SECTION 2 - SPECIFICATIONS
2.1 The GOW-MAC SERIES 5900 DID is a rugged, compact gas chromatograph designed
for the analysis of low concentrations of impurities in base gases.
2.2 The right-hand section of the SERIES 5900 contains the electronics necessary for
the proper operation of the gas chromatograph. A 7” LCD digital touchscreen will
control the DID high voltage, column and detector temperatures, valves and signal
output. Contained within the internal housing is the signal processing and support
electronics for column oven and DID.
2.3 The center section of the chromatograph houses the column oven, a purged housing
which contains the pneumatically actuated valves and the DID detector, and bellows
meteringvalvestocontrolthecolumns,detectorandpurgegas.Allinletttingsand
columnttingsareVCRtype.Therearealsoexitportsformeasuringcolumnow,
purgeowandsampleowwhichare1/8"compressionttings.
There are a maximum of four pneumatically actuated valves mounted inside a purged
ho u si n g. Re fe r to t h e sys te m o w d i ag r am a n d va lve s e c t i o n to d ete r m i ne w h i c h va lve s
are included in your Series 5900.
2.4 SpecicationsfortheSeries5900DIDGC
Dimensions:19.5"Wx25"Dx121/2"H(48.26cmx63.5cmx31.75cm)
Weight: Net: 50 lbs. (22.68 kg)
Power Requirements: Series 5900: 105-125 VAC, 60 Hz, 1100 W
Series 5902: 200-240 VAC, 50 Hz, 1100 W
Fuse: Series 5900: 10 amp
Series 5902: 5 amp
Column Oven
Dimension:10"Wx41/2"Dx71/2"H(25.64cmx11.54cmx19.23cm)
Temperature Range: ambient to 300 °C
Display Readout: 7” LCD Digital Touchscreen
Temperature Control: digital programmable PID controller, solid state, time
proportioning, RTD input, direct reading, ambient to 300 °C
on display screen
Detector
Type: Discharge Ionization
Sensitivity: 5 ppb CH4
Carrier Gas: Grade 6 purity helium (He)
Discharge Gas: Grade 6 purity helium (He)
Temperature Range: ambient to 120 °C