte 2
OXXX XXXX
Where O is the byte Order bit, and X is the binary data. If O is 0, the byte is the MSB. If O is 1, the byte is the
LSB
See the table below for examples on how to use binary commands.

INTEGRA User Manual Revision 1.1 7
How to Use Binary Commands
Example 1: Out of Scale Condition when Using *CEU or *CTU
If the value of these bytes is 0xFE7F, an out of scale condition exists.
INTEGRA is measuring 151 mJ in a 30 0mJ scale. The data sent by INTEGRA will be: 0x40B6
Decode this as follows.
1. Look at bit 7 of each byte to determine the high and low bytes.
2. Keep bits 0 to 6 of each byte. Shift the High byte left by 7 bits (multiply by 128)
3. Add the high and low bytes
4. Divide the result by the full scale value, 16382.
5. Multiply the result by the set scale, 300mJ.
The two data bytes are 0x40 and 0xB6. In Binary they are: 0100 0000 and 1011 0110.
The byte order bit is 0 for the high byte and 1 for the low byte. The high byte is therefore 0x40, and the low
byte is 0xB6.
The data is the lower 7 bits of each byte, or high byte of 0x40 and low byte of 0x36.
Shifting the high byte 7 by its results in 0x40 x 128 = 0x2000.
Adding this value to the low byte results in 0x2036, or 8246 decimal: 8246 / 16382 * 300mJ = 151mJ.
If you send *ceu INTEGRA will send continuous data with 9 bytes per pulse.
If you send *ctu INTEGRA will send the current measurement with 9 bytes per pulse.
The 9 bytes are decoded as follows:
Byte 8
Always 0X02, or STX. Let the host know this is the start of data.
Byte 7
The scale index, or with 0x80. This is done so that it can never be the STX or ETX byte. So if the
scale was set to 29, then this byte would be hex (29) = 0x1D. Or this with 0x80 and the value
sent is 0x9D.
Byte 6
The upper 7 data bits of the energy, or with 0x80. If the pulse is over range, this byte is 0xFE.
Byte 5
The lower 7 data bits of the energy, or with 0x80. If the pulse is over range, this byte is 0x7F.
Byte 4
The upper 7 bits of the pulse period timer or with 0x80. (28 bits total)
Byte 3
The next 7 bits of the pulse period timer or with 0x80. (28 bits total)
Byte 2
The next 7 bits of the pulse period timer or with 0x80. (28 bits total)
Byte 1
The lower 7 bits of the pulse period timer or with 0x80. (28 bits total)
Byte 0
Always 0X03, or ETX. Let the host know this is the end of data.
Example 2: *CEU and *CTU with Out of Scale
INTEGRA is measuring 151 mJ in a 300 mJ scale. The pulse frequency is 1531 Hz. The data sent by
INTEGRA will be: 0x0297A0B68080FABC03.
Decode this as follows.
1. The valid data is between the 0x20 (Start of Text) and 0x03 (End of Text) codes.
2. Valid data is 0x97A0B68080FABC
3. The first byte is the scale, or’d with 0x80
4. The second and third bytes are the data, each or’d with 0x80.
5. The remaining 4 bytes are the pulse period counts, each or’d with 0x80
Valid data is 0x97A0B68080FABC
The scale byte is 0x97. Mask off bit 7 resulting in 0x17, or 23 decimal. The scale is 23, or 300 mJ.
The energy data bytes are 0xA0B6. If the value of these bytes is 0xFE7F, an out of scale condition exists.
If no out of scale condition exits, mask off bit 7 of each byte: 1010 0000 1011 0110 ≥ 0010 0000 0011 0110
Resulting in 0x2036, or 8246 decimal: 8246 / 16382 * 300 mJ = 151 mJ.
The pulse period bytes are 0x8080FABC. Mask off bit 7 of each byte resulting in 0x003D3C, or 15676 decimal.
The period timer is based on a 24E6 Hz clock, so the period is found as:
If you send *ceu it will be 15676 counts / 24E6 Counts per second = 653.17us.
If you send *ctu the pulse frequency will be 1 / 653.17us = 1531Hz.

INTEGRA User Manual Revision 1.1 8
4.5.LIST OF SERIAL COMMANDS FOR THE INTEGRA (SUMMARY)
#
Command Name
Command
Description
DISPLAY
01
Set Scale
SCS
Manually sets the scale
02
Set Scale Up
SSU
Changes scale to the next higher scale
03
Set Scale Down
SSD
Changes scale to the next lower scale
04
Get Current Scale Index
GCR
Returns scale index between 0 and 41
05
Set Autoscale
SAS
Sets the autoscale
06
Get Autoscale
GAS
Returns autoscale status
07
Display Valid Scale
DVS
Displays the valid scales for the connected head
08
Set Trigger Level
STL
Sets the internal trigger level when measuring pulse energy
09
Get Trigger Level
GTL
Returns trigger level value
10
Get Measure Mode
Display
GMD
Returns the current measure mode on INTEGRA
MEASUREMENT
Data Acquisition
11
Query Current Value
CVU
Gets the value currently displayed on the screen
12
Send Continuous
Transmission of Data
CAU
Sends the values in ASCII to the serial port with the data
sampling setting
13
Send Continuous Value
with Period
CEU
Sends continuous value with period
14
Send Current Value with
Frequency
CTU
Sends current value with frequency
15
Stop the CAU Command
CSU
Stops the CAU Command
16
Query New Value Ready
NVU
Determines if a new reading is available
17
Get Laser Frequency
GRR
Sends the laser rep rate frequency in ASCII to the serial port
18
Set Binary Joulemeter
Mode
SS1
Sets the binary joulemeter mode or ASCII mode
19
Get Binary Joulemeter
Mode
GBM
Returns the binary joulemeter mode or ASCII mode
Setup
20
Set Personal Wavelength
Correction
PWC
Specifies the wavelength
21
Get Wavelength
GWL
Returns the wavelength in nm
Control
22
Set Anticipation
ANT
Turns the anticipation on or off
23
Get Anticipation Status
GAN
Returns the anticipation status
24
Noise Suppression
AVG
Applies the noise suppression algorithm
25
Set Zero Offset
SOU
Zeroes the reading
26
Clear Zero Offset
COU
Undoes the zeroing of the reading
27
Get Zero Offset
GZO
Returns the zero offset status
28
Set Diode Zero Offset
SDZ
Zeroes the reading for all the scales for a photodiode
29
Set User Multiplier
MUL
Sets the multiplier value
30
Get User Multiplier
GUM
Returns the current multiplier value
31
Set User Offset
OFF
Sets the offset value
32
Get User Offset
GUO
Returns the current offset value
33
Set Single Shot Energy
Mode
SSE
Sets the Single Shot Energy mode
34
Set Attenuator
ATT
Sets the attenuator
35
Get Attenuator
GAT
Returns the attenuator status
INSTRUMENT AND DETECTOR INFORMATION
36
Query Version
VER
Gets firmware version of the monitor
37
Query Status
STS
Retrieves the detector information and monitor settings
38
Query Extended Status
ST2
Returns the extended status

INTEGRA User Manual Revision 1.1 9
The Serial Commands Format is:
All text commands must begin with a trig character (*) and DO NOT end with a line-feed or a carriage-return. All
parameters must NOT have a space between the command and the list of parameters, nor between the
parameters themselves. The characters do not have to be capitals and mixed case is ok. Replies to all text mode
commands are also in text mode, and end with a carriage-return and line-feed.
*MUL+8 character numerical value Ex: « *MUL1.000000 » or « *MUL-1.34e-3 » or «*MUL0.000543 »
*OFF+8 character numerical value like above
*STL+4 character numerical value like « *STL10.2 » or « *STL0.22 »
*SSE1 / *SSE0 : Single Shoot Energy ON or OFF
*ATT1 / *ATT0: Attenuator ON / OFF (When available)
4.6.LIST OF SERIAL COMMANDS FOR THE INTEGRA (COMPLETE)
This section describes all the commands in ASCII output mode, but keep in mind, it is also valid for Binary Mode
as described in section 4.4.
4.6.1. Display
01 - Set Scale
This command is used to force the display of the current data into a specific scale. The lower scale is always
zero, the higher scales can be found in the table below. The Autoscale mode applies the best scale for the
current values in real time. The parameter must be one of the identifiers in the table below and have 2 digits.
Command
Parameters
Answer
SCS
Range index
Range Identifiers:
Index
Value
Index
Value
00
1 picowatt or picojoule
21
30 milliwatts or millijoules
01
3 picowatts or picojoules
22
100 milliwatts or millijoules
02
10 picowatts or picojoules
23
300 milliwatts or millijoules
03
30 picowatts or picojoules
24
1 watt or joule
04
100 picowatts or picojoules
25
3 watts or joules
05
300 picowatts or picojoules
26
10 watts or joules
06
1 nanowatt or nanojoule
27
30 watts or joules
07
3 nanowatts or nanojoules
28
100 watts or joules
08
10 nanowatts or nanojoules
29
300 watts or joules
09
30 nanowatts or nanojoules
30
1 kilowatt or kilojoule
10
100 nanowatts or nanojoules
31
3 kilowatts or kilojoules
11
300 nanowatts or nanojoules
32
10 kilowatts or kilojoules
12
1 microwatt or microjoule
33
30 kilowatts or kilojoules
13
3 microwatts or microjoules
34
100 kilowatts or kilojoules
14
10 microwatts or microjoules
35
300 kilowatts or kilojoules
15
30 microwatts or microjoules
36
1 megawatt or megajoule
16
100 microwatts or microjoules
37
3 megawatts or megajoules
17
300 microwatts or microjoules
38
10 megawatts or megajoules
18
1 milliwatt or millijoule
39
30 megawatts or megajoules
19
3 milliwatts or millijoules
40
100 megawatts or megajoules
20
10 milliwatts or millijoules
41
300 megawatts or megajoules
Default: Autoscale.

INTEGRA User Manual Revision 1.1 10
Example
The following example sets the scale to 3 nanowatts or nanojoules:
Command: *SCS08
Answer:
02 - Set Scale Up
This command is used to force the display of the current data into a higher scale.
Command
Parameters
Answer
SSU
None
03 - Set Scale Down
This command is used to force the display of the current data into a lower scale.
Command
Parameters
Answer
SSD
None
04 - Get Current Scale Index
This command returns the scale index between 0 and 41. Please refer to Set Scale command (SCS) details for
the complete scale index table.
Command
Parameters
Answer
GCR
None
Index from 0 to 41
Example
Command: *GCR
Answer: Range : 10 \r\n
05 - Set Autoscale
This command is used to force the display into autoscale.
Command
Parameters
Answer
SAS
1: On
0: Off

INTEGRA User Manual Revision 1.1 11
06 - Get Autoscale
This command returns whether or not the autoscale option is activated.
Command
Parameters
Answer
GAS
None
1: On
0: Off
Example
Command: *GAS
Answer: Autoscale : 1 \r\n
07 - Display Valid Scale
This command is used to display all of the valid scales the connected head supports. The scales are displayed
in scale index. Please refer to the Set Scale section for the table correspondence.
Command
Parameters
Answer
DVS
None
The valid scale index.
The following example is for a UP19K INTEGRA, which can have the following scales:
100 mW
300 mW
1 W
3 W
10 W
30 W
Example
Command: *DVS
Answer: [22] : 100.0 m <CR> <LF>
[23] : 300 m <CR> <LF>
[24] : 1.000 <CR> <LF>
[25] : 3.00 <CR> <LF>
[26] : 10.00 <CR> <LF>
[27] : 30.0 <CR> <LF>
08 - Set Trigger Level
This command sets the internal trigger level when using the device in energy mode.
Command
Parameters
Answer
STL
Trigger Level (in percentage) must
be 4 numerical values
Default: 2%
The value should be set between 0.1 and 99.9.

INTEGRA User Manual Revision 1.1 12
Example
Command: *STL15.4 (15.4%)
*STL0.20 (.2%)
Answer:
09 - Get Trigger Level
This command returns the trigger level in %. The value is between 0.1% and 99.9%.
Command
Parameters
Answer
GTL
None
Returns the trigger level in %.
Example
Command: *GTL
Answer: Trigger Level : 2.0 \r\n
10 –Get Measure Mode Display
This command returns the INTEGRA’s measurement mode. Depending on the head, it can be Power Mode in
W, Power Mode in dBm (DBM), Energy Mode in J or Single Shot Energy Mode in J (SSE).
Command
Parameters
Answer
GMD
None
POWER = 0
ENERGY = 1
SSE = 2
DBM = 6
Example
Command: *GMD
Answer: Mode : 0 \r\n
4.6.2. Data Acquisition
11 - Query Current Value
This command is used to query the value that is currently being displayed by the monitor. The value is displayed
in watts or in joules.
Command
Parameters
Answer
CVU
None
Current value

INTEGRA User Manual Revision 1.1 13
Example
For example, a 12 milliwatts reading would be displayed like this:
Command: *CVU
Answer: 0.012 <CR> <LF>
12 - Send Continuous Transmission of Data
This command is used to send data to the serial port according to the data sampling setting. The maximum
transfer speed is 200 Hz.
Command
Parameters
Answer
CAU
None
Data in ASCII
13 - Send Continuous Value with Period
INTEGRA will send continuous energy data and the pulse period in timer counts. They are comma separated.
Command
Parameters
Answer
CEU
None
Continuous value with pulse period
in timer counts
Example
This example is for a 32 Hz laser:
Command: *CEU
Answer: 5.066E-01,32.0<CR> <LF>
….
14 - Sent Current Value with Frequency
INTEGRA will send the current measurement and the pulse repetition rate in Hz. They are comma separated.
Command
Parameters
Answer
CTU
None
Current value with frequency
Example
This example is for a 32 Hz laser:
Command: *CTU
Answer: 5.066E-01,32.0<CR> <LF>

INTEGRA User Manual Revision 1.1 14
15 - Stop the CAU Command
This command is used to stop the real time transfer enabled by the CAU Command.
Command
Parameters
Answer
CSU
None
16 - Query New Value Ready
This command is used to check whether a new value is available from the device. Though optional, its use is
recommended when doing single pulse operations.
Command
Parameters
Answer
NVU
Available/Not Available <CR> <LF>
1/0
Example
Command: *NVU <enter>
Answer: New Data Not Available
<CR> <LF>
Note that the Query Current Value command will return the current values from the device even if they have not
been updated since the last query.
17 - Get Laser Frequency
This command is used to get the laser frequency.
Command
Parameters
Answer
GRR
None
Data in ASCII

INTEGRA User Manual Revision 1.1 15
18 - Set Binary Joulemeter Mode
This command is used to set the monitor in binary or ASCII mode. Refer to section 4.4 for the INTEGRA binary
mode description.
Command
Parameters
Answer
SS1
0= ASCII
1= Binary
Example
Command: *SS11
Answer: The INTEGRA is set in binary
mode
19 - Get Binary Joulemeter Mode
This command returns whether or not the binary joulemeter mode is activated for serial communication. Refer to
section 4.4 for the INTEGRA binary mode description.
Command
Parameters
Answer
GBM
None
1: On
0: Off
Example
Command: *GBM
Answer: Binary Joulemeter Mode :
0 \r\n
4.6.3.Setup
20 - Set Personal Wavelength Correction
This command is used to specify the wavelength in nm being used on the detector. The EEPROM in the detector
contains measured spectral data for a wide range of wavelengths. If the wavelength input by the user is different
from the predefined list of wavelengths on the device, a custom value is interpolated. Specifying zero as a
wavelength or providing an out-of-bound value as a parameter restores the default settings. A valid value is set
between the lowest and highest wavelengths supported by the device, and it should not be a floating point value.
The input parameter must have 5 digits. If the desired wavelength does not have 5 digits you must enter a zero-
padded number. For example, to set the wavelength at 514 nm, you must enter 00514.
Command
Parameters
Answer
PWC
Wavelength
Default: Calibration wavelength, (typically 1064 nm, varies with the detector model)
Table of contents
Other Gentec-EO Measuring Instrument manuals

Gentec-EO
Gentec-EO Original Series User manual

Gentec-EO
Gentec-EO PRONTO-50 Series User manual

Gentec-EO
Gentec-EO PRONTO-Si User manual

Gentec-EO
Gentec-EO TUNER User manual

Gentec-EO
Gentec-EO PRONTO-50 Series User manual

Gentec-EO
Gentec-EO SDX-1239 User manual

Gentec-EO
Gentec-EO TPM-300-CE User manual

Gentec-EO
Gentec-EO PRONTO-250-PLUS User manual

Gentec-EO
Gentec-EO PRONTO-250 User manual

Gentec-EO
Gentec-EO Pronto Series User manual