Instruction Manual
HI 93713
Phosphate Low Range
ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
HI 93713 is warranted for two years against defects in
workmanshipandmaterialswhenusedforitsintended purpose and
maintainedaccordingtoinstructions.
Thiswarrantyislimitedtorepairorreplacementfreeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required, contactthedealerfromwhomyou purchased the
instrument.Ifunderwarranty,reportthemodelnumber,dateofpurchase,
serialnumberand the natureofthefailure.Ifthe repair isnotcovered
by the warranty, you will be notified of the charges incurred. If the
instrumentistobereturnedtoHannaInstruments,firstobtainaReturned
GoodsAuthorization Numberfrom theCustomer Servicedepartment
and then send it with shipment costs prepaid. When shipping any
instrument,make sureitisproperly packagedforcompleteprotection.
Tovalidate yourwarranty,filloutandreturnthe enclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrights are reserved. Reproductionin whole or inpart is prohibited
withoutthewrittenconsent of the copyrightowner,HannaInstruments
Inc.,Woonsocket,RhodeIsland, 02895 , USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
The HI 93713 meter measures phosphate (PO43–) content in
water, wastewater and seawater in the 0.00 to 2.50 mg/L
(ppm) range.
Themeter usesan exclusivepositive-lockingsystem toensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in powder form and are supplied in
packets. The amount of reagent is precisely dosed to ensure
maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading
TypicalEMC ±0.01 mg/L
Deviation
Light Source Light Emitting Diode @ 890 nm
Method Adaptationof the ascorbic acid method.
The reaction between phosphate and
the reagent causes a blue tint in the
sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.)
REQUIRED REAGENTS
Code Description Quantity
HI 93713-0 Powder reagent 1 packet
REAGENT SETS
HI93713-01 Reagents for 100 tests
HI93713-03 Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximum concentrationreadable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
V
Warranty
General Description
Preliminary Examination
Specifications Display Code Guide