The following is a summary of display messages. If your meter displays an
error message, please follow the actions for the error message as described
in the following table. If the problem persists, please contact GlucoRx
Customer care on 01483 755133.
DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
NOTE
●
If you wish to exit memory mode before the last result being displayed, press
the M button for 3 seconds.
●
When the memory is full, the oldest test result will be replaced by the newest
test result.
●
The control solution results are NOT stored in the memory. Only blood glucose
results will be stored.
Viewing Results
Your Meter stores the 999 most recent blood glucose test results in its
memory. You can review the test results with these easy steps.
STEP 1. Enter the memory mode
With the meter turned off, press Mtwice. The first
test result will appear, indicating you are in
memory mode.
STEP 2.
When using the meter for the first time “---”will
appear, indicating that there are no test results in
the memory.
STEP 3. Recall test results
If you continue to press M, you can then review
the last 999 tests in the memory.
STEP 4. Exit the memory mode
After displaying the last test result in memory,
press the Magain. The meter displays "End" and
then turns off.
Strip Vial Embedding and Replacement
1. When you use your meter first time, the meter
is designed to embed into the strip vial cap at
an upward angle. If the meter is embedded
well, you will hear a“click” afterwards.
2. Open the strip vial cap to take out a new strip
for blood testing.
3. After all the strips have been used up, detach
the meter from the strip vial a downward
angle; dispose the empty strip vial. Replace
with a new strip vial.
MESSAGE WHAT IT MEANS
< 1.1 mmol/L
> 33.3 mmol/L
SYMBOL INFORMATION
Symbol Referent
In vitro diagnostic medical device
Consult instructions for use
Temperature limitation
Use by
Batch code
Manufacturer
Serial number
Authorised representative in the European Community
Caution, consult accompanying documents
CE mark
Humidity limitation
Collection for electrical and electronic equipment
MESSAGE WHAT IT MEANS WHAT TO DO
Appears when the
battery is too low.
Replace the batteries
immediately.
Repeat with a new test strip.
System operation range is
10°C to 40°C (50°F to 104°F).
Repeat the test after the
meter and test strip are in the
above temperature range.
Repeat the test with a new
test strip. If the meter still
does not work, please
contact GlucoRx Customer
care on 01483 755133
for assistance.
Appears when a used
test strip is inserted.
Appears when ambient
temperature is above or
below system operation
range.
Problems with the meter.
Appears when test strip is
removed while counting
volume.
Review the instructions and
repeat test with a new strip.
If the problem persists,
please contact GlucoRx
Customer care
on 01483 755133 for help.
Model No.: TD-4283
Dimension & Weight: 43.5 (L) x 39.3 (W) x 24.2 (H) mm
Weight: 18 g (including battery)
Power Source: one 3V CR2032 lithium battery (for at least 500 measurements)
Display: LCD
Memory: 999 memory sets
Automatic detection of electrode insertion
Auto sample loading detection
Automatic reaction time count-down
Auto switch-o after 3 minutes without action
Temperature warning
Operating Condition:
10°C to 40°C (50°F to 104°F), below 85% R.H.
Meter Storage/ Transportation Conditions:
-20°C to 60°C (-4°F to 140°F), below 95% R.H.
Strip Storage/ Transportation conditions:
2°C to 32°C (35.6°F to 89.6°F), below 85% R.H.
Measurement Units: xed mmol/L
Measurement Range: 1.1–33.3 mmol/L
Expected service life: 5 years
This device has been tested to meet the electrical and safety requirements of:
IEC/EN 61010-1, IEC/EN 61010-2-101, EN 61326-1, IEC/EN 61326-2-6.
SPECIFICATIONS
TaiDoc Technology Corporation
B1-7F., No. 127, Wugong 2nd Rd.,
Distributed by GlucoRx Ltd.
4 Wintonlea, Monument Way West, Woking, Surrey, GU21 5EN, UK
Website: www.glucorx.co.uk
Customer careline: 01483 755133
Wugu Dist., New Taipei City 248, Taiwan (R.O.C.)
www.taidoc.com
MedNet GmbH
Borkstraβe 10, 48163 Mϋnster, Germany For self-testing
NOTE
●
Replacing the battery does not aect the test results stored in the memory.
●
As with all small batteries, these batteries should be kept away from children. If
swallowed, promptly seek medical assistance.
●
Battery might leak chemicals if unused for a long time. Remove the battery if
you are not going to use the device for an extended period ( i.e. 3 months or
more).
●
Properly dispose of the battery according to your local environmental
regulations.
Your meter comes with one 3V CR2032 lithium battery. If the low battery
symbol ” ” appears on the screen, this indicates that the battery is low
and it is time to change the battery.
E-b error appears:
The power is not enough to do a test. Please change the battery
immediately.
Caring for Your Meter
Cleaning
● To clean your meter exterior, wipe it with a disinfecting wipe
(Example:Micro-Kill Plus™) to clean exposed surfaces thoroughly and
remove any visible dirt, blood or any other body fluid with the wipe. Then
dry the device with a soft, dry and clean cloth. Do NOT rinse your meter with
water.
● Do NOT use organic solvents to clean your meter.
Meter Storage
● Storage conditions: -20°C to 60°C (-4°F to 140°F), below 95% relative
humidity.
● Always store or transport your meter in its original storage case.
● Avoid dropping and heavy impact.
● Avoid direct sunlight and high humidity.
Meter Disposal
Your used meter should be treated as contaminated that may carry a risk of
infection during measurement. The batteries in your used meter should be
removed and the meter disposed in accordance with local environmental
regulations.
Your meter falls outside the scope of the European Directive
2002/96/EC-Directive on waste electrical and electronic equipment (WEEE).
Replacing the Battery
To replace the battery, make sure that your meter is turned o.
1. Open the battery cover in an anti-clockwise direction and lift it up to remove.
2. Remove the old battery and replace with one new 3V CR2032 lithium battery.
3. Close the battery cover. If the battery is inserted correctly, you will hear a
"beep" afterwards.
BATTERY
MAINTENANCE
We strongly recommend that you perform AST ONLY at the following times:
●
In a pre-meal or fasting state (more than 2 hours since the last meal).
●
Two hours or more after taking insulin.
●
Two hours or more after exercise.
Do NOT use AST if:
●
You think your blood glucose is low.
●
You are unaware of hypoglycaemia.
●
You are testing for hyperglycaemia.
●
Your AST results do not match the way you feel.
●
Your routine glucose results often uctuate.
PLEASE
AFFIX
STAMP
GlucoRx Ltd
4 Wintonlea
Monument Way West
Woking
Surrey
GU21 5EN
United Kingdom
TERMS & CONDITIONS OF WARRANTY
1. GlucoRx warrants this product to be free of defects in workmanship and materials within the said warranty period
on the Warranty Certificate.
2. During the warranty period if this product is found to be defective, you may send it with the Warranty Certificate to
our office for warranty service. GlucoRx will then repair or replace defective parts, or exchange the whole product as
GlucoRx may choose, with no additional charge to the original owner. After such repair, replacement or exchange,
the product will be warranted for the remaining time of the warranty period.
3. This warranty is valid only if the Warranty Certificate and Warranty Registration Card are duly completed with date
of obtaining product, serial number and if the Warranty Registration Card is sent to our office (or you register on
our website www.glucorx.co.uk) no later than 6 months from the date obtained.
4. This warranty is void if this product has been repaired or serviced by an unauthorised person. This warranty does
not cover defects caused by misuse, abuse, accident, tampering, lack of reasonable care, fire or any other acts
beyond human control.
5. Except as stated in the above paragraphs, GlucoRx disclaim all other warranties, including those of merchantability
of fitness for a particular purpose with respect to the use of this product. GlucoRx shall not be liable for any direct,
consequential or incidental damages arising out of the use or inability to use this product.