Section 1 Sensor membrane specifications
Specifications are subject to change without notice.
Specification Membrane 2956A Membrane 29552A
Recommended applications Trace measurement High concentration (> 1 mg/l)
Material PFA PTFE
Thickness [μm] 25 50
Calibration gas Span gas or air
Dissolved measurement range 0 ppb to 50 ppm 0 ppb to 200 ppm
Accuracy1
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 0.4 ppb,
or ±1 Pa
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 20 ppb,
or ± 4 Pa
Expected current in air @ 1 bar 25°C
[μA] 25.3 6.5
Temperature compensation range – 5 to 45° C
Temperature measuring range – 5 to 100° C
Response time225 sec. 6 min.
Recommended minimum liquid flow
rate3 [mL/min] 35041004
Recommended minimum linear flow
rate3 [cm/sec] 30 10
Section 2 General information
In no event will the manufacturer be liable for damages resulting from any improper use of product or
failure to comply with the instructions in the manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product
including, without limitation, direct, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user is solely responsible to identify
critical application risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
1At room temperature (25 °C).
2Response time at 25°C for a 90% signal change
3Liquid flow through an ORBISPHERE 32001 flow chamber, with protection cap and no grille
4These flow rates take into account the decomposition of ozone in the tubing between the line
and the flow chamber (theoretical flow rates in the absence of decomposition would be 10 times
less)
English 3