Giltron 9851 User manual

사 용 설 명 서- 영문
MULTI-FUNCTION DATALOGGER
WITH PRINTER
9851, Temp./RH logging printer

INDEX
Introduction
Material Supplied
Features
Comparison table
Meter setting
Single measurement
Multiple measurement
Automatic logging
Diagram
9851 Temp./RH logging printer
Software
File
Mode
Port
Commad
Print
1
1
9
13
14
15
15
16
17
18
...............................
...................................
.........................
.........................
..........................
........................
...............................
.......................
........................................
.......................................
.......................................
...................................
...................................
.......................
..........................
Specification sheet
Trouble shooting
7
8
1
2
.........
3
4
5
6
.........................
.....................
Top cover of meter & Accessory
Meter keypad

This meter designed with three measurement modes:
1)Single point measurement
2)Multiple points measurement ( Manually record)
3)Automatically Logging
INTRODUCTION
MATERIAL SUPPLIED
Features
Check for damaged or missing parts in your meter before starting.
Thank you for purchasing this multiple function printer! This unit
has been developed to meet your max.satisfaction by its user-
friendly design. Please review the entire manual for a complete
overview of how to operate this meter.
The complete meter set should contain
User friendly interface
RS232 cable and software enable to link with PC to download
Backlight function
Tripod mountable for long time use
Power off time selectable
Big Dot matrix LCD
Powered by 4pcs AA batteries or 9V adaptor
Measuring/Programming ....Anywhere, anytime
1
& upload
1. Meter
2. 4pcs AAbatteries
3. Printing Thermo-paper
4. Operation manual
5. Carry Case
6. RH probe (Acalibration data small note enclosed)
7. RS232 cable
8. Software CD
Printing anywhere, anytime

COMPARISON TABLE
Measured parameters of each meter:
MODEL FUNCTION PARAMETERS
2
Temperature,DP
Humidity, Wet Bulb
9851 Hygro-thermometer
Please refer to below for the difference of each measurement mode.
MEASUREMENT DESCRIPTION MEMORY POINT
Single Point
Measurement
Single point
measurement No Memory
Multiple Point
Measurement
File name editable
multiple point
measurement 99 points
Automatic
Logging Programmable
datalogging function 3000 points
Comparison table of multiple points and datalogging measurements
Editable or
Default with date
and time (Note 1)
Default (Note 2)
Press "ON/OFF" key to
measure and store
by pressing specified
keys
MEASURING /
SAMPLING
Automatically measure
and store according to
pre-set parameters
(Note 3)
MULTIPLE POINT
MEASUREMENT DATALOGGING
RECORD 99 points
FILE
NAME
3000 points

Note 1: Eachrecord is designed to show withcurrent date and
time if userdoesn't edit any.
For example if the filename is "05-06 09:21:51" means thedate
is 6th May(or 5th June, based on yourdate mode
setting )and thetime is 09:21:51.
Note 2: Eachrecord is designed to show withthe current date
and time.
Note 3: a)To set upBegin-Date,Start-Time,End-Date,Suspend-
time and samplerate from the meter.
b)Each meter startsto record from Begin-Date & timewith
specified sample rateuntil Suspend-time.
c)Automatically start againnext day from Start-Time until
End-Date.
d)Logging stops recordingwhen End-Date or max.
memory points isachieved.
e)Logging can bestopped and start again with thesame
setting.
3
ACCESSORY The temperature and humidity
sensors are built in theprobe.
Please keep the small piece
of noted enclosed with each
probe for calibration purpose.
The following illustrationis the connectors for 9851.
TOP COVER -CONNECTOR

METER KEYPAD-WITH PRINTER
4
ON/OFF
12
5
8
0
3
6
9
4
7
F2 F3 F4F1
SYM.
GHI
PQRS
DEF
MNO
WXYZ
ABC
JKL
TUV
LINE
FEED
Based on different functions, the
top cover connectorand color are
different. It is dark greenfor 9851.
TopCover
LCD
Keypad
Backlight
Up and Downkeys
Paper Roll cover
Welcome
MEAS MEM LOG SET
Function keys
Corresponding
1SYM stands for"1,*,.,:,$,+,-,="
2ABC stands for"2,a,b,c,A,B,C"
3DEF stands for"3,d,e,f,D,E,F"
4GHI stands for"4,g,h,i,G,H,I"
5JKL stands for"5,j,k,l,J,K,L"
6MNO stands for"6,m,n,o,M,N,O"
7PQRS stands for"7,p,q,r,s,P,Q,R,S"
8TUV stands for"8,t,u,v,T,U,V"
9WXYZ stands for"9,w,x,y,z,W,X,
Y,Z"
0 stands for"0 and space"
DC adaptor port
RS232 Port

LCD Cont.(1-5): LCD brightness. From darkest 1 to brightest 5
Prn Cont. (1-9): Printing darkness. From lightest 1 to heaviest 9
Unit: The unit could be metric or imperial / or
Auto Off: The selectable time frame is 1 to 20 min.
CF
Set Clock: To choose the date mode and set your local time.
Date mode:MM-DD-YY or DD-MM-YY or YY-MM-DD
Set ID: If choose "Disable", the ID will not be printed out.
ID: To edit an ID of this printer , the ID will be printed out at
printing mode.
Descriptions :
Select item: To select which measured parameters need to be
showed on the LCD. The unit of each parameter
could be selected from here as well.
Pressing or key to shift the cursor.
Pressing EDIT to enter modification mode.
Pressing PRN to enter printing mode.
Pressing NEXT/BACK to enter next or previous page.
Pressing EXIT to return to main menu
METER SETTING
Before measuring, pleasecheck the meter setting first toconfirm
the basic settingsof the meter are what youneed.
Toenter each command, just press thecorresponding F1 to F4
key.
F2 F3 F4F1
Welcome
MEAS MEM LOG SET
SET
F4 :Meter Setting
5

There are threemeasurement modes :
1. Single measurement(MEAS)
2. Multiple measurement(MEM)and
3. Autologging (LOG).
Press F1 (MEAS) to proceed single measurement
After entering measurement mode , press F4 (PRN) to enter
printing mode. The printing can be stopped any time by press
F1(STOP or EXIT) key while it is printing .
External temp. & RH probe calibration introduction is in the
Single measurement mode .
Definition
SINGLE MEASUREMENT
MEAS
F1 :Single Measurement
F2 F3 F4F1
Welcome
MEAS MEM LOG SET
6
Operating the singlemeasurement as a general meter.
The printing contrastcan be changed by pressing HIGHor
LOW key .

MULTIPLE MEASUREMENT
This is tomanually record what you measured withreal time
, and filename (a note or identification forthe record ) is edit-
able to memorizeor recognize .
There are total99 memorize capacities in this measurement.
Press MEAS to measure the parameters.
Select F1(ABORT) or F4(SAVE) to exit the measurement
or memory the record .
Press F3 ( Edit ) to edit the file name of recorded value .
While editing the file name, select from the keys 1SYM, 2ABC,
3DEF, 4GHI...., Press and hold one of the key to select from
the displayed letters by releasing the key .
Press CLR shortly to delete one data or press for over 2 seconds
to delete the whole memory.
Before printing, select the printing range by pressing EDIT
key,then press START to print.
Definition
7
F2 :Multiple measurement
MEM
F2 F3 F4F1
Welcome
MEAS MEM LOG SET

AUTOMATIC LOGGING
Press SET to set needed setting first. The Date mode is
based on the mode you set under SET (see page 5). The
sampling rate is from 1 to 7200 seconds. "Expect" means
total memory of this meter, "Remain" means how many
memory left in meter.
Press START to begin the logging function. While logging:
Press VIEW to see the data ( more than one ) in the screen,
or press MEAS to see a real-time data,
or press STOP to quit logging.
Press P-PG or N-PG to review previous or next 100 points.
After stopping, press START to begin again. The sampling rate
and previous record are remained if the previous setting
has not been changed.
"Suspend" is the stop time of each day during setting date.
If you want to record 24 hours a day, you have to set :
00:00:00 as "Start" , and 23:59:59 as "Suspend" .
Definition
F2 F3 F4F1
Welcome
MEAS MEM LOG SET
F3 :Datalogging
LOG
This is toautomatically record what you set upbefore
measurement with realtime, first to set records start/stop
date and time,sample rate , memory points. The screen
content shows Expectmemory points and Remain memory
points while insetting.
8

EDIT PRN
EXIT NEXT
Set Clock:
MM-DD-YY HH:MM:SS
Set ID:
- - : :20
06 30 05 18 29
Disable
ID:
User Name
EDIT PRN
EXIT NEXT
LCD Cont.(1-5):
Prn Cont. (1-9):
Auto Off:
3
5
20mins
Enable
EXIT EDIT BACK
Temp
Select Item
PRN
MEAS MEM LOG SET
Welcome
EXIT Cali PRN
Temp:
Remarks:
9851 Hygo-Thermometer Setting/ Single measurement
1.Press CAL to enter calibration
mode. See Page 10 for the details
9
REC
EXIT
Temp:
C
RH
WBT
DP
60.8% 17.4C
20.0C
Temp:25.5C
RH
DP
WBT
%
C
C
Max : 20.5C
Min: 20.0C
Avg: 20.2C
Curr.: 20.3C
Temp REC
NEXT
EXIT
Temp:
Max : 60.5%
Min: 60.0%
Avg: 60.2%
Curr.: 60.3%
RH REC
NEXT
2.Under the " REC" function, " Curr." means the current value.
3.There are total 4 sub display under REC function. They are
temperature, humidity, dew point and wet bulb. Press F4 to
choose the display you need.
EXIT LXW MANUAL
RH Cali Mode Select
INPUT
Input probe cali data
Press F2
Calibrate manually
Press F3

MEAS MEM LOG SET
Welcome
9851 Hygro-Thermometer CalibrationProcedure
EXIT LXW EDIT
SAVE
View Cali Data
Long press
EXIT LXW MANUAL
RH Cali Mode Select
INPUT
Zcal : -7340019 E-4
Scal : 5399497 E-5
RH 32.8 %
EXIT Cali PRN
Temp:
REC
RH
WBT
DP
60.8% 17.4C
20.0C
Temp:25.5C
Input probe cali data
Press F2
Calibrate manually
Press F3
ABORT LXW ENTER
SAVE
Input Cali Data
RH Manual Calibration
RH 60.8%
RH Manual Calibration
RH 75.3 %
RH Manual Calibration
ABORT
Remarks:
1. Be sure to keep a calibration
note enclosed in the package
This's for future RH probe cali-
bration use
2. While keyin the calibration value,
3. PressABORT anytime to exit without save or pressEND to
save and exitif you don't want to waitfor 30 minutes.
4. The LCD value willflash while doingmanual 33% and 75%
calibration until theprocedure finished.
5. Suggest tocalibrate at stable room temp. environment.
Flashing for 30mins
Flashing for 30mins
No need to use
salt bottles Need to use
salt bottles
(Need to plug probeinto salt bottles)
10
Note:
"Input cali.data" methodis for RH
probe replacement.
"Manual calibration" methodis for
RH probe replacement& old probe
re-calibration.
Zcal : E-4
Scal : 5399497 E-5
-7340019
EXIT 75.3
32.8
(Put probe intocorresponding salt bottles
for 30 minsbefore press F2or F3)
END
ABORT END
END
refer to thevalue indicated on
the note. The values areslope and offset value ofeach probes.
Temp. 23.5 C
Temp. 23.5 C
Temp. 23.5 C

MEAS MEM LOG SET
Welcome
EXIT MEAS EDIT NEXT
01:
EXIT CLR PRN BACK
ABORT SAVE
01:
06-30 23:56:49
EXIT MEAS EDIT NEXT
Remarks:
9851 Hygro-Thermometer Multiplemeasurement
Note:
After saving #01memory, press
for next measurement.Then,
#02 with acursor will be
highlighted.
11
02:
01:
Temp:
RH:
DP:
WBT:
02:
RH:
DP:
WBT:
Temp:
Temp:
RH
WBT
DP
60.8% 17.4C
20.0C
Temp:25.5C
02:
RH: 60.8%
DP: 17.4C
WBT: 20.0C
Temp: 25.5C
01: ROOM1
ABORT
CLR BACK ENTER
4ghiGHI
RH: 60.8%
DP: 17.4C
WBT: 20.0C
Temp: 25.5C
01:
1.Press MEAS tosee a real-time measurement values,
then pressABORT without saving, press SAVE to store
2.Press CLR shortlyto delete one data or pressfor over 2 seconds
to delete thewhole memory.
3.In Edit mode,press ENTER to save the descriptionor press
abort to leavewithout save. Press CLR to clearthe description
or press BACKto previous text.

EDIT VIEW
EXIT NEXT
Begin:
Start:
End:
Suspend:
Rate:
Expect:
Remain:
30-06-05
02:28:55
02-07-05
03:28:55
60Sec(s)
3000Point(s)
3000Point(s)
Begin:
Start:
30-06-05
02:28:55
CLR PRN
START BACK
End:
Suspend:
Rate:
Expect:
Remain:
02-07-05
03:28:55
60Sec(s)
3000Point(s)
3000Point(s)
START
EXIT SET NEXT
STOP VIEW
Logging...
MEAS ESC
0001: 06-30 02:28:55
0002:
STOP ESC
MEAS
N-PG PRN
P-PG BACK
9851 Hygro-Thermometer Datalogging
1.Before logging, make sure you have set up the parameters
by pressing SET key , otherwise no data will be displayed.
MEAS MEM LOG SET
Welcome
12
0001:
0002:
Temp:
RH:
DP:
WBT:
Temp: 25.5C
RH: 60.8%
DP: 17.4C
WBT: 20.0C
0001:
0002:
Temp:
RH:
DP:
WBT:
0001: 06-30 02:28:55
Temp:
RH
WBT
DP
60.8% 17.4C
20.0C
Temp:25.5C

SOFTWARE
Material & O/S
Software CD and RS232 cable with D-sub plug are needed.
Operating system need to be WIN98/2000/NT/XP or above.
New Save Load Print Upload
file names
Download
Logger
Download
Memory
Download
All
The enclosed softwareis a quick tool for youto download the
memorized data toPC for further analysis or uploadthe pre-edited
file names tometers, this may save your settingtime.
FILE
New: To create a new file.
13
Load: To open a saved file.
Save: To save current file.
Save as: Tosave current fileas new filename.
Print: To print all data or loggingdata or memory data.
Exit: To withdraw the software
File

14
Print
Load
Mode-Memory
Mode-Logger
MODE
There are twodata modes. One is Memory, the other one is
Logger.
Toselect the mode from main menuor quick selection window.
In logger mode,you may select from "Goto" tochoose logged
data range. There are 100points in onepage.

Command
15
PORT
Select the correctCom. port to ensure the meteris
communicated with PC.I
For most PCs,the Com. port number is COM1.
There are 8COM ports for selection in thissoftware.
n connecting mode,"PC Mode"
will appear onthe LCD and "COM #" &Edit will show
on the leftbottom corner of screen.
Port
There are fourcommands in this software.
-Download measured loggingdata from meter to PC.
-Download measured memorydata from meter to PC.
-Download measured logging& memory data from meter toPC.
-Upload pre-edited filenames from PC to meter.
Toedit the file names, click leftkey of mouse twice quickly
, then thecursor will flash to indicate thecolumn is ready for
editing.
Before uploading ordownloading, make sure you choose the
correct download mode. If you choose "Download memory
data" but executedownloading "Logger data ", you willhave to
switch the modeto have the correct display.
COMMAND

Select the datayou want to print : There are 3selection :
1. Print onlyLogger Data
2. Print onlyMemory Data
3. Print all( Both Logger Data and MemoryData in sequel )
Example: (If you choose Print all from 9811 IR meter )
Page1
<< DataLogging Printer Report Date:07-12-2005 TIME:17:26:30 >>
Memory Data:
STEP DESCRIPTION Temp
======= ============== ======
1 07-12 10:46:24 26.2 C
2 07-12 10:46:30 26.5 C
3 07-12 10:46:33 25.8 C
4 07-12 10:46:38 25.9 C
5 07-12 10:46:48 26.1 C
.............................
Logger Data:
STEP DESCRIPTION Temp
======== ============== ======
1 07-12 13:00:00 28.5 C
2 07-12 13:10:00 28.4 C
3 07-12 13:20:00 28.0 C
4 07-12 13:30:00 27.8 C
5 07-12 13:40:00 28.1 C
...............................
PRINT
16

1. Power onbut no display
a)Make sure thetime of pressing "ON/OFF" key ismore
than 200 miniseconds(ms).
b)Check the batteryare in place and make surethey are
at good contactand correct polarity.
c)Replace with newbatteries and try again .
d)Move the batteriesfor ten seconds ,then replace back
again.
2. Display disappear
a)Check whether thelow battery indicator is displayed
on or beforedisplay disappears. If yes , replacewith
new batteries .
TROUBLE SHOOTING
17
3.E 2. Problem:The value is underflow.
4.E 3. Problem: The value is overflow
5.E 4. Problem: Calculated source value occur errorevent
6.E11. Problem: RH calibrationerror
7.E33. Problem: MeasureHardware Error.
길 우 트 레 이 딩
서울특별시 영등포구 선유로33길 15, 703호
(양평동 3가, 에이스테크노타워)
02)2636-0009(대)팩스02)2636-4753
[email protected] www.gilwoo.co.kr
Table of contents
Popular Data Logger manuals by other brands

Huawei
Huawei FusionModule 1000A20 Initial Configuration Parameter Manual

Elitech
Elitech RCW-400A user manual

Pego
Pego PLUSR 200 EXPERT Use and maintenance manual

Microtronics
Microtronics myDatalogEASY V3 user manual

Canary Systems
Canary Systems MLMux4 user guide

AEMC instruments
AEMC instruments Simple Logger II Series quick start guide