Item # Contents
601024 ODO200M meter, manual, USB cable, 9V battery,
transport case, probe with 1 meter cable
601025 ODO200M meter, manual, USB cable, 9V battery,
transport case, probe with 4 meter cable
601026 ODO200M meter, manual, USB cable, 9V battery,
transport case, probe with 10 meter cable
INITIAL INSPECTION
Carefully unpack the instrument and accessories and inspect for
damage. Compare received parts with items on the packing list. If any
parts are damaged or missing, contact YSI Customer Service at 800-
897-4151 (+1 937 767-7241) or the authorized YSI distributor where
the instrument was purchased.
PRECAUTIONS
THE PROBE
Sensor caps last longer if stored in a moist environment and kept clean
and free of scratches in the paint layer. Erratic readings can result from
damaged or dirty caps. If unstable readings occur from large scratches
in the paint or dye layer, replace the sensor cap. See the Maintenance
and Troubleshooting section of this manual for information on how to
clean or replace the sensor cap. The average sensor cap replacement
interval is 12 to 18 months, although they may last longer.
Avoid substances that may damage probe materials such as
concentrated acid, caustics and strong solvents. Probe materials
include urethane (cable), Xenoy (probe body), acrylic (sensor cap) and
acetal and stainless steel (probe guard).
To prevent the sensor cap from drying out, store the probe in the gray
calibration/storage sleeve with a moistened, clean sponge. In the event
the sensor cap dries out, rehydrate it following the instructions in the
Maintenance and Troubleshooting section of this manual.
INITIAL SETUP
Throughout the manual, the term “probe” refers to the end of the cable
where the sensor is located, the term “sensor” refers to the Optical
Dissolved Oxygen sensing portion of the cable/probe assembly, and
the term “sensor cap” refers to the removable sensing cap that is
replaced about once per year (Figure 1).