* Valox® is a registered trademark of SABIC Innovative Plastics
Chlorine Digital (S3L) Wiring
• Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.
• Refer to the wiring diagram above to connect the 2650 cable to the terminals on the 8630 Chlorine Transmitter.
• Maximum cable length is 30 m (100 ft).
• Cable should have noise immunity for the sensor signal of up to 30 m (100 ft).
• Cable capacitance must be less than 30 pF/foot.
35.0 mm
(1.38 in.)
61.0 mm
(2.40 in.)
115.8 mm
(4.56 in.)
Gnd
I/O
V+
15
14
13
Signal Ground
Digital (S3L) data
+5 VDC
White
Red
Black
Shield
Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090
For the most up-to date information, please refer to our website at www.gfsignet.com
3-2650.090 Rev. 10 01/19 English © Georg Fischer Signet LLC 2019
General
Compatible Sensors: All Signet Amperometric DryLoc Sensors
Compatible Instruments: Signet 3-8630-3P Chlorine Transmitter
Mounting: DryLoc connection
Materials: Valox® (PBT)*
Cable: 4.6 m (15 ft) 3 conductor shielded, 22 AWG
Shipping Weight: 0.64 kg (1.41 lb)
Performance
Electronics Accuracy: < 5 nA or 1% of reading, whichever is
greater @ 25 °C over full input range
Temperature: ± 1.0 °C (PT1000) over full operation range
(when calibrated at ambient temperature)
System Response: 500 ms (update rate)
Operational Range: ± 450 nA
Resolution: 0.1 nA
Electrical
Input Specifi cations:
Sensor: Raw signal
Temperature: PT1000 RTD
Output Specifi cations:
Digital (S3L): Serial ASCII, TTL level 9600 bps
Max. Cable Length: 30 m (100 ft)
Power Supply Input:
Digital (S3L) mode: 5 to 6.5 V ± 10%, 3 mA max
Environmental
Storage Temperature: –20 °C to 85 °C (–4 °F to 185 °F)
Operating Temperature: 0 °C to 85 °C (32 °F to 185 °F).
Electronics only
Relative Humidity: 0 to 95%, non-condensing
(no electrode connected)
Enclosure Requirements: NEMA 4X/IP65 with electrode
connected
Wiring
• English
• Deutsch
• Français
• Español
• 中文
Specifi cations Standards and Approvals
CE
RoHS compliant
Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and
OHSAS 18001 for Occupational Health and Safety.
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received,
including interference that may cause undesired operation.
China RoHS (Go to www.gfsignet.com for details)
Dimensions