Gima 31002 User manual

PROFESSIONAL MEDICAL PRODUCTS
MICROSCOPIO BIOLOGICO
BIOLOGICAL MICROSCOPE
MICROSCOPE BIOLOGIQUE
BIOLOGISCHES MIKROSKOP
MICROSCOPIO BIOLÓGICO
MICROSCÓPIO BIOLÓGICO
31002
M31002-M-Rev.3.09.20
Importato da / Imported by / Importé de /
Importiert von / Importado de / Importado de:
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
Fabbricante / Manufacturer /
Fabricant /
Hersteller / Fabricante / Fabricante
:
NINGBO YONGXIN OPTICS CO., LTD.
No. 169 Mujin Road, Hi-tech Park Ningbo, China
Made in China
Modello / Model / Modèle
Vorlage / Modelo / Modelo N-126

15
Index
SAFETY NOTE......................................................................................................................................16
MAINTENANCE AND CARE................................................................................................................16
1. COMPONENTS NAME ....................................................................................................................17
2. ASSEMBLY.......................................................................................................................................18
2.1 Assembly Diagram....................................................................................................................18
2.2 Assembly procedure.................................................................................................................19
3. ADJUSTMENT AND OPERATION .................................................................................................20
3.1 Adjustment set diagram...........................................................................................................20
3.2 Adjustment and Operation.......................................................................................................22
4. TECHNICAL SPECIFICATIONS ......................................................................................................24
4.1 Main Specications .................................................................................................................24
4.2 Eyepiece and objectives ..........................................................................................................25
5. TROUBLESHOOTING GUIDE .........................................................................................................26
5.1 Optical System ..........................................................................................................................26
5.2 Mechanical System...................................................................................................................27
5.3 Electrical System ......................................................................................................................27
Safety Note
1. Open the box carefully to avoid the accessories, like lens, dropping to ground or being damaged.
2. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking
environment. Make sure the stage is at, horizontal and rm enough.
3. When moving the microscope, carefully carry it with the handle and the base.
4. Make sure the instrument is earthed, to avoid lighting strike.
5. For safety, be sure the main switch is in “O” (off) state and cut off the power supply before replacing
the bulb or the fuse. If you replace the bulb during use or right after use, allow the lamp bulb and the
lamp house to cool completely before touching.
(Designated bulb: 1W S-LED)
6. Check the input voltage: be sure the input voltage which is signed in the back of the microscope is
consistent with the power supply voltage, or it will bring a serious damage to the instrument.
7. Always use the power cord provided by Gima.
8. The electrical equipment of the microscope should be discard as electronic waste.
Maintenance and Care
1. All the lenses have been adjusted properly; do not dismount them by yourself please.
2. The nosepiece and coarse and ne focusing parts are so delicate that it is forbidden to disassemble
them carelessly by yourself.
3. Keep the instrument clean, and do not pollute the optical element when wiping away the dust on the
instrument.
4. The contaminations on the prism, like ngerprints and oil smudges, could be gently wiped with a
piece of soft cloth or tissue paper, gauze which has been immersed in pure alcohol or ether. (Note
that the alcohol and ether are highly ammable, do keep them away from the re or potential sources
of electrical sparks, and use them in a drafty room as possible as you can.)
5. Do not attempt to use organic solvents to clean the microscope components other than the glass
components. To clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral deter-
gent.
6. When using, if the microscope is splashed by liquid, cut off the power at once, and wipe away the
splash.
ENGLISH

16
7. Do not disassemble any parts of the microscope, as this will affect the function or reduce the perfor-
mance of the microscope.
8. Place the instrument in a cool, dry position. When not using the microscope, keep it covered with a
dust cover. Make sure the lamp socket is cool before covering the microscope.
1. Components Name
Eyepiece Binocular viewing head
Nosepiece
Objective
Condenser
Stage
Illumination
Main body
ENGLISH

17
2. Assembly
2.1 Assembly Diagram
The following gure shows the installation sequence of the components. The number in the gure
shows the assembly steps.
• Before installing, be sure every components is clean, do not score any parts or glass surface.
• Keep well with hexagon wrench provided. When changing the components, you will need it again.
10X or 16X Eyepiece
Binocular viewing head
Objective
Main body
Power cord
2
4
1
3
ENGLISH

18
2.2 Assembly procedure
Installing binocular viewing head (g. 1-2)
Insert the binocular viewing head into the microscope head and turn it to a proper position, then tight it.
g. 1 g. 2
Installing the eyepiece (g. 3-4)
Insert the eyepiece into the eyepiece tube until they are against each other (g. 4).
g. 3 g. 4
Installing the objective (g. 5-6)
1. Adjusting the coarse focus knob until the support device of the mechanical stage reaches its low limit
position.
2. Screw the lowest magnication objective into the nosepiece from the left or the right side, then
revolve the nosepiece clockwise and mount other objectives by the sequence of low to high magni-
cation.
Installing objective this way will make the change of magnication to be easier during using.
• Clean the objective regularly, for lens is susceptible to dust.
• When operating, use 10× magnication objective to search and focus specimen rstly, then replace
with higher magnication objective if necessary.
• When replacing the objective, slowly turn the nosepiece until you hear “clicked”, that means the ob-
jective is in place.
ENGLISH

19
g. 5 g. 6
Connecting the Power Cord (g. 7-9)
The cable and cords are vulnerable when bent or twisted, never subject the power cord to ex-
cessive force.
1. Set the main switch 1 to “0” (off) state before connecting the power cord.
2. Insert the plugs 2into the power jack 3of the microscope safely.
3. Plug the power cord 4into the power supply receptacle 5. Make sure the connection is well.
• Do use the supplied power cord all the time. If lost or damaged, select the same standard cord,
please.
• A wide range of voltage, like 100V~240V, is acceptable for this microscope.
g. 7 g. 8
Replacing the fuse (g. 8-9)
Do remember to set the main switch 1to the
state “0” (OFF) and unplug the power cord be-
fore replacing the fuse. Rotate the fuse out of the
holder 3, replace with a new fuse, then rotate it
back to the holder again.
g. 9
12
3
4 5
ENGLISH

20
3. Adjustment and Operation
3.1 Adjustment set diagram
g. 10
Portrait Adjustment Knob
Aperture Iris Diaphragm
adjustment
Lateral Adjustment Knob
Right Coarse Focus Knob
Main Switch Tension Adjustment Collar
Right Fine Focus Knob
ENGLISH

21
g. 11
Left Fine Focus Knob Left Coarse Focus Knob Brightness
Adjustment Knob
Clocking Set
Condenser Focusing Knob
Interpupillar Distance Indicator
ENGLISH

22
3.2 Adjustment and Operation
Brightness Adjustment (g. 12-13)
1. Connect the power cord and set the main switch 1to “ – ”state (ON).
2. Turning the brightness adjustment knob 2.
g. 12 g. 13
Placing the Specimen (g. 14)
1. Place the specimen the center of the mechanical stage and use the stage clips to clamp it.
2. Turn the portrait and lateral adjustment knob 1 of the mechanical ruler, move the specimen to the
required position.
• Be careful when changing the objective. If you nish the observation with the short working distance
objective, and want to change another one, becareful of not letting the objective touch the specimen.
g. 14 g. 15
Focusing the Specimen (g. 15)
1. Focus the specimen with 10x objective. To avoid the objective touching the specimen during focus-
ing, you should raise the mechanical stage to let the specimen close to the objective at rst, then
slowly separate them to bring the specimen to focus.
Turn the coarse focus knob 1conversely to lower the specimen and search images in the 10×ocular
simultaneously, and then use the ne knob 2to focus. After that, you can replace with other magni-
cation objectives safely, and focus without the risk of damaging the specimen.
• To make the observation more convenient, you can use the locking set 3to x the stage in a vertical
direction.
1 2
2 1 3
ENGLISH
1

23 ENGLISH
Condenser Adjustment (g. 16)
Turn the condenser focus knob 1to move the
condenser up and down. Raise the condenser
when using the high magnication objective,
and descend it when using the low magnica-
tion one.
1. Focus the specimen with 10x objective.
2. Adjust the condenser focus knob 1to get a
clear image of the eld iris diaphragm.
3. Turn the condenser centering knobs 2to
center the image of eld iris diaphragm in the
eld of view.
• The condenser and the objective are coaxial.
It has been adjusted before leaving factory, so
the user needn’t to adjust them by self
• The highest position of the condenser has
been adjusted too. It also needn’t any user’s
operation. (The top surface of the condenser
is 0.03 mm - 0.4 mm lower than the stage top
surface.)
Aperture Iris Diaphragm Adjustment (g. 17)
Turn the aperture iris diaphragm stick 1to adjust
the aperture iris diaphragm.
• The aperture iris diaphragm is designed for
the adjustment of the numerical aperture, not
for the brightness.
Generally, setting the aperture iris diaphragm
to 70- 80% of the N.A. of the objective in use
will provide an image with good contrast. If
you want to observe the image of the aperture
iris diaphragm, remove one eyepiece and look
through the tube. You will see a dark circle en-
croaching on the bottom of the tube.
Adjusting the Interpupillary Distance
(g. 18)
The interpupillary distance range: 47mm-78mm
While looking through the eyepieces, move both
eyepieces round until the left and right elds of
view coincide completely.
g. 16
g. 17
g. 18
2
1
1

24
Adjusting the Diopter (g. 19)
1. Turn the eyepiece 1to adjust the diopter while
looking through it.
• The diopter range of the eyepiece is ±5 di-
opter. The number aligned to the line on the
viewing head is the diopter in use.
Adjusting the tension adjustment collar (g.
20)
Turn the tension adjustment collar 1with your
ngers. When the collar is turned in the direction
of the arrow, the tension of the coarse adjust-
ment knob 2increases. Turning the collar in the
opposite direction decreases the tension.
If the nosepiece descends on its own or if the
specimen gets out of focus quickly even when it
is brought into focus using the ne adjustment
knob 3, it means the tension of the coarse ad-
justment knob is too low. Turn the collar in the
direction of the arrow to increase the tension.
4. Technical Specications
4.1 Main Specications
g. 20
Optical System Limited optical system/Innite optical system
Viewing Head Seidentopf Binocular Head, 30° Inclined,
Interpupillary Distance 47-78 mm
Eyepiece WF 10X/20 and P16X/11
Nosepiece Backward Quadruple Nosepiece
Objective Achromatic Objective/Innite Semi-plan Objective 4×, 10×, 40×, 100×
Focusing Coaxial Coarse and Fine Adjustment,
Moving Range 0.002mm, Fine Division 20mm
Condenser Abbe Condenser, NA1.25
Stage Rackless Double Layers Mechanical Stage 150×139mm,
Moving Range 75×52mm
Illumination 1W S-LED
21 3
g. 19
1
ENGLISH

25 ENGLISH
4.2 Eyepiece and objectives
Achromatic Objective
Innite Semi-plan Objective
Eyepiece
Total Magnication
Magnication Numerical
Aperture NA
Focal Length
(mm)
Focal distance
(mm)
Working
Length (mm)
Objective
4× 0,10 0,17 31,05 37,5 dry
10× 0,25 0,17 17,13 7,316 dry
40× 0,65 0,17 4,65 0,632 dry
100× 1,25 0,17 2,906 0,198 oil
Magnication Numerical
Aperture NA
Focal Length
(mm)
Focal distance
(mm)
Working
Length (mm)
Objective
4× 0,10 0,17 45 16.8 dry
10× 0,25 0,17 18 5.8 dry
40× 0,65 0,17 4.5 1.43 dry
100× 1,25 0,17 1.8 0.13 oil
Eyepiece Magnication Focal Length (mm) Field of View (mm)
Wide eld eyepiece 10× 24,95 Φ18
Wide eld eyepiece 10× 25 Φ20
Wide eld eyepiece 16× 15.58 Φ11
Eyepiece 10× 10× 10× 10×
Objective 4× 10× 40× 100×
Total Magnication 40× 100× 400× 1000×
Eyepiece 16× 16× 16× 16×
Objective 4× 10× 40× 100×
Total Magnication 64× 160× 640× 1600×

26
TROUBLE CAUSE SOLUTION
1. The edge of the eld of
view is dark or the bright-
ness is not uniform
The nosepiece is not in the locat-
ed position (objective and light
path not coaxial)
Locate the nosepiece properly
where it clicks
The surface of the lamp becomes
black
Change a new lamp bulb
A lens (the objective, condenser,
eyepiece or collector) is dirty.
Clean it thoroughly
2. Dirt or dust is visible in the
eld of view
Dirt/dust on the specimen Replace with a clean specimen
Dirt/dust on the eyepieces Clean them
3. Visibility is poor
Image is not sharp;
Contrast is poor;
Details are indistinct
Specimen is not covered Add cover glass on it
The thickness of the cover glass is
not suitable
Use standard cover glass with
thickness of 0.17mm
Specimen is placed reversely Turn it over
Dry objective has oil on it. (espe-
cially for 40X objectives)
Wipe the oil
A lens (the objective, condenser,
eyepiece or collector) is dirty.
Clean it
Immersion oil is not used with the
100x objective
Use specied oil
Air bubbles existed in the immer-
sion oil
Eliminate the bubble
The aperture iris diaphragm is
stopped down too far
Adjust the aperture iris dia-
phragm properly
Dirt or dust on the eyepiece Clean it
4. One side of image is
blurred
Condenser is not properly cen-
tered
Center the condenser with the
centering screw
The nosepiece is not properly en-
gaged
Engage the nosepiece properly
The specimen is not clamped Clamp it with stage clips
5. The brightness is not
enough
The aperture iris diaphragm is too
small
Adjust it properly
The condenser is too low Adjust it properly
A lens (the objective, condenser,
eyepiece or collector) is dirty.
Clean it
5. Troubleshooting Guide
5.1 Optical System
ENGLISH

27 ENGLISH
5.2 Mechanical System
5.3 Electrical System
TROUBLE CAUSE SOLUTION
1. Field of view of one eye
does not match that of the
other
Interpupillary distance is incorrect Adjust interpupillary distance
2. Observation is tiring The diopter is not proper Adjust the diopter properly
The brightness of the illumination
is not proper for eyes
Adjust the lamp voltage
TROUBLE CAUSE SOLUTION
1. The bulb cannot light No power supply Check the power cord connec-
tion
The pin of the bulb doesn’t insert
properly
Insert the pin deeply
The bulb broken Replace with a new one
2. The bulb burns out
suddenly
The bulb is not the specied one;
The voltage is too high
Use the specied bulb; lower the
voltage
3. The illumination is
not bright enough
The voltage is too low Raise the voltage
4. Image icks The bulb is about to burn out Replace with a new one
The bulb is not inserted deeply Check its connection
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specic recycling point for electric and elec-
tronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Symbols
Caution: read instructions (warnings)
carefully Follow instructions for use
Product complies with European
Directive Date of manufacture
Keep away from sunlight Keep in a cool, dry place
Product code Lot number
WEEE disposal
Other manuals for 31002
1
Table of contents
Other Gima Microscope manuals