
9
GOW-MAC Instrument Co. Model 21-080 Gas Leak Detector Rev. 0, 0623
3-2-7 Tolookforaleak,movethesampleprobeopeningtothetargetareaortting.
NOTE: to locate the smallest leaks on HI range, the user should wear a glove on the
hand holding the probe to isolate the user from the probe inlet.
3-2-8 Leak indication. Segments on the bottom row of the display appear when a leak
is detected. The number of segments from 1 to 16 illuminate in proportion to the
intensity of the signal which is proportional to the concentration of the sampled gas.
4-0 Leak Checking Procedure
CAUTION
EXTREME CARE SHOULD BE EXERCISED
WHEN THE PROBE IS USED AROUND
ELECTRICAL HEATERS, SWITCHES, ETC.
4-1 Care must be exercised to avoid exhaling of breath near the probe or instrument
housing itself. The carbon dioxide (CO2) exhaled can be drawn into the sample probe
or the reference air inlet of the leak detector and falsely indicate a leak.
4-2 If helium (He) is being checked, a leak indicates polarity NEG (negative). Most other
gases or vapors indicate POS (positive polarity). Polarity is on a function of the
sample (leak) gas's thermal conductivity relative to ambient air reference.
4-3 When checking for leaks, it is IMPERATIVEthatttingsorsuspectedleakareas
are DRY. If the area has previously been checked with a water based bubble type
solution and HAS NOT BEEN DRIED, low readings, as well as possible detector
contamination may occur. For example, the thermal conductivity of water vapor
willcounteractthesignalfromheliumleaks.Theneteectwillbefalselowleak
indication.
MOISTURE
DO NOT ALLOW MOISTURE TO BE DRAWN
INTO THE PROBE!
4-4 Using the probe, carefully move the tip around and over suspected joints, seals,
seams, or other areas where leaks occur. The probe must be moved slowly to detect
small leaks.
4-5 A leak is indicated by illumination of display segments and by sound if the alarm
is enabled. Remove the probe from the area, allow the segments to settle to zero
(rezero if necessary), and return probe to the suspected location to verify the leak.
4-6 For more detailed information on leak detector sensitivity, refer Section 8-0
Sensitivity.