GENUV MG-06 User manual

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
1. Product Features and Applications
2. Product Specification and Configuration
3. Product Installation
4. Operation
5. Relay Operation method
6. Relative Reaction Curves of UV Sensors
7.A/S Request in Case of Product Failure
8. Notes
Instruction Manual
MG-06
(UV Radiometer)
Ver.1.0

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
Page 1
Fig. 1 UV Radiometer 6.0
1. Product Features andApplication
1) Features
Display is 3 types : Relative Power (RP), Accumulative Time (AT)
Output type : Relay contact
Standard panel : DIN 48 * 48 (LED display)
2) Applications
UV Lamp Monitoring / Water Sterilizer / Air Cleaner/ UV Hardener / UV Irradiator
2. Product Specification and Configuration
1) Display Panel
- Panel Size: 48 ×48 ×85.5 ㎣, Panel cutting size : 45 ×45 ㎟(Tolerance : -0, + 0.5 ㎜)
- Power Supply: 85 ~ 265 VAC (50/60 ㎐), Consumption : ≤3 W
- Operating Temperature: 0~50 ℃, Operating Humidity: 35 ~ 85 % RH
- Relay : 250AC 3A, 30VDC 3A.
(1) Display Panel (4) Panel Cutting Size(mm)
(2) Size of Display Panel
[ Front view ]
2) Sensor Probe
Brown Line : #3, Black Line : #4, Blue Line : #5
Information of Sensor Probe is in the enclosed Certificate of Quality (CQ)
3) Power cable and output cable are not included.
(3) Size of Display Panel
[ Side view ]
48.00
48.00
8.20
44.00
85.50
45.00
(-0,+0.5)
45.00
(-0,+0.5)

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
Page 2
3. Production Installation
1) Mounting the Sensor Probe
a. Operation temperature of Sensor Probe is -30 ~ 85 ℃(-22 ~ 185℉)
b. UV sensor of Sensor Probe and UV light source should be fixed in set distance to do verticality.
Once a mounting distance has been determined, make sure to use the same distance for any additional sensors used
for other lamps; doing so will provide a more accurate comparison of irradiance among different lamps.
c. In LW series, Sensor Probe connects after close teflon tape 3~4 times in screw page.
d. After mounting and positioning the sensor probe, make sure the window of Sensor Probe is clean and dry.
If it is not clean, then gently wipe it off with a lint free swab.
2) Mounting the Display Panel
a. Mount the Display Panel in a temperature & humidity of less than 50 ℃(122℉) & 35 ~ 85 %RH.
b. 3 wires of Sensor Probe connect to #3, #4 and #5 of Display.
[3 : +5V (Brown), 4 : GND (Black), 5 : Vin (Blue)].
c. Power cable connect to #9 and #10 of Display. (Power is AC voltage. Be careful!)
d. Alarm for UV lamp running overtime #1 and #2 of Display.
e. Relay can use the 5A / 230VAC, and its criterion is the setting value of C-01.
#8 is N/O (Normal Open), #6 is N/C (Normal Close) and # 7 is Comm (Common).
If you want to get the short signal under 70 %, you should set 70 in C-01and connect the each wires
in #7 and #8 of Display.
Fig. 2 Connection Diagram of UV Radiometer 6 Fig. 3 Connection picture of UV Radiometer 6
Sensor Probe
AC Power 9
10
3(Brown)
4(Black)
5(Blue)

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
4. Operation
1) Setting
2) Panel Instruction
3) Keys Function
a. Set Key
1. 100% resetting, Clear timer to zero.
2. Setting parameter.
3. Page down and shifting the setting the bit.
b. Page up Key : Pressing [Set Key] + [Page down Key] together can enter or quit the parameter displaying page.
c. Parameters adjustment steps and displaying page :
②, ③is fault page, which will appear when the E-01 and E-02 occurring.
Page 3
Mark
Name
Range
Application
C-01
Lower limited
value of the UV-
low Alarm
0~100%
According the actual requirement, set the lamp low limited
warming line. In case the inspected value lower than last value,
warning up and the delay be actived.
C-02
0~19990 hours
After set the lamp using time. In actually run time, when the
totally run time exceeds the original setting time. Warning up
and delay be actived.
ALM
GUVD-MG06S
%
RUN
Hs
SET
Set Key Page down Key
Run Time
Relative power(%)
RUN Light
Alarm Light
ALM
GUVD-MG06S
%
RUN
Hs
SET
ALM
GUVD-MG06S
%
RUN
Hs
SET
ALM
GUVD-MG06S
%
RUN
Hs
SET
①②③

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
ALM
GUVD-MG06S
%
RUN
Hs
SET
ALM
GUVD-MG06S
%
RUN
Hs
SET
4). Process of Parameter Resetting to 100%
*When you set the relative power 100% , the AT time will set to 0 hrs.
e.x) Setting of C-01
a. Press [Set Key] + [Page down Key] together, enter picture ④.
b. Press [Set Key] entezzr picture ⑤. And then press the [Page down Key] can reduce the flashing number.
c. Press [Sett Key] second time, will appear picture ⑥, that flashing bit is shifted.
d. Press [Page down Key] flashing number will be reduce by one press. Press [Set Key] to chose the digit, then press
[Page down Key] chose the number. Fix it by press [Set Key] 3seconds.
e. Press shift back to picture ④.
f. Back from parameter setting : One way is, press [Set Key] + [Page down Key] together, the other way is leave it for 1minute.
g. Parameters adjustment steps for C-02 same as above C-01 setting steps.
Page 4
ALM
GUVD-MG06S
%
RUN
Hs
SET
ALM
GUVD-MG06S
%
RUN
Hs
SET
Press the set key last for 3sec.
ALM
GUVD-MG06S
%
RUN
Hs
SET
④⑤⑥
⑦⑧
ALM
GUVD-MG06S
%
RUN
Hs
SET
ALM
GUVD-MG06S
%
RUN
Hs
SET
Press the set key
last for 5sec.

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
5) Alarm codes explanation
Page 5
Mark
Name
Application
Reason and solution
E-01
UV low limited
warning
As the UV relative
value down to 50% of
original setting, device
be warning.
1. Quartz glass on the top of sensor is dirty. Cleaning it.
2. The irradiance of the lamp maybe weak, please
replace the UV lamp.
E-02
UV lighting
period exceeded
warning
As the UV run time
exceed to original
setting run-time,
device be warning.
Replace lamp
5. Relay Operation method
1) Wiring diagram
Fig. 5 Wiring diagram of UV Radiometer 6
a. Connect AC power Terminal #9 and #10 and connect Sensor cable to terminal #5 (Blue wire), #3 (Brown wire), #4 (Black wire).
Please refer to Fig. 6.
b. Relay output terminal is (#6 ~ #8),
If you want to use relay output please use with #7 and #8 (N/O , Normal open , in normal case the terminal status is open, but if the signal is ON,
the terminal will be closed. OR #6 and #7 (N/C , Normal Close, in normal case the terminal status is close, but if the signal is ON,
the terminal will be Open).
In normal status the relay will contact #6 and #7 and in abnormal status the relay will close #7 and #8
If you use large capacity ballast , please use capacitor (220㎊/2㎸) as Fig. 7.
1
3
4
5
2
6
8
9
10
7
Time Alarm
NO
+5VDC
GND
UV Sensor
NC
UV low Alarm C
NO
Power
85~265VAC
50/60Hz

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
Fig. 6 Connect diagram of MG-06
Red light Green light
Red light
ballast
Back side of MG-06 , Numeric is terminal Number.
5 4 3 2 1
Ballast and light connection
AC power
Fig. 7 Wiring diagram of Green light and Red light
(Red light means abnormal , Green light means normal)
Power cable Power cable
Green light
ballast
Capacitor
220㎊/2kV
Page 6
In normal status , the relay will connect #6 and #7 , the Green light will be ON and Red light will be off
In abnormal status , the relay will connect #7 and #8 , the Green light will be of and Red light will be ON.
Sensor Probe
AC Power
Supply 9
10
3
4
5 10 9 8 7 6
6. Relative Reaction Curves of UV Sensors
Fig. 4 Relative Responsivity Curve of UV Sensor
(1) UVV Sensor (2) UVC Sensor

GNF - 722 - 12 (0) Genicom Co., Ltd. A4(297×210㎟)
Page 7
8. Notes
1) CAUTION
TURN ALL POWER OFF. NEVER EXPOSE EYES OR SKIN TO UV LIGHT FROM ANY SOURCE
WEAR GLOVES, FACE SHIELD/GLASSES(PER ANSI Z87.1)
AND COVER ALL EXPOSED SKIN. DO NOT TOUCH LAMP GLASS WITHOUT GLOVES.
2) NOTE
Read this entire instruction sheet before starting the installation.
7.A/S Request in Case of Product Failure
1) Should any failure is found in product, please call the sales company or customer center for A/S.
2) Product warranty period is 1 year from the date of procurement with no charge.
However, failure which is caused by user’s misuse or carelessness within warrant period or any failure after
the warrant period shall be chargeable for it’s A/S.
3) Product inquiry and on-line customer service
Tel : +82-42-862-3982, Fax : +82-42-862-2982
E-mail : uvsensor@geni-uv.com
Web site : http://www.geni-uv.com