Giropes GI400i User manual

GI410 LED
GI410 LCD
.4.8_20/08/2020
EN
INDUSTRIAL WEIGHING INDICATOR
GI 400 - GI 410
USER MANUAL
El fabricante se reserva el derecho de modificar sin previo aviso las características de sus productos para introducir mejoras técnicas o cumplir con nuevas
regulaciones oficiales./Le constructeur se réserve le droit de modifier les caractéristiques de ses produits en vue d’y apporter des améliorations techniques
ou de respecter de nouvelles réglamentations./The manufacturer reserves the right to modify the specifications of its products in order to make technical
improvements or comply with new regulations.
GI410 LED
GI400 LED
GI400 LCD
GI410 LCD
Pol. Empordà Internacional Calle F. Parcela 15-16
17469 VILAMALLA - (Girona) SPAIN
T. (34) 972 527 212 - F. (34) 972 527 211

2
EN
1. MECHANICAL DESCRIPTION
1.1 FRONT PANEL
1.2 BACK PANEL
2. CONNECTION
2.1 TEST PHASE
2.2 DISPLAY OF THE SERIAL NUMBER
2.3 SETTING THE INITIAL ZERO
3. EXECUTABLE FUNCTIONS
3.1 SETTING THE SCALE TO ZERO
3.2 WEIGHT
3.3 TARE, CONSECUTIVE TARAS, GROSS-NET
3.4 BLOCKING/UNBLOCKING TARE
3.5 MANUAL TARE
3.6 MEMORIZED TARAS TLU
3.7 ACCUMULATION (ONLY WORKS WITH THE PRINT ENABLED PROTOCOL)
3.8 WEIGHT TOTALIZATION
4. OPERATION MODES (USER MENU)
4.1 CALCULATION OF THE AVERAGE UNIT WEIGHT CALAUU
4.2 PIECE COUNTING AUU
4.3 CHECKWEIGHER CHECKW
4.4 CHECKWEIGHER SIMPLE CCt SI
4.5 IT SHOWS THE WEIGHT WITH ONE MORE DECIMAL, FOR ACURACY BY 10
4.6 ACCUMULATION ACCUMU
4.7 SUBTOTAL SUBTOT
4.8 TOTALIZATION
4.9 CODE TCODE
4.10 NUMBER OF TICKET NTIC
4.11 WEIGHT MODE WEIGH
4.12 DOSAGE DOSIFI
4.13 LIMITS LIMIT
4.14 ACTIVITY CONTROL 2 REL
4.15 DATA DISPLAY MEMORIA DSD
4.16 MEMORIZED TARAS TLU
4.17 ANALOG OUTPUT (ONLY FOR LED DEVICES)
4.18 FUNCTION MENU FUNCIO
5. SPECIAL FUNCTIONS OF THE NUMERICAL KEYBOARD
6. NO METROLOGICAL TECHNICAL MENU
6.1.MENU PARAME : COM1, COM2, RS485 Y COM4
6.2 ETHERNET MENU
6.3 CONFIGURATION MENU
6.4 ABT VM
6.5 DEFAUL
7. PRINTER
7.1 TICKET FORMATS GI400
8. ERROR MESSAGES
9. CONNECTIONS
10. DSD MEMORY
11. REMOVAL OF ELECTRONIC EQUIPMENT
12. OPTIONAL BATTERY
13. WARRANTY
4.19 RELISA RELISA
USER MANUAL GI400 / GI410 EN
8.1 ERRORS OF BUILT-IN PRINTER G410 Print

3
1. MECHANICAL DESCRIPTION
1.1 FRONT PANEL
LED
LCD
SIGNS
1 Signal of limits (checkweigher)
2 Signal of level of battery (work with battery optional)
3 Signal of zero
4 Signal of steady weight
5 Gross weight
6 Net weight
7 Weighing range
8 Piece counting
9 Units of measure
10 Numerical keyboard (only 410 models)
11 Signal of relays (only for LCD)
7
2
3 4 5 6 11 1 9 8
USER MANUAL GI400 / GI410 EN

4
KEYBOARD
KEY DEFINITION
Short Press: Access to User Menus
Long Press: Technical Menu
Short Press: Do zero
Long Press: Remove tare
Short Press: Show Gross/Net
Long Press: Set tare
Short Press: Do the tare
Long Press: Introduce tare manually
Short Press: Weight totalization and total accumulation print
Long Press: Exit
Short Press: Print / Manual shipping weight
Long Press: Selector of unit (for lb version only), entry to the functions menu
(for other versions).
Switch on / Switch off
CURSOR KEYS (Only works in the editing menus)
Move editable digit to right side
Move editable digit to left side
Increases selected digit
Decreases selected digit
TECLAS RECHAZO Y ACEPTAR
Erase the current value
Hold down to go back.
Validate the selected value
NUMERICAL KEYBOARD (ONLY 410 MODELS)
Erase key
FKey to change numbers to characters and vice versa.
(by pressing enter in the functions menu)
USER MANUAL GI400 / GI410 EN

5
1.2 BACK PANEL
INOX IP54 INOX IP65
5
12
34 12 45
1Power connector IEC PG9 for power cable
2Connector RS232-Com1 (SubD 9 male tracks) PG7 for RS232-COM1
3Connector RS232-Com2 (SubD 9 male tracks) PG7 for RS232-COM2
4Connector Load cell (SubD 9 female tracks) PG7 for load cell cable
5Calibration switch Calibration switch
ABS IP54 ABS IP65
12 34
5
1Power connector IEC PG9 for power cable
2Connector RS232-COM1 (SubD 9 male tracks) PG7 for RS232-COM1
3Connector RS232-COM2 (SubD 9 male tracks) PG7 for RS232-COM2
4Connector Load cell (SubD 9 female tracks) PG7 for load cell cable
5Calibration switch Calibration switch
INOX IP68 GI410 PRINT
1 2 3 4 o 5
1PG7 for additional port Power connector IEC
2PG9 for power cable Connector Load cell (SubD 9 female tracks)
3PG7 for load cell cable Connector RS232-Com1 (SubD 9 male tracks)
4PG7 for RS232-COM1 Connector RS232-Com2 (SubD 9 male tracks)
5PG7 for RS232-COM2 Ethernet Connection (opcional)
6Pressostatic membrane Calibration switch
12345 6
6
USER MANUAL GI400 / GI410 EN

6
2. CONNECTION
Connect the terminal to the platform through a DB9 connector. To switch on the terminal you must press for some seconds the key . To
switch off the terminal it is necessary to press for some seconds the key ; but to switch off the terminal completely it can only be done by
disconnected the terminal from the electrical connection.
2.1 TEST PHASE
DISPLAY DESCRIPTION
In first place it appears the model of the equipment. (91410=GI410) o (91411 =GI411)
91410 o 91411
LO.003 Next it will show the version of the program
8.8.8.8.8.8 Then it will appear all LEDs and display segments turned ON. (only LED version)
u1.010 After that appears the version of the user’s software
9rAvEd This is followed by the message
9.804 Then displays the value of gravity set
n calib Then the message number of calibrations, followed by the corresponding value appears.
9 3.008 Firmware Internal version
p x.xxx Finnally it appears on the screen the weighing software version
2.2 DISPLAY OF THE SERIAL NUMBER
If during the phase of test display the user presses the key the device is going to show the serial number. Which you will also find printed
on the back label.
001234
2.3 SETTING THE INITIAL ZERO
When connecting the indicator it will start the setting of zero of the scale, to do so the following conditions are necessary:
1. Steady weight (luminous sign of steady weight on;
2. Weight value not less than -10% of the Max. in relation to the zero of calibration of the scale;
3. Weight value not higher than 10% of the Max. in relation to the zero of calibration of the scale.
During the development of the operation of zero setting the display is going to show the following message:
ZEro
If the weight is steady but out of the range of zero the display is going to blink for about 1 second more or less.
N.B. The user can’t do any operation until the setting of zero has been done
The precision of the setting of zero is inside and not out of the division: 0,25e
USER MANUAL GI400 / GI410 EN

7
3. EXECUTABLE FUNCTIONS
3.1 SETTING THE SCALE TO ZER0
When the user presses the key assumes the metrological function of weight zeroing, in the modality described below, in reference to
the current regulations.
1.Steady weight (the bright sign of steady weight is on);
2. Weight value not less than -2% of the Max. in relation to the zero of calibration of the scale;
3. Weight value not higher than 2% of the Max. in relation to the zero of calibration of the scale.
The brilliant sign indicates that the function has been done.
3.2 WEIGHT
The device switches on a led of stabilibty when a weight is placed on the platform and when the weight is steady. When the scale is unloaded
the device switches one the zero and stability leds.
3.3 TARE, SUCCESSIVE TARAS, GROSS-NET
To do a tare the user must press the key . When this key is pressed the indicator is going to show immediately the net weight, and the
Net led switches on.
If the user wants to do another tare, s/he needs to press the key again.
To know the gross weight (container plus contents) the user must press the key . On the display the led Net switches off and the sign of
gross weight B/G switches on.
If the user presses this key again the device is going to go back to the net weight and the led of Net is going to switch on again.
3.4 BLOCKING/UNBLOCKING TARE
If the user wants to block the used tare in the previous section, so that it does not disappear when going through zero, the user must press for
some seconds, at least 2, the key tare . The value of tare remains blocked. The display informs the operation showing “Tar.Blo”.
To unblock the tare the user must keep pressed 2s the key . The display is going to indicate this action by showing the message“Tar.
DES” on the screen.
To program the viewer to block the tare automatically, you have to enter the technical menu. Select “Config”, and in “Config” in “Autarb”. We
can put yes or no to the automatic tare block (“Autarb”).
3.5 MANUAL TARE
The terminal allows the incorporation of a manual tare by pressing during two seconds the key . The user must use cursor, escape and
enter keys for input new tare value. If procedure before is used, tare value will keep loocked till you unlock it.
3.6 MEMORIZED TARAS TLU
The visor has 5 memorized tares: TLU 1, TLU 2, TLU 3, TLU 4, TLU 5.
Before you can use this function you must edit and activate the desired tare from the User menu/TLU.
The tare value is locked until it is unlocked with the procedure seen in blocking / unblocking tare.
3.7 ACCUMULATION (ONLY WORKS WITH PRINT PROTOCOL ACTIVATED)
The user can not use this function without activating it first. To do it, the user must go to the menu “ModE”. To go to the menu the user must
follow the steps described in the section number 4 and watch the procedure in the section 4.6
Each time that accumulation is done, the message Acc appears on screen.
This message disappears after some seconds and it shows again the value of the last weight placed on the platform. If the user has a printer
connected to the terminal, the printer prints the weight with selected format.
Note: The accumulation can work manually or automatically
USER MANUAL GI400 / GI410 EN

8
3.7.1 SUCCESSIVE MANUAL ACCUMULATIONS
Select the accumulation and printing in manual mode.
Place the weigh on the scale, press key when stable indication turns on.
If we want to accumulate again, press the key again.
The indicator will continue memorizing if: the scale has been discharged previously or tare is done or addicional weigh has been added.
3.7.2 SUCCESSIVE AUTO ACCUMULATIONS
Select the accumulation in automatic mode and enter the number of accumulations to totalize, the printing in stable, stable with zero pass or
stable to the discharge.
Place the weigh on the scale, when it takes stability it will accumulate.
The indicator will continue memorizing if: the scale has been discharged previously or tare is done or placed on the platform on additional
weight until reaching the number of accumulations programmed or totalized.
3.8 WEIGHT TOTALIZATION
If one of accumulation modes is active, once all desired weights had been accumulated and the scale is in zero the user wants to know the
total of the weights s/he must press the key . The terminal is going to show the total number of weights and the total net weight.
•If there is a printer defined in the parameters of rs232 and connected, a ticket is going to be printed.
(For ticket copies see menu “Config” / “Print” / “Copy”)
4. OPERATION MODES (USER MENU)
DISPLAY DESCRIPTION AND SEQUENCE OF USE
• The user can access to this menu once by pressing the key
• The display is going to show the message "ModE ” on the screen and next it will appear the first available function.
MODE
• The user must use the keys described bellow to select the desired function.
- Move to the right side the edition
digit. - Move to the left side the edition digit.
- Inside a menu increases the value
of the selected number.
- It goes to the previous function.
- Scrolls to the back menu.
- Within the menu decreases the selec-
ted number
- Inside a menu, goes back
memorizing the value set.
- It selects the current function.
- If the user keeps this key pressed s/he can exit from the
function without memorizing the value set (function ESC).
- With short press, returns to zero the current value.
USER MANUAL GI400 / GI410 EN

9
THE AVAILABLE FUNCTIONS ARE:
CAL AUU CALCULATION OF THE AVERAGE UNIT WEIGHT
AUU PIECE COUNTING
CHECt CHECKWEIGHER
CCt SI CHECKWEIGHER SIMPLE
by 10 IT SHOWS THE WEIGHT WITH ONE MORE DECIMAL FOR MORE ACCURACY
ACCUMU ACCUMULATION
Subtot SUBTOTAL
t CodE SET PRODUCT CODE
n tiC SET TICKET NUMBER
WEIGHt WEIGHT MODE
doSiFi DOSAGE (only with additional board 4I/40 (relay) (OPTIONAL)
LiMit LIMITS (only with additional board 4I/4O (relay) (OPTIONAL)
2 rEL ZERO AND WEIGHT (only with additional board (relay) OPTIONAL)
dSd FISCAL MEMORY (only with additional board ALIBI) (OPTIONAL)
tLU MEMORIZED TARAS TLU
AnALog ANALOG OUTPUTS (Only with analog board OPTIONAL)
FUnCio FUNCTION MODE
4.1 CALCULATION OF THE AVERAGE UNIT WEIGHT CAL AUU
This function calculates the unit weight of the pieces through the piece counting program. To calculate the average unit weight the user must pla-
ce a number of known pieces on the platform and the press the key . The screen of the display is going to show the message SP.xxxx.
The user must then introduce, using the keys, the number of pieces that make the sample.
To calculate and register the value of the average unit weight the user must press the key .
The terminal shows- during two seconds- the message “WEiUni” on the screen, followed by the unit weight of the piece. After that, the indicator
goes back and shows the number of pieces in the scale.
If the user presses the key and there is no weight on the platform or the weight is inferior to the necessary to calculate it on the screen,
the display is going to show the message “add”. This message indicates that more pieces are needed so that the calculation can be done.
4.2 PIECE COUNTING AUU
This function is used to do the calculation of the pieces loaded on the platform. It does so by taking the programmed average weight (PMU).
The indicator asks for the PMU by showing the message Px.xxxx. The user must introduce the value of the unit weight using the cursor keys.
Press key to confirm.
NOTE: The parts count mode remains active until we access the weight mode.
4.3 CHECKWEIGHER CHECt¡
This function activates the mode “CHECK WEIGHER” or mode goes-not-goes .
The equipment is ready to manage automatically up to 4 zones through limits of weight for each one of them.
When the user accesses the menu CHECt the first parameter s/he must configures is the menu tArgEt: this menu programs the value of the
target of weight that the user wants to achieve. Is the central value of OK zone.
After that the user must choose the type of TARGET between TTYPE1 and TTYPE2.
Relisa RELISA
USER MANUAL GI400 / GI410 EN

10
4.3.1 TARGET TYPE 1 T.TYPE 1
If the user chooses the T. TYPE 1 it allows to choose the inferior limits and the superior ones with a porcentage of the TARGET value.
Exemple:
-PC. 010 (We chose how lower limit 10% below the Target)
PC. 010 (We chose how upper limit 10% above the Target)
LOW
LED: yellow LED
LCD: - Yellow backlit
OK
LED: green LED
LCD: OK green backlit
HIGHT
LED: red LED
LCD: + red backlit
TARGET 1000 g
HOW LOWER 900 g UPPER LIMIT 1100 G
HOW LOWER = target -10% of Target=
1000 - [10%(1000)]=900
UPPER LIMIT = target +10% of Target=
1000 + [10%(1000)]=1100
4.3.2 TARGET TYPE 2 T.TYPE 2
If the user chooses the T.TYPE 2, it allows to choose the inferior and superior limits directly: L (lower) XXXXX y H (Hight) XXXXX
Exemple:
L 800
H 1200
LOW OK HIGHT
TARGET 1000 g
HOW LOWER 800 g HIGH LIMIT 1200 g
Once they are configured, the TARGET value and the LIMITS, the program ask the user to configure the limit of the ALARM.
The alarm programs the value wich any value above this one is going to activate the alarm sign, and 3 lights are going to bright.
Finally the user must choose the alarm sound. He can choose between bEEPt or bEEPM.
USER MANUAL GI400 / GI410 EN

11
4.3.3 TYPE BIP bEEPt
First we choose the area in which we want the beep to sound, this will occur when entering the selected zone:
no tAr: When it’s not in the target range.
LoU: Below the target range.
tGtrnG : Target range.
HiGH : Above the target range.
4.3.4 BIP MODE bEEPM
Then choose the kind of beep: B off/long/short
4.3.5 YES/NO ACTIVE
Finally asks us if we want or not to activate the function checkweigher, in the case of not activating it is memorized the configuration that
we have chosen previously.
Note: The Checkweigher mode remains active until we access the weight mode
4.4 CHECKWEIGHER SIMPLE CCt SI
This function activates the “CHECK-WEIGHER SIMPLE” mode or simple pass-not-pass mode. In the case of having indicators with numeric
keypad, we can also activate it through F + 9 + 9
The device is prepared to automatically manage up to 3 zones by two weight limits.
When accessing the CCt SI menu the first parameter to be configured is the lower limit. The screen briefly shows InF: and remains showing
L00000. With the cursor keys or the membrane keyboard, we enter the value of the lower limit and validate with
The screen will briefly shows SuP: and remains showing H00000. With the cursor keys or the membrane keypad, we enter the value of the upper
limit and validate with The screen briefly shows CWS.on and returns to weighing mode with the simple checkweigher function activated
4.5 IT SHOWS THE WEIGHT WITH ONE MORE DECIMAL MORE FOR ACCURACY bY 10
With the selection of this function the indicator displays one more decimal in the weight.
Once this function is authorized there is an activation of the display. In this situation the less significant number of the display gets activated
and it indicates the value per 1/10 of the division of the verification. The decimal moves but the display does not increase a digit.
With the device connected, the display of the data of weight does not respect the metrological indication. It is for this reason that while it
is activated all the communication with the series channel is inhibited. The display remains active for a period of approximately 6 seconds,
after this time the terminal, automatically, sets the display in normal weight. Being out of legal metrology, the device keeps extra accuracy
till user press key.
NOTE:
The Checkweigher Simple - mode remains active until set the
weight mode again. In case of having indicators with numeric
keyboard, it can also be deactivated by pressing F+9+8
USER MANUAL GI400 / GI410 EN

12
4.6 ACCUMULATION ACCUMU
This function carries out the process of auto accumulation and ticket printing
TO MAKE THE AUTO-ACCUMULATION FUNCTION WORK, IT IS NECESSARY TO SELECT FIRST THE PROTOCOL PRINT IN COM1
OR COM2, AS IT IS DESCRIBED IN THE SECTION 5 “PARAME” (POINT 6.1)
DISPLAY STEPS
ACUMUL
•Press the key
the display is going to show one of the legends detailed below.
ACUMUL
ACCoFF ACCMAn ACCAUt
The user must choose them with the keys and , the selected option is memorized
by pressing the key .
DESCRIPTION: ACCOFF, ACCAUT AND MANACC, AND CONFIGURATION
ACC.oFF ACUMULAR OFF
The accumulation mode is deactivated
ACC.AUt AUTO ACCUMULATION
The accumulation mode is automatic. In the moment of detecting steady weight the device does
the accumulation automatically. Once this function is validated the terminal shows on its display
the message (nA xxxx). The user must introduce the desired value of the number of accumula-
tions using the following keys:
If this value is DIFFERENT FROM 0 the system continues with the accumulation of weights until
it reaches the prearranged value. If the programmed value is 0, the weights are accumulated until
the user decides to press the key (limit 9999). (Totalitzation)
ACC.MAn ACCUMULATION MANUALLY
The mode of accumulation is done manually by pressing the key .The weight must be
stable and must change from accumulation before. It will accumulate till the user press .
4.7 SUBTOTAL SUBTOT
With this function the user can see the number of accumulated weights and the total weight up to the actual moment. The screen displays first
the number of weights and next it automatically shows the total accumulated weight. This function does not entail the elimination of the number
of weights and the total of weight.
4.8 TOTALITZATION
By pressing key device shows totalitzation value. If one printer mode is selected, the total ticket will be printed. By pressing again
key will print a copy. Making a new accumulation after the memory will be erased.
4.9 CODE TCODE
This function allows the introduction of a code which is added to the printed ticket. To introduce this code the user must use the keys speci-
fied below.
THIS OPTION IS ONLY VALID USING PROTOCOL PRINT.
USER MANUAL GI400 / GI410 EN

13
4.10 NUMBER OF TICKET NTIC
With this function the user can print in the ticket a ticket number which is going to increase automatically at the end of the weight, (number of
delivery or number of ticket). Use the cursor keys for change the value and enter escape to exit or memorize..
4.11 WEIGHT MODE WHEiGt
The user access the normal mode of weighing again.
4.12 DOSAGE dOSIFI
PRECISE PLATE ADDED 4I/4O(relay)
With 4 inputs and 4 outputs. The dosing function is only available when the plate is correctly connected to the display.
The card has four relays with potential-free output up to 0.5A 125VAC or 2A 30VDC. With them you can check remote instruments such as relays, signal
lamps and PLC. The card also has four inputs. The 4 inputs are emulated to the keys of the equipment when they are in dosage mode. The dosing function
remains active until it is switched off.
The equipment is ready to dose automatically with two different possibilities: 1 product at two speeds or 2 products at one speed, with unload. The device
allows the pogramming, for each porduct, of the weight of the value of the thickness or product 1, and the value of the weight when the device works at
slow speed (that is the value of the smooth) or the value of the product 2.The device can also program a zone of unload where it is defined the value of the
dosage( to correct the blocking of the system because there is product left in the weighing zone).
When the equipment starts working it automatically functions with the last formula it has worked with. This allows the device to start again with its daily work.
When the user gets into the menu to program the dosage “doSiFi”, the following menu is going to appear on the screen:
doSiFi
reset YES / NO yes-no (It sets the process to zero, it also places all the weights to ZERO)
Prod1 W WEIGHT thick weight to program or weight of the product 1
Prod1 t TIME time of lack of material, if the weight on the screen does not change in this space of time,
it activates the relais of alarm.
Prod 2W WEIGHT smooth weight to program or weight of the product 2
Prod 2t TIME Time of lack of material
2 discw WEIGHT Maximum waste weight
TIME Time of zero, when the zero is activated a time to wait for the falling of the material
is also activated
2 disct
tArEC AUTO/NO Make, or not, auto tare when dosage starts
EnAbLE YES / NO
RESET
By selecting YES, all parameters are set zero.
By selecting NO, you can access directly to the parameters to modify them.
PROD1
It edits the values of the formula for the product 1. If the weight is programmed to 0, the equipment is going to dosify the product 2.
•W (Weight): The desired weight for this product is programmed.
•T (Timer): Time of lack of material, if the weight on the screen does not vary in this space of time, it is going to activate the alarm (RL4)
PROD2
It edits the values of the formula for the product 2. If the weight is programmed at 0, the equipment will dosify only the product 1.
•W (Weight): The desired weight for this product this product added to product 1 weight is programmed.
•T (Timer): Time of lack of material, if the weight on the screen does not vary in this space of time, it is going to activate the relais of
alarm. (RL4)
USER MANUAL GI400 / GI410 EN

14
Z.DISC
•W (weight): This option programs the weight within which the equipment determines that the dosage has been finished, (maximum of
possible waste weight that may stay without the alarm ringing)
•T (timer): Time waste of material, if the weight on the screen does not vary in this space of time, it is going to activate the relais of
alarm (RL4).
ENABLE
It exits from the menu of dosage and asks to the user if he wants to get start doing the dosage. To start doing the dosage appears on the
display the legend - dos ON - The keys will do the functions specified below:
Note:
The value of timers must be bigger than 000,0 s. for assure the capacity of activate the alarm (RL4). If you want that
RL4 (alarm) will never be activated, keep timers to 0s.
ENTRANCE EXT. KEY INDICATOR FUNCTION
INPUT 1 START key
INPUT 2 STOP Key
INPUT 3 RESTART key
INPUT 4
ABORT key / exits from the DOSI function and goes back to
the normal mode. You need to press two times if you use the
indicator key.
Note: In the event that the alarm relay (RL4) is activated, press the key and then restart key to continue the
process where it was interrupted.
To dispense a single product, it is necessary to leave the Prod 1W to 0 and set the weight to be dosed in the Prod
2W
4.13 LIMITS LIMIT
THEY ONLY WORKS WITH THE OPTIONAL RELAYS BOARD
The equipment is able and thought to manage automatically up to 4 limits through registers of weight for each one of them.
These registers are called
•Start
•Low Limit (Low.L.)
•High Limit (High. L.)
•Alarm
The associated relays is always activated when the weight of the scale is placed between the two limits.
0
Start
i
Low limit
nf (inferior)
High Limit
Alarm
OL
relés OFF RL 1 RL 2 RL 3 RL 4 RL 4
OK
As each relay is in fact a commute circuit, taking the normaly closed contacts, the user is going to obtain the inverse sign of the previous figu-
re. When we go to the function of the programming of limits (LiMitE), the following menu is going to appear on the screen:
USER MANUAL GI400 / GI410 EN

15
DISPLAY
RESET SI Goes to next parameter without erase the information.
It places at 0 all the relays so that it deactivates all of them.
NO
START It programs the value, every value above this one is going to activate the relays of the L.INF, RELAYS 1 (S
xx.xxx)
L.LoU It programs minimum value, every value below this one is going to activate the relays of the L:INF, RELAYS 1
(L xx.xxx)
OK This value can not be programmed but it activates the RELAYS 2 whenever the value of the weight is superior
to Low L and inferior to the High L
HiGH It programs the high value, every value above this one is going to activate the relays of the L.SUP, RELAYS 3
(H xx.xxx)
ALARM it programs the high value, every value above this one is going to activate the relays of alarm, RELAYS 4 (A
xx.xxx)
ENABLE Exit from the menu limits and start to apply them if YES is choosed. If we choose NO exit from the menu and
doesn’t apply the limits stored on memory.
NOTES: THE FUNCTION OF THE LIMITS IS AUTOMATIC WHEN IT EXITS FROM THE ZERO RANGE. THE LIMIT
FUNCITON CONTINUES ACTIVATE TILL USER DESACTIVATES IT.
4.14 ACTIVITY CONTROL 2 rEL
REQUIRES ADDITIONAL PLATE 4E/4S(relés)
It is a variant of the limit function, but involving only two relays, deactivating the first when exiting the value “Starts" and the second disen-
gaging once we reach the value "target"
2 rEL rESEt Select Yes/No to restore values to 0..
Start (00,000) Deactivation Value 1st Relay
target (00,000) Deactivation Value 2nd Relay
EnAbLE Select Yes/No to activate the function
4.15 DATA VISUALISATION MEMORIA dSd
•Accessing to DSD menu (only available if DSD memory is installed) the indicator give access to DSD file to see the weights stored with
fiscal memory process.
•The indicator will show the last code inuse, with cursor keys you can input the desired value.
•On last line it shows the current weight.
•Then, operator must input the weighing number about he/she want fet data of brut and tare.
•If data is not stored, it will show an error message, if all is correct, it will show weighing data.
For exit from data menu, keep pressed key some seconds.
During data viewing “Tx” function from PC will not work.
USER MANUAL GI400 / GI410 EN

16
4.16 MEMORIZED TARAS TLU tLU
•The visor has 5 memorized tares: TLU 1, TLU 2, TLU 3, TLU 4, TLU 5.
•Before you can use this function you must edit and activate the desired tare from the user menu/TLU
•The tare value is locked until it is unlocked with the procedure seen in lock/unlock Tare.
SELECT: To select the tare that we want to activate, this has to be previously created in the editing section
EDIT: To create the tares up to a maximum of 5.
DELETE: To erase all memorized tares.
4.17 ANALOG OUTPUT AnALoG
ACCESS TO THE MENU IS POSSIBLE ONLY WHEN THE CARD IS CONNECTED
This card allows the exit of 4-20mA o 0-20mA or 0...+10V o 2…+10V, proportional to the gross weight or the net weight in the scale.
The exit of the current is active and supplies the necessary voltage.
The analog exit is actualized every 50ms and acquires the appropriate value of the weight, converted into in that moment. It is for this
reason that if the filter in the weight is decelerated, the analogical exit is also decelerated.
To do the setting of the parameters, the user must access to:
AnALog
ActivA Select Yes/No to turn functionality on or off
4-20mA Requesting full scale (P. MAX) in kg that will deliver 10V.
0-10V REQUESTING FULL SCALE (U MAX) IN KG THAT WILL DELIVER 10V.
ADJUST
An Min Output for negative weight
An.Zer Output for zero weight
An MAX Departure for the fund allocated scale (U MAX)
We have to activate the AnALog function before we start setting
The setup process begins by selecting the operating mode, either 4-20mA or 0-10V depending on the type of output you want. At this point
you must specify the full scale U MAX which must be equal to or less than the full scale of the scale.
When the back of the scale is confirmed, the device starts the establishment of values of the analog exit, or the values of the digital/analog
converter are introduced (comprised between 0 and 65535), for these values it is appropriated a determined value of exit in voltage or
current.
In this configuration the keys of the device acquire the following functions:
An Min ( N. xxxxx)
It establishes the minimum value of the analog exit. With this value we adjust so that the equipment gives a value below 4 mA or 0v. It
also allows us to program the magnitude of the signal when the value loaded on the platform is below zero, belonging to the condition
underload.
An.2 Er( z. xxxxx)
It establishes the magnitude of the analog exit, be that in current or voltage, when the scale displays the minimum weight, which is
going to belong to the Zero of the scale.
An MAX ( M. xxxxx)
It establishes the magnitude of the analog exit, be that in current or voltage, when the scale displays the maximum weight, which is
going to belong to the previous FS programmed.
Press to access the menú ACTIVA.
NOTE:
When you have an indicator with Numeric-Keyboard, you can
use the Fast-access to U-MAX with the Keys F+8+0
USER MANUAL GI400 / GI410 EN

17
EXAMPLE
We are going to take as example a hopper scale with a back of scale of 300kg. Of this scale we only use the range up to 200kg for the exit
of 20mA as this is the maximum capacity of the hopper.
In F.S we will introduce 200kg and we will adjust the value of An_Max so that in the exit we can register 20mA with a multimeter or an
automaton. Next we are going to go to the parameter An_Zer and adjust the value to obtain 4mA in the exit; with this we are going to have a
perfect measuring range of 200kg and 16mA. In the parameter An_Min we are going to introduce a value so that we can read the negatives
to up to the 3mA. In the case that the automaton does not allow it, we will adjust the zero to 5mA. The measuring range is going to be of
200kg and 15mA.
4.18 FUNCTION MENU FUnCio
Access the FUnCio menu tree, in the next point (5. SPECIAL FUNCTIONS OF THE NUMERIC KEYPAD) it is explained, how it works from
the operation of the displays with numeric keyboard, for which FUnCio is designed.
In the diagram in point 5, the left part corresponds to the options structure that we access with this menu.
4.19 RELISA rELiSA
This Menu is related to the Protocol rELiSA.
Here you can input the fixed variable of Protocol rELiSA.
5. SPECIAL FUNCTIONS OF THE NUMERICAL KEYBOARD
(FOR MODELS WITH KEYBOARD)
You can enter the following menu by pressing and holding the F key. You can also access each function directly using the direct keys.
The numeric keypad allows us to enter the values of the different menus of the viewfinder much faster than with the keys with arrows, which
still continue working.
Using the numeric keypad it is possible to memorize a maximum of: 400 items of type of weighing (File1), 400 descriptive items (File2),
10 tares and 1000 weight.
Pr. File 1 Filxxx (Select File 1) DIRECT KEYS FUNCTION
CODE XXXXXX
+ +
Edit item1 menu
DESCRI 24 alphanumeric characters
+ + Activates the item 1
TARA TXXXXX
TYPE
WEIGHT
+ + Deactivates the item 1
CHECKW TARGET TXXXXX
LOW LXXXXX
HIGHT HXXXXX
ALARM AXXXXX
AUU PXXXXX
EANCODE 12 alphanumeric characters (NO FUNCTION)
USER MANUAL GI400 / GI410 EN

18
Pr. File 2 Filxxx (Select File 2) DIRECT KEYS FUNCTION
CODEE XXXXXX
+ + Edit item 2 menu
DESCRI 24 alphanumeric characters
+ + Activates the item 2
+ + Deactivates the item 2
PrTare Select tare Number or edit DIRECT KEYS FUNCTION
TARAXX
+ +
Edit pretare menu
TXXXXX
+ + Activates tare
+ + Deactivates tare
WEIGHT DIRECT KEYBOARDS UNCTION
VISUAL inUnun 000000 (internal weighing number)
+ +
Displays the
information of the
weights
IT1num
IT2num
W Net
+ +
Sets the type of warning
when filling out the 80%
of the weighing buffer.
When reaching the 100%
the indicator displays the
error message OveBuf
W Brut
W Tare
N Piece
DATE
CONFIG (BFULLX) 0 No warning
1 Bip
2 Bip+ BUFFULL Message
USER MANUAL GI400 / GI410 EN

19
TOTAL DIRECT KEYS FUNCTION
FILE 1 Filxxx
+ +
It shows the total of the
weighing associated to
the item 1
Net
Gross
Tare
+ +
It shows the totals of
the weighing
associated to the item2
Pieces
TotWeigh
FILE 2 Filxxx
Net
Gross
Tare
Pieces
TotWeigh
LIST DIRECT KEYS FUNCTION
TOTAL E SinBor
+ +
Prints the weight total
with or without deleting
the memory
ConBor
FILTRO IT1XXX
+ +
Prints the lists of the
filtered weighing, per
item 1, item 2 and date
IT2XXX
DATENO/DATESI INICIO
FIN
+ + Prints the filtered
information per item 2
FILE 2 Filxxx
+ + Prints the filtered
information per item 1
FILE 1 Filxxx
DELATE To access the erasing functions it is necessary to enter a
password, by default 0000
DIRECT KEYS FUNCTION
FILE1 FilXXX ALLIT1
+ +
Deletes item by item the totals of
the item1 selected (TOTIT1) or
the item1 completely (ALLIT1)
TOTIT1
FILE2 FilXXX ALLIT2
+ +
Deletes item by item the totals of
the item2 selected (TOTIT2) or
the item2 completely (ALLIT2)
TOTIT2
+ + Deletes a tare scheduled before
TARAPR TARAXX
+ + Deletes the stored table of
memorized weighing
WEIGHT
+ +
General deleting, it deletes all
the information in item1, item2,
tares and table of weighing.
GENERATES NO
YES
CAPASS Change the password
SELECT Allows us to activate the memorized items or tares
USER MANUAL GI400 / GI410 EN

20
CHANGE THE NAME OF THE ITEM1 AND THE ITEM 2
Enter the technical no metrological menu, inside the menu CONFIG, select the option CODES.
Choose the items that you need to change the name from FILE1, FILE2.
Write the new name of the item in the numerical keyboard.
EXAMPLE
In FILE1 we write “lorry” and in FILE2 we write “product”, we press the ENTER key during each operation to confirm.
EDIT THE DATA OF THE ITEM OF THE TYPE OF WEIGHING (ITEM1)
From normal weighing mode press F+1+0 (one after the other). Write the number of item 1 we want to edit
Now we introduce the desired data:
DISPLAY DESCRIPCIÓN
CODE numerical code of the item
DESCRI description of the item with a maximum of 24 alphanumerical characters
TARE value of the tare item
TYPE to select the type of item WEIGHT, CHECKW or PMU, in the case of Checkweigh and PMU the user must
set its parameters
EANCODE code EAN of the product
Once each one of the parameters are confirmed with the key , the user must press the key , the screen will display briefly the
message SAVED and next it will return to the weighing mode. .
EDIT DATA OF THE DESCRIPTIVE ITEM (ITEM2)
Once the user is in the normal weighing mode, press the keys + + ( one after the other).
The screen is going to display the message PRO (the first three digits of the “product”), followed by 000.
Write now the number of the item2 that the user wants to edit, for example PR000 1 and press the key ENTER, the screen is going to display
the menu described beFore.
Introduce now the desired data:
DISPLAY DESCRIPTION
CODE numerical code of the item
DESCRI description of the item with a maximum of 24 alphanumerical characters.
Once each one of the parameters have been confirmed with the key the user must press the key , the screen will briefly display
the message SAVED and next I will return to its weighing mode.
EDIT TARA DATA
Once the user is in the normal weighing mode of the device, press the keys + + ( one after the other).
The screen is going to show the message TARE00. Write the number of tare that the user wants to edit, for exampleTARE01 , and then press
the key the screen is going to display the messageT00.000,, so that the user can introduce the value of tare.
Once it is set, the user must press the key to confirm the operation, the screen will briefly display the message SAVED and it is also
going to show again the message TARE01 so that the user can choose another tare to edit. Once the user has edited all the tares needed, the
user must press the key during 3 seconds so that the device goes back to its normal weighing mode.
NOTE:
When you have an indicator with Numeric-Keyboard,
you can use the Fast-access F+8+1
to input Item1 and Item2 without store them before
USER MANUAL GI400 / GI410 EN