Instruction Manual
HI 93711
Free & Total Chlorine
ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTY
HI93711iswarrantedfortwoyears against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarrantyislimitedtorepairorreplacementfreeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required, contactthedealerfromwhomyou purchased the
instrument.Ifunderwarranty,reportthe model number, date of pur-
chase,serialnumberandthenatureof the failure. If the repair is not
coveredbythe warranty,youwillbenotified ofthechargesincurred.If
theinstrumentistobereturnedtoHannaInstruments,first obtain a
ReturnedGoodsAuthorizationNumberfrom the Customer Service de-
partmentand thensenditwithshipment costsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturnthe enclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrights are reserved. Reproductionin whole or inpart is prohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket,Rhode Island, 02895 , USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
The HI 93711 meter measures the free and total chlorine
(Cl2) content in water and wastewater.
Themeter usesan exclusivepositive-locking systemtoensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
SPECIFICATIONS
Range FreeCl20.00 to 2.50 mg/L
TotalCl20.00 to 3.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
TypicalEMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the EPA recommended
DPD method 330.5. The reaction
between the chlorine and the DPD
reagentcausesa pinktint inthesample.
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
POWDER:
Code Unit Description Quantity
HI 93701-0 FreeCl2DPD 1 packet
HI 93711-0 TotalCl2DPD 1 packet
LIQUID:
Code Unit Description Quantity
HI 93701A-T Free and Total Cl2DPD1indicator 3 drops
HI 93701B-T Free and Total Cl2DPD1buffer 3 drops
HI 93701C Total Cl2only DPD3solution 1 drop
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximum concentrationreadable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDE
V