manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Gossen MetraWatt
  6. •
  7. Temperature Controllers
  8. •
  9. Gossen MetraWatt METRATHERM IR BASE User manual

Gossen MetraWatt METRATHERM IR BASE User manual

Operating Instructions
METRATHERM IR BASE
Temperature Sensor 3-447-091-03
2/12.20
1
1 Safety Precautions
Observe this documentation, in particular all included safety infor-
mation, in order to protect yourself and others from injury, and to
prevent damage to the instrument.
• Carefully and completely read and adhere to these oper-
ating instructions.
The document can be found at http://www.gossenme-
trawatt.com. Retain this document for future reference.
• Carefully and completely read and adhere to the product
documentation for the utilized measuring instrument.
• The instrument is an accessory and may only be used
with instruments that specify it as accessory.
• The instrument may only be used for the tests/measure-
ments described in the associated instrument‘s docu-
mentation.
• Tests/measurements may only be performed by a quali-
fied electrician, or under the supervision and direction of
a qualified electrician. The user must be instructed by a
qualified electrician concerning performance and evalua-
tion of the tests/measurements.
• Wear suitable and appropriate personal protective
equipment (PPE) whenever working with the instrument.
• The instrument may only be used as long as it’s in good
working order.
Inspect the instrument before use. Pay particular atten-
tion to damage, broken insulation or kinked cables.
• If the instrument doesn’t function flawlessly, remove it
from operation and secure it against inadvertent use.
• If the instrument is damaged during use, e.g. through
falling, remove it from operation and secure it against
inadvertent use.
• Do not use the instrument after long periods of storage
under unfavorable conditions (e.g. humidity, dust or
extreme temperature).
• Do not use the instrument after extraordinary stressing
due to transport.
• The instrument must not be exposed to direct sunlight.
• Only use the instrument in compliance with the speci-
fied technical data and under the specified conditions
(ambient conditions, IP protection class, measuring cat-
egory, etc.).
• Do not use the instrument if the battery compartment
cover has been removed.
Otherwise you risk touching dangerous voltages.
• The instrument is a class 2 laser product that emits laser
radiation which can cause of eye or optic damage.
Avoid direct eye exposure. Avoid eye exposure through
reflection off surfaces.
2 Applications
Please read this important information!
2.1 Intended Use / Use for Intended Purpose
The METRATHERM IR BASE is a non contact temperature mea-
surement accessory instrument. It may only be used with instru-
ments that specify it as accessory.
The temperature is measured by pointing the METRATHERM IR
BASE at the surface to be measured, and reading the temperature
on the display of the instrument it is connected with. It automati-
cally compensates for temperature deviations from ambient tem-
perature.
Safety of the operator, as well as that of the instrument, is only
assured when it’s used for its intended purpose.
2.2 Use for Other than Intended Purpose
Using the instrument for any purposes other than those described
in these instrument operating instructions is contrary to use for
intended purpose.
2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property dam-
age, personal injury or consequential damage resulting from
improper or incorrect use of the product, in particular due to failure
to observe the product documentation. Furthermore, all guarantee
claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH assume any liability for data
loss.
2.4 Opening the Instrument / Repairs
The instrument may only be opened by authorized, trained per-
sonnel in order to ensure flawless, safe operation and to assure
that the guarantee isn’t rendered null and void. Even original
replacement parts may only be installed by authorized, trained
personnel.
Unauthorized modifications to the instrument are prohibited.
If it can be ascertained that the instrument has been opened by
unauthorized personnel, no guarantee claims can be honored by
the manufacturer with regard to personal safety, measuring accu-
racy, compliance with applicable safety measures or any conse-
quential damages.
3 The Instrument
3.1 Scope of Delivery
Please check for completeness.
1 METRATHERM IR BASE temperature sensor
(Z680A)
1 9 V battery
1 Operating instructions (this document)
© Gossen Metrawatt GmbH
Edited in Germany • Subject to change, errors excepted • PDF version available on the Internet
All trademarks, registered trademarks, logos, product names, and company names are the property of their respective owners.
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone +49 911 8602-111
Fax +49 911 8602-777
E-mail [email protected]
www.gossenmetrawatt.com
Gossen Metrawatt GmbH METRATHERM IR BASE
3.2 Symbols on the Instrument
4 Technical Data
Power
Ambient Conditions
Electrical Safety
Electromagnetic Compatibility (EMC)
Mechanical Design
Measurements
5 Preparation
The METRATHERM IR BASE is powered by a 9 V battery. One
battery is included in delivery; replacements must be provided.
You have to install the battery before first use and replace it later if
depleted. If the battery runs low, the low battery LED on the instru-
ments lights up.
The measurement unit can be °C or °F. The unit is set by an inter-
nal switch located in the battery compartment. If you want to
change the unit later on, also follow the steps below but without
replacing the battery
Required tools: Philips screwdriver.
The instrument is switched off.
1. Place the instrument face down on a stable surface.
2. Loosen and remove the screw from the battery compartment
cover.
3. Unlock and remove the battery compartment cover.
4. Set the internal °C/°F switch to the desired unit.
5. Connect the 9 V battery to the battery snap.
6. Insert the 9 V battery into the battery compartment, making
sure that the wires are not damaged (e.g. squished).
7. Place the battery compartment cover onto the battery com-
partment and press it back into place.
8. Secure the battery compartment cover with the screw.
The instrument is provided with power and ready for operation.
6 Operation
The METRATHERM IR BASE is an accessory. It‘s operation is
described in the documentation of the instruments it may be used
with.
General Information for Using the Instrument
When using the instrument, make sure that the target is larger than
the instrument's spot size. The smaller the target, the closer you
should be to it. When accuracy is critical, make sure the target is
at least twice as large as the spot size. Refer to the chart printed
on the back of the instrument for more information.
The instrument automatically compensates for temperature devia-
tions from ambient temperature. However, the IR sensor needs
time (up to 30 minutes) to adjust to temperature changes.
7 CE Declaration
The instrument fulfills all requirements of applicable EU directives
and national regulations. We confirm this with the CE mark. The
CE Declaration of Conformity is available upon request.
Warning concerning a point of danger
(attention, observe documentation!)
Warning! Laser radiation
European conformity marking
The instrument and the battery may not be disposed of
with household trash.
Current and further information is available on our web-
site at http://www.gossenmetrawatt.com under the
search terms “WEEE” and “environmental protection”.
Power supply: 9 V battery, type NEDA 1604 / IEC 6F22
Operating temperature: 0 … +50 °C
Storage temperature: −20 … +60 °C
Relative humidity: max. 80 %, no condensation allowed
Elevation: max. 2000 m
Measuring category: Cat O
Pollution degree: 2
Protection class: 0 per IEC 61140
Interference emission: EN 61326-1 class A
Interference immunity: EN 61326-1
EN 61326-2-1
Protection: Housing: IP40
per DIN VDE 0470-1/EN 60529
Housing : 164 × 50 × 40 cm (W × H × D)
Weight: 182 g
Temperature range –30 … +550 °C
Output 1 °C or 1 °F (per 1 mV)
Accuracy
±2 % of reading or ± 2 °C (4 °F), whichever is
greater.
Accuracy is specified for ambient tempera-
ture range of +18 … +28 °C.
Emissivity 0.95 fixed value
!
Field of View
Distance/spot size = approx. 8:1 ratio
Object under test must be at least 2.54 cm.
(Calculated at 90 % energy on the focal
point.)
Laser power Less than 1 mW
wavelength 630… 670 nm
Spectral response 6 … 14 μm (wavelength)

This manual suits for next models

1

Other Gossen MetraWatt Temperature Controllers manuals

Gossen MetraWatt R2700 User manual

Gossen MetraWatt

Gossen MetraWatt R2700 User manual

Gossen MetraWatt R2500 User manual

Gossen MetraWatt

Gossen MetraWatt R2500 User manual

Popular Temperature Controllers manuals by other brands

75F Dynamic Airflow Balancing quick start guide

75F

75F Dynamic Airflow Balancing quick start guide

Icon Electronics D3-TCT1 operating instructions

Icon Electronics

Icon Electronics D3-TCT1 operating instructions

RCS UTR 9442 operating manual

RCS

RCS UTR 9442 operating manual

Airstream TC5-2V4SA user manual

Airstream

Airstream TC5-2V4SA user manual

Elko iNels RFTI-10B quick start guide

Elko

Elko iNels RFTI-10B quick start guide

AKO AKO-16523-V manual

AKO

AKO AKO-16523-V manual

PMA KS800 operating instructions

PMA

PMA KS800 operating instructions

Honeywell T991A installation guide

Honeywell

Honeywell T991A installation guide

Lennox CL50-RoofTop Installation operating & maintenance manual

Lennox

Lennox CL50-RoofTop Installation operating & maintenance manual

Hitachi ATW-ITS-01 installation manual

Hitachi

Hitachi ATW-ITS-01 installation manual

oventrop Unibox T RTL operating instructions

oventrop

oventrop Unibox T RTL operating instructions

Siemens RDU50U Technical instructions

Siemens

Siemens RDU50U Technical instructions

KIOUR TOUCH RSD2 Operations

KIOUR

KIOUR TOUCH RSD2 Operations

oventrop Unibox TQ RTL vario operating instructions

oventrop

oventrop Unibox TQ RTL vario operating instructions

Seltron ACC30 manual

Seltron

Seltron ACC30 manual

Gira 391 operating instructions

Gira

Gira 391 operating instructions

Athena 18 series instruction manual

Athena

Athena 18 series instruction manual

Friedrich FRESHAIRE PVH09K3FA Service & parts manual

Friedrich

Friedrich FRESHAIRE PVH09K3FA Service & parts manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.