Chromium VI High Range Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 723 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six packets of Chromium VI HR Reagent
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 723
Technical specifications:
Range 0 to 999 ppb
Resolution 1 ppb
Accuracy ±5 ppb ±4% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of the
ASTM Manual of Water and Environmental
Technology, D1687-92
, Diphenylcarbohydrazide method.
The reaction between chromium VI and the reagent
causes a purple tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected]
or view our worldwide contact list at www.hannainst.com.
www.ponpe.com
[email protected]
•Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of HI 723-25 reagent. Replace
the cap and shake vigorously for about
10 seconds. Place the cuvette back into the
meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 6 minutes and press the
button.
•The instrument displays concentration in ppb
of chromium VI. The meter automatically turns
off after 10 minutes.
INTERFERENCE
Interference may be caused by:
Vanadium above 1 ppm: to remove the interference
wait 10 minutes before reading
Mercury (Hg2
2+/Hg2+) above 200 mg/L
Molybdenum (Mo6+) above 200 mg/L
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 723-25 Reagent for 25 Chromium VI High Range tests
OTHER ACCESSORIES
HI 723-11 Chromium VI HR Certified Standard Kit
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731321 Glass cuvettes (4 pcs.)
HI 731225 Cuvette black cap for checker HC (4 pcs.)
HI 731353 Cuvette seal cap for checker HC (4 pcs.)
HI 740028 1.5V AAA batteries (4 pcs.)
HI 93703-50 Cuvette cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead
the instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
IST723 03/13
Measurement procedure: Accessories:Tips for an accurate measurement
[email protected]
www.ponpe.com

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 93719 User manual

Hanna Instruments

Hanna Instruments HI 93719 User manual

Hanna Instruments C 216 User manual

Hanna Instruments

Hanna Instruments C 216 User manual

Hanna Instruments HI 8010 User manual

Hanna Instruments

Hanna Instruments HI 8010 User manual

Hanna Instruments HI9813-51 User manual

Hanna Instruments

Hanna Instruments HI9813-51 User manual

Hanna Instruments HI 96832 User manual

Hanna Instruments

Hanna Instruments HI 96832 User manual

Hanna Instruments HI 2400 User manual

Hanna Instruments

Hanna Instruments HI 2400 User manual

Hanna Instruments HI 7635 User manual

Hanna Instruments

Hanna Instruments HI 7635 User manual

Hanna Instruments HI 96739 User manual

Hanna Instruments

Hanna Instruments HI 96739 User manual

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments HI97721 User manual

Hanna Instruments

Hanna Instruments HI97721 User manual

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI 93702 User manual

Hanna Instruments

Hanna Instruments HI 93702 User manual

Hanna Instruments HI 96801 User manual

Hanna Instruments

Hanna Instruments HI 96801 User manual

Hanna Instruments HI99164 User manual

Hanna Instruments

Hanna Instruments HI99164 User manual

Hanna Instruments HI97753 User manual

Hanna Instruments

Hanna Instruments HI97753 User manual

Hanna Instruments HI 91410 User manual

Hanna Instruments

Hanna Instruments HI 91410 User manual

Hanna Instruments HI708 User manual

Hanna Instruments

Hanna Instruments HI708 User manual

Hanna Instruments HI 5315 User manual

Hanna Instruments

Hanna Instruments HI 5315 User manual

Hanna Instruments HI 96707C User manual

Hanna Instruments

Hanna Instruments HI 96707C User manual

Hanna Instruments HI733 User manual

Hanna Instruments

Hanna Instruments HI733 User manual

Hanna Instruments HI96719C User manual

Hanna Instruments

Hanna Instruments HI96719C User manual

Hanna Instruments HI 2210 User manual

Hanna Instruments

Hanna Instruments HI 2210 User manual

Hanna Instruments HI97737 User manual

Hanna Instruments

Hanna Instruments HI97737 User manual

Hanna Instruments HI781 User manual

Hanna Instruments

Hanna Instruments HI781 User manual

Popular Measuring Instrument manuals by other brands

Fluke 80K-6 instruction sheet

Fluke

Fluke 80K-6 instruction sheet

MARTINDALE AV85 instruction manual

MARTINDALE

MARTINDALE AV85 instruction manual

TPI 712 manual

TPI

TPI 712 manual

Ocean Optics QE Pro Series Installation and operation manual

Ocean Optics

Ocean Optics QE Pro Series Installation and operation manual

Ruskin TDP05K Technical bulletin

Ruskin

Ruskin TDP05K Technical bulletin

Dexcom G5 Mobile Getting started guide

Dexcom

Dexcom G5 Mobile Getting started guide

JDS Uniphase JD7105B manual

JDS Uniphase

JDS Uniphase JD7105B manual

AFRISO HydroFox DMU 08 operating instructions

AFRISO

AFRISO HydroFox DMU 08 operating instructions

Fluke 714C user manual

Fluke

Fluke 714C user manual

Waycon LLD-150-PROF2 manual

Waycon

Waycon LLD-150-PROF2 manual

Asclepion QuadroStar 532/980 user manual

Asclepion

Asclepion QuadroStar 532/980 user manual

DeWalt DW0802 instructions

DeWalt

DeWalt DW0802 instructions

FRER C15WPM100 operating manual

FRER

FRER C15WPM100 operating manual

Rofin SC x10 operating manual

Rofin

Rofin SC x10 operating manual

PURE Avalon 300R Connect manual

PURE

PURE Avalon 300R Connect manual

Thermo Scientific Orion Star A113 instruction sheet

Thermo Scientific

Thermo Scientific Orion Star A113 instruction sheet

Spirax Sarco MTI10 Insertion Installation and maintenance instructions

Spirax Sarco

Spirax Sarco MTI10 Insertion Installation and maintenance instructions

Renishaw Primo Radio Part Setter installation guide

Renishaw

Renishaw Primo Radio Part Setter installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.